Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Open accessAbstract
Photoplethysmography (PPG) is a low-cost, noninvasive optical technique that uses change in light transmission with changes in blood volume within tissue to provide information for cardiovascular health and fitness. As remote health and wearable medical devices become more prevalent, PPG devices are being developed as part of wearable systems to monitor parameters such as heart rate (HR) that do not require complex analysis of the PPG waveform. However, complex analyses of the PPG waveform yield valuable clinical information, such as: blood pressure, respiratory information, sympathetic nervous system activity, and heart rate variability. Systems aiming to derive such complex parameters do not always account for realistic sources of noise, as testing is performed within controlled parameter spaces. A wearable monitoring tool to be used beyond fitness and heart rate must account for noise sources originating from individual patient variations (e.g., skin tone, obesity, age, and gender), physiology (e.g., respiration, venous pulsation, body site of measurement, and body temperature), and external perturbations of the device itself (e.g., motion artifact, ambient light, and applied pressure to the skin). Here, we present a comprehensive review of the literature that aims to summarize these noise sources for future PPG device development for use in health monitoring.
AI snapshot
Written from the paper’s full text.
TL;DR
This review synthesizes literature on sources of inaccuracy in photoplethysmography (PPG) for continuous cardiovascular monitoring, grouping them into individual variation, physiological processes, and external perturbations. It argues that advanced PPG parameters require accounting for these noise sources across diverse populations and real-world conditions, not just controlled fitness/heart-rate settings.
Key findings
- Noise sources are divided into three main categories: individual variation (skin tone, obesity, age, gender), physiology (respiration, venous pulsation, body site, local body temperature), and external perturbation (motion artifact, ambient light, applied pressure).
- Skin tone: a 535 nm wrist device showed error of 1.04 BPM for “fair” skin vs 10.90 BPM for “dark” skin; Polar OH1 bias increased with darker skin (p < 0.001), while Bent et al. found no statistically significant skin-tone relationship but a mean absolute error of approximately 9 BPM across all skin tones.
- Obesity: Monte Carlo simulations predicted a 40% loss of PPG signal amplitude on the wrist as simulated radial artery depth increased from 2.5 to 3.5 mm; obesity is also linked to changes in capillary density, blood flow, SpO2, trans epidermal water loss, and skin thickness.
- Motion and pressure: Basis Peak vs Fitbit Charge HR correlations were r = 0.92 vs r = 0.83 overall, and during high-intensity events r = 0.77 vs r = 0.58; Apple Watch 4 had MAE of 2.7 BPM at rest and 4.6 BPM while walking. Applied pressure from 0–15 kPa produced over 300% variation in the b/a ratio, dropping to less than 5% at optimal pressure; a force regulator reduced coefficient of variation from 50.9% to 1.8%.
Methods
- Design: comprehensive literature review; no new experimental data are reported.
- Data/sample: synthesizes published human studies, in silico modeling, commercial and FDA-cleared/approved PPG devices, and prior reviews; examples include studies with 256 subjects for skin tone, 70 subjects for Polar OH1, 56 patients for wearable comparison, and 93 individuals for age-related PPG features.
- Main analysis: narrative synthesis categorizing noise sources into individual variation, physiology, and external factors, with summary tables for age-induced PPG changes, temperature effects, and adaptive filters for motion artifacts.
Limitations
- The paper states that few noise sources had a comprehensive solution, understanding of how conditions such as obesity affect skin and cardiovascular monitoring is still developing, and future research should incorporate larger studies inclusive of diverse populations.
- Evident from the text: as a narrative review, it does not report a systematic search strategy or quantitative meta-analysis; many cited studies are small, some lack statistical analyses, and findings conflict (e.g., Bent et al. vs other skin-tone studies; inconsistent skin thickness data).
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 283 cited sources checked · 0 supported · 0 flagged
- Findingnone given
PPG-based RCIM devices that are FDA cleared or approved and can accurately and consistently record clinical parameters in a sufficiently diverse population for true health monitoring are scarce.
“PPG-based RCIM devices that are U.S. Food and Drug Administration (FDA) cleared or approved and can accurately and consistently record clinical parameters in a sufficiently diverse population for true health monitoring are scarce.”
Quote in the paperSubstantial claim - Findingnone given
Publications that experimentally and explicitly demonstrate the effect of obesity or BMI on the PPG waveform are limited.
“While, to the authors’ knowledge, publications that experimentally and explicitly demonstrate the effect of obesity or BMI on the PPG waveform are limited, we will explore works which suggest that there would be a substantial effect [56,59,60,61,62,63,64,65,66,67,68,69,70].”
Quote in the paperNot a substantial claim0 of 10 sources checked- cites [56] Gibney, 2010, Skin and subcutaneous adipose layer thickness in adults with diabetes at sites used for insulin injections: Implications for needle length recommendations, Curr. Med. Res. Opin. source
Not checked in the free run.
- cites [59] Elgendi, 2012, On the analysis of fingertip photoplethysmogram signals, Curr. Cardiol. Rev. source
Not checked in the free run.
- cites [60] Iacopi, 2020, Using Skin Bioengineering to Highlight How Weight and Diabetes Mellitus Modify the Skin in the Lower Limbs of Super-Obese Patients, Diabetes Metab. Syndr. Obes. Targets Ther. source
Not checked in the free run.
- cites [61] Chin, 1999, Increased cutaneous blood flow but impaired post-ischemic response of nutritional flow in obese children, Atherosclerosis source
Not checked in the free run.
- cites [62] Francischetti, 2011, Skin capillary density and microvascular reactivity in obese subjects with and without metabolic syndrome, Microvasc. Res. source
Not checked in the free run.
- cites [63] Czernichow, 2010, Microvascular dysfunction in healthy insulin-sensitive overweight individuals, J. Hypertens. source
Not checked in the free run.
- cites [64] Debbabi, 2006, Increased skin capillary density in treated essential hypertensive patients, Am. J. Hypertens. source
Not checked in the free run.
- cites [65] Ijzerman, 2004, Impaired microvascular function in obesity: Implications for obesity-associated microangiopathy, hypertension, and insulin resistance, Circulation source
Not checked in the free run.
- cites [66] Kapur, 2013, Obesity is associated with a lower resting oxygen saturation in the ambulatory elderly: Results from the cardiovascular health study, Respir. Care source
Not checked in the free run.
- cites [67] Petrofsky, 2015, The effect of BMI on oxygen saturation at rest and during mild walking, J. Appl. Med. Sci.
Not checked in the free run.
- Findingreasoning
While many sources of noise had documented potential solutions, few had a comprehensive solution.
“While many sources of noise had documented potential solutions, few had a comprehensive solution and even our understanding of how conditions such as obesity affect the skin and cardiovascular monitoring are still developing.”
Quote in the paperSubstantial claim - Findingreasoning
A review of the literature shows that the difficulty in creating an RCIM PPG device partially originates from the many sources of noise that researchers encounter.
“A review of the literature shows that the difficulty in creating such a device partially originates from the many sources of noise that researchers encounter.”
Quote in the paperSubstantial claim - Findingown data
Sources of noise can be divided primarily into three categories: individual variation, physiological processes, and external perturbation.
“As shown in Table 7, we found that sources of noise can be divided primarily into three categories: individual variation, physiological processes, and external perturbation.”
Quote in the paperSubstantial claim
- Premisenone given
Systems aiming to derive complex parameters from PPG do not always account for realistic sources of noise, as testing is performed within controlled parameter spaces.
“Systems aiming to derive such complex parameters do not always account for realistic sources of noise, as testing is performed within controlled parameter spaces.”
Quote in the paperSubstantial claim - Premisenone given
The difficulty in determining features in PPG devices lies in the numerous sources of noise that can impede the output of the PPG.
“The difficulty in determining features in PPG devices lies in the numerous sources of noise that can impede the output of the PPG.”
Quote in the paperSubstantial claim - Premisecitation
Variation in the quasi-DC component of PPG is due to many factors: optical properties of the tissue, average blood volume, respiration, vasomotor activity, vasoconstrictor waves, Traube Hering Meyer waves, and thermoregulation.
“The variation in the quasi-DC component is due to many factors: the optical properties of the tissue, average blood volume, respiration, vasomotor activity, vasoconstrictor waves, Traube Hering Meyer waves, and thermoregulation [4,5,6,7,8,9,10,11,12,13,14].”
Quote in the paperNot a substantial claim0 of 10 sources checked- cites [4] Burton, 1940, A study of the adjustment of peripheral vascular tone to the requirements of the regulation of body temperature, Am. J. Physiol. Leg. Content source
Not checked in the free run.
- cites [5] Burton, 1939, The range and variability of the blood flow in the human fingers and the vasomotor regulation of body temperature, Am. J. Physiol. Leg. Content source
Not checked in the free run.
- cites [6] Hertzman, 1942, The vasomotor components in the vascular reactions in the finger to cold, Am. J. Physiol. Leg. Content source
Not checked in the free run.
- cites [7] Hertzman, 1942, The reactions of the digital artery and minute pad arteries to local cold, Am. J. Physiol. Leg. Content source
Not checked in the free run.
- cites [8] Hertzman, A.B., and Flath, F. (1963). Continuous Simultaneous Registration of Sweating and Blood Flow in a Small Skin Area, Saint Louis University. source
Not checked in the free run.
- cites [9] Hertzman, 1940, Applications of photoelectric plethysmography in peripheral vascular disease, Am. Heart J. source
Not checked in the free run.
- cites [10] Hertzman, 1942, The absence of vasoconstrictor reflexes in the forehead circulation. Effects of cold, Am. J. Physiol. Leg. Content source
Not checked in the free run.
- cites [11] Hyndman, 1971, Spontaneous rhythms in physiological control systems, Nature source
Not checked in the free run.
- cites [12] Fiser, 1978, Spectral analysis of resting variability of some circulatory parameters in man, Physiol. Bohemoslov.
Not checked in the free run.
- cites [13] Harness, 1989, Low-frequency photoplethysmograph signals, Clin. Phys. Physiol. Meas. source
Not checked in the free run.
- Premisecitation
There are many reports of inaccuracy in these devices.
“Additionally, there are many reports of inaccuracy in these devices [33].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [33] Ramella-Roman, J.C. (2020, January 1–6). Design of wearable devices for diverse populations (Conference Presentation). Proceedings of the Optics and Biophotonics in Low-Resource Settings VI, San Francisco, CA, USA. source
Not checked in the free run.
- Interpretationreasoning
These many sources of noise create limitations in the application of PPG to derive advanced physiological parameters.
“These many sources of noise create limitations in the application of PPG to derive advanced physiological parameters.”
Quote in the paperSubstantial claim - Premisecitation
An individual's skin tone, and thus Fitzpatrick category, is correlated to the amount of eumelanin in their epidermis.
“An individual’s skin tone, and thus Fitzpatrick category, is correlated to the amount of eumelanin in their epidermis [38].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [38] Thody, 1991, Pheomelanin as well as eumelanin is present in human epidermis, J. Investig. Dermatol. source
Not checked in the free run.
- Premisecitation
The majority of findings find that dermal capillary density does negatively correlate on average with increasing BMI.
“In adults, while some studies show that at rest there is no significant change in dermal capillary density, the majority of findings find that dermal capillary density does negatively correlate on average with increasing BMI [62,63,64,65].”
Quote in the paperSubstantial claim0 of 4 sources checked- cites [62] Francischetti, 2011, Skin capillary density and microvascular reactivity in obese subjects with and without metabolic syndrome, Microvasc. Res. source
Not checked in the free run.
- cites [63] Czernichow, 2010, Microvascular dysfunction in healthy insulin-sensitive overweight individuals, J. Hypertens. source
Not checked in the free run.
- cites [64] Debbabi, 2006, Increased skin capillary density in treated essential hypertensive patients, Am. J. Hypertens. source
Not checked in the free run.
- cites [65] Ijzerman, 2004, Impaired microvascular function in obesity: Implications for obesity-associated microangiopathy, hypertension, and insulin resistance, Circulation source
Not checked in the free run.
- Premisereasoning
For those with a lighter skin tone, green light enables a higher signal-to-noise ratio for determining heart rate.
“For those with a lighter skin tone, this enables a higher signal-to-noise ratio for determining heart rate: the primary parameter derived by PPG.”
Quote in the paperNot a substantial claim - Premisecitation
Bent et al. found no statistically significant relationship between measured heart rate from wearable devices and an ECG reference across skin tones.
“By collecting data from 56 patients (34 female, 22 males, 18–54 years old, and at least 8 participants in each Fitzpatrick classification) when they are at rest and exercising (elevating heart rate to 50% of maximum via a treadmill), it was found that there was no statistically significant relationship between measured heart rate from the wearable and an ECG (Bittium Faros 180) reference.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [46] Bent, 2020, Investigating sources of inaccuracy in wearable optical heart rate sensors, NPJ Digit. Med. source
Not checked in the free run.
- Premisecitation
Fallow et al. found that mean modulation (perfusion index, AC/DC) was significantly lower in Fitzpatrick V individuals than others for all wavelengths (p < 0.001).
“Mean modulation, often termed the perfusion index, defined as the ratio of AC/DC, was significantly lower in Fitzpatrick V individuals than others for all wavelengths (p < 0.001).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [52] Fallow, 2013, Influence of skin type and wavelength on light wave reflectance, J. Clin. Monit. Comput. source
Not checked in the free run.
- Premisecitation
Lower absorption of melanin using light sources at higher wavelengths can improve the signal for individuals with a high Fitzpatrick classification.
“The lower absorption of melanin using light sources at higher wavelengths can improve the signal for individuals with a high Fitzpatrick classification.”
Quote in the paperSubstantial claim - Premisecitation
Numerous other studies that stratify results against Fitzpatrick classification identify the same trend: errors in determining heart rate from wearable PPG devices that use primarily green light are increased in individuals with dark skin tones due to high absorption caused by increased amounts of epidermal melanin.
“In addition to these examples, numerous other studies that stratify results against Fitzpatrick classification identify the same trend; errors in determining heartrate from wearable PPG devices that use primarily green light as their source is increased in individuals with dark skin tones due to the high absorption caused by increased amounts of epidermal melanin [42,43,44,45].”
Quote in the paperSubstantial claim0 of 4 sources checked- cites [42] Wang, 2016, Algorithmic principles of remote PPG, IEEE Trans. Biomed. Eng. source
Not checked in the free run.
- cites [43] Jeanne, 2013, Robust pulse rate from chrominance-based rPPG, IEEE Trans. Biomed. Eng. source
Not checked in the free run.
- cites [44] Shao, 2016, Simultaneous monitoring of ballistocardiogram and photoplethysmogram using a camera, IEEE Trans. Biomed. Eng. source
Not checked in the free run.
- cites [45] Addison, 2018, Video-based heart rate monitoring across a range of skin pigmentations during an acute hypoxic challenge, J. Clin. Monit. Comput. source
Not checked in the free run.
- Interpretationreasoning
The variation of BMI across individuals is a potential source of variation for PPG measurements.
“The variation of BMI across individuals is thus a potential source of variation for PPG measurements.”
Quote in the paperSubstantial claim - Premisecitation
The depth of the target vessel and intervening pulsating arterioles may vary across individuals and within individuals with body location and due to factors such as level of obesity.
“Furthermore, one component of optimizing source/detector separation is the depth of the target vessel and intervening pulsating arterioles, which may vary across individuals and within individuals with body location and due to factors such as level of obesity [56].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [56] Gibney, 2010, Skin and subcutaneous adipose layer thickness in adults with diabetes at sites used for insulin injections: Implications for needle length recommendations, Curr. Med. Res. Opin. source
Not checked in the free run.
- Interpretationreasoning
The effect of obesity on dermal capillary density could lead to a decrease in the dominant 'DC' component of a PPG waveform due to the increased blood volume of the obese.
“This effect of obesity could lead to a decrease in the dominant “DC” component of a PPG waveform due to the increased blood volume of the obese [61,62,63,64].”
Quote in the paperSubstantial claim0 of 4 sources checked- cites [61] Chin, 1999, Increased cutaneous blood flow but impaired post-ischemic response of nutritional flow in obese children, Atherosclerosis source
Not checked in the free run.
- cites [62] Francischetti, 2011, Skin capillary density and microvascular reactivity in obese subjects with and without metabolic syndrome, Microvasc. Res. source
Not checked in the free run.
- cites [63] Czernichow, 2010, Microvascular dysfunction in healthy insulin-sensitive overweight individuals, J. Hypertens. source
Not checked in the free run.
- cites [64] Debbabi, 2006, Increased skin capillary density in treated essential hypertensive patients, Am. J. Hypertens. source
Not checked in the free run.
- Premisecitation
There are many components to a PPG signal and its derivatives that literature records having variation as a function of age.
“There are many components to a PPG signal and its derivatives that literature records having variation as a function of age [83].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [83] Ahn, 2017, New aging index using signal features of both photoplethysmograms and acceleration plethysmograms, Healthc. Inform. Res. source
Not checked in the free run.
