Insulin Pump Occlusions: For Patients Who Have Been Around the (Infusion) Block

Open access
Klonoff, David C., Freckmann, Guido, Heinemann, Lutz·2017·Journal of Diabetes Science and Technology

AI snapshot

Written from the paper’s full text.

TL;DR

This editorial reviews insulin pump and infusion set occlusions, including causes, frequency, detection delays, prediction algorithms, and prevention. It concludes that occlusions are common, often undetected or detected too late, and require better patient education and more robust predictive alarms.

Key findings

  • Occlusions can cause hyperglycemia from failed insulin delivery and hypoglycemia if increased infusion rates are used to overcome an occlusion and noninfused insulin is suddenly delivered after relief.
  • In a 2001 study, approximately 66% (38 of 58) subjects with type 1 diabetes using an insulin pump had an episode of hyperglycemia over 24 weeks, and 36% of episodes were caused by an infusion set occlusion. In a 2002 study of 146 subjects with type 1 diabetes, 25%, 36%, and 22% of those using insulin aspart, lispro, and buffered regular insulin, respectively, reported four or more clogs or blockages; only 9% (15/158), 6% (5/81), and 7% (9/136) coincided with a hyperglycemic episode.
  • Time from occlusion to alarm ranged from 1.5 hours to 24 hours, with most combinations at 2-4 hours; the time was approximately double at an infusion rate of 0.5 U/hour versus 1.0 U/hour. Zisser reported that glucose rises by approximately 1 mg/dl/minute for the first 30 minutes of occlusion, and by extrapolation an occlusion typically does not alarm until glucose has risen by approximately 120-240 mg/dl.
  • A prediction algorithm tested in nine subjects achieved 50% sensitivity and 66% specificity; real-time activation could have reduced time spent >180 mg/dl by 29% (from 82.7 to 58.8 hours/week/patient). In prevention studies, a novel dual-ported catheter had fewer silent occlusions than a traditional set for manual insertion (3/22 [13.6%] vs 12/24 [50%]; P = .012) and mechanical insertion (2/24 [8.3%] vs 9/25 [36%]; P = .037); in a second study, flow interruption events occurred in 3/117 (2.6%) versus 12/118 (10.2%), a 75% risk reduction (95% CI, 20-92%; P = .030).

Methods

  • Design: editorial/narrative review of insulin pump occlusions, covering causes, frequency and significance, detection, prediction, and prevention.
  • Data: cited clinical studies, laboratory/nonclinical comparisons, in vitro pump experiments, and in silico simulations, including studies in type 1 diabetes and healthy adults without diabetes.
  • Analysis: qualitative synthesis of published evidence; no new primary data or systematic review methodology are reported.

Limitations

  • Stated: a definite link between occlusion frequency and elevation of A1C has not been demonstrated, and whether or which occlusion-detection patents are currently used by pump manufacturers cannot be determined.
  • Evident: as an editorial/narrative review, it does not use a systematic search or formal quality assessment, and it synthesizes heterogeneous evidence including small clinical studies, in vitro experiments, and in silico simulations.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1177/1932296817700545

Cites

16 of 16 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Claims the paper makes

19 of 44 cited sources checked · 5 supported · 1 flagged · accuracy not yet measured

  • Findingcitation
    § Prediction

    A failure detection algorithm applied to 100 in silico pump patients with type 1 diabetes detected a total occlusion or dislocation in 75% of patients within 63 minutes with a 10% false positive rate.

    “A failure detection algorithm applied to 100 in silico pump patients with type 1 diabetes detected a total occlusion or dislocation in 75% of patients within 63 minutes with a 10% false positive rate.16”
    Quote in the paperNot a substantial claim1 of 1 sources checked
    • cites [16] 16. Facchinetti A, Del Favero S, Sparacino G, Cobelli C. An online failure detection method of the glucose sensor-insulin pump system: improved overnight safety of type-1 diabetic subjects. IEEE Trans Biomed Eng. 2013;60(2):406-416. [DOI] [PubMed] [Google Scholar] source

      The Performance Evaluation section says this paper tested its algorithm retrospectively on 18 weeks of data from 9 patients, not 100 simulated patients. The Results report 50% sensitivity and failures identified 2.52 ± 1.91 days before the recorded event, rather than 75% detected within 63 minutes; they do not report a 10% false-positive rate. The Discussion mentions a separate approach tested on 100 simulated patients, but does not give the claimed performance figures. (LLM, conf=0.89)

      Checked by a reasoning model — may be wrong.

      Experimental — accuracy not yet measured
  • Premisecitation

    A definite link between the frequency of insulin pump occlusions and elevation of A1C has not been demonstrated.

