A Comparison of Time Delay in Three Continuous Glucose Monitors for Adolescents and Adults

Open access
Sinha, Manasi, McKeon, Katherine M., Parker, Savan, Goergen, Laura G., Zheng, Hui, El-Khatib, Firas H., Russell, Steven J. · 2017 · Journal of Diabetes Science and Technology
Cited by 49

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Abstract

Background: The physiologic delay in glucose diffusion from the blood to the interstitial fluid and instrumental factors contribute to the delay between changes in plasma glucose (PG) and measurements made by continuous glucose monitors (CGMs). This study compared the duration of this delay for three CGMs. Methods: A total of 24 healthy adolescent and adult subjects with type 1 diabetes wore three CGM devices simultaneously for 48 hours: Dexcom G4 Platinum, Abbott Navigator, and Medtronic Enlite. The time delay between PG and CGM-estimated plasma glucose (CGMG) was estimated by comparing time-shifted CGMG with reference PG taken every 15 minutes. Results: The delay estimated by our approach was larger for the Navigator than for the G4 Platinum in adolescents (7.7 ± 1.1 versus 5.6 ± 0.9 min, P = .0396) and adults (10.9 ± 1.1 versus 8.1 ± 0.7 min, P = .0107). The delay was nominally longer for the Navigator than for the Enlite in both the adolescent (7.7 ± 1.1 versus 4.3 ± 1.0 min, P = .0728) and adult (10.9 ± 1.1 versus 8.3 ± 0.9 min, P = .111) populations, but these differences were not statistically significant. There was no difference in the delay between G4 Platinum and Enlite. Adolescents had shorter delays than adults for all three devices. There was a significant correlation between longer delay and increasing age for the G4 Platinum and Navigator. Conclusions: There are differences in the estimated PG to CGMG time delays between CGM devices in the same subjects. The delay between PG and CGMG is smaller for adolescents than for adults. The PG–to–CGMG time delay is influenced by both instrument and host factors.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1177/1932296817704443

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 2 references

Cited by

See all 49

Metadata

Authors
Sinha, M., McKeon, K. M., Parker, S., Goergen, L. G., Zheng, H., El-Khatib, F. H., & Russell, S. J.
Venue
Journal of Diabetes Science and Technology
Published
2017-05-01
Volume
11
Issue
6
Pages
1132-1137
ISSN
1932-2968 (ISSN-L)
OpenAlex
W2611014035
Cited by
49
References
2
FWCI
3.34
Type
journal article
Language
en
Open access
green

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: A Comparison of Time Delay in Three Continuous Glucose Monitors for Adolescents and Adults. https://app.cochat.ai/papers/10.1177/1932296817704443
The paper
Sinha, M., McKeon, K. M., Parker, S., Goergen, L. G., Zheng, H., El-Khatib, F. H., & Russell, S. J. (2017). A Comparison of Time Delay in Three Continuous Glucose Monitors for Adolescents and Adults. Journal of Diabetes Science and Technology. https://doi.org/10.1177/1932296817704443

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.