Robust Fault Detection System for Insulin Pump Therapy Using Continuous Glucose Monitoring
Open accessAbstract
Background The popularity of continuous subcutaneous insulin infusion (CSII), or insulin pump therapy, as a way to deliver insulin more physiologically and achieve better glycemic control in diabetes patients has increased. Despite the substantiated therapeutic advantages of using CSII, its use has also been associated with an increased risk of technical malfunctioning of the device, which leads to an increased risk of acute metabolic complications, such as diabetic ketoacidosis. Current insulin pumps already incorporate systems to detect some types of faults, such as obstructions in the infusion set, but are not able to detect other types of fault such as the disconnection or leakage of the infusion set. Methods In this article, we propose utilizing a validated robust model-based fault detection technique, based on interval analysis, for detecting disconnections of the insulin infusion set. For this purpose, a previously validated metabolic model of glucose regulation in type 1 diabetes mellitus (T1DM) and a continuous glucose monitoring device were used. As a first step to assess the performance of the presented fault detection system, a Food and Drug Administration-accepted T1DM simulator was employed. Results Of the 100 in silico tests (10 scenarios on 10 subjects), only two false negatives and one false positive occurred. All faults were detected before plasma glucose concentration reached 300 mg/dl, with a mean plasma glucose detection value of 163 mg/dl and a mean dete…
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In an experimental full IIS occlusion with five different insulin pumps, the lowest and highest volumes of uninjected insulin when the alarm sounded were recorded by the two patch pumps; the three traditional pumps performed in the midrange between the two patch pumps.
“The lowest and highest volumes of uninjected insulin when the alarm sounded were recorded by the two patch pumps; the three traditional pumps performed in the midrange between the two patch pumps.”
From Insulin Pump Occlusions: For Patients Who Have Been Around the (Infusion) Block (2017) · cites this paper as [14]
The patch pump had the fastest alarm threshold of the four pumps tested.
“The patch pump had the fastest alarm threshold of the four pumps.14”
From Insulin Pump Occlusions: For Patients Who Have Been Around the (Infusion) Block (2017) · cites this paper as [14]
No citation lists for this paper
No citation lists for this paper yet — checked again daily.
References 29 · Cited by 52
- Authors
- Herrero, P., Calm, R., Vehí, J., Armengol, J., Georgiou, P., Oliver, N., & Tomazou, C.
- Affiliations
- Imperial College London (GB)
- Universitat de Girona (ES)
- Institut d'Informàtica i Aplicacions, Universitat de Girona, Campus de Montilivi, Girona, Spain
- Charing Cross Hospital (GB)
- Corresponding author
- Herrero, Pau
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2012-09
- Volume
- 6
- Issue
- 5
- Pages
- 1131-1141
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC3570848
- OpenAlex
- W1996070385
- Cited by
- 52
- References
- 29
- Citations by year
- 2024: 3 · 2023: 4 · 2022: 1 · 2021: 3 · 2020: 2 · 2019: 5 · 2018: 7 · 2017: 6 · 2016: 7 · 2015: 4 · 2014: 5 · 2013: 4 · 2012: 1
- FWCI
- 3.59
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Wellcome Trust
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Engineering › Control and Systems Engineering › Fault Detection and Control Systems
- Physical Sciences › Engineering › Biomedical Engineering › Intravenous Infusion Technology and Safety
- Keywords
- insulin pump therapy, continuous glucose monitoring, fault detection, interval analysis, model-based fault detection, type 1 diabetes, sensor-augmented pump, model uncertainty quantification
- MeSH terms
- Adult, Algorithms, Computer Simulation, Diabetes Mellitus, Type 1, Diagnostic Errors, Humans, Hyperglycemia, Hypoglycemia, Hypoglycemic Agents, Insulin, Insulin Infusion Systems, Medication Errors, Models, Biological, Models, Theoretical, Reference Values, Blood Glucose Self-Monitoring, Equipment Failure Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review