Accuracy and Precision of the Axis-Shield Afinion Hemoglobin A1c Measurement Device
The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
Background: The Afinion HbA1c (Axis-Shield) is a newer point-of-care device for measurement of hemoglobin A1c (A1C) using a boronate affinity method unlike the more commonly used DCA immunoassay method (Siemens Medical Solutions Diagnostics). The Afinion's accuracy and precision, when compared with high-performance liquid chromatography (HPLC) and DCA methods, have not been established in pediatric practice settings. Methods: Capillary blood was collected from 700 subjects with diabetes mellitus at seven Pediatric Diabetes Consortium sites. Each subject's A1C was measured locally using Afinion and DCA devices, and by a central laboratory (University of Minnesota) using a Tosoh HPLC method. In addition, repeated measurements on six whole blood samples provided by the National Glycohemoglobin Standardization Program (NGSP) were taken at three clinical centers using the Afinion and DCA methods and centrally using the Tosoh HPLC method to assess the precision of each device. Results: The coefficient of variation for measurements of whole blood samples for precision analysis was 2% for Afinion, 3% for DCA, and 1% for HPLC. In the patient samples measured at the seven clinic sites, the Afinion generated higher A1C results than the HPLC (mean difference = +0.15; p < 0.001), while the DCA produced lower values (mean difference = −0.19; p < 0.001). The absolute differences with HPLC were similar for the Afinion and DCA (median 0.2%) with a slight advantage for the Afinion when compared with DCA ( p < 0.001 by rank test). The DCA tended to read lower than HPLC, particularly at high A1C levels ( p < 0.001), while the Afinion's accuracy did not vary by A1C. Conclusions: When compared to the central laboratory HPLC method, the differences between the results of the Afinion and DCA devices are clinically insignificant, and the Afinion and DCA have similar accuracy and precision when used in pediatric practice settings.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“The primary outcome was HbA 1c , measured with the Axis-Shield Afinion point-of-care analyzer (18).”
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Lenters-Westra, E., & Slingerland, R. J. (2010) · Clinical Chemistry · doi:10.1373/clinchem.2009.130641 · Cited by 228
Research works printed but not linked: 10
- 10.2337/diacare.26.4.1158 · DOI · Paper page
- 10.2337/dc06-1909 · DOI · Paper page
- 10.1111/j.1399-543X.2005.00088.x · DOI · Paper page
- 10.1089/dia.2010.0065 · DOI
- 10.1177/096228029900800204 · DOI · Paper page
- 10.1007/s005920070028 · DOI
- Guerci B for the DCA Study Group., 1997, Diabetes Metab., 23, 195
- 10.1515/CCLM.2008.228 · DOI · Paper page
- 10.1016/j.cca.2010.09.004 · DOI
- 10.1515/CCLM.2011.101 · DOI
Cited by
Diagnostic point-of-care tests in resource-limited settings
Drain, P. K., Hyle, E. P., Noubary, F., Freedberg, K. A., Wilson, D., Bishai, W. R., Rodriguez, W., & Bassett, I. V. (2014) · The Lancet Infectious Diseases · doi:10.1016/s1473-3099(13)70250-0 · Cited by 744
Pyatak, E. A., Carandang, K., Vigen, C. L. P., Blanchard, J., Diaz, J., Concha-Chavez, A., Sequeira, P. A., Wood, J. R., Whittemore, R., Spruijt-Metz, D., et al. (2018) · Diabetes Care · doi:10.2337/dc17-1634 · Cited by 116
Self-citationHirst, J. A., McLellan, J. H., Price, C. P., English, E., Feakins, B. G., Stevens, R. J., & Farmer, A. J. (2017) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2016-0303 · Cited by 102
Larsen, S. C., Horgan, G., Mikkelsen, M.-L. K., Palmeira, A. L., Scott, S., Duarte, C., Santos, I., Encantado, J., O'Driscoll, R., Turicchi, J., et al. (2020) · PLOS Medicine · doi:10.1371/journal.pmed.1003168 · Cited by 17
“Blood lipids and HbA1c were measured in fasted state (approximately 10 hours) by a finger-prick test using the point-of-care instrument Alere Afinion AS100 analyser (Abbott Laboratories, Abbott Park, IL) [19,25,26].”
Method
Metadata
- Authors
- Wood, J. R., Kaminski, B. M., Kollman, C., Beck, R. W., Hall, C. A., Yun, J. P., Cengiz, E., Haller, M. J., Hassan, K., Klingensmith, G. J., & Tamborlane, W. V.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2012-03
- Volume
- 6
- Issue
- 2
- Pages
- 380-386
- ISSN
- 1932-2968
- Cited by
- 44
- References
- 1
- Language
- en
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Accuracy and Precision of the Axis-Shield Afinion Hemoglobin A1c Measurement Device. https://app.cochat.ai/papers/10.1177/193229681200600224
- The paper
- Wood, J. R., Kaminski, B. M., Kollman, C., Beck, R. W., Hall, C. A., Yun, J. P., Cengiz, E., Haller, M. J., Hassan, K., Klingensmith, G. J., & Tamborlane, W. V. (2012). Accuracy and Precision of the Axis-Shield Afinion Hemoglobin A1c Measurement Device. Journal of Diabetes Science and Technology. https://doi.org/10.1177/193229681200600224
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.