- Premisecitation
Yousef et al. found that there is no significant increase in pulse magnitude measured at the index finger with an increase in age.
“In a study with participants ranging in age from 30 to 60+, Yousef et al. found that there is no significant increase in pulse magnitude measured at the index finger with an increase in age.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [88] Yousef, 2012, The analysis of PPG morphology: Investigating the effects of aging on arterial compliance, Meas. Sci. Rev. source
Not checked in the free run.
- Interpretationreasoning
Aging leads to various anatomical and physiological changes that impact the ability to use PPG to assess cardiovascular health.
“Aging leads to various anatomical and physiological changes that impact the ability to use PPG to assess cardiovascular health.”
Quote in the paperSubstantial claim - Premisecitation
Overall, the morbidly obese exhibit a reduction in water, measured as an increase in the TEWL compared to normal BMI subjects.
“Overall, the morbidly obese exhibit a reduction in water, measured as an increase in the TEWL compared to normal BMI subjects [69].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [69] Rodrigues, 2017, Excessive weight favours skin physiology-up to a point: Another expression of the obesity paradox, Skin Pharmacol. Physiol. source
Not checked in the free run.
- Premisecitation
Caceres and Hasan found that TTW decreases with age.
“It was found that TTW decreases with age [89].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [89] Hasan, 2015, A New First-Derivative Related Index to Assess Pulse Wave Transit Time from a Photoplethysmographic Waveform: Age Dependence and Agreement with Normative Data, Int. J. Bioinform. Biomed. Eng.
Not checked in the free run.
- Premisecitation
Allen and Murray found that an increase in pulse wave velocity with age is related to a diminished dicrotic notch.
“Corroborating with other literature, an increase in pulse wave velocity with age is related to a diminished dicrotic notch.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [90] Allen, 2003, Age-related changes in the characteristics of the photoplethysmographic pulse shape at various body sites, Physiol. Meas. source
Not checked in the free run.
- Premisecitation
In a population of 134 healthy, Caucasian subjects (median age 43, total range: 13-72), there was a statistically significant decrease in PTT at the toes (r2 = 0.48), fingers (r2 = 0.26), and ears (r2 = 0.15).
“By analyzing the change in PTT as a function of age within a population of 134 healthy, Caucasian subjects (median age 43, total range: 13–72), it was found that there is a statistically significant decrease at the toes (r2 = 0.48), fingers (r2 = 0.26), and ears (r2 = 0.15) [91,92].”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [91] Allen, 2002, Age-related changes in peripheral pulse timing characteristics at the ears, fingers and toes, J. Hum. Hypertens. source
Not checked in the free run.
- cites [92] 2011, Non-invasive estimate of blood glucose and blood pressure from a photoplethysmograph by means of machine learning techniques, Artif. Intell. Med. source
Not checked in the free run.
- Premisecitation
The parameters attributed to Ahn et al. yielded a statistically significant, albeit poor, correlation with age.
“The parameters listed in Table 3 attributed to Ahn et al. yielded a statistically significant, albeit poor, correlation with age [83].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [83] Ahn, 2017, New aging index using signal features of both photoplethysmograms and acceleration plethysmograms, Healthc. Inform. Res. source
Not checked in the free run.
- Premisecitation
As arteries age, the tunica intima and tunica media layers within the arteries thicken with an increase in the number and density of collagen.
“As arteries age, the tunica intima and tunica media layers within the arteries thicken with an increase in the number and density of collagen [78].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [78] Knight, 2017, Anatomy and physiology of ageing 1: The cardiovascular system, Nurs. Times
Not checked in the free run.
- Premisecitation
As age increases, PTT decreases.
“PTT is inversely related to PWV—the same authors note in a different work that as age increases, PTT decreases.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [91] Allen, 2002, Age-related changes in peripheral pulse timing characteristics at the ears, fingers and toes, J. Hum. Hypertens. source
Not checked in the free run.
- cites [92] 2011, Non-invasive estimate of blood glucose and blood pressure from a photoplethysmograph by means of machine learning techniques, Artif. Intell. Med. source
Not checked in the free run.
- Interpretationreasoning
The aforementioned literature provides strong evidence that obesity, when assessed in combination with a subject's metabolic state, body location, gender and age, will likely influence PPG systems and can manifest as significant noise sources.
“We believe that the aforementioned literature provides strong evidence that obesity, when assessed in combination with a subject’s metabolic state, body location, gender and age will likely influence PPG systems and can manifest themselves as significant noise sources.”
Quote in the paperNot a substantial claim - Premisecitation
Ferdinando et al. were able to identify five classes of obesity from PPG waveforms using 17 parameters derived from the decomposition of the PPG waveform into five lognormal functions.
“Using k-nearest neighbor and support vector machines, Ferdinando et al. were able to identify five classes of obesity from PPG waveforms using 17 parameters derived from the decomposition of the PPG waveform into five lognormal functions.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [75] Liang, 2018, A new, short-recorded photoplethysmogram dataset for blood pressure monitoring in China, Sci. Data source
Not checked in the free run.
- cites [76] Ferdinando, H., Huotari, M., and Myllylä, T. (2019, January 23–27). Photoplethysmography signal analysis to assess obesity, age group and hypertension. Proceedings of the 2019 41st Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Berlin, Germany. source
Not checked in the free run.
- Premisecitation
Oxygen saturation of hemoglobin has consistently been shown to be inversely associated with BMI across various populations.
“Lastly, oxygen saturation of hemoglobin has consistently been shown to be inversely associated with BMI across various populations [66,67,73].”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [66] Kapur, 2013, Obesity is associated with a lower resting oxygen saturation in the ambulatory elderly: Results from the cardiovascular health study, Respir. Care source
Not checked in the free run.
- cites [67] Petrofsky, 2015, The effect of BMI on oxygen saturation at rest and during mild walking, J. Appl. Med. Sci.
Not checked in the free run.
- cites [73] Littleton, 2017, The effects of obesity on lung volumes and oxygenation, Respir. Med. source
Not checked in the free run.
- Premisecitation
Boonya-ananta et al. used Monte Carlo simulations that predicted a 40% loss of PPG signal amplitude due to increased skin thickness in obese individuals, specifically on the wrist as the simulated radial artery increases in depth from 2.5 to 3.5 mm.
“For example, the increase in skin thickness affects the vessel depth, and Boonya-ananta et al. used Monte Carlo simulations that predicted a 40% loss of PPG signal amplitude due to this effect in obese individuals, specifically on the wrist as the simulated radial artery increases in depth from 2.5 to 3.5 mm [74].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [74] Hirt, 2019, Skin changes in the obese patient, J. Am. Acad. Dermatol. source
Not checked in the free run.
- Interpretationreasoning
These age-related changes can manifest themselves in variations observed across PPG waveforms either in the shape or amplitude of the waveform.
“Overall, these changes can manifest themselves in variations observed across PPG waveforms either in the shape or amplitude of the waveform.”
Quote in the paperSubstantial claim - Premisecitation
One observation regarding the skin of the obese which may serve to increase PPG signal intensity is the increase in trans epidermal water loss (TEWL) as BMI increases.
“One observation regarding the skin of the obese which may serve to increase PPG signal intensity is the increase in trans epidermal water loss (TEWL) as BMI increases [69].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [69] Rodrigues, 2017, Excessive weight favours skin physiology-up to a point: Another expression of the obesity paradox, Skin Pharmacol. Physiol. source
Not checked in the free run.
- Premisecitation
The cross-linking of collagen fibers, along with fractured and fatigue elastin, leads to a loss of compliance and increased artery stiffness.
“The cross-linking of these fibers, along with fractured and fatigue elastin, leads to a loss of compliance and increased artery stiffness [78].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [78] Knight, 2017, Anatomy and physiology of ageing 1: The cardiovascular system, Nurs. Times
Not checked in the free run.
- Premisecitation
Once adolescence is reached, skin is known to thin as age increases.
“Once adolescence is reached, skin is known to thin as age increases [70,80,81,82].”
Quote in the paperNot a substantial claim0 of 4 sources checked- cites [70] Derraik, J.G., Rademaker, M., Cutfield, W.S., Pinto, T.E., Tregurtha, S., Faherty, A., Peart, J.M., Drury, P.L., and Hofman, P.L. (2014). Effects of age, gender, BMI, and anatomical site on skin thickness in children and adults with diabetes. PLoS ONE, 9. source
Not checked in the free run.
- cites [80] Theeten, 2017, High frequency ultrasound to assess skin thickness in healthy adults, Vaccine source
Not checked in the free run.
- cites [81] Laurent, 2007, Echographic measurement of skin thickness in adults by high frequency ultrasound to assess the appropriate microneedle length for intradermal delivery of vaccines, Vaccine source
Not checked in the free run.
- cites [82] Firooz, 2017, The influence of gender and age on the thickness and echo-density of skin, Skin Res. Technol. source
Not checked in the free run.
- Interpretationnone given
The literature has shown that obesity dramatically affects physiological factors associated with PPG signal intensity and quality, including capillary density, capillary recruitment, blood flow, SpO2, TEWL, and skin thickness.
“Overall, the literature has shown that obesity dramatically affects physiological factors associated with PPG signal intensity and quality, including capillary density, capillary recruitment, blood flow, SpO2, TEWL, and skin thickness.”
Quote in the paperNot a substantial claim - Premisecitation
e/a decreases as age increases.
“| e/a | Decrease | Takazawa et al., 1998 Elgendi et al., 2012 | [29,59] |”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [29] Takazawa, 1998, Assessment of vasoactive agents and vascular aging by the second derivative of photoplethysmogram waveform, Hypertension source
Not checked in the free run.
- cites [59] Elgendi, 2012, On the analysis of fingertip photoplethysmogram signals, Curr. Cardiol. Rev. source
Not checked in the free run.
- Premisecitation
Dicrotic notch shape decreases as age increases.
“| Dicrotic notch shape | Decrease | Allen et al., 2003 | [90] |”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [90] Allen, 2003, Age-related changes in the characteristics of the photoplethysmographic pulse shape at various body sites, Physiol. Meas. source
Not checked in the free run.
- Premisecitation
PTT decreases as age increases.
“| PTT | Decrease | Allen et al., 2002 Monte-Moreno, 2011 | [91,92] |”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [91] Allen, 2002, Age-related changes in peripheral pulse timing characteristics at the ears, fingers and toes, J. Hum. Hypertens. source
Not checked in the free run.
- cites [92] 2011, Non-invasive estimate of blood glucose and blood pressure from a photoplethysmograph by means of machine learning techniques, Artif. Intell. Med. source
Not checked in the free run.
- Premisecitation
Inflection point area decreases as age increases.
“| Inflection point area | Decrease | Ahn et al., 2017 | [83] |”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [83] Ahn, 2017, New aging index using signal features of both photoplethysmograms and acceleration plethysmograms, Healthc. Inform. Res. source
Not checked in the free run.
- Premisecitation
Magnitude of d decreases as age increases.
“| Magnitude of d | Decrease | Ahn et al., 2017 | [83] |”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [83] Ahn, 2017, New aging index using signal features of both photoplethysmograms and acceleration plethysmograms, Healthc. Inform. Res. source
Not checked in the free run.
- Premisecitation
TTW decreases as age increases.
“| TTW | Decrease | Cáceres et al., 2015 | [89] |”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [89] Hasan, 2015, A New First-Derivative Related Index to Assess Pulse Wave Transit Time from a Photoplethysmographic Waveform: Age Dependence and Agreement with Normative Data, Int. J. Bioinform. Biomed. Eng.
Not checked in the free run.
- Premisecitation
Reflection index increases as age increases.
“| Reflection index | Increase | Ahn et al., 2017 | [83] |”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [83] Ahn, 2017, New aging index using signal features of both photoplethysmograms and acceleration plethysmograms, Healthc. Inform. Res. source
Not checked in the free run.
- Premisecitation
PWV increases as age increases.
“| PWV | Increase | Cáceres et al., 2015 | [89] |”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [89] Hasan, 2015, A New First-Derivative Related Index to Assess Pulse Wave Transit Time from a Photoplethysmographic Waveform: Age Dependence and Agreement with Normative Data, Int. J. Bioinform. Biomed. Eng.
Not checked in the free run.
- Premisecitation
Artery compliance decreases as age increases.
“| Artery compliance | Decrease | Knight et al., 2017 Allen et al., 2002 | [78,90] |”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [78] Knight, 2017, Anatomy and physiology of ageing 1: The cardiovascular system, Nurs. Times
Not checked in the free run.
- cites [90] Allen, 2003, Age-related changes in the characteristics of the photoplethysmographic pulse shape at various body sites, Physiol. Meas. source
Not checked in the free run.
- Interpretationcitation
This work suggests that the decrease in vessel compliance will contribute more to variations in a PPG waveform than other effects of aging.
“This work suggests that the decrease in vessel compliance will contribute more to variations in a PPG waveform than other effects of aging [88].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [88] Yousef, 2012, The analysis of PPG morphology: Investigating the effects of aging on arterial compliance, Meas. Sci. Rev. source
Not checked in the free run.
- Premisecitation
Reckelhoff reported the mean blood pressure was 6-10 mmHg higher in males than in pre-menopausal females.
“Regarding blood pressure, Reckelhoff reported the mean blood pressure was 6–10 mmHg higher in males than in pre-menopausal females.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [101] Reckelhoff, 2001, Gender differences in the regulation of blood pressure, Hypertension source
Not checked in the free run.
- Premisecitation
Diastolic peak amplitude decreases as age increases.
“| Diastolic peak amplitude | Decrease | Ahn et al., 2017 | [83] |”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [83] Ahn, 2017, New aging index using signal features of both photoplethysmograms and acceleration plethysmograms, Healthc. Inform. Res. source
Not checked in the free run.
- Premisecitation
Monte-Moreno utilized machine learning to determine blood pressure from a PPG obtained from subjects ranging from 9-80 years old.
“In 2010, Monte-Moreno utilized machine learning to determine blood pressure from a PPG obtained from subjects ranging from 9–80 years old.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [92] 2011, Non-invasive estimate of blood glucose and blood pressure from a photoplethysmograph by means of machine learning techniques, Artif. Intell. Med. source
Not checked in the free run.
- Interpretationreasoning
These physiological baseline differences can skew PPG signals along with interpretation algorithms for the determination of cardiovascular health per gender.
“These physiological baseline differences can skew PPG signals along with interpretation algorithms for the determination of cardiovascular health per gender.”
Quote in the paperSubstantial claim - Premisecitation
Slope of line between b and d increases as age increases.
“| Slope of line between b and d | Increase | Ahn et al., 2017 | [83] |”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [83] Ahn, 2017, New aging index using signal features of both photoplethysmograms and acceleration plethysmograms, Healthc. Inform. Res. source
Not checked in the free run.
- Premisecitation
c/a decreases as age increases.
“| c/a | Decrease | Takazawa et al., 1998 Elgendi et al., 2012 | [29,59] |”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [29] Takazawa, 1998, Assessment of vasoactive agents and vascular aging by the second derivative of photoplethysmogram waveform, Hypertension source
Not checked in the free run.
- cites [59] Elgendi, 2012, On the analysis of fingertip photoplethysmogram signals, Curr. Cardiol. Rev. source
Not checked in the free run.
- Premisecitation
Systolic rising edge slope decreases as age increases.
“| Systolic rising edge slope | Decrease | Allen et al., 2003 | [90] |”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [90] Allen, 2003, Age-related changes in the characteristics of the photoplethysmographic pulse shape at various body sites, Physiol. Meas. source
Not checked in the free run.
- Premisecitation
Skin thickness decreases as age increases.
“| Skin thickness | Decrease | Derraik et al., 2014 Van Mulder et al., 2017 Farage et al., 2013 | [70,80,93] |”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [70] Derraik, J.G., Rademaker, M., Cutfield, W.S., Pinto, T.E., Tregurtha, S., Faherty, A., Peart, J.M., Drury, P.L., and Hofman, P.L. (2014). Effects of age, gender, BMI, and anatomical site on skin thickness in children and adults with diabetes. PLoS ONE, 9. source
Not checked in the free run.
- cites [80] Theeten, 2017, High frequency ultrasound to assess skin thickness in healthy adults, Vaccine source
Not checked in the free run.
- cites [93] Farage, 2013, Characteristics of the aging skin, Adv. Wound Care source
Not checked in the free run.
- Premisecitation
Systolic time decreases as age increases.
“| Systolic time | Decrease | Ahn et al., 2017 | [83] |”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [83] Ahn, 2017, New aging index using signal features of both photoplethysmograms and acceleration plethysmograms, Healthc. Inform. Res. source
Not checked in the free run.
- Premisecitation
d/a decreases as age increases.
“| d/a | Decrease | Takazawa et al., 1998 Elgendi et al., 2012 | [29,59] |”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [29] Takazawa, 1998, Assessment of vasoactive agents and vascular aging by the second derivative of photoplethysmogram waveform, Hypertension source
Not checked in the free run.