    “Insulin pump occlusions are felt by many diabetes professionals to impair blood glucose control,1 although a definite link between occlusion frequency and elevation of A1C has not been demonstrated.2,3”
    Quote in the paperNot a substantial claim1 of 2 sources checked
    • cites [2] 2. van Bon AC, Bode BW, Sert-Langeron C, DeVries JH, Charpentier G. Insulin glulisine compared to insulin aspart and to insulin lispro administered by continuous subcutaneous insulin infusion in patients with type 1 diabetes: a randomized controlled trial. Diabetes Technol Ther. 2011;13(6):607-614. [DOI] [PubMed] [Google Scholar] source
    • cites [3] 3. Pfützner A, Sachsenheimer D, Grenningloh M, et al. Using insulin infusion sets in CSII for longer than the recommended usage time leads to a high risk for adverse events: Results from a prospective randomized crossover study. J Diabetes Sci Technol. 2015 (6):1292-1298. [DOI] [PMC free article] [PubMed] [Google Scholar] source

      The Results report HbA1c by infusion-set wear time, while the Discussion mentions catheter occlusions without relating their frequency to HbA1c. Our copy does not assess that relationship or state whether a definite link has been demonstrated (searched: occlusion frequency, HbA1c, and glycemic control). (LLM, conf=0.72)

      Experimental — accuracy not yet measured
  • Findingcitation
    § Detection

    When an insulin pump is shut off, blood glucose level rises by approximately 1 mg/dl/minute for the first 30 minutes of occlusion.

    “Zisser demonstrated in 2008 that when an insulin pump is shut off blood glucose level rises by approximately 1 mg/dl/minute for the first 30 minutes of occlusion.”
    Quote in the paperSubstantial claim1 of 1 sources checked
    • cites [12] 12. Zisser H. Quantifying the impact of a short-interval interruption of insulin-pump infusion sets on glycemic excursions. Diabetes Care. 2008;31(2):238-239. [DOI] [PubMed] [Google Scholar] source

      The Results report that sensor glucose rose by 18 ± 58 mg/dL over a two-hour pump suspension, supporting a rise after insulin delivery stops. The paper does not report a rise of approximately 1 mg/dL per minute during the first 30 minutes of occlusion or attribute that finding to Zisser in 2008. Page numbers were not available in our copy. (LLM, conf=0.85)

      Experimental — accuracy not yet measured
  • Findingcitation
    § Detection

    The time from occlusion to alarm was approximately double for an infusion rate of 0.5 U/hour than with an infusion rate of 1.0 U/hour.

    “The time from occlusion to alarm was approximately double for an infusion rate of 0.5 U/hour than with an infusion rate of 1.0 U/hour.11”
    Quote in the paperSubstantial claim1 of 1 sources checked
    • cites [11] 11. van Bon AC, Dragt D, DeVries JH. Significant time until catheter occlusion alerts in currently marketed insulin pumps at two basal rates. Diabetes Technol Ther. 2012;14(5):447-448. [DOI] [PubMed] [Google Scholar] source

      The Research Design and Methods passage reports glucose changes after a 30-minute pump disconnection, not the time from an occlusion to an alarm. Searches for the 0.5 and 1.0 U/hour comparison and for occlusion-alarm timing found no passage addressing the claim in our copy; the indexed passages have no page labels. (LLM, conf=0.76)

      Experimental — accuracy not yet measured
  • Interpretationcitation

    Hypoglycemia can occur if an occlusion is due to kinking or compression and the insulin infusion rate is increased to overcome the occlusion and maintain normoglycemia, because after the occlusion is relieved a relatively large volume of noninfused insulin is suddenly delivered.

    “hypoglycemia if an occlusion is due to kinking or compression and if the insulin infusion rate is increased to overcome the occlusion and maintain normoglycemia, then after the occlusion is relieved, a relatively large volume of noninfused insulin that was not infused during the occlusion will suddenly be delivered.4”
    Quote in the paperSubstantial claim1 of 1 sources checked
    • cites [4] 4. Dumont-Fillon D, Tahriou H, Conan C, Chappel E. Insulin micropump with embedded pressure sensors for failure detection and delivery of accurate monitoring. Micromachines. 2014;5(4):1161-1172. [Google Scholar] source

      Split into 2 atoms — 0 confirmed, 2 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.

      Experimental — accuracy not yet measured
  • May combine two claims. 7
  • Classified as probably not a substantial claim. 4
  • Not yet classified. 1
  • May combine two claims.

