Skin and subcutaneous adipose layer thickness in adults with diabetes at sites used for insulin injections: Implications for needle length recommendations
Open accessAbstract
OBJECTIVE: During subcutaneous insulin therapy, inadvertent intramuscular (IM) injections may increase pain and/or adversely affect glucose control. The most appropriate needle length for patients depends on skin thickness (ST) and the distance to muscle fascia. ST and subcutaneous adipose layer thickness (SCT) were measured in adults with diabetes. RESEARCH DESIGN AND METHODS: A total of 388 US adults with diabetes (in three BMI subgroups: <25, 25-29.9, and >or=30 kg/m(2)) with diverse demographic features were evaluated. Each subject had ultrasound measurements of ST and SCT at four injection sites. RESULTS: Subjects had BMI 19.4-64.5 kg/m(2), age 18-85 years; 40% Caucasian, 25% Asian, 16% Black, 14% Hispanic; 28% type 1 diabetes. Mean ST (+/-95% CI) was: arm 2.2 mm (2.2, 2.3), thigh 1.9 mm (1.8, 1.9), abdomen 2.2 mm (2.1, 2.2) and buttocks 2.4 mm (2.4, 2.5). Multivariate analyses showed body site, gender, BMI, and race are statistically significant factors for ST but effects were small. Thigh ST was <0.6 mm thinner than the buttocks. Differences of 10 kg/m(2) account for 0.2 mm ST variation. Mean SCT was: arm 10.8 mm (10.2, 11.3), thigh 10.4 mm (9.8, 10.9), abdomen 13.9 mm (13.2, 17.7) and buttocks 15.4 mm (14.7, 16.2). Females had 5.1 mm greater SCT. Differences of 10 kg/m(2) account for 4 mm SCT variation. ADVERSE EVENTS: A few mild hypo- or hyperglycemia events, unrelated to study procedure, were detected and treated before subject discharge from study visits. LIMITATIONS: Only adults in the US were studied; some measurements could not be obtained on every subject, at every injection site. CONCLUSIONS: Injection site ST does not differ by clinically significant degrees in demographically diverse adults with diabetes; SCT has a wider range. Needles >or=8 mm, inserted perpendicularly, may frequently enter muscle in limbs of males and those with BMI <25 kg/m(2). With 90 degrees insertion, needles 4-5 mm enter the subcutaneous tissue with minimal risk of IM injection in virtually all adults. These data will assist recommending appropriate length needles for subcutaneous insulin injections in adults.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Needles as short as 4 mm, injected at a 90 angle, enter the subcutaneous space with minimal risk of intramuscular injection in most adults.
“Needles as short as 4 mm, injected at a 90 angle, enter the subcutaneous space with minimal risk of intramuscular injection in most adults (92).”
The depth of the target vessel and intervening pulsating arterioles may vary across individuals and within individuals with body location and due to factors such as level of obesity.
“Furthermore, one component of optimizing source/detector separation is the depth of the target vessel and intervening pulsating arterioles, which may vary across individuals and within individuals with body location and due to factors such as level of obesity [56].”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [56]
Publications that experimentally and explicitly demonstrate the effect of obesity or BMI on the PPG waveform are limited.
“While, to the authors’ knowledge, publications that experimentally and explicitly demonstrate the effect of obesity or BMI on the PPG waveform are limited, we will explore works which suggest that there would be a substantial effect [56,59,60,61,62,63,64,65,66,67,68,69,70].”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [56]
Skin thickness is directly correlated to BMI and can dampen PPG signal amplitude.
“Perhaps most detrimental to the PPG waveform is skin thickness, as it is directly correlated to BMI and can dampen PPG signal amplitude [56,70].”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [56]
The citation lists haven’t been fetched yet
References 32 · Cited by 283
- Authors
- Gibney, M. A., Arce, C. H., Byron, K. J., & Hirsch, L. J.
- Affiliations
- Becton Dickinson (United States) (US)
- Corresponding author
- Hirsch, Laurence J.
- ORCID
- Hirsch, Laurence J.
- Venue
- Current Medical Research and Opinion
- Publisher
- Informa Healthcare
- Published
- 2010-05-04
- Volume
- 26
- Issue
- 6
- Pages
- 1519-1530
- ISSN
- 0300-7995 (ISSN-L)
- 1473-4877
- OpenAlex
- W2132319148
- Cited by
- 283
- References
- 32
- Citations by year
- 2026: 12 · 2025: 10 · 2024: 22 · 2023: 21 · 2022: 25 · 2021: 27 · 2020: 17 · 2019: 18 · 2018: 18 · 2017: 14 · 2016: 18 · 2015: 17 · 2014: 24 · 2013: 10 · 2012: 17
- FWCI
- 9.65
- Type
- journal article
- Language
- en
- Open access
- green
- figshare.com
- Licence
- cc-by
- Locations
- Current Medical Research and Opinion (publishedVersion)
- PubMed (publishedVersion)
- Figshare (submittedVersion, cc-by)
- Figshare (cc-by)
- Funders
- U.S. Food and Drug Administration
- NIH Clinical Center
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Emergency Medicine › Intramuscular injections and effects
- Health Sciences › Medicine › Dermatology › Skin Diseases and Diabetes
- Keywords
- skin thickness, subcutaneous fat thickness, insulin injection site, needle length, subcutaneous insulin therapy, intramuscular injection, body mass index, type 1 diabetes, type 2 diabetes, injection site pain, glycemic control, sex differences, racial differences, arms, thigh, abdomen, buttocks
- MeSH terms
- Adolescent, Adult, Aged, Aged, 80 and over, Diabetes Mellitus, Type 1, Equipment Design, Female, Humans, Hypoglycemic Agents, Injections, Intradermal, Insulin, Male, Middle Aged, Needles, Skin, Ultrasonography, United States, Body Mass Index, Subcutaneous Fat, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review