- cites [59] Elgendi, 2012, On the analysis of fingertip photoplethysmogram signals, Curr. Cardiol. Rev. source
Not checked in the free run.
- Premisecitation
Capillary recruitment decreases as age increases.
“| Capillary Recruitment | Decrease | Leveque et al., 1984 | [94]. |”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [94] Leveque, 1984, In vivo studies of the evolution of physical properties of the human skin with age, Int. J. Dermatol. source
Not checked in the free run.
- Premisecitation
Monte-Moreno's work correlated with both systolic and diastolic blood pressure with an r2 of approximately 0.90.
“While this work utilized a population of individuals with good cardiovascular health, it was found to correlate with both systolic and diastolic blood pressure, with an r2 of approximately 0.90 [92].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [92] 2011, Non-invasive estimate of blood glucose and blood pressure from a photoplethysmograph by means of machine learning techniques, Artif. Intell. Med. source
Not checked in the free run.
- Premisecitation
Magnitude of c increases as age increases.
“| Magnitude of c | Increase | Ahn et al., 2017 | [83] |”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [83] Ahn, 2017, New aging index using signal features of both photoplethysmograms and acceleration plethysmograms, Healthc. Inform. Res. source
Not checked in the free run.
- Premisecitation
b/a increases as age increases.
“| b/a | Increase | Ahn et al., 2017 | [83] |”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [83] Ahn, 2017, New aging index using signal features of both photoplethysmograms and acceleration plethysmograms, Healthc. Inform. Res. source
Not checked in the free run.
- Premisecitation
Diastole variability in the PPG signal is related to peripheral venous pulsation.
“Shelley et al. demonstrates diastole variability in the PPG signal is related to peripheral venous pulsation [116].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [116] Shelley, 1993, The detection of peripheral venous pulsation using the pulse oximeter as a plethysmograph, J. Clin. Monit. source
Not checked in the free run.
- Premisecitation
PPG optical signal modulation by respiration is most commonly manifested as baseline and amplitude modulation.
“PPG optical signal modulation by respiration is most commonly manifested as baseline and amplitude modulation [117].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [117] Charlton, 2016, An assessment of algorithms to estimate respiratory rate from the electrocardiogram and photoplethysmogram, Physiol. Meas. source
Not checked in the free run.
- Premisecitation
Nowara et al. report insignificant differences in blood volume pulse SNR in iPPG between males and females.
“Nowara et al. report insignificant differences in blood volume pulse SNR in iPPG (imaging PPG) between males and females [112].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [112] Nowara, E.M., McDuff, D., and Veeraraghavan, A. (2020, January 14–19). A Meta-Analysis of the Impact of Skin Tone and Gender on Non-Contact Photoplethysmography Measurements. Proceedings of the IEEE/CVF Conference on Computer Vision and Pattern Recognition Workshops, Seattle, WA, USA. source
Not checked in the free run.
- Premisecitation
The difference in compliance between the arterial and venous systems, as well as the lower-pressure gradient, translates to the relative amplitude of the AC venous component of the PPG being smaller than the AC arterial component.
“The difference in compliance between the arterial and venous systems, as well as the lower-pressure gradient, translates to the relative amplitude of the AC venous component of the PPG being smaller than the AC arterial component [125].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [125] Shelley, 2005, The effect of venous pulsation on the forehead pulse oximeter wave form as a possible source of error in Spo2 calculation, Anesth. Analg. source
Not checked in the free run.
- Premisecitation
Respiration and cardiac output are inherently linked, as an increase in respiration rate can directly affect the variation of heart rate through the nervous system reduced inhibitory control.
“Physiologically, respiration and cardiac output are inherently linked, as an increase in respiration rate can directly affect the variation of heart rate through the nervous system reduced inhibitory control [118].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [118] Meredith, 2012, Photoplethysmographic derivation of respiratory rate: A review of relevant physiology, J. Med. Eng. Technol. source
Not checked in the free run.
- Premisecitation
Respiration rate extraction is more precise when performed on ECG data as opposed to PPG.
“However, the results indicate that respiration rate extraction is more precise when performed on ECG data as opposed to PPG.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [117] Charlton, 2016, An assessment of algorithms to estimate respiratory rate from the electrocardiogram and photoplethysmogram, Physiol. Meas. source
Not checked in the free run.
- Premisecitation
Algorithms operating on the time domain provided more accurate results than frequency domain extraction.
“Interestingly, algorithms operating on the time domain provided more accurate results than frequency domain extraction [117].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [117] Charlton, 2016, An assessment of algorithms to estimate respiratory rate from the electrocardiogram and photoplethysmogram, Physiol. Meas. source
Not checked in the free run.
- Premisecitation
Modulation of the baseline due to respiration presents itself as a superposition of the PPG signal corresponding to the cardiac cycle and a lower-frequency sinusoidal waveform.
“Modulation of the baseline due to respiration presents itself as a superposition of the PPG signal corresponding to the cardiac cycle and a lower-frequency sinusoidal waveform [117].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [117] Charlton, 2016, An assessment of algorithms to estimate respiratory rate from the electrocardiogram and photoplethysmogram, Physiol. Meas. source
Not checked in the free run.
- Premisecitation
RIFV induces a phase shift between each cycle by elongating or squeezing each wave.
“RIFV induces a phase shift between each cycle by elongating or squeezing each wave.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [115] Dehkordi, 2018, Extracting instantaneous respiratory rate from multiple photoplethysmogram respiratory-induced variations, Front. Physiol. source
Not checked in the free run.
- cites [121] Addison, 2012, Developing an algorithm for pulse oximetry derived respiratory rate (RR oxi): A healthy volunteer study, J. Clin. Monit. Comput. source
Not checked in the free run.
- Premisecitation
Respiration signal amplitude has the strongest influence on respiratory-induced variation of the PPG signal.
“From experimentation results, it is concluded that respiration signal amplitude has the strongest influence on respiratory-induced variation of the PPG signal.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [122] Li, 2010, Comparison of respiratory-induced variations in photoplethysmographic signals, Physiol. Meas. source
Not checked in the free run.
- Premisecitation
Dehkordi et al. and Addison et al. identified three significant variations to the PPG signal by respiratory rate: respiratory-induced intensity variation (RIIV), respiratory-induced amplitude variation (RIAV), and respiratory induced frequency variation (RIFV).
“Dehkordi et al. and Addison et al. identified the three significant variations to the PPG signal by respiratory rate after analyzing PPG from 139 healthy adults (67–18 years old) and the Capnobase dataset [115,121]. These variations are respiratory-induced intensity variation (RIIV), respiratory-induced amplitude variation (RIAV), and respiratory induced frequency variation (RIFV), as shown in FIG”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [115] Dehkordi, 2018, Extracting instantaneous respiratory rate from multiple photoplethysmogram respiratory-induced variations, Front. Physiol. source
Not checked in the free run.
- cites [121] Addison, 2012, Developing an algorithm for pulse oximetry derived respiratory rate (RR oxi): A healthy volunteer study, J. Clin. Monit. Comput. source
Not checked in the free run.
- Interpretationreasoning
The presence of the venous pulsation associated peak in the diastolic phase of the PPG waveform can present errors with identification and quantification of the dicrotic notch and diastolic features.
“The presence of the venous pulsation associated peak in the diastolic phase of the PPG waveform can present errors with identification and quantification of the dicrotic notch and diastolic features.”
Quote in the paperSubstantial claim - Premisecitation
Shifts in blood volume in the venous system due to changes in respiratory behavior will cause a corresponding change to the baseline amplitude of the PPG signal as blood volume increases and decreases.
“Shifts in blood volume in the venous system due to changes in respiratory behavior will cause a corresponding change to the baseline amplitude of the PPG signal as blood volume increases and decreases [120].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [120] Ysehak Abay, T., Shafqat, K., and Kyriacou, P.A. (2019). Perfusion Changes at the Forehead Measured by Photoplethysmography during a Head-Down Tilt Protocol. Biosensors, 9. source
Not checked in the free run.
- Premisecitation
Changes to diastolic amplitude due to the presence of the venous waveform can increase diastolic phase amplitude up to 40% of the total alternating amplitude of the PPG signal.
“It appears that changes to diastolic amplitude due to the presence of the venous waveform can increase diastolic phase amplitude up to 40% of the total alternating amplitude of the PPG signal.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [116] Shelley, 1993, The detection of peripheral venous pulsation using the pulse oximeter as a plethysmograph, J. Clin. Monit. source
Not checked in the free run.
- Premisecitation
When comparing the accuracy of commercial devices, a significantly higher device measurement error is seen in males than females for all devices.
“When comparing the accuracy of commercial devices as done by Shcherbina et al., a significantly higher device measurement error is seen in males than females for all devices [109].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [109] Shcherbina, A., Mattsson, C.M., Waggott, D., Salisbury, H., Christle, J.W., Hastie, T., Wheeler, M.T., and Ashley, E.A. (2017). Accuracy in wrist-worn, sensor-based measurements of heart rate and energy expenditure in a diverse cohort. J. Pers. Med., 7. source
Not checked in the free run.
- Premisecitation
RIIV modulates the DC baseline of the PPG curve superimposing the wave on top of a low frequency sinusoid.
“RIIV modulates the DC baseline of the PPG curve superimposing the wave on top of a low frequency sinusoid.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [115] Dehkordi, 2018, Extracting instantaneous respiratory rate from multiple photoplethysmogram respiratory-induced variations, Front. Physiol. source
Not checked in the free run.
- cites [121] Addison, 2012, Developing an algorithm for pulse oximetry derived respiratory rate (RR oxi): A healthy volunteer study, J. Clin. Monit. Comput. source
Not checked in the free run.
- Premisereasoning
Pulse oximetry values were both lower in cases where CVP contributed to the signal.
“The pulse oximetry values were both lower in cases where CVP contributed to the signal.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [125] Shelley, 2005, The effect of venous pulsation on the forehead pulse oximeter wave form as a possible source of error in Spo2 calculation, Anesth. Analg. source
Not checked in the free run.
- cites [127] Sami, 1991, Central venous pulsations associated with a falsely low oxygen saturation measured by pulse oximetry, J. Clin. Monit. source
Not checked in the free run.
- Premisecitation
The lower-frequency wave manifestation in the total PPG signal can be attributed to the venous vascular system.
“It is reported that the lower-frequency wave manifestation in the total PPG signal can be attributed to the venous vascular system [119].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [119] Alian, 2014, Photoplethysmography, Best Pract. Res. Clin. Anaesthesiol. source
Not checked in the free run.
- Premisecitation
Venous pressure can add low-frequency noise to the PPG waveform, often between 0 and 0.5 Hz, contributing to baseline oscillations.
“Venous pressure can also add low-frequency noise to the PPG waveform, the same effects seen by respiration, which is caused by venous anatomy [129,130,131]. Often between 0 and 0.5 Hz, this noise contributes to be baseline oscillations.”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [129] Walton, 2010, Measuring venous oxygenation using the photoplethysmograph waveform, J. Clin. Monit. Comput. source
Not checked in the free run.
- cites [130] Nilsson, 2007, Combined photoplethysmographic monitoring of respiration rate and pulse: A comparison between different measurement sites in spontaneously breathing subjects, Acta Anaesthesiol. Scand. source
Not checked in the free run.
- cites [131] Nilsson, 2003, Macrocirculation is not the sole determinant of respiratory induced variations in the reflection mode photoplethysmographic signal, Physiol. Meas. source
Not checked in the free run.
- Premisecitation
Respiration is one of the most significant sources of error in heart rate measurements using PPG.
“This is one of the most significant sources of error in heart rate measurements using PPG, even though respiratory rate is the most sensitive vital sign, often used as an indicator of clinical deterioration [117].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [117] Charlton, 2016, An assessment of algorithms to estimate respiratory rate from the electrocardiogram and photoplethysmogram, Physiol. Meas. source
Not checked in the free run.
- Premisecitation
As respiratory rate increases, the respiratory signal decreases, indicating that higher rates of respiration leads to lower fluctuations in respiratory influence changes to PPG.
“Moreover, as the respiratory rate increases, the respiratory signal decreases, indicating that higher rates of respiration leads to lower fluctuations in respiratory influence changes to PPG [122].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [122] Li, 2010, Comparison of respiratory-induced variations in photoplethysmographic signals, Physiol. Meas. source
Not checked in the free run.
- Premisecitation
RIAV causes significant changes in each peak amplitude.
“RIAV causes significant changes in each peak amplitude.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [115] Dehkordi, 2018, Extracting instantaneous respiratory rate from multiple photoplethysmogram respiratory-induced variations, Front. Physiol. source
Not checked in the free run.
- cites [121] Addison, 2012, Developing an algorithm for pulse oximetry derived respiratory rate (RR oxi): A healthy volunteer study, J. Clin. Monit. Comput. source
Not checked in the free run.
- Premisecitation
The skin of the foot of obese patients is about 7 degrees Celsius higher than the foot of nonobese patients.
“For example, Iacopi et al. found the skin of the foot of obese patients to be about 7 degrees Celsius higher than the foot of nonobese patients, and thus caution should be noted if PPG is measured at the foot [60].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [60] Iacopi, 2020, Using Skin Bioengineering to Highlight How Weight and Diabetes Mellitus Modify the Skin in the Lower Limbs of Super-Obese Patients, Diabetes Metab. Syndr. Obes. Targets Ther. source
Not checked in the free run.
- Interpretationcitation
External temperature has a much more significant effect on the PPG waveform than body-site temperature.
“However, many of these studies suggested that external temperature has a much more significant effect on the PPG waveform.”
Quote in the paperNot a substantial claim - Premisecitation
A drop in 40 degrees Celsius contributed to a 12 dB drop in PPG signal amplitude.
“Askaraian et al. reported that while measuring PPG with a submerged finger of 20 healthy volunteers aged 18 to 80 years old, they found that a drop in 40 degrees Celsius contributed to a 12 dB drop in PPG signal amplitude [157].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [157] Askarian, B., Jung, K., and Chong, J.W. (2019). Monitoring of Heart Rate from Photoplethysmographic Signals Using a Samsung Galaxy Note8 in Underwater Environments. Sensors, 19. source
Not checked in the free run.
- Interpretationreasoning
Anatomical variations in parameters such as skin thickness and blood basal perfusion will lead to changes in the AC and DC amplitude, and the duration of a PPG waveform.
“Anatomical variations in parameters such as skin thickness and blood basal perfusion will lead to changes in the AC and DC amplitude, and the duration of a PPG waveform.”
Quote in the paperSubstantial claim - Premisecitation
Locations on the head and finger provide the greatest PPG amplitude.
“Overall, it has been found that locations on the head and finger provide the greatest PPG amplitude [95,145].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [95] Hartmann, 2019, Quantitative comparison of photoplethysmographic waveform characteristics: Effect of measurement site, Front. Physiol. source
Not checked in the free run.
- cites [145] Lee, Y., Han, H., and Kim, J. (2008, January 14–17). Influence of motion artifacts on photoplethysmographic signals for measuring pulse rates. Proceedings of the 2008 International Conference on Control, Automation and Systems, Seoul, Korea.
Not checked in the free run.
- Premisecitation
There is a reduction in PPG Root Mean Square values as ambient temperatures went from warm to cold temperatures.
“Khan et al. investigated the effects of finger temperature on PPG signal on 20 healthy adult volunteers (24.5 ± 4.1 years of age), reporting a reduction in PPG Root Mean Square values as ambient temperatures went from warm to cold temperatures [154].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [154] Khan, 2015, Investigating the effects of temperature on photoplethysmography, IFAC Pap. Online source
Not checked in the free run.
- Premisecitation
Local skin surface temperature changes affect PPG components.
“Jeong et al. found that local skin surface temperature changes affect PPG components of 16 healthy subjects aged 23–30 years (26 ± 2.1) and BMI ranging between 20 and 26 (23 ± 4.8), hence recommending that temperature be monitored in order to reliably evaluate cardiovascular parameters [158].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [158] Jeong, 2014, Effects of skin surface temperature on photoplethysmograph, J. Healthc. Eng. source
Not checked in the free run.
- Premisecitation
Many photodetectors are subject to creating artifacts from sensor-tissue motion and sensor deformation changes.
“For example, many of the photodetectors are subject to creating artifacts from sensor-tissue motion and sensor deformation changes [160].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [160] Wijshoff, 2012, Reducing motion artifacts in photoplethysmograms by using relative sensor motion: Phantom study, J. Biomed. Opt. source
Not checked in the free run.
- Premisecitation
PPG amplitude showed a direct response to skin temperature elevation, especially in the finger skin.
“Lindberg et al. found that PPG amplitude showed a direct response to skin temperature elevation, especially in the finger skin of 10 Caucasian young adults (aged 22–30) using three different arrangements of PPG probes with different source/detector separations [153].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [153] Lindberg, 1991, Photoplethysmography, Med. Biol. Eng. Comput. source
Not checked in the free run.
- Premisecitation
There are significant reductions in PPG pulse wave amplitudes (p < 0.0001) between healthy volunteers and individuals with systemic sclerosis both at rest and after cooling the finger to 16 degrees Celsius.