    Findingcitation
    § Detection

    In a study of five brands of insulin pumps, the time from occlusion to alarm ranged from 1.5 hours to 24 hours, with most combinations in the range of 2-4 hours.

    “The time from occlusion to alarm ranged from 1.5 hours to 24 hours; most combinations were in the range of 2-4 hours.”
    Quote in the paperSubstantial claim1 of 1 sources checked
    • cites [11] 11. van Bon AC, Dragt D, DeVries JH. Significant time until catheter occlusion alerts in currently marketed insulin pumps at two basal rates. Diabetes Technol Ther. 2012;14(5):447-448. [DOI] [PubMed] [Google Scholar] source

      The Research Design and Methods section describes a 30-minute pump disconnection study involving three pump makes, not a measurement of occlusion-to-alarm times. Searches for occlusion alarms and the claimed 1.5–24-hour and 2–4-hour ranges found no passage addressing those figures in our copy. (LLM, conf=0.76)

      Experimental — accuracy not yet measured
  • May combine two claims.

    Interpretationcitation
    § Conclusions

    Occlusions of insulin infusion sets often occur without an alarmed warning, or the warning is provided too late.

    “often without an alarmed warning or this provided too late”
    Quote in the paperSubstantial claim3 of 3 sources checked
    • cites [1] 1. Gibney M, Xue Z, Swinney M, Bialonczyk D, Hirsch L. Reduced silent occlusions with a novel catheter infusion set (BD FlowSmart): results from two open-label comparative studies. Diabetes Technol Ther. 2016;18(3):136-143. [DOI] [PMC free article] [PubMed] [Google Scholar] source
    • cites [11] 11. van Bon AC, Dragt D, DeVries JH. Significant time until catheter occlusion alerts in currently marketed insulin pumps at two basal rates. Diabetes Technol Ther. 2012;14(5):447-448. [DOI] [PubMed] [Google Scholar] source

      The Introduction and Conclusions discuss deliberate infusion-set disconnections and their effects on glucose, not occlusions. Searches for occlusion alarms and delayed warnings found no passage addressing whether occlusions occur without an alarm or whether alarms arrive too late. Page numbers were not supplied for these passages. (LLM, conf=0.76)

      Experimental — accuracy not yet measured
    • cites [17] 17. Cescon M, DeSalvo DJ, Ly TT, et al. Early detection of infusion set failure during insulin pump therapy in Type 1 diabetes. J Diabetes Sci Technol. 2016;10(6):1268-1276. [DOI] [PMC free article] [PubMed] [Google Scholar] Articles from Journal of Diabetes Science and Technology are provided here courtesy of Diabetes Technology Society source

      The Discussion (page not available) says pump mechanisms may detect a full occlusion only after some time and do not detect infusion-site infections, supporting the concern about late or absent warnings for some failures. It does not establish that occlusions often occur without an alarm. (LLM, conf=0.84)

      Experimental — accuracy not yet measured
  • May combine two claims.

    Findingcitation
    § Prevention

    In the first study (n=25), significantly fewer silent occlusions occurred with the novel IIS than with the traditional IIS for both manual (3 of 22 [13.6%] vs 12 of 24 [50%]; P = .012) and mechanical (2 of 24 [8.3%] vs 9 of 25 [36%]; P = .037) insertions.

    “In the first study, significantly fewer silent occlusions occurred with the novel IIS than with the traditional IIS for both manual (3 of 22 [13.6%] vs 12 of 24 [50%]; P = .012) and mechanical (2 of 24 [8.3%] vs 9 of 25 [36%]; P = .037) insertions.”
    Quote in the paperSubstantial claim1 of 1 sources checked
    • cites [1] 1. Gibney M, Xue Z, Swinney M, Bialonczyk D, Hirsch L. Reduced silent occlusions with a novel catheter infusion set (BD FlowSmart): results from two open-label comparative studies. Diabetes Technol Ther. 2016;18(3):136-143. [DOI] [PMC free article] [PubMed] [Google Scholar] source
  • May combine two claims.

    Findingcitation
    § Detection

    In an experimental full IIS occlusion with five different insulin pumps, the lowest and highest volumes of uninjected insulin when the alarm sounded were recorded by the two patch pumps; the three traditional pumps performed in the midrange between the two patch pumps.

    “The lowest and highest volumes of uninjected insulin when the alarm sounded were recorded by the two patch pumps; the three traditional pumps performed in the midrange between the two patch pumps.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [14] 14. Borot S, Franc S, Cristante J, et al. Accuracy of a new patch pump based on a microelectromechanical system (MEMS) compared to other commercially available insulin pumps: results of the first in vitro and in vivo studies. J Diabetes Sci Technol. 2014;8(6):1133-1141. [DOI] [PMC free article] [PubMed] [Google Scholar] source
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Causes of Occlusion

    The patent literature contains information about occlusion detection systems from at least 400 patents from 1971 through 2016.

