Clinical Evaluation of a New, Lower Pain, One Touch Lancing Device for People With Diabetes: Virtually Pain-Free Testing and Improved Comfort Compared to Current Lancing Systems
Open accessThe DOI matches its Crossref record; 11 of the 11 references it cites were found in a registry or an index.
Abstract
Background: Patients worldwide use lancing devices to perform self-monitoring of blood glucose (SMBG). Continued innovation remains important to minimize pain associated with finger sticking to facilitate patient adherence to SMBG regimens. The original One Touch Delica lancer (OTD) reduced the pain of testing. This study investigated a completely re-engineered lancing system, called One Touch Delica Plus (OTDP), intended to further minimize pain while improving the patient experience. Methods: One hundred and three subjects (type 1 or 2 diabetes) experienced finger sticks with both lancing systems. Device depth and blood volume (BV) generation were determined during visit 1. These parameters were used during pain assessments in visit 2. Patient feedback was also recorded using surveys. Results: Finger-stick testing was less painful using OTDP compared to original OTD irrespective of whether 30 (−48.2 mm, −30.4 mm, P < 0.0001) or 33 gauge (−42.6 mm, 30.5 mm, P = 0.0004) lancets were used. All 103 subjects generated the required BV (≥0.4 µL) per protocol. At equivalent depth penetration, OTDP generated higher average BVs than OTD using 30 (1.58 µL, 1.10 µL) or 33 gauge lancets (1.39 µL, 0.98 µL). Subjects responded that testing with OTDP was virtually pain free (78%) and that testing with OTDP was less painful than using their current lancing systems (79%) or original OTD (67%). Conclusion: OTDP demonstrated lower pain than the original OTD lancing system and participants agreed that testing with OTDP was virtually pain free and less painful than using their current lancing systems.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Certain meter system characteristics, such as the use of lancing devices that are less painful (21) and the ability to reapply blood to a strip with an insufficient initial sample, or meters with integrated speech that can read aloud glucose levels for visually impaired individuals (22), may also be beneficial to people with diabetes (23) and may make BGM less burdensome to perform.”
From 7. Diabetes Technology: Standards of Care in Diabetes—2025 (2025) · cites this paper as [21]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
International Diabetes Federation 2017
Carracher, A. M., Marathe, P. H., & Close, K. L. (2018) · Journal of Diabetes · doi:10.1111/1753-0407.12644 · Cited by 13,418
Introduction: Standards of Medical Care in Diabetes—2018
(2018) · Diabetes Care · doi:10.2337/dc18-sint01 · Cited by 472
Finger Pricking and Pain: A Never Ending Story
Heinemann, L. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200526 · Cited by 148
Adherence of self-monitoring of blood glucose in persons with type 1 diabetes in Sweden
Moström, P., Ahlén, E., Imberg, H., Hansson, P.-O., & Lind, M. (2017) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2016-000342 · Cited by 110
Clinical Utility of SMBG: Recommendations on the Use and Reporting of SMBG in Clinical Research
Schnell, O., Barnard, K., Bergenstal, R., Bosi, E., Garg, S., Guerci, B., Haak, T., Hirsch, I. B., Ji, L., Joshi, S. R., et al. (2015) · Diabetes Care · doi:10.2337/dc14-2919 · Cited by 36
Grady, M., Mitchell, A. P., Pynes, M. K., Katz, L. B., & Ginsberg, B. H. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814533172 · Cited by 32
Self-citationHeinemann, L., & Boecker, D. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500420 · Cited by 28
Comparison of Lancing Devices for Self-Monitoring of Blood Glucose regarding Lancing Pain