“They report significant reductions in PPG pulse wave amplitudes (p < 0.0001) between the groups both at rest and after cooling the finger to 16 degrees Celsius [156].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [156] Hahn, 1999, Local cold exposure test with a new arterial photoplethysmographic sensor in healthy controls and patients with secondary Raynaud’s phenomenon, Microvasc. Res. source
Not checked in the free run.
- Interpretationreasoning
Beyond within-individual variation and changing physiologies, external factors can significantly impact RCIM PPG.
“Thus, beyond variation found within individuals and changing physiologies, external factors can also significantly impact RCIM PPG.”
Quote in the paperSubstantial claim - Premisecitation
During sedentary moments, respiration rate, thermoregulation, and sympathetic nervous system activity cause the DC baseline to wander.
“During sedentary moments, respiration rate, thermoregulation, and sympathetic nervous system activity cause the DC baseline to wander [162].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [162] Maeda, 2011, Relationship between measurement site and motion artifacts in wearable reflected photoplethysmography, J. Med. Syst. source
Not checked in the free run.
- Interpretationnone given
Different anatomies demonstrate PPGs with varying waveforms.
“Overall, the literature has demonstrated that different anatomies demonstrate PPGs with varying waveforms.”
Quote in the paperSubstantial claim - Interpretationreasoning
Due to the presence of a distinct dicrotic notch and low skin thickness, the finger is likely to be a successful anatomy that can be used with consistency.
“Due to the presence of a distinct dicrotic notch and low skin thickness, the finger is likely to be a successful anatomy that can be used with consistency.”
Quote in the paperNot a substantial claim - Premisecitation
When motion is cyclical or periodic, heart rate tracking devices can mistake the cyclical motion for the cardiac cycle, causing false readings.
“Heart rate tracking devices will pick up on the cyclical motion and mistake it for the cardiac cycle, causing false readings as well [46].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [46] Bent, 2020, Investigating sources of inaccuracy in wearable optical heart rate sensors, NPJ Digit. Med. source
Not checked in the free run.
- Premisecitation
Significant changes in DC and AC amplitudes of the PPG pulse indicate that mild cold exposure has a substantial effect on finger blood circulation.
“Significant changes in DC and AC amplitudes of the PPG pulse indicate that the mild cold exposure has a substantial effect on finger blood circulation.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [159] Zhang, 2006, The effect of local mild cold exposure on pulse transit time, Physiol. Meas. source
Not checked in the free run.
- Premisecitation
Drift current varies directly with temperature when photodiodes are used in photovoltaic mode.
“Drift current varies directly with temperature when photodiodes are used in photovoltaic mode [161].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [161] Optoelectronics, O. (2020, August 06). Temperature and Bias Effects. Available online: http://www.osioptoelectronics.com/technology-corner/frequently-asked-questions/temperature-bias-effects.aspx.
Not checked in the free run.
- Premisecitation
The non-pulsating component of PPG is a function of the average blood volume, respiration, the sympathetic nervous system, and thermoregulation.
“Thus, while the pulsating components of PPG are related to arterial blood volume, the non-pulsating component is a function of the average blood volume, respiration, the sympathetic nervous system, and thermoregulation [151].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [151] Allen, 2007, Photoplethysmography and its application in clinical physiological measurement, Physiol. Meas. source
Not checked in the free run.
- Premisecitation
PPG waveforms are largely similar between bilateral anatomies at the ear, finger, and toe at rest.
“Allen and Murray have noted that PPG waveforms are largely similar between bilateral anatomies at the ear, finger, and toe at rest, according to their study featuring 116 (68 male, 48 female) healthy Caucasian subjects with a median age of 43 [90].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [90] Allen, 2003, Age-related changes in the characteristics of the photoplethysmographic pulse shape at various body sites, Physiol. Meas. source
Not checked in the free run.
- Premisecitation
The hand and the face had significantly higher (p < 0.01) perfusion values than the trunk, upper, and lower limb.
“It was found that the hand and the face had significantly higher (p < 0.01) perfusion values than the trunk, upper, and lower limb.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [144] Tur, 1983, Basal perfusion of the cutaneous microcirculation: Measurements as a function of anatomic position, J. Investig. Dermatol. source
Not checked in the free run.
- Premisecitation
The finger was shown to have a significantly lower (p < 0.001) reflection index.
“Finally, the finger was shown to have a significantly lower (p < 0.001) reflection index.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [95] Hartmann, 2019, Quantitative comparison of photoplethysmographic waveform characteristics: Effect of measurement site, Front. Physiol. source
Not checked in the free run.
- Premisecitation
There exists tremendous variation even within a given body site caused by factors such as age, gender, BMI, sun-damage, and experimental methodology, leading to coefficients of variation up to 40%.
“However, there exists tremendous variation even within a given body site caused by factors such as age, gender, BMI, sun-damage, and experimental methodology, leading to coefficients of variation up to 40% [139,140,141].”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [139] Dahlqvist, 2018, Vibrotactile and thermal perception and its relation to finger skin thickness, Clin. Neurophysiol. Pract. source
Not checked in the free run.
- cites [140] Olsen, 1995, High-frequency ultrasound characterization of normal skin. Skin thickness and echographic density of 22 anatomical sites, Skin Res. Technol. source
Not checked in the free run.
- cites [141] Martin, 1992, Effects of skin thickness and skinfold compressibility on skinfold thickness measurement, Am. J. Hum. Biol. source
Not checked in the free run.
- Premisecitation
The finger yielded a significant peak point position, meaning a shorter systolic rising edge (p < 0.001 for all sites except under the wrist, which had p = 0.04).
“The finger yielded a significant peak point position, meaning a shorter systolic rising edge (p < 0.001 for all sites except under the wrist, which had p = 0.04).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [95] Hartmann, 2019, Quantitative comparison of photoplethysmographic waveform characteristics: Effect of measurement site, Front. Physiol. source
Not checked in the free run.
- Premisecitation
The wrist significantly (p < 0.01) had a greater full-width half-max than the finger.
“It was found that the wrist significantly (p < 0.01) had a greater full-width half-max than the finger.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [26] Rajala, 2018, Comparison of photoplethysmogram measured from wrist and finger and the effect of measurement location on pulse arrival time, Physiol. Meas. source
Not checked in the free run.
- Premisecitation
The forearm had a high respiratory rate spectral power content compared to anatomies such as the finger, which had signal primarily derived from heart rate.
“It was found that the forearm had a high respiratory rate spectral power content compared to anatomies such as the finger, which had signal primarily derived from heart rate [130].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [130] Nilsson, 2007, Combined photoplethysmographic monitoring of respiration rate and pulse: A comparison between different measurement sites in spontaneously breathing subjects, Acta Anaesthesiol. Scand. source
Not checked in the free run.
- Premisecitation
PPG reproducibility is affected by cold exposure and acclimatization after middle finger submersion in 9 ± 2 °C water for 2 min.
“Zhang et al. demonstrated how PPG reproducibility is affected by cold exposure and acclimatization after middle finger submersion in 9 ± 2 °C water for 2 min and comparison with index finger as a reference [159].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [159] Zhang, 2006, The effect of local mild cold exposure on pulse transit time, Physiol. Meas. source
Not checked in the free run.
- Premisecitation
All trackers had a significant difference from ECG during low-intensity events.
“All had a significant difference during low-intensity events.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [167] Dooley, 2017, Estimating accuracy at exercise intensities: A comparative study of self-monitoring heart rate and physical activity wearable devices, JMIR mHealth uHealth source
Not checked in the free run.
- Premisecitation
There was a significant difference between research-grade and consumer-grade wearables.
“There was a significant difference between research-grade and consumer-grade wearables.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [46] Bent, 2020, Investigating sources of inaccuracy in wearable optical heart rate sensors, NPJ Digit. Med. source
Not checked in the free run.
- Premisecitation
With this algorithm, Karlen et al. were able to distinguish where motion artifacts occurred in a beat-to-beat manner.
“With this algorithm, Karlen et al. were able to distinguish where motion artifacts occurred in a beat-to-beat manner.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [168] Karlen, 2012, Photoplethysmogram signal quality estimation using repeated Gaussian filters and cross-correlation, Physiol. Meas. source
Not checked in the free run.
- Premisecitation
The upper arm had the lowest artifact ratio among upper arm, forearm, wrist, and finger.
“They found the upper arm had the lowest artifact ratio, which is the ratio of the magnitude of the PPG signal after a motion and the magnitude of the PPG signal before the motion.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [162] Maeda, 2011, Relationship between measurement site and motion artifacts in wearable reflected photoplethysmography, J. Med. Syst. source
Not checked in the free run.
- Interpretationreasoning
During intermediate and low intensity motions, accuracy is dependent on the device.
“This conclusion showed that during intermediate and low intensity motions, the accuracy is dependent on the device.”
Quote in the paperSubstantial claim - Premisecitation
The mean absolute error (MAE) for all motion phases was better for consumer wearables than research wearables.
“The mean absolute error (MAE) for all motion phases was better for the consumer wearables than the research wearables.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [46] Bent, 2020, Investigating sources of inaccuracy in wearable optical heart rate sensors, NPJ Digit. Med. source
Not checked in the free run.
- Premisecitation
The green sensor had a lower artifact ratio for all locations compared to the IR sensor.
“The green sensor also had a lower artifact ratio for all locations compared to the IR sensor.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [162] Maeda, 2011, Relationship between measurement site and motion artifacts in wearable reflected photoplethysmography, J. Med. Syst. source
Not checked in the free run.
- Premisecitation
During the high intensity running stage, some of the lowest MAPE were from the Mio Alpha (0.82%), TomTom Runner Cardio (0.97%), and Basis Peak (3.28%).
“They also reported that during the high intensity running stage, they saw some of the lowest MAPE from the MA (0.82%), the TT (0.97%), and the BP (3.28%).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [166] Stahl, 2016, How accurate are the wrist-based heart rate monitors during walking and running activities? Are they accurate enough?, BMJ Open Sport Exerc. Med. source
Not checked in the free run.
- Interpretationcitation
Motion artifacts can affect the acquired PPG signal differently based on sensor location and wavelength of light used.
“Motion artifacts can affect the acquired signal differently based on the location of the sensor along with what wavelength of light is used.”
Quote in the paperSubstantial claim - Premisecitation
The wearable device and activity condition are significantly correlated.
“Bent et al. confirmed that the wearable device and activity condition are significantly correlated in their study [46].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [46] Bent, 2020, Investigating sources of inaccuracy in wearable optical heart rate sensors, NPJ Digit. Med. source
Not checked in the free run.
- Premisecitation
With longer wavelength, red had a significantly bigger ΔSNR than green or blue.
“With the longer wavelength, red had a significantly bigger ΔSNR than green or blue.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [164] Lee, J., Matsumura, K., Yamakoshi, K.-I., Rolfe, P., Tanaka, S., and Yamakoshi, T. (2013, January 3–7). Comparison between red, green and blue light reflection photoplethysmography for heart rate monitoring during motion. Proceedings of the 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Osaka, Japan.
Not checked in the free run.
- Premisecitation
Karlen et al. created an algorithm with 96.21% sensitivity and 99.2% positive predictive value for good pulses using repeated Gaussian filters and cross-correlation.
“Karlen et al. created an algorithm with 96.21% sensitivity and 99.2% positive predictive value for good pulses using repeated Gaussian filters and cross-correlation [168].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [168] Karlen, 2012, Photoplethysmogram signal quality estimation using repeated Gaussian filters and cross-correlation, Physiol. Meas. source
Not checked in the free run.
- Premisecitation
The Basis Peak performed better relative to ECG (r = 0.92, p < 0.0001) than the Fitbit Charge HR (r = 0.83, p < 0.0001).
“Overall, they discovered that the BP performed better, in relationship to the ECG, (r = 0.92, p < 0.0001) than the FB (r = 0.83, p < 0.0001).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [165] Jo, 2016, Validation of biofeedback wearables for photoplethysmographic heart rate tracking, J. Sports Sci. Med.
Not checked in the free run.
- Interpretationown data
Motion artifacts cause varying accuracies depending on motion intensity and the type of heart rate monitor used.
“All the studies have shown that motion artifacts cause varying accuracies depending on motion intensity and the type of heart rate monitor used.”
Quote in the paperSubstantial claim - Premisecitation
The highest correlation coefficient between the PPG signal and a chest ECG was that of the green sensor on the upper arm.
“The highest correlation coefficient between the PPG signal and a chest ECG was that of the green sensor on the upper arm.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [162] Maeda, 2011, Relationship between measurement site and motion artifacts in wearable reflected photoplethysmography, J. Med. Syst. source
Not checked in the free run.
- Interpretationcitation
All the parameters could be used to eliminate sections of the PPG signal that were corrupted with motion artifacts.
“All the parameters could then be used to eliminate sections of the PPG signal that were corrupted with motion artifacts.”
Quote in the paperSubstantial claim - Interpretationreasoning
These detection methods will help keep heart rate and SpO2 measurement error lower.
“This will help to keep heart rate and SpO2 measurement error lower.”
Quote in the paperSubstantial claim - Interpretationreasoning
Each of these sensors is used to detect and then reduce or eliminate noise so a clean signal can be reconstructed.
“Each of these sensors is used to detect and then reduce or eliminate noise so a clean signal can be reconstructed.”
Quote in the paperNot a substantial claim - Interpretationreasoning
The use of adaptive filters and secondary sensors has been shown to decrease the error in heart rate and SpO2 measurements.
“This has been shown to decrease the error in heart rate and SpO2 measurements.”
Quote in the paperNot a substantial claim - Premisecitation
Quadratic phase was present in corrupted signals.
“The quadratic phase was present in corrupted signals.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [173] Krishnan, R., Natarajan, B., and Warren, S. (April, January 31). Analysis and detection of motion artifact in photoplethysmographic data using higher order statistics. Proceedings of the 2008 IEEE International Conference on Acoustics, Speech and Signal Processing, Las Vegas, NV, USA. source
Not checked in the free run.
- Premisecitation
Secondary sensors can be integrated into PPG sensors to detect motion, including accelerometers, gyroscopes, piezoelectric transducers, or another wavelength of light as a motion reference.
“As implemented by consumer-grade HR monitors, secondary sensors can also be integrated into PPG sensors to detect motion. These include accelerometers [175,176,177,178,179], gyroscopes [179,180], piezoelectric transducers [181], or the utilization of another wavelength of light as a motion reference [182,183].”
Quote in the paperNot a substantial claim0 of 7 sources checked- cites [175] Gibbs, P., and Asada, H.H. (2005, January 8–10). Reducing motion artifact in wearable bio-sensors using MEMS accelerometers for active noise cancellation. Proceedings of the 2005 American Control Conference, Portland, OR, USA.
Not checked in the free run.
- cites [176] Khan, 2015, A robust heart rate monitoring scheme using photoplethysmographic signals corrupted by intense motion artifacts, IEEE Trans. Biomed. Eng. source
Not checked in the free run.
- cites [177] Han, H., Kim, M.-J., and Kim, J. (2007, January 23–26). Development of real-time motion artifact reduction algorithm for a wearable photoplethysmography. Proceedings of the 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Lyon, France. source
Not checked in the free run.
- cites [178] Zhang, 2015, Photoplethysmography-based heart rate monitoring in physical activities via joint sparse spectrum reconstruction, IEEE Trans. Biomed. Eng. source
Not checked in the free run.
- cites [179] Casson, 2016, Gyroscope vs. accelerometer measurements of motion from wrist PPG during physical exercise, ICT Express source
Not checked in the free run.
- cites [180] Lee, 2018, Motion artifact cancellation in wearable photoplethysmography using gyroscope, IEEE Sens. J. source
Not checked in the free run.
- cites [181] Lee, 2019, Motion artifact identification and removal from wearable reflectance photoplethysmography using piezoelectric transducer, IEEE Sens. J. source
Not checked in the free run.
- Interpretationreasoning
With increased applied pressure, the AC amplitude begins to increase due to a decrease in optical path length through the tissue and the approach of the transmural pressure to zero.
“With increased applied pressure, the AC amplitude begins to increase due to a decrease in optical path length through the tissue and the approach of the transmural pressure to zero.”
Quote in the paperNot a substantial claim - Interpretationreasoning
Different vasculature will occlude at different amounts of pressure, contributing to perturbations in the signal.
“Different vasculature will occlude at different amounts of pressure, also contributing to perturbations in the signal.”
Quote in the paperNot a substantial claim - Premisecitation
Grabovskis et al. reported a less than 3% coefficient of variation within a subject at the single measurement site when determining the optimal pressure needed to determine an unloaded artery.
“Grabovskis et al. reported a less than 3% coefficient of variation within a subject at the single measurement site when determining the optimal pressure needed to determine an unloaded artery.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [209] Grabovskis, A., Marcinkevics, Z., Rubenis, O., Rubins, U., and Lusa, V. (2012, January 16–19). Photoplethysmography system for blood pulsation detection in unloaded artery conditions. Proceedings of the Biophotonics: Photonic Solutions for Better Health Care III, Brussels, Belgium. source
Not checked in the free run.
