Background: We investigated wearable components of the Dexcom G6 continuous glucose monitoring (CGM) System in simulated therapeutic and diagnostic radiologic procedures. Methods: G6 transmitters were loaded with simulated glucose data and attached to sensors. Sets of sensor/transmitter pairs were exposed to x-rays to simulate a radiotherapeutic procedure and to radiofrequency (RF) and magnetic fields to simulate diagnostic magnetic resonance imaging (MRI). The x-ray simulation provided a cumulative dose of 80 Gy. The MRI simulation used RF fields oscillating at 64 or 128 MHz and magnetic fields of 1.5 or 3 T. During the MRI simulation, displacement force, induced heating, and induced currents were measured. After the simulations, bench tests were used to assess data integrity on the transmitters and responsiveness of sensors to various concentrations of aqueous glucose. Results: Glucose concentrations reported by sensor/transmitter pairs after undergoing x-irradiation or a simulated MRI exam were similar to those from control (unexposed) devices. During the 3 T MRI simulation, the devices experienced a displacement force of 306 g, which was insufficient to dislodge the sensor/transmitter from the substrate, RF-induced heating of <2°C, and an induced current of <16 pA. Data stored on the transmitters prior to the MRI simulation remained intact. Conclusion: Wearable components of the G6 CGM System retain basic functionality and data integrity after exposure to simulated therapeutic and diagnostic radiologic procedures. The devices are unlikely to be affected by x-irradiation used in typical imaging studies. Simulated MRI procedures create displacement force, minimal heating, and current in sensor/transmitter pairs.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
State of Type 1 Diabetes Management and Outcomes from the T1D Exchange in 2016–2018
Foster, N. C., Beck, R. W., Miller, K. M., Clements, M. A., Rickels, M. R., DiMeglio, L. A., Maahs, D. M., Tamborlane, W. V., Bergenstal, R., Smith, E., et al. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0384 · Cited by 1,907
Six-Month Randomized, Multicenter Trial of Closed-Loop Control in Type 1 Diabetes
Brown, S. A., Kovatchev, B. P., Raghinaru, D., Lum, J. W., Buckingham, B. A., Kudva, Y. C., Laffel, L. M., Levy, C. J., Pinsker, J. E., Wadwa, R. P., et al. (2019) · New England Journal of Medicine · doi:10.1056/nejmoa1907863 · Cited by 1,030
Cited by
Continuous glucose monitors and automated insulin dosing systems in the hospital consensus guideline
Galindo, R. J., Umpierrez, G. E., Rushakoff, R. J., Basu, A., Lohnes, S., Nichols, J. H., Spanakis, E. K., Espinoza, J., Palermo, N. E., Awadjie, D. G., et al. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820954163 · Cited by 170
Galindo, R. J., Aleppo, G., Klonoff, D. C., Spanakis, E. K., Agarwal, S., Vellanki, P., Olson, D. E., Umpierrez, G. E., Davis, G. M., & Pasquel, F. J. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820932903 · Cited by 155
“26 Simulated MRI conditions, however, created displacement force, minimal heating, and current in sensor/transmitter pairs.”
Sherr, J. L., Heinemann, L., Fleming, G. A., Bergenstal, R. M., Bruttomesso, D., Hanaire, H., Holl, R. W., Petrie, J. R., Peters, A. L., & Evans, M. (2022) · Diabetes Care · doi:10.2337/dci22-0018 · Cited by 136
“In at least one simulation it was found that CGM can stay in place during radiographic and MRI procedures [38, 39].”
BackgroundBaker, M., Musselman, M. E., Rogers, R., & Hellman, R. (2022) · American Journal of Health-System Pharmacy · doi:10.1093/ajhp/zxab456 · Cited by 27
“X-rays, however, are unlikely to affect Dexcom G6 components, and the device may be worn during scans.(28) As this patient population received frequent chest X-rays, communication with the radiology team was important.”
Background
Metadata
- Authors
- Thomas, C., Welsh, J. B., Lu, S., & Gray, J. M.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2020-04-22
- Volume
- 15
- Issue
- 4
- Pages
- 781-785
- ISSN
- 1932-2968 (ISSN-L)
- OpenAlex
- W3016327970
- Cited by
- 37
- References
- 2
- FWCI
- 3.98
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Safety and Functional Integrity of Continuous Glucose Monitoring Components After Simulated Radiologic Procedures. https://app.cochat.ai/papers/10.1177/1932296820920948
- The paper
- Thomas, C., Welsh, J. B., Lu, S., & Gray, J. M. (2021). Safety and Functional Integrity of Continuous Glucose Monitoring Components After Simulated Radiologic Procedures. Journal of Diabetes Science and Technology. https://doi.org/10.1177/1932296820920948
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.