Kocher, S., Tshiananga, J. K. T., & Koubek, R. D. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300517 · Cited by 23
Ito, T., Kamoi, K., Minagawa, S., Kimura, K., & Kobayashi, A. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400420 · Cited by 16
Low discomfort and pain associated with intensified insulin therapy in children and adolescents
Karges, B., Muche, R., Moritz, M., Riegger, I., Debatin, K., Heinze, E., Wabitsch, M., & Karges, W. (2008) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2007.11.007 · Cited by 14
Cited by
7. Diabetes Technology: Standards of Medical Care in Diabetes—2022
American Diabetes Association Professional Practice Committee (2022) · Diabetes Care · doi:10.2337/dc22-s007 · Cited by 358
“are less painful (9) and the ability to reap-”
Background7. Diabetes Technology: Standards of Care in Diabetes—2025
American Diabetes Association Professional Practice Committee, ElSayed, N. A., McCoy, R. G., Aleppo, G., Balapattabi, K., Beverly, E. A., Briggs Early, K., Bruemmer, D., Echouffo-Tcheugui, J. B., Ekhlaspour, L., et al. (2025) · Diabetes Care · doi:10.2337/dc25-s007 · Cited by 348
7. Diabetes Technology: Standards of Care in Diabetes — 2023
ElSayed, N. A., Aleppo, G., Aroda, V. R., Bannuru, R. R., Brown, F. M., Bruemmer, D., Collins, B. S., Hilliard, M. E., Isaacs, D., Johnson, E. L., et al. (2023) · Diabetes Care · doi:10.2337/dc23-s007 · Cited by 301
7. Diabetes Technology: Standards of Care in Diabetes—2024
American Diabetes Association Professional Practice Committee, ElSayed, N. A., Aleppo, G., Bannuru, R. R., Bruemmer, D., Collins, B. S., Ekhlaspour, L., Hilliard, M. E., Johnson, E. L., Khunti, K., et al. (2024) · Diabetes Care · doi:10.2337/dc24-s007 · Cited by 281
7. Diabetes Technology: Standards of Medical Care in Diabetes—2021
American Diabetes Association (2021) · Diabetes Care · doi:10.2337/dc21-s007 · Cited by 202
“Certainmeter systemcharacteristics, suchas the use of lancing devices that are less painful (5) and the ability to reapply blood to a stripwith an insufficient initial sample,”
Background
Metadata
- Authors
- Grady, M., Lamps, G., Shemain, A., Cameron, H., & Murray, L.
- Affiliations
- LifeScan Scotland Ltd., Inverness, UK
- APS America, LLC., Marietta, GA, USA
- Synexus USA, Pinellas Park, FL, USA
- Corresponding author
- Grady, Mike
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2019-07-17
- Volume
- 15
- Issue
- 1
- Pages
- 53-59
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC7783011
- OpenAlex
- W2960188884
- Cited by
- 18
- References
- 11
- Citations by year
- 2026: 1 · 2025: 1 · 2024: 3 · 2023: 5 · 2022: 2 · 2021: 4 · 2020: 2
- FWCI
- 1.62
- Type
- journal article
- Language
- en
- Open access
- green
- europepmc.org
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- LifeScan
- APS America
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Health Sciences › Medicine › Dermatology › Skin Diseases and Diabetes
- Keywords
- self-monitoring of blood glucose, diabetes mellitus, blood volume, patient comfort, type 1 diabetes, type 2 diabetes
- MeSH terms
- Blood Specimen Collection, Diabetes Mellitus, Humans, Pain, Pain Measurement, Blood Glucose Self-Monitoring
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Clinical Evaluation of a New, Lower Pain, One Touch Lancing Device for People With Diabetes: Virtually Pain-Free Testing and Improved Comfort Compared to Current Lancing Systems. https://app.cochat.ai/papers/10.1177/1932296819856665
- The paper
- Grady, M., Lamps, G., Shemain, A., Cameron, H., & Murray, L. (2021). Clinical Evaluation of a New, Lower Pain, One Touch Lancing Device for People With Diabetes: Virtually Pain-Free Testing and Improved Comfort Compared to Current Lancing Systems. Journal of Diabetes Science and Technology. https://doi.org/10.1177/1932296819856665
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.