- Interpretationcitation
Interference from ambient light can be effectively compensated by optical shielding of the transducer, selective filtering of noise outside the PPG bandwidth, and correlated double sampling.
“This can be effectively compensated by various methods like optical shielding of the transducer, selective filtering of noise outside the PPG bandwidth, and with correlated double sampling.”
Quote in the paperNot a substantial claim - Premisecitation
When pressures are kept within the range of 5-15 kPa, there is less than 2% error in the SpO2 measurements.
“Utilizing this modality, they demonstrated that, when pressures are kept within the range of 5–15 kPa, there is less than 2% error in the SpO2 measurements.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [215] Liu, C., Correia, R., Ballaji, H.K., Korposh, S., Hayes-Gill, B.R., and Morgan, S.P. (2018). Optical fibre-based pulse oximetry sensor with contact force detection. Sensors, 18. source
Not checked in the free run.
- Interpretationreasoning
Interference of ambient light with PPG signals can lead to inaccurate heart rate estimation.
“The interference of ambient light with the PPG signals can lead to an inaccurate heart rate estimation.”
Quote in the paperSubstantial claim- Rests on an assumption: Ambient light intensity is generally multiple orders of magnitude larger than the pulsatile (AC) component of the PPG, which can lead to signal saturation.
- Premisecitation
With a contact force of 0.1-0.8 N applied at the finger, the PTT measured from the R peak of an ECG and the peak of the second derivative of the PPG increased significantly (p = 0.014) until the estimated transmural pressure approached −0.1 N, after which the PTT remained roughly constant.
“Regarding PTT, Teng and Zhang discovered that, with a contact force of 0.1–0.8 N applied at the finger, the PTT measured from the R peak of an ECG and the peak of the second derivative of the PPG increased significantly (p = 0.014) until the estimated transmural pressure approached −0.1 N, after which the PTT remained roughly constant [208].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [208] Teng, 2006, The effect of applied sensor contact force on pulse transit time, Physiol. Meas. source
Not checked in the free run.
- Interpretationreasoning
Ambient light rejection is important to preserve the efficacy of PPG sensors.
“Thus, ambient light rejection is important to preserve the efficacy of PPG sensors [197,198].”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [197] Sazonov, E. (2014). Wearable Sensors: Fundamentals, Implementation and Applications, Elsevier.
Not checked in the free run.
- cites [198] Kim, J., Lee, T., Kim, J., and Ko, H. (2015, January 25–29). Ambient light cancellation in photoplethysmogram application using alternating sampling and charge redistribution technique. Proceedings of the 2015 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Milan, Italy.
Not checked in the free run.
- Rests on an assumption: Ambient light intensity is generally multiple orders of magnitude larger than the pulsatile (AC) component of the PPG, which can lead to signal saturation.
- Interpretationnone given
The accuracy of PPG measurements could be limited due to interference of environment noise such as ambient light.
“The accuracy of PPG measurements could be limited due to the interference of environment noise such as ambient light.”
Quote in the paperNot a substantial claim- Rests on an assumption: Ambient light intensity is generally multiple orders of magnitude larger than the pulsatile (AC) component of the PPG, which can lead to signal saturation.
- Premisecitation
With low applied external pressure from the sensor, the waveform exhibits lower SNR with a low AC amplitude primarily because of a longer optical path length through the tissue and a lower reflectance of photons due to high tissue absorption.
“With a low applied external pressure from the sensor, the waveform exhibits lower SNR with a low AC amplitude primarily because of a longer optical path length through the tissue and a lower reflectance of photons due to high tissue absorption [203].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [203] Teng, 2007, Theoretical study on the effect of sensor contact force on pulse transit time, IEEE Trans. Biomed. Eng. source
Not checked in the free run.
- Premisecitation
For the PTT measured from 50% of the pulse amplitude, there was a significant increase in PTT (p = 0.038) before reaching a constant at −0.1 N transmural pressure.
“For the PTT measured from 50% of the pulse amplitude, there was a significant increase in PTT (p = 0.038) before reaching a constant at −0.1 N transmural pressure.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [208] Teng, 2006, The effect of applied sensor contact force on pulse transit time, Physiol. Meas. source
Not checked in the free run.
- Interpretationreasoning
Selective filtering of unwanted signals above and below the 0.5 to 4 Hz frequency range is a method to eliminate noise due to ambient light.
“Selective filtering of unwanted signals (above and below 0.5 to 4 Hz frequency range) is another method followed to eliminate the noise due to ambient light.”
Quote in the paperNot a substantial claim - Interpretationreasoning
Integrating a solution to produce constant contact pressure or being able to measure the pressure in PPG devices will produce more consistent and repeatable measurements.
“Therefore, integrating a solution to produce constant contact pressure or being able to measure the pressure in PPG devices will produce more consistent and repeatable measurements.”
Quote in the paperSubstantial claim - Premisecitation
In 1991, commercial pulse oximeters had measurement error caused by ambient light.
“In 1991, it was found that commercial pulse oximeters had measurement error caused by ambient light [199].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [199] Ralston, 1991, Potential errors in pulse oximetry: I. Pulse oximeter evaluation, Anaesthesia source
Not checked in the free run.
- Interpretationreasoning
Variations in applied pressure influence the resulting PPG waveform, causing an increase or decrease in amplitude along with a shift in offset.
“Discrepancies in PPG signals can arise where variations in applied pressure influence the resulting waveform. This can result in an increase or decrease in amplitude along with a shift in offset.”
Quote in the paperSubstantial claim - Interpretationreasoning
Inaccurate PPG signal acquisition due to varying contact pressures can lead to inaccurate secondary measurements.
“Overall, the above studies show that inaccurate PPG signal acquisition due to varying contact pressures can lead to inaccurate secondary measurements.”
Quote in the paperSubstantial claim - Premisecitation
The arrangement of the source and detector in Patterson et al.'s flexible PPG sensor contributes to eliminating interference of ambient noise to the signal.
“Apart from its electronic system design, the arrangement of the source and detector in this sensor also contributes to eliminate the interference of ambient noise to the signal.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [201] Patterson, J.A., McIlwraith, D.C., and Yang, G.-Z. (2009, January 3–5). A flexible, low noise reflective PPG sensor platform for ear-worn heart rate monitoring. Proceedings of the 2009 Sixth International Workshop on Wearable and Implantable Body Sensor Networks, Berkeley, CA, USA. source
Not checked in the free run.
- Premisecitation
When the transmural pressure reaches zero, the AC amplitude reaches its maximum.
“When the transmural pressure reaches zero, the AC amplitude reaches its maximum [204,205].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [204] Yamakoshi, 1982, New oscillometric method for indirect measurement of systolic and mean arterial pressure in the human finger. Part 1: Model experiment, Med. Biol. Eng. Comput. source
Not checked in the free run.
- cites [205] Forster, 1986, Oscillometric determination of diastolic, mean and systolic blood pressure—A numerical model, J. Biomech. Eng. source
Not checked in the free run.
- Premisecitation
Patterson et al.'s opto-electronic modulation scheme helps eliminate ambient light noise through multiplexing and sampling of light level by keeping LEDs off for a fixed time period and later subtracting it out from the desired signals.
“The opto-electronic modulation scheme in this flexible PPG sensor helps to eliminate ambient light noise through the multiplexing technique and sampling of light level by keeping the LEDs off for a fixed time period and later subtracting it out from the desired signals.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [201] Patterson, J.A., McIlwraith, D.C., and Yang, G.-Z. (2009, January 3–5). A flexible, low noise reflective PPG sensor platform for ear-worn heart rate monitoring. Proceedings of the 2009 Sixth International Workshop on Wearable and Implantable Body Sensor Networks, Berkeley, CA, USA. source
Not checked in the free run.
- Interpretationcitation
Modifications in the signal processing method at the signal amplifier can successfully reject DC photocurrent in PPG sensors.
“Modifications in the signal processing method at the signal amplifier can also successfully reject DC photocurrent in PPG sensors.”
Quote in the paperNot a substantial claim
- Classified as probably not a substantial claim. 40
- May combine two claims. 102
- The quote only partly states this claim. 1
Probably not a substantial claim: may not pass S3 (load-bearing).
Findingnone givenAll six identified devices with FDA status appear to be able to provide patients and providers with data for oxygen saturation (SpO2), respiration rate, and pulse rate.
“All six identified devices with FDA status appear to be able to provide patients and providers with data for oxygen saturation (SpO2), respiration rate, and pulse rate.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingnone givenIn some cases, FDA approval is only for when the wearable is used within a software suite or healthcare framework, as with the Samsung Gear S2.
“In some cases, FDA approval is only for when the wearable is used within a software suite or healthcare framework. This is the case with the Samsung Gear S2, which is FDA cleared to monitor heart rate toward detecting atrial fibrillation when done with the LIVMOR Halo™ Detection System.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based).
Findingnone givenTo the authors' knowledge, there is no work that comprehensively summarizes the literature surrounding these sources of noise and how they affect the waveform of the PPG and its derivatives.
“To the authors’ knowledge, there is no work that comprehensively summarizes the literature surrounding these sources of noise and how they affect the waveform of the PPG and its derivatives.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Findingnone givenTo the authors' best knowledge, literature does not exist that demonstrates the relationship between skin thickness and obesity for the finger, a common location for pulse oximeters.
“To the authors’ best knowledge, literature does not exist that demonstrates this relationship for the finger, a common location for pulse oximeters.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingnone givenThe temperature of the body site where PPG data are collected can influence the resultant waveform amplitude and the time between waveforms.
“It has been shown that the temperature of the body site where PPG data are being collected can influence the resultant waveform amplitude, and even the time between waveforms. This is summarized in Table 5.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S2 (checkable), S4 (based).
Findingnone givenOur understanding of how conditions such as obesity affect the skin and cardiovascular monitoring is still developing.
“While many sources of noise had documented potential solutions, few had a comprehensive solution and even our understanding of how conditions such as obesity affect the skin and cardiovascular monitoring are still developing.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
InterpretationreasoningPPG can fill the gap in preventative cardiovascular monitoring if a sufficiently accurate and precise device is developed.
“PPG can fill this gap if a sufficiently accurate and precise device is developed.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.
Premisenone givenPPG systems developed for remote and wearable use are typically for general wellness and fitness, which precludes them from being prescribed for medical use at home.
“PPG systems developed for remote and wearable use are typically for general wellness and fitness. This precludes it from being prescribed for medical use at home.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing), S4 (based). May combine two claims.
Interpretationnone givenThe indications for use in these devices are very significant advancements in remote monitoring, but still lag in the potential prognostic capabilities of the PPG.
“The indications for use in these devices are very significant advancements in remote monitoring, but still lag in the potential prognostic capabilities of the PPG.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationRCIM has proven to be a promising route to deliver preventative care by reducing both the death rate and burdens placed on the healthcare system.
“Remote and continuous/intermittent monitoring (RCIM) has proven to be a promising route to deliver preventative care by reducing both the death rate and burdens placed on the healthcare system [1,2,3].”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [1] Wilkoff, 2008, HRS/EHRA expert consensus on the monitoring of cardiovascular implantable electronic devices (CIEDs): Description of techniques, indications, personnel, frequency and ethical considerations: Developed in partnership with the Heart Rhythm Society (HRS) and the European Heart Rhythm Association (EHRA); and in collaboration with the American College of Cardiology (ACC), the American Heart Association (AHA), the European Society of Cardiology (ESC), the Heart Failure Association of ESC (HFA), and the Heart Failure Society of America (HFSA). Endorsed by the Heart Rhythm Society, the source
Not checked in the free run.
- cites [2] Roger, 2011, Heart disease and stroke statistics—2011 update: A report from the American Heart Association, Circulation source
Not checked in the free run.
- cites [3] Jencks, 2009, Rehospitalizations among patients in the Medicare fee-for-service program, N. Engl. J. Med. source
Not checked in the free run.
May combine two claims.
PremisecitationThe blood pressure cuff, ECG patch, and Holter monitor are not often used for long-term remote monitoring.
“The blood pressure cuff, ECG patch, and Holter monitor are also not often used for long-term remote monitoring [24].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [24] Bansal, 2018, Portable out-of-hospital electrocardiography: A review of current technologies, J. Arrhythmia source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Premisenone givenHardware and software within the device itself can contribute to the noise.
“In addition, the hardware and software within the device itself can contribute to the noise.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationThese tests are rarely administered preventatively, despite research which concludes that preventative testing could reduce deaths by up to 25%.
“Additionally, these tests are rarely administered preventatively, despite research which concludes that preventative testing could reduce deaths by up to 25% [1].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [1] Wilkoff, 2008, HRS/EHRA expert consensus on the monitoring of cardiovascular implantable electronic devices (CIEDs): Description of techniques, indications, personnel, frequency and ethical considerations: Developed in partnership with the Heart Rhythm Society (HRS) and the European Heart Rhythm Association (EHRA); and in collaboration with the American College of Cardiology (ACC), the American Heart Association (AHA), the European Society of Cardiology (ESC), the Heart Failure Association of ESC (HFA), and the Heart Failure Society of America (HFSA). Endorsed by the Heart Rhythm Society, the source
Not checked in the free run.
Probably not a substantial claim: may not pass S4 (based). May combine two claims.
Premisenone givenThese sources of error pertain to variation within and across individuals (e.g., skin tone, obesity, age, and gender), physiology (e.g., respiration, venous pulsation, body site of measurement, and body temperature), and external perturbations of the device itself (e.g., motion artifact, ambient light, and applied pressure to the skin).
“These sources of error pertain to variation within and across individuals (e.g., skin tone, obesity, age, and gender), physiology (e.g., respiration, venous pulsation, body site of measurement, and body temperature), and external perturbations of the device itself (e.g., motion artifact, ambient light, and applied pressure to the skin).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationPTT, used to determine PWV and estimate blood pressure without a cuff, requires extraction of the time delay from two PPG waveforms or from an ECG and PPG waveform.
“Additionally, some parameters in the literature such as pulse transit time (PTT), which is used to determine pulse wave velocity (PWV) and estimate blood pressure without a cuff, requires extraction of the time delay from two PPG waveforms or from an ECG and PPG waveform [31,32].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [31] Nabeel, 2018, Arterial blood pressure estimation from local pulse wave velocity using dual-element photoplethysmograph probe, IEEE Trans. Instrum. Meas. source
Not checked in the free run.
- cites [32] Nabeel, P., Karthik, S., Joseph, J., and Sivaprakasam, M. (2017, January 8–10). Experimental validation of dual PPG local pulse wave velocity probe. Proceedings of the 2017 IEEE International Symposium on Medical Measurements and Applications (MeMeA), Rochester, MN, USA. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Premisenone givenNumerous non-FDA cleared/approved fitness devices exist and can estimate heart rate by quantifying the number of systolic peaks in a period of time, but a single parameter limits the amount of extractable information.
“Numerous non-FDA cleared/approved fitness devices exist and can estimate heart rate by quantifying the number of systolic peaks in a period of time, but a single parameter limits the amount of extractable information.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based).
Premisenone givenComplex analyses of the PPG waveform yield valuable clinical information such as blood pressure, respiratory information, sympathetic nervous system activity, and heart rate variability.
“However, complex analyses of the PPG waveform yield valuable clinical information, such as: blood pressure, respiratory information, sympathetic nervous system activity, and heart rate variability.”
Quote in the paperNot a substantial claimMay combine two claims.
Premisenone givenThe literature has demonstrated the potential diagnostic and prognostic strength for the PPG; however, the PPG features can only be utilized if the waveform is of a high quality with high signal-to-noise ratio (SNR).
“Overall, the literature has demonstrated the potential diagnostic and prognostic strength for the PPG; however, the PPG features can only be utilized if the waveform is of a high quality with high signal-to-noise ratio (SNR).”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationBeyond fitness and heart rate monitoring, the primary medical use of PPG has been focused on obtaining information about the cardiovascular system towards cardiovascular disease (CVD) diagnosis and treatment.
“Beyond fitness and heart rate monitoring, the primary medical use of the PPG has been focused on obtaining information about the cardiovascular system towards cardiovascular disease (CVD) diagnosis and treatment [17,18].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [17] Elgendi, 2019, The use of photoplethysmography for assessing hypertension, NPJ Digit. Med. source
Not checked in the free run.
- cites [18] Vakily, 2017, A system for continuous estimating and monitoring cardiac output via arterial waveform analysis, J. Biomed. Phys. Eng.
Not checked in the free run.
Classified as probably not a substantial claim.