    “The patent literature contains information about occlusion detection systems from at least 400 patents from 1971 through 2016.5”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [5] 5. Sims Deltec, Inc. Occlusion detection system for an infusion pump. December 9, 1997. Available at: https://www.google.com/patents/US5695473. Accessed February 28, 2017. source
  • Not yet classified.

    Premise
    § Frequency and Significance of Pump Occlusions

    In that study, the median occlusion rate was 0% for both groups.

    “The median occlusion rate was 0% for both groups”
    Quote in the paperNot yet classified0 of 1 sources checked
    • cites [8] 8. Hoogma RP, Schumicki D. Safety of insulin glulisine when given by continuous subcutaneous infusion using an external pump in patients with type 1 diabetes. Horm Metab Res. 2006;38(6):429-433. [DOI] [PubMed] [Google Scholar] source
  • May combine two claims.

    Interpretationcitation
    § Detection

    Occlusions are not necessarily noted immediately and interruptions of insulin flow during infusion often appear to not trigger occlusion alarms because it takes time to build up sufficient back pressure until an occlusion is detected and an alarm sounds.

    “Occlusions are not necessarily noted immediately and interruptions of insulin flow during infusion often appear to not trigger occlusion alarms10 as it takes time to build up sufficient back pressure until an occlusion is detected and an alarm sounds.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [10] 10. Evert AB, Bode BW, Buckingham BA, et al. Improving patient experience with insulin infusion sets: practical guidelines and future directions. Diabetes Educ. 2016;42(4):470-484. [DOI] [PubMed] [Google Scholar] source
  • Classified as probably not a substantial claim.

    Premisecitation
    § Frequency and Significance of Pump Occlusions

    For the multivariate statistical analysis, P was set at the .025 level for treatment effect.

    “For the multivariate statistical analysis, P was set at the .025 level for treatment effect.2”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [2] 2. van Bon AC, Bode BW, Sert-Langeron C, DeVries JH, Charpentier G. Insulin glulisine compared to insulin aspart and to insulin lispro administered by continuous subcutaneous insulin infusion in patients with type 1 diabetes: a randomized controlled trial. Diabetes Technol Ther. 2011;13(6):607-614. [DOI] [PubMed] [Google Scholar] source
  • May combine two claims.

    Premisecitation
    § Frequency and Significance of Pump Occlusions

    Percentages of subjects with at least one unexplained hyperglycemia and/or IIS occlusion were not significantly different between GLU and ASP (68.4% vs 62.1%, P = .04) and between GLU and LIS (68.4% vs 61.3%, P = .03).

    “Percentages of subjects with at least one unexplained hyperglycemia and/or IIS occlusion were not significantly different between GLU and ASP (68.4% vs 62.1%, P = .04) and between GLU and LIS (68.4% vs 61.3%, P = .03).”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [2] 2. van Bon AC, Bode BW, Sert-Langeron C, DeVries JH, Charpentier G. Insulin glulisine compared to insulin aspart and to insulin lispro administered by continuous subcutaneous insulin infusion in patients with type 1 diabetes: a randomized controlled trial. Diabetes Technol Ther. 2011;13(6):607-614. [DOI] [PubMed] [Google Scholar] source
  • Classified as probably not a substantial claim.

    Premisecitation
    § Frequency and Significance of Pump Occlusions

    In a follow-up study comparing the occlusion rate of pumps delivering GLU against those of both ASP and LIS, 256 subjects with type 1 diabetes received three sequences of insulin analogs, each for 13 weeks for a total of 39 weeks.

    “In a follow-up study comparing the occlusion rate of pumps delivering GLU against those of both APS and LIS, 256 subjects with type 1 diabetes received three sequences of insulin analogs, each for 13 weeks for a total of 39 weeks.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [2] 2. van Bon AC, Bode BW, Sert-Langeron C, DeVries JH, Charpentier G. Insulin glulisine compared to insulin aspart and to insulin lispro administered by continuous subcutaneous insulin infusion in patients with type 1 diabetes: a randomized controlled trial. Diabetes Technol Ther. 2011;13(6):607-614. [DOI] [PubMed] [Google Scholar] source
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation
    § Frequency and Significance of Pump Occlusions

    At the end of that study, each subject had used all three analogs.