InterpretationreasoningA wearable monitoring tool to be used beyond fitness and heart rate must account for noise sources originating from individual patient variations (e.g., skin tone, obesity, age, and gender), physiology (e.g., respiration, venous pulsation, body site of measurement, and body temperature), and external perturbations of the device itself (e.g., motion artifact, ambient light, and applied pressure to,
“A wearable monitoring tool to be used beyond fitness and heart rate must account for noise sources originating from individual patient variations (e.g., skin tone, obesity, age, and gender), physiology (e.g., respiration, venous pulsation, body site of measurement, and body temperature), and external perturbations of the device itself (e.g., motion artifact, ambient light, and applied pressure to”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationreasoningThese results indicate that overall, mean modulation goes down with increasing Fitzpatrick scale, and with exercise one would expect greater blood volume changes causing a higher signal for the more absorbing blue/green wavelengths, at least for the source to detector distances used in the study.
“These results indicate that overall, the mean modulation goes down with increasing Fitzpatrick scale, and with exercise one would expect greater blood volume changes causing a higher signal for the more absorbing blue/green wavelengths, at least for the source to detector distances used in the study.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationMohapatra et al. found that with a multiwavelength PPG device, perfusion index, pulsatile strength, and SNR all increased when data collected with the 590 nm LED were analyzed.
“With 20 subjects ranging from Fitzpatrick II to Fitzpatrick IV, the perfusion index (AC/DC), pulsatile strength, and SNR were all found to increase when data collected with the 590 nm LED were analyzed.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [51] Mohapatra, P., Preejith, S., and Sivaprakasam, M. (2017, January 22–25). A novel sensor for wrist based optical heart rate monitor. Proceedings of the 2017 IEEE International Instrumentation and Measurement Technology Conference (I2MTC), Torino, Italy. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationObesity, determined by having a body mass index greater than 30, affects 40% of the United States and leads to cardiovascular disease, hypertension, and type 2 diabetes.
“Obesity, determined by having a body mass index greater than 30, affects 40% of the United States and leads to cardiovascular disease, hypertension, and type 2 diabetes [57].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [57] Poirier, 2006, Obesity and cardiovascular disease: Pathophysiology, evaluation, and effect of weight loss: An update of the 1997 American Heart Association Scientific Statement on Obesity and Heart Disease from the Obesity Committee of the Council on Nutrition, Physical Activity, and Metabolism, Circulation source
Not checked in the free run.
May combine two claims.
InterpretationreasoningThe wavelength range needs to be optimized for both skin tone and blood absorption, particularly as more advanced parameters are derived from PPG signals, because absorption of green light by melanin in individuals with a darker skin tone limits light penetration to the subcutaneous tissue where the blood is located.
“However, the wavelength range needs to be optimized for both skin tone and blood absorption, particularly as more advanced parameters are derived from PPG signals, as the absorption of green light by melanin in individuals with a darker skin tone limits the light penetration to the subcutaneous tissue where the blood is located.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationMendelson et al. found that relative amplitude of the AC component to the entire signal (AC + DC) increased as LED/photodiode spacing increased from 4 to 11 mm, although no statistical analyses were presented.
“It was found that relative amplitude of the AC component to the entire signal (AC + DC) increased as LED/photodiode spacing increased from 4 to 11 mm, although no statistical analyses were presented.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [53] Mendelson, 1988, Noninvasive pulse oximetry utilizing skin reflectance photoplethysmography, IEEE Trans. Biomed. Eng. source
Not checked in the free run.
May combine two claims.
PremisecitationPreejith et al.'s device had an error of 1.04 BPM for 'fair' skinned individuals and 10.90 BPM for 'dark' skinned participants, a greater than 10 times increase in absolute error.
“There was an error of 1.04 beats per minute (BPM) for “fair” skinned individuals and an error of 10.90 BPM for “dark” skinned participants, citing the lack of usability of their device for dark skinned individuals.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [40] Preejith, S., Alex, A., Joseph, J., and Sivaprakasam, M. (2016, January 12–14). Design, development and clinical validation of a wrist-based optical heart rate monitor. Proceedings of the 2016 IEEE International Symposium on Medical Measurements and Applications (MeMeA), Benevento, Italy. source
Not checked in the free run.
May combine two claims.
PremisecitationBlood flow regulation and oxygen saturation are both known to deviate with respect to obesity and BMI.
“Blood flow regulation and oxygen saturation are both known to deviate with respect to obesity and BMI [71].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [71] Limberg, 2016, Peripheral Blood Flow Regulation in Human Obesity and Metabolic Syndrome, Exerc. Sport Sci. Rev. source
Not checked in the free run.
May combine two claims.
PremisecitationPreejith et al. found a significant direct correlation between SNR of heart rate and skin tone using a 535 nm LED, indicating that a darker skin tone yields higher error in the measurement.
“Preejith et al. confirmed that skin tone matters when using green light only. The analysis of their PPG biosensor (single source 535 nm LED) showed a significant direct correlation SNR of the heart rate with skin tone, indicating that a darker skin tone yields higher error in the measurement [40].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [40] Preejith, S., Alex, A., Joseph, J., and Sivaprakasam, M. (2016, January 12–14). Design, development and clinical validation of a wrist-based optical heart rate monitor. Proceedings of the 2016 IEEE International Symposium on Medical Measurements and Applications (MeMeA), Benevento, Italy. source
Not checked in the free run.
May combine two claims.
PremisereasoningMany PPG devices use green light (~550 nm), primarily driven by the relatively high absorption spectrum of hemoglobin in this range, which can give a strong pulsatile signal with changes in blood volume.
“However, many PPG devices use green light (~550 nm). The decision to use green light in most PPG systems is primarily driven by the relatively high absorption spectrum of hemoglobin in this range (Figure 3b), which is the main absorber in blood and thus can potentially give a strong pulsatile signal with changes in blood volume.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationObesity can lead to changes in skin thickness, blood flow, and oxygen saturation, affecting the optical properties of skin and the distance light has to travel to reach a target vessel.
“In particular, obesity can lead to changes in skin thickness, blood flow, and oxygen saturation. This affects the optical properties of skin in addition to the distance light has to travel to reach a target vessel or vessels [48,59,60].”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [48] Jacques, 2013, Optical properties of biological tissues: A review, Phys. Med. Biol. source
Not checked in the free run.
- cites [59] Elgendi, 2012, On the analysis of fingertip photoplethysmogram signals, Curr. Cardiol. Rev. source
Not checked in the free run.
- cites [60] Iacopi, 2020, Using Skin Bioengineering to Highlight How Weight and Diabetes Mellitus Modify the Skin in the Lower Limbs of Super-Obese Patients, Diabetes Metab. Syndr. Obes. Targets Ther. source
Not checked in the free run.
May combine two claims.
PremisecitationIndividuals with obesity experience increased cutaneous blood flow to meet the oxygenation needs of tissue, but the blood flow of adipose tissue generally decreases with obesity both after a meal and during a fasting state.
“Individuals with obesity experience increased cutaneous blood flow to meet the oxygenation needs of tissue. However, the blood flow of adipose tissue generally decreases with obesity both after a meal and during a fasting state [61,72].”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [61] Chin, 1999, Increased cutaneous blood flow but impaired post-ischemic response of nutritional flow in obese children, Atherosclerosis source
Not checked in the free run.
- cites [72] Altintas, 2016, In vivo reflectance-mode confocal microscopy assessments: Impact of overweight on human skin microcirculation and histomorphology, J. Biomed. Opt. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThese often-overlooked disparities with diversity (e.g., skin tone and obesity) are now becoming more documented in the literature.
“These often-overlooked disparities with diversity (e.g., skin tone and obesity) are now becoming more documented in the literature [34,35,36].”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [34] Colvonen, P.J., DeYoung, P.N., Bosompra, N.-O.A., and Owens, R.L. (2020). Limiting Racial Disparities and Bias for Wearable Devices in Health Science Research, Oxford University Press US. source
Not checked in the free run.
- cites [35] Hidalgo, 2021, Critical care trainees call for pulse oximetry reform, Lancet Respir. Med. source
Not checked in the free run.
- cites [36] Rodriguez, 2021, Synthetic Photoplethysmography (PPG) of the radial artery through parallelized Monte Carlo and its correlation to Body Mass Index (BMI), Sci. Rep. source
Not checked in the free run.
May combine two claims.
Premisenone givenEumelanin has high absorbance with a peak in the ultraviolet wavelength (220 nm) and a steady decay through the visible wavelength region.
“This is due to the high absorbance of eumelanin with a peak in the ultraviolet wavelength (220 nm) and a steady decay through the visible wavelength region.”
Quote in the paperSubstantial claimMay combine two claims.
PremisereasoningThe absorption of epidermal melanin is much higher in the visible region of light and much lower in the near infrared region (NIR), so NIR light will travel further through pigmented skin.
“Since the absorption of epidermal melanin is much higher in the visible region of light and much lower in the near infrared region (NIR), the NIR range of light will travel further through pigmented skin.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationHermand et al. found that bias (mean difference between the Polar OH1 and ground truth) increased with darker skin (p < 0.001), and heart rate accuracy was positively correlated to skin tone (p < 0.05).
“In determining heart rate, it was found that bias (defined as the mean difference between the OH1 and ground truth) increased with darker skin (p < 0.001), and heart rate accuracy was positively correlated to skin tone (p < 0.05).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [41] Hermand, 2019, Validation of a photoplethysmographic heart rate monitor: Polar OH1, Int. J. Sports Med. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationThe wearable devices used in Bent et al. utilized red and near infrared light, which can more easily penetrate the epidermis.
“Additionally, the wearable devices used in this study utilized red and near infrared light, which can more easily penetrate the epidermis.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [46] Bent, 2020, Investigating sources of inaccuracy in wearable optical heart rate sensors, NPJ Digit. Med. source
Not checked in the free run.
May combine two claims.
PremisecitationFor Fitzpatrick IV individuals, the perfusion index increased between 1.2 and 7.1 times with the use of 590 nm light, pulsatile strength increased by 1.1 to 3.1 times, and SNR increased by 1.3 to 2.6 times, although no statistical significance analyses were performed.
“For Fitzpatrick IV individuals, the perfusion index increased between 1.2 and 7.1 times with the use of the 590 light, while pulsatile strength increased by 1.1 to 3.1 times and SNR increased by 1.3 to 2.6 times, although no statistical significance analyses were performed [51].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [51] Mohapatra, P., Preejith, S., and Sivaprakasam, M. (2017, January 22–25). A novel sensor for wrist based optical heart rate monitor. Proceedings of the 2017 IEEE International Instrumentation and Measurement Technology Conference (I2MTC), Torino, Italy. source
Not checked in the free run.
May combine two claims.
PremisecitationIn the resting condition, green light had larger mean modulation (p < 0.001) than other wavelengths, and after exercise, blue and green had greater SNR ratios than red or infrared (p < 0.001).
“In the resting condition, green light had larger mean modulation (p < 0.001) than other wavelengths, and after exercise, blue and green had greater SNR ratios than red or infrared (p < 0.001).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [52] Fallow, 2013, Influence of skin type and wavelength on light wave reflectance, J. Clin. Monit. Comput. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
InterpretationreasoningThe lack of increase in error observed by Bent et al. may be due to the already large error present in the results (mean absolute error approximately 9 BPM across all skin tones).
“It is possible that the lack of increase in error observed is due to the already large error present in the results reported—the mean absolute error is approximately 9 BPM across all skin tones.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationreasoningMost of the listed changes are due to the cardiovascular system, as arteries change how they respond to a cardiac pulse as the vessel ages.
“Most of the listed changes are due to the cardiovascular system, as arteries change how they respond to a cardiac pulse as the vessel ages.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationSkin thickness is directly correlated to BMI and can dampen PPG signal amplitude.
“Perhaps most detrimental to the PPG waveform is skin thickness, as it is directly correlated to BMI and can dampen PPG signal amplitude [56,70].”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [56] Gibney, 2010, Skin and subcutaneous adipose layer thickness in adults with diabetes at sites used for insulin injections: Implications for needle length recommendations, Curr. Med. Res. Opin. source
Not checked in the free run.
- cites [70] Derraik, J.G., Rademaker, M., Cutfield, W.S., Pinto, T.E., Tregurtha, S., Faherty, A., Peart, J.M., Drury, P.L., and Hofman, P.L. (2014). Effects of age, gender, BMI, and anatomical site on skin thickness in children and adults with diabetes. PLoS ONE, 9. source
Not checked in the free run.
May combine two claims.
PremisecitationThe ratio b/a has been found to correlate positively to age, while c/a, d/a and e/a have all been correlated negatively to age.
“The ratio b/a has been found to correlate positively to age, while c/a, d/a and e/a have all been correlated negatively to age [29,59].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [29] Takazawa, 1998, Assessment of vasoactive agents and vascular aging by the second derivative of photoplethysmogram waveform, Hypertension source
Not checked in the free run.
- cites [59] Elgendi, 2012, On the analysis of fingertip photoplethysmogram signals, Curr. Cardiol. Rev. source
Not checked in the free run.
May combine two claims.
PremisecitationKeratinocytes become shorter, water content decreases, lipid content decreases, and collagen synthesis and turnover decrease with age.
“Keratinocytes become shorter, water content decreases, lipid content decreases, and most notably, collagen synthesis and turnover decrease [93].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [93] Farage, 2013, Characteristics of the aging skin, Adv. Wound Care source
Not checked in the free run.
May combine two claims.
InterpretationreasoningThis change in skin thickness decreases the necessary propagation of light to obtain a PPG signal, which can contribute to an increase in PPG signal strength, along with capillary depletion.
“This change in skin thickness decreases the necessary propagation of light to obtain a PPG signal, which can contribute to an increase in PPG signal strength, along with capillary depletion.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenThe increased absorption of deoxygenated hemoglobin compared to that of oxygenated hemoglobin in the far-red region could decrease SNR of PPG measurements that use those wavelengths of light.
“The increased absorption of deoxygenated hemoglobin compared to that of oxygenated hemoglobin in the far-red region could decrease SNR of PPG measurements that use those wavelengths of light.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisereasoningPPG principally measures the response/elasticity of arteries to blood flow, and thus will change if the properties governing artery compliance change.
“PPG principally measures the response/elasticity of arteries to blood flow, and thus will change if the properties governing artery compliance changes.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationLeveque et al. found a decrease in skin thickness and a positive, direct relationship between PPG amplitude and age, which they attribute to capillary depletion.
“A Holtain skin caliper applied to the dorsal forearm found a decrease in skin thickness, and an infrared PPG was used to determine PPG amplitude. The authors attribute the positive, direct relationship between PPG amplitude and age to capillary depletion, due to similar skin thicknesses observed between older participants and children, despite an increased PPG amplitude in the older participants.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [94] Leveque, 1984, In vivo studies of the evolution of physical properties of the human skin with age, Int. J. Dermatol. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationAhn et al. extracted features from fingertip-based PPG and the second derivative PPG (APG) and correlated them to chronological age to determine what they defined as the vascular age index.
“Using data from 93 individuals of various ages, Ahn et al. extracted features from fingertip-based PPG and the second derivative PPG, acceleration plethysmography (APG, Figure 4), and correlated them to chronological age in order to determine what they defined as the vascular age index [83].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [83] Ahn, 2017, New aging index using signal features of both photoplethysmograms and acceleration plethysmograms, Healthc. Inform. Res. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationHartmann et al. determined that there is no statistically significant change in PPG amplitude as a function of age.
“Another study by Hartmann et al. determined that there is no statistically significant change in PPG amplitude as a function of age; however, this work featured 36 individuals ranging in age from only 33 to 58.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [95] Hartmann, 2019, Quantitative comparison of photoplethysmographic waveform characteristics: Effect of measurement site, Front. Physiol. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Premisenone givenThe integumentary system also responds to aging.
“However, the integumentary system also responds to aging.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationSuzuki and Oguri used artificial intelligence to determine blood pressure from a cuffless monitor and incorporated age into their algorithm.
“Suzuki and Oguri similarly used artificial intelligence to determine blood pressure from a cuffless monitor and incorporated age into their algorithm [96].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [96] Suzuki, S., and Oguri, K. (2009, January 2–6). Cuffless blood pressure estimation by error-correcting output coding method based on an aggregation of adaboost with a photoplethysmograph sensor. Proceedings of the 2009 Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Hilton Minneapolis, MI, USA. source
Not checked in the free run.
May combine two claims.
PremisecitationGender differentiation in average blood pressure and average heart rate indicate that baseline differences in vascular and cardiovascular parameters must be considered to determine cardiovascular health between males and females, which also extends into analyses of the PPG waveform.
“Phuong and Maibach noted that gender differentiation in average blood pressure and average heart rate indicate that baseline differences in vascular and cardiovascular parameters must be considered to determine cardiovascular health between males and females, which also extends into analyses of the PPG waveform [98].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [98] Phuong, C., and Maibach, H.I. (2016). Gender differences in skin. J. Textb. Aging Skin, 1729–1755. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.
InterpretationreasoningThis leads to dramatic changes in PWV and subsequently PTT.
“This leads to dramatic changes in PWV and subsequently PTT.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationProctor et al. reported that the average heart rate for males was 70-72 BPM and the average for females was 78-85 BPM in 16 endurance-trained men and 14 endurance-trained women.
“Proctor et al. reported that the average heart rate for males was 70–72 BPM and the average for females was 78–85 BPM in 16 endurance-trained men and 14 endurance-trained women [99].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [99] Proctor, D.N., Beck, K.C., Shen, P.H., Eickhoff, T.J., Halliwill, J.R., and Joyner, M.J. (1998). Influence of age and gender on cardiac output-V o 2 relationships during submaximal cycle ergometry. J. Appl. Physiol. source
Not checked in the free run.
May combine two claims.
PremisecitationArtificial intelligence and machine learning have been used to determine blood pressure from PPG and mitigate the effects age have on the estimate.
“Artificial intelligence and machine learning have been used to determine blood pressure from PPG and mitigate the effects age have on the estimate.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [92] 2011, Non-invasive estimate of blood glucose and blood pressure from a photoplethysmograph by means of machine learning techniques, Artif. Intell. Med. source
Not checked in the free run.
- cites [96] Suzuki, S., and Oguri, K. (2009, January 2–6). Cuffless blood pressure estimation by error-correcting output coding method based on an aggregation of adaboost with a photoplethysmograph sensor. Proceedings of the 2009 Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Hilton Minneapolis, MI, USA. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationAn algorithm that incorporates age into systolic and diastolic blood pressure determination was created.
“An algorithm that incorporates age into systolic and diastolic blood pressure determination was created.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [92] 2011, Non-invasive estimate of blood glucose and blood pressure from a photoplethysmograph by means of machine learning techniques, Artif. Intell. Med. source
Not checked in the free run.
May combine two claims.
PremisecitationSkin thickness is a gender-dependent variable known to affect signal strength, and is observed to be higher in men than women across all age ranges.
“Skin thickness is another gender-dependent variable known to affect signal strength (as mentioned previously), and is observed to be higher in men than women across all age ranges [97,103].”
Quote in the paperNot a substantial claim0 of 2 sources checkedMay combine two claims.
PremisecitationSignificant differences in artery diameter lead to significantly different radial artery flow rates, with a flow rate of 21 ± 4 and 10 ± 1 mL/min for males and females, respectively.
“These significant differences in artery diameter lead to significantly different radial artery flow rates as well, with a flow rate of 21 ± 4 and 10 ± 1 mL/min for males and females, respectively [102].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [102] Joannides, 2002, Influence of vascular dimension on gender difference in flow-dependent dilatation of peripheral conduit arteries, Am. J. Physiol. Heart Circ. Physiol. source
Not checked in the free run.
May combine two claims.
PremisecitationPost-menopausal females have higher blood pressure than males, which can be attributed to arterial stiffness.
“However, post-menopausal females have higher blood pressure than males, which can be attributed to arterial stiffness [101].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [101] Reckelhoff, 2001, Gender differences in the regulation of blood pressure, Hypertension source
Not checked in the free run.
May combine two claims.
PremisecitationMales have a 15-30% increase in heart mass compared to females resulting in females' hearts having to beat faster to maintain the same output.
“Males have a 15–30% increase in heart mass compared to females resulting in females’ hearts having to beat faster to maintain the same output [100].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [100] Prabhavathi, 2014, Role of biological sex in normal cardiac function and in its disease outcome—A review, J. Clin. Diagn. Res. JCDR
Not checked in the free run.
May combine two claims.
PremisecitationPhysiological differences between men and women extend into cardiovascular health and are noticeable in the PPG waveform.
“The physiological differences between men and women extend into cardiovascular health and are thus noticeable in the PPG waveform [97,98].”
Quote in the paperSubstantial claim0 of 2 sources checkedProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationShuster et al. measured skin thickness in 90 Caucasian men and 107 women and found that forearm skin thickness is greater in men than women, but no statistics were presented to explore statistical significance.
“Shuster et al. measured skin thickness in 90 Caucasian men and 107 women and found that forearm skin thickness is greater in men than women, but no statistics were presented to explore statistical significance [103].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [103] Shuster, 1975, The influence of age and sex on skin thickness, skin collagen and density, Br. J. Dermatol. source
Not checked in the free run.
May combine two claims.
PremisecitationPost-puberty, both males and females have an increase in arterial compliance, determining no gender-dependent differences between the measurement.
“Post-puberty, both males and females have an increase in arterial compliance, determining no gender-dependent differences between the measurement.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [104] Ahimastos, 2003, Gender differences in large artery stiffness pre-and post puberty, J. Clin. Endocrinol. Metab. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationFor respiratory measurements, respiratory synchronous variation in the PPG signal is irrespective of gender, and thus no action is required.
“For respiratory measurements, Nilsson et al. have reported that respiratory synchronous variation in the PPG signal is irrespective of gender, and thus no action is required [111].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [111] Nilsson, 2006, Age and gender do not influence the ability to detect respiration by photoplethysmography, J. Clin. Monit. Comput. source
Not checked in the free run.
May combine two claims.
PremisecitationDehghanojamahalleh et al. demonstrated variations in PPG morphology attributed to gender differences, with direct influence of gender variation only significantly different at the upper peripheral measurement site such as the hand and fingers as opposed to lower extremities such as the ankle and feet.
“Dehghanojamahalleh et al. demonstrated variations in PPG morphology attributed to gender differences [105]. Interestingly, direct influence of gender variation was only significantly different at the upper peripheral measurement site such as the hand and fingers as opposed to lower extremities such as the ankle and feet.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [105] Dehghanojamahalleh, 2019, Sex-Related Differences in Photoplethysmography Signals Measured From Finger and Toe, IEEE J. Transl. Eng. Health Med. source
Not checked in the free run.
May combine two claims.
PremisecitationFor females, PWV stayed roughly the same, while there was an increase in PWV for males.
“However, for females, PWV stayed roughly the same, while there was an increase in PWV for males.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [104] Ahimastos, 2003, Gender differences in large artery stiffness pre-and post puberty, J. Clin. Endocrinol. Metab. source
Not checked in the free run.
May combine two claims.
PremisecitationDuring prepuberty, females have lower arterial compliance and higher central and peripheral PWV yielding stiffer arteries compared to males.
“It was discovered during prepuberty, females have lower arterial compliance and higher central and peripheral PWV yielding stiffer arteries compared to males.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [104] Ahimastos, 2003, Gender differences in large artery stiffness pre-and post puberty, J. Clin. Endocrinol. Metab. source
Not checked in the free run.
May combine two claims.
PremisecitationThe finger is arterially dominated while the forearm has a larger venous component.
“The finger is arterially dominated while the forearm has a larger venous component [120].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [120] Ysehak Abay, T., Shafqat, K., and Kyriacou, P.A. (2019). Perfusion Changes at the Forehead Measured by Photoplethysmography during a Head-Down Tilt Protocol. Biosensors, 9. source
Not checked in the free run.
May combine two claims.
PremisecitationPulse wave propagation delay between the genders indicates baseline differences in arterial stiffness with women displaying lower degrees of pulse arrival latency delay indicating higher vascular stiffness.
“Pulse wave propagation delay between the genders indicates baseline differences in arterial stiffness with women displaying lower degrees of pulse arrival latency delay indicating higher vascular stiffness [106,107].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [106] Mukkamala, 2015, Toward ubiquitous blood pressure monitoring via pulse transit time: Theory and practice, IEEE Trans. Biomed. Eng. source
Not checked in the free run.
- cites [107] Guajardo, 2018, Sex differences in vascular dysfunction and cardiovascular outcomes: The cardiac, endothelial function, and arterial stiffness in ESRD (CERES) study, Hemodial. Int. source
Not checked in the free run.
May combine two claims.
PremisecitationThe venous system contributes periodicity to the PPG signal, but this is often considered noise.
“The venous system similarly contributes periodicity to the PPG signal, but this is often considered noise.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [116] Shelley, 1993, The detection of peripheral venous pulsation using the pulse oximeter as a plethysmograph, J. Clin. Monit. source
Not checked in the free run.
May combine two claims.
PremisecitationShelley et al. and Sami et al. both demonstrated the influence of the venous signal on SpO2, showing that forehead and earlobe pulse oximeters produced a significantly biased variable component of the PPG determined to originate from central venous pressure (CVP).
“Shelley et al. and Sami et al. both demonstrated the influence of the venous signal on SpO2, showing that forehead and earlobe pulse oximeters (respectively) produced a significantly biased variable component of the PPG determined to originate from central venous pressure (CVP) [125,127].”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [125] Shelley, 2005, The effect of venous pulsation on the forehead pulse oximeter wave form as a possible source of error in Spo2 calculation, Anesth. Analg. source
Not checked in the free run.
- cites [127] Sami, 1991, Central venous pulsations associated with a falsely low oxygen saturation measured by pulse oximetry, J. Clin. Monit. source
Not checked in the free run.
May combine two claims.
PremisecitationVenous pulsations have been shown to artificially inflate the systolic peak amplitude and interfere with the rising systolic edge of the typical PPG waveform.
“They have been shown to artificially inflate the systolic peak amplitude and interfere with the rising systolic edge of the typical PPG waveform; an important parameter used in cardiovascular health determination [116].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [116] Shelley, 1993, The detection of peripheral venous pulsation using the pulse oximeter as a plethysmograph, J. Clin. Monit. source
Not checked in the free run.
May combine two claims.
PremisecitationNijland et al. confirmed in fetal lambs that reflectance pulse oximetry SpO2 readings were significantly lower than fiberoptic SaO2 values, and when the vessel was coagulated, this difference became negligible.
“Nijland et al. also confirmed these results in fetal lambs, where reflectance pulse oximetry SpO2 readings were significantly lower than fiberoptic SaO2 values. However, when the vessel was coagulated, this difference became negligible [128].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [128] Nijland, 1996, Reflectance pulse oximetry in fetal lambs: Subcutaneous vessels and vasoconstriction affect its reliability, J. Clin. Monit. source
Not checked in the free run.
May combine two claims.
PremisecitationDicrotic notch duration at the finger was not significantly different from measurements under the wrist and forehead, but was different from the earlobe, arm, and wrist.
“Dicrotic notch duration at the finger was not significantly different from measurements under the wrist and forehead, but was different from the earlobe, arm, and wrist.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [95] Hartmann, 2019, Quantitative comparison of photoplethysmographic waveform characteristics: Effect of measurement site, Front. Physiol. source
Not checked in the free run.
May combine two claims.
PremisecitationEliminating venous contributions to PPG by application of contact pressure may eliminate the arterial pulsation in hypotensive patients as well as increase pressure-induced injury risk to the measurement site.
“However, it is possible that this will eliminate the arterial pulsation in hypotensive patients as well as increase pressure-induced injury risk to the measurement site [125].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [125] Shelley, 2005, The effect of venous pulsation on the forehead pulse oximeter wave form as a possible source of error in Spo2 calculation, Anesth. Analg. source
Not checked in the free run.
May combine two claims.
PremisecitationFor silicon photodiodes, the absorption coefficient increases with temperature, so the detectors absorb less light at higher temperatures, inducing an apparent shift in PPG amplitude.
“For silicon photodiodes, the absorption coefficient increases with temperature, and thus the detectors will absorb less light at higher temperatures, inducing an apparent shift in PPG amplitude [161].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [161] Optoelectronics, O. (2020, August 06). Temperature and Bias Effects. Available online: http://www.osioptoelectronics.com/technology-corner/frequently-asked-questions/temperature-bias-effects.aspx.
Not checked in the free run.
May combine two claims.
PremisecitationThe back of the ear and earlobe yielded significantly higher PPG signal amplitude than the hand and postauricular region, but LDV values and comparisons to the fingertips and the rest of the face were not significant.
“The back of the ear and earlobe yielded significantly higher PPG signal amplitude than the hand and postauricular region (i.e., neck behind the ear), but LDV values and comparisons to the fingertips and the rest of the face were not significant.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [144] Tur, 1983, Basal perfusion of the cutaneous microcirculation: Measurements as a function of anatomic position, J. Investig. Dermatol. source
Not checked in the free run.
May combine two claims.
PremisecitationThe fingertips have the thinnest skin, followed by the forearm, dorsal hand, cheek, and forehead.
“Overall, the literature has shown that the fingertips have the thinnest skin, followed by the forearm, dorsal hand, cheek, and forehead [139].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [139] Dahlqvist, 2018, Vibrotactile and thermal perception and its relation to finger skin thickness, Clin. Neurophysiol. Pract. source
Not checked in the free run.
May combine two claims.
PremisecitationAllen et al. reported a 2.6-times increase in PPG amplitude, 93% decrease in LDF flux, and median decrease of 0.089 degrees Celsius following a deep inspiratory gasp of air.
“They reported a 2.6-times increase in PPG amplitude, 93% decrease in LDF flux, and median decrease of 0.089 degrees Celsius [150].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [150] Allen, 2002, Microvascular blood flow and skin temperature changes in the fingers following a deep inspiratory gasp, Physiol. Meas. source
Not checked in the free run.
May combine two claims.
PremisecitationAll these motions range between 0.1 and 20 Hz, which is within the frequency range for heart rate (1-4 Hz).
“All these motions range between 0.1 and 20 Hz, which is within the frequency range for heart rate (1–4 Hz) [163].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [163] Zong, C., and Jafari, R. (2015, January 25–29). Robust heart rate estimation using wrist-based PPG signals in the presence of intense physical activities. Proceedings of the 2015 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Milan, Italy. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationFor all wavelengths of light, the forehead is superior to the cheeks and the whole face for determining heart rate.
“Fallet et al. found via iPPG that for all wavelengths of light, the forehead is superior to the cheeks and the whole face for determining heart rate.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [146] Fallet, S., Moser, V., Braun, F., and Vesin, J.-M. (2016, January 11–14). Imaging photoplethysmography: What are the best locations on the face to estimate heart rate?. Proceedings of the 2016 Computing in Cardiology Conference (CinC), Vancouver, BC, Canada. source
Not checked in the free run.
May combine two claims.
InterpretationreasoningBoth micro-motions and macro-motions cause more significant fluctuations to the PPG signal than sedentary motions.
“Adjusting positions or tapping a finger can be considered a micro-motion, while a macro-motion is performing an exercise such as walking or jogging—both types of motion cause more significant fluctuations to the signal than sedentary motions.”
Quote in the paperNot a substantial claim- Rests on an assumption: Motion artifacts are picked up by PPG sensors and cause fluctuations within the collected signal.
May combine two claims.
PremisecitationThe amount of pressure applied is often underreported and can manipulate the resultant PPG waveform.
“Grabovskis et al. noted that the amount of pressure applied is often underreported and can manipulate the resultant PPG waveform [132].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [132] Grabovskis, 2013, Effect of probe contact pressure on the photoplethysmographic assessment of conduit artery stiffness, J. Biomed. Opt. source
Not checked in the free run.
May combine two claims.
PremisecitationPTT is also similar bilaterally at the ears and finger, while a small difference exists at the toes.
“Additionally, PTT is also similar bilaterally at the ears and finger, while a small difference exists at the toes [91].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [91] Allen, 2002, Age-related changes in peripheral pulse timing characteristics at the ears, fingers and toes, J. Hum. Hypertens. source
Not checked in the free run.
May combine two claims.
InterpretationcitationPhysiological temperature changes due to thermoregulation may impact a signal, and studies detailing changes caused by external temperatures also illustrate significant PPG impact.
“Thus, while physiological temperature changes due to thermoregulation may impact a signal, studies detailing changes caused by external temperatures also illustrate significant PPG impact.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim.
PremisecitationPPG quality is improved with warm temperatures.
“The authors conclude that PPG quality is improved with warm temperatures, yet do not mention the effect of thermoregulatory responses [155].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [154] Khan, 2015, Investigating the effects of temperature on photoplethysmography, IFAC Pap. Online source
Not checked in the free run.
May combine two claims.
InterpretationreasoningMotion artifacts within a PPG signal can skew results for SpO2, heart rate, and other PPG-dependent measurements, creating false alarms or inaccurate readings.
“With the addition of motion artifacts within a PPG signal, results for SpO2, heart rate, and other PPG dependent measurements can be skewed. This can create false alarms or inaccurate readings.”
Quote in the paperSubstantial claim- Rests on an assumption: Motion artifacts are picked up by PPG sensors and cause fluctuations within the collected signal.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThere is a consistent increase in signal amplitude in the wrist PPG compared to the finger PPG.
“Additionally, the authors noted a consistent increase in signal amplitude in the wrist PPG compared to the finger PPG.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [26] Rajala, 2018, Comparison of photoplethysmogram measured from wrist and finger and the effect of measurement location on pulse arrival time, Physiol. Meas. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationThe appropriate amount of pressure is such that the b/a ratio of the SDPPG is equal to 0.70, meaning that the external pressure on the arterial wall is equal to the mean arterial pressure.
“They determine that the appropriate amount of pressure is such that the b/a ratio of the SDPPG is equal to 0.70. This means that the external pressure on the arterial wall is equal to the mean arterial pressure [132].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [132] Grabovskis, 2013, Effect of probe contact pressure on the photoplethysmographic assessment of conduit artery stiffness, J. Biomed. Opt. source
Not checked in the free run.
May combine two claims.
PremisecitationGreen was the best option due to its higher correlation coefficient, but blue is an alternative due to its significantly similar ΔSNR as green.
“It was determined that green was the best option due to its higher correlation coefficient, but blue is an alternative option due to its significantly similar ΔSNR as green.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [164] Lee, J., Matsumura, K., Yamakoshi, K.-I., Rolfe, P., Tanaka, S., and Yamakoshi, T. (2013, January 3–7). Comparison between red, green and blue light reflection photoplethysmography for heart rate monitoring during motion. Proceedings of the 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Osaka, Japan.
Not checked in the free run.
May combine two claims.
PremisecitationRed and IR have deeper penetration depths than blue and green, making them prone to more motion artifacts such as factors that directly affect the DC baseline.
“With their longer wavelengths, red and IR have deeper penetration depths compared to blue and green, making them prone to more motion artifacts such as the factors that directly affect the DC baseline [162,164].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [162] Maeda, 2011, Relationship between measurement site and motion artifacts in wearable reflected photoplethysmography, J. Med. Syst. source
Not checked in the free run.
- cites [164] Lee, J., Matsumura, K., Yamakoshi, K.-I., Rolfe, P., Tanaka, S., and Yamakoshi, T. (2013, January 3–7). Comparison between red, green and blue light reflection photoplethysmography for heart rate monitoring during motion. Proceedings of the 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Osaka, Japan.
Not checked in the free run.
May combine two claims.
PremisecitationThe Apple Watch had the best MAPE (1.14-6.70%), while Fitbit Charge HR and Garmin Forerunner 225 had MAPE of 2.38-16.99% and 7.87-24.38%, respectively.
“In another study, performed by Dooley et al., the comparison was made between the Apple Watch (AP), FB, and Garmin Forerunner 225 (GF), where the AP was considered the best in terms of MAPE (1.14–6.70%), while the FB and the GF had a MAPE of 2.38–16.99% and 7.87–24.38%, respectively [167].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [167] Dooley, 2017, Estimating accuracy at exercise intensities: A comparative study of self-monitoring heart rate and physical activity wearable devices, JMIR mHealth uHealth source
Not checked in the free run.
May combine two claims.
PremisecitationThe Apple Watch (p = 0.84) and Garmin Forerunner 225 (p = 0.35) had no significant difference in HR readings from the ECG during vigorous intensity.
“The AP (p = 0.84) and GF (p = 0.35) had no significant difference in HR readings from the ECG during vigorous intensity.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [167] Dooley, 2017, Estimating accuracy at exercise intensities: A comparative study of self-monitoring heart rate and physical activity wearable devices, JMIR mHealth uHealth source
Not checked in the free run.
May combine two claims.
PremisecitationSukor et al.'s algorithm had an accuracy of 83 ± 11%, a sensitivity of 89 ± 10%, and a specificity of 77 ± 19%.
“They found their algorithm to have an accuracy of 83 ± 11%, a sensitivity of 89 ± 10 %, and a specificity of 77 ± 19%.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [169] Sukor, 2011, Signal quality measures for pulse oximetry through waveform morphology analysis, Physiol. Meas. source
Not checked in the free run.
May combine two claims.
PremisecitationWhen comparing their algorithm to an ECG, Sukor et al. were able to lower the error in heart rate readings to 0.49 ± 0.66 bpm, whereas without the algorithm, the error was 7.23 ± 5.78 bpm.
“When comparing their algorithm to an ECG, they were able to lower the error in heart rate readings to 0.49 ± 0.66 bpm, whereas without the algorithm, the error was 7.23 ± 5.78 bpm.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [169] Sukor, 2011, Signal quality measures for pulse oximetry through waveform morphology analysis, Physiol. Meas. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.
InterpretationreasoningDepending on the exercise, each tracker will have a different accuracy, which could be attributed to the different noise-canceling algorithms used on board.
“This shows that, depending on the exercise, each tracker will have a different accuracy, which could be attributed to the different noise-canceling algorithms used on board.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationThe Apple Watch 4 had the highest accuracy with an MAE of 2.7 BPM at rest and 4.6 BPM during the walking phase.
“This study also showed that the Apple Watch 4 had the highest accuracy with an MAE of 2.7 BPM at rest and 4.6 BPM during the walking phase,”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [46] Bent, 2020, Investigating sources of inaccuracy in wearable optical heart rate sensors, NPJ Digit. Med. source
Not checked in the free run.
May combine two claims.
PremisecitationDuring low-intensity events, the Basis Peak provided accurate heart rate readings (r = 0.84, p < 0.05), while the Fitbit Charge HR had a moderate correlation (r = 0.73, p < 0.05).
“During the low-intensity events, the BP provided accurate heart rate readings (r = 0.84, p < 0.05), while the FB had a moderate correlation (r = 0.73, p < 0.05).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [165] Jo, 2016, Validation of biofeedback wearables for photoplethysmographic heart rate tracking, J. Sports Sci. Med.
Not checked in the free run.
May combine two claims.
PremisecitationWhen going from a resting phase to a low-intensity walking phase, the correlation coefficient went down but slowly rebounded when moved into a more intense walking phase.
“When going from a resting phase to a low-intensity walking phase, they found that the correlation coefficient went down but had slowly rebounded when moved into a more intense walking phase.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [166] Stahl, 2016, How accurate are the wrist-based heart rate monitors during walking and running activities? Are they accurate enough?, BMJ Open Sport Exerc. Med. source
Not checked in the free run.
May combine two claims.
PremisecitationHigh-intensity events caused the Basis Peak to have a weaker correlation (r = 0.77, p < 0.05), whereas the Fitbit Charge HR had the weakest correlation (r = 0.58, p < 0.05).
“High-intensity events caused the BP to have a weaker correlation (r = 0.77, p < 0.05), whereas the FB had the weakest correlation (r = 0.58, p < 0.05).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [165] Jo, 2016, Validation of biofeedback wearables for photoplethysmographic heart rate tracking, J. Sports Sci. Med.
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenAside from location and wavelength, the intensity of motion plays a role in how accurate heart rate monitoring is.
“Aside from location and wavelength, the intensity of motion also plays a role in how accurate heart rate monitoring is.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationThe mean absolute percentage error (MAPE) of the Basis Peak and Fitbit Charge HR were 3.6% and 6.2%, respectively.
“These results align with an assessment done by Stahl et al., where they measured the mean absolute percentage error values (MAPE) of BP and the FB to be 3.6% and 6.2%, respectively [166].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [166] Stahl, 2016, How accurate are the wrist-based heart rate monitors during walking and running activities? Are they accurate enough?, BMJ Open Sport Exerc. Med. source
Not checked in the free run.
May combine two claims.
PremisereasoningAccelerometers and gyroscopes are the standards, but when trying to measure micro motions such as finger tapping, they do not always pick up on the motion because the wrist remains still.
“Accelerometers and gyroscopes are the standards, but when trying to measure micro motions such as finger tapping, accelerometers and gyroscopes do not always pick up on the motion because the wrist remains still.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [179] Casson, 2016, Gyroscope vs. accelerometer measurements of motion from wrist PPG during physical exercise, ICT Express source
Not checked in the free run.
- cites [180] Lee, 2018, Motion artifact cancellation in wearable photoplethysmography using gyroscope, IEEE Sens. J. source
Not checked in the free run.
May combine two claims.
PremisecitationChong et al. achieved an accuracy of 93.7% during a daily motion experiment and saw a significant reduction in error for SpO2 and heart rate measurements.
“They were able to achieve an accuracy of 93.7% during a daily motion experiment, and they saw a significant reduction in error for SpO2 and heart rate measurements.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [170] Chong, 2014, Photoplethysmograph signal reconstruction based on a novel hybrid motion artifact detection–reduction approach. Part I: Motion and noise artifact detection, Ann. Biomed. Eng. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
InterpretationreasoningThe discussed detection methods are all real-time ways to detect and eliminate sections of acquired PPG signals that contain corrupted areas due to motion artifacts.
“The discussed detection methods are all real-time ways to detect and eliminate sections of acquired PPG signals that contain corrupted areas due to motion artifacts.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationSelvaraj et al. used kurtosis and Shannon Entropy to determine motion artifacts and achieved an accuracy of 88.8% in a laboratory setting.
“Kurtosis and Shannon Entropy were the measurements used by Selvaraj et al. to determine motion artifacts, and they were able to perform an accuracy of 88.8% in a laboratory setting [174].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [174] Selvaraj, N., Mendelson, Y., Shelley, K.H., Silverman, D.G., and Chon, K.H. (September, January 30). Statistical approach for the detection of motion/noise artifacts in Photoplethysmogram. Proceedings of the 2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Boston, MA, USA. source
Not checked in the free run.
May combine two claims.
InterpretationreasoningWith the use of adaptive filters and secondary sensors, motion artifacts can be identified and eliminated, or identified and reconstructed to restore the original PPG signal void of all motion artifacts.
“With the use of adaptive filters and secondary sensors, motion artifacts can be identified and eliminated, or they can be identified and reconstructed to restore the original PPG signal, void of all motion artifacts.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationKrishnan et al. discovered that skew and kurtosis measurements in the time domain were higher when there was corruption, and frequency-domain kurtosis was smaller when there was corruption.
“They discovered skew and kurtosis measurements in the time domain to be higher when there was corruption, and the frequency-domain kurtosis was smaller when there was corruption.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [173] Krishnan, R., Natarajan, B., and Warren, S. (April, January 31). Analysis and detection of motion artifact in photoplethysmographic data using higher order statistics. Proceedings of the 2008 IEEE International Conference on Acoustics, Speech and Signal Processing, Las Vegas, NV, USA. source
Not checked in the free run.
May combine two claims. The quote only partly states this claim.
PremisecitationCoucerio et al. achieved a specificity of 92.7% and a sensitivity of 82.7% using time-domain and period domain analysis of 26 features.
“They were able to achieve a specificity of 92.7% and a sensitivity of 82.7%.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [172] Couceiro, R., Carvalho, P., Paiva, R.P., Henriques, J., and Muehlsteff, J. (September, January 28). Detection of motion artifacts in photoplethysmographic signals based on time and period domain analysis. Proceedings of the 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society, San Diego, CA, USA. source
Not checked in the free run.
May combine two claims.
PremisecitationKarlen et al. achieved a positive predictive value of 96.68% and a sensitivity of 98.93% for mobile PPG measurements.
“They were able to achieve a positive predictive value of 96.68% and a sensitivity of 98.93%.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [171] Karlen, W., Ansermino, J.M., and Dumont, G. (September, January 28). Adaptive pulse segmentation and artifact detection in photoplethysmography for mobile applications. Proceedings of the 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society, San Diego, CA, USA. source
Not checked in the free run.
May combine two claims.
PremisecitationBoth the piezoelectric transducer and another wavelength of light can pick up finer motions such as finger tapping.
“Both the piezoelectric transducer and another wavelength of light can pick up these finer motions [181,182].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [181] Lee, 2019, Motion artifact identification and removal from wearable reflectance photoplethysmography using piezoelectric transducer, IEEE Sens. J. source
Not checked in the free run.
- cites [182] Zhang, Y., Song, S., Vullings, R., Biswas, D., Simões-Capela, N., van Helleputte, N., van Hoof, C., and Groenendaal, W. (2019). Motion artifact reduction for wrist-worn photoplethysmograph sensors based on different wavelengths. Sensors, 19. source
Not checked in the free run.
May combine two claims.
PremisecitationWithout force regulation, the coefficient of variation between five posture stages was 50.9%; the addition of the force regulator dropped the coefficient of variation to 1.8%.
“Without force regulation, the coefficient of variation between five posture stages was calculated at 50.9%. The addition of the force regulator dropped the coefficient of variation to 1.8%.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [212] Sim, 2018, A contact-force regulated photoplethysmography (PPG) platform, AIP Adv. source
Not checked in the free run.
May combine two claims.
PremisecitationWith an applied pressure range of 0-15 kPa, there was a variation of over 300 percent in the b/a ratio; when an optimal pressure was reached, the variation coefficient dropped to less than 5 percent.
“Grabovskis et al. reported with an applied pressure range of 0–15 kPa, there was a variation of over 300 percent in the b/a ratio [209]. When an optimal pressure was reached, the variation coefficient dropped to less than 5 percent.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [209] Grabovskis, A., Marcinkevics, Z., Rubenis, O., Rubins, U., and Lusa, V. (2012, January 16–19). Photoplethysmography system for blood pulsation detection in unloaded artery conditions. Proceedings of the Biophotonics: Photonic Solutions for Better Health Care III, Brussels, Belgium. source
Not checked in the free run.
May combine two claims.
PremisecitationFor shorter wavelengths, the occlusion pressure was lower, as more superficial arterioles with low mean arterial pressure were in the light path at these wavelengths.
“For shorter wavelengths, the occlusion pressure was lower, as more superficial arterioles with low MAP were in the light path at these wavelengths.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [210] Spigulis, J., Gailite, L., Erts, R., and Lihachev, A. (2007, January 17–19). Contact probe pressure effects in skin multi-spectral photoplethysmography. Proceedings of the European Conference on Biomedical Optics, Munich, Germany. source
Not checked in the free run.
May combine two claims.
InterpretationcitationThe change in these fundamental features can affect other measurements such as the b/a ratio, the relationship between the frequency response of a PPG signal and a blood pressure signal, or the pulse transit time (PTT).
“The change in these fundamental features can affect other measurements such as the b/a ratio, derived from the a and b peaks of the second derivative of a PPG signal, used to characterize arterial stiffness [132], the relationship between the frequency response of a PPG signal and a blood pressure signal [207], or the pulse transit time (PTT) which is used as an indicator for fluctuating stiffness”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [132] Grabovskis, 2013, Effect of probe contact pressure on the photoplethysmographic assessment of conduit artery stiffness, J. Biomed. Opt. source
Not checked in the free run.
- cites [207] Hsiu, 2011, Effects of different contacting pressure on the transfer function between finger photoplethysmographic and radial blood pressure waveforms, Proc. Inst. Mech. Eng. Part H J. Eng. Med. source
Not checked in the free run.
May combine two claims.
InterpretationreasoningAs external pressure exceeds the point where transmural pressure is no longer zero, the vessel begins to occlude, and the AC amplitude begins to decrease until there is no longer a signal where the vessel is occluded.
“As the external pressure exceeds the point where the transmural pressure is no longer zero, the vessel begins to occlude, and the AC amplitude begins to decrease until there is no longer a signal where the vessel is occluded.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationAt 60 mmHg, there were the greatest R2 values for four out of five harmonics, and with different applied pressures, the R2 values ranged from 0.13 to 0.77 for the first harmonic.
“At 60 mmHg, there were the greatest R2 values for four out of five harmonics, and with the different applied pressures, the R2 values ranged from 0.13 to 0.77 for the first harmonic.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [207] Hsiu, 2011, Effects of different contacting pressure on the transfer function between finger photoplethysmographic and radial blood pressure waveforms, Proc. Inst. Mech. Eng. Part H J. Eng. Med. source
Not checked in the free run.
May combine two claims.
PremisecitationWith an applied force at the finger of 0.2-1.8 N, there was an increase in AC amplitude and AC/DC ratio until the parameters reached a peak and then decreased, while the pulse area decreased, and the DC offset increased.
“As reported by Teng and Zhang, with an applied force at the finger of 0.2–1.8 N, there was an increase in AC amplitude and AC/DC ratio until the parameters reached a peak and then decreased, while the pulse area decreased, and the DC offset increased [206].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [206] Teng, 2004, The effect of contacting force on photoplethysmographic signals, Physiol. Meas. source
Not checked in the free run.
May combine two claims.
PremisecitationFor the PTT calculated at 90% of the pulse amplitude, there was a similar increasing trend for the PTT but without a significant difference (p = 0.107), and the PTT leveled out to a constant value for a transmural pressure of around 0.1 N.
“Lastly, for the PTT calculated at 90% of the pulse amplitude, there was a similar increasing trend for the PTT but without a significant difference (p = 0.107). The PTT leveled out to a constant value for a transmural pressure of around 0.1 N.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [208] Teng, 2006, The effect of applied sensor contact force on pulse transit time, Physiol. Meas. source
Not checked in the free run.
May combine two claims.
PremisecitationRhee et al.'s finger-ring system applied a skin pressure of about 75 mmHg and, due to its compliance, held steady the applied pressure while the finger was in different positions.
“This band would apply a skin pressure of about 75 mmHg and, due to its compliance, it would hold steady the applied pressure while the finger was in different positions.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [213] Rhee, 2001, Artifact-resistant power-efficient design of finger-ring plethysmographic sensors, IEEE Trans. Biomed. Eng. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Interpretationnone givenVarious attempts are made in PPG sensing circuitry to effectively reject DC photocurrent.
“Various attempts are also made in the PPG sensing circuitry level to effectively reject DC photocurrent.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationPPG signals are