    “At the end of the study they had each used all three analogs.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [2] 2. van Bon AC, Bode BW, Sert-Langeron C, DeVries JH, Charpentier G. Insulin glulisine compared to insulin aspart and to insulin lispro administered by continuous subcutaneous insulin infusion in patients with type 1 diabetes: a randomized controlled trial. Diabetes Technol Ther. 2011;13(6):607-614. [DOI] [PubMed] [Google Scholar] source
  • May combine two claims.

    Premisecitation
    § Causes of Occlusion

    Occlusion detection by infusion pumps may depend on detecting increased pressure in the tubing, increased radial diameter of the tubing, lack of a temperature gradient of heated fluid, lack of optical signals from an inline piston, or an unexplained rise in the concentration of interstitial fluid glucose.

    “Occlusion detection by infusion pumps may depend on detecting increased pressure in the tubing, increased radial diameter of the tubing, lack of a temperature gradient of heated fluid, lack of optical signals from an inline piston, or an unexplained rise in the concentration of interstitial fluid glucose.5”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [5] 5. Sims Deltec, Inc. Occlusion detection system for an infusion pump. December 9, 1997. Available at: https://www.google.com/patents/US5695473. Accessed February 28, 2017. source
12 of 12
  • The paper relies on this without support.

    Most occlusion detection technology is proprietary.

    “Most occlusion detection technology is proprietary.”
  • The paper relies on this without support.

    In either gradual or acute occlusion, a predictive algorithm can analyze the rise of pressure and deliver an occlusion alarm.

    “In either gradual or acute occlusion, a predictive algorithm can analyze the rise of pressure and deliver an occlusion alarm.”
  • The paper relies on this without support.

    Increasing back pressure is generally detected by measuring the maximum pressure and the shape of the pressure wave for each pump stroke in the reservoir and/or IIS to look for evidence of an increased pressure pattern indicative of an occlusion.

    “Increasing back pressure is generally detected by measuring the maximum pressure and the shape of the pressure wave for each pump stroke in the reservoir and/or IIS to look for evidence of an increased pressure pattern indicative of an occlusion.”
  • The authors state this as an assumption.

    The relationship of glucose increase to time without insulin remains linear.

    “assuming that the relationship of glucose increase to time without insulin remains linear”
  • The paper relies on this without support.

    Such a rise in glycemia can be difficult to correct later by correction boluses.

    “Such a rise in glycemia can be difficult to correct later by correction boluses.”
  • The authors state this as an assumption.

    The patient would have acknowledged the alert and reacted with appropriate actions.

    “assuming that the patient would have acknowledged the alert and reacted with appropriate actions”
  • The paper relies on this without support.

    If a patient delays IIS replacement beyond the manufacturer’s recommended interval for changing these IIS, then there is increased risk of site inflammation or infection.

    “If a patient delays IIS replacement beyond the manufacturer’s recommended interval for changing these IIS, then there is increased risk of site inflammation or infection.”
  • The paper relies on this without support.

    This type of problem will be manifested by recurrent hyperglycemia in spite of steady or increasing does of insulin with or often without an alarm.

    “This type of problem will be manifested by recurrent hyperglycemia in spite of steady or increasing does of insulin with or often without an alarm.”
8 of 8

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Cites

Flagged first, then by citations

1 of the 17 references couldn’t be matched to a known paper.

Cited by

Sorted by citations

Authors
Klonoff, D. C., Freckmann, G., & Heinemann, L.
Affiliations
  • Mills Peninsula Health Services (US)
  • Universität Ulm (DE)
  • Science & Co GmbH, Düsseldorf, Germany
Venue
Journal of Diabetes Science and Technology
Published
2017-03-30
Volume
11
Issue
3
Pages
451-454
ISSN
1932-2968 (ISSN-L)
OpenAlex
W2599130127
Cited by
41
References
17
Citations by year
2026: 6 · 2025: 1 · 2024: 8 · 2023: 6 · 2022: 3 · 2021: 5 · 2020: 2 · 2019: 6 · 2018: 3 · 2017: 1
FWCI
2.24
Type
journal article
Language
en
Open access
Locations
  • Journal of Diabetes Science and Technology (publishedVersion)
  • PubMed (publishedVersion)
  • PubMed Central (submittedVersion)
Topics
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
  • Physical Sciences › Engineering › Biomedical Engineering › Intravenous Infusion Technology and Safety
  • Health Sciences › Health Professions › Emergency Medical Services › Patient Safety and Medication Errors
Keywords
diabetes self-management, patient education, insulin delivery devices
MeSH terms
Diabetes Mellitus, Type 1, Equipment Failure, Humans, Hypoglycemic Agents, Insulin, Insulin Infusion Systems

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-13.

Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-13.

Cited sources are checked by gpt-6-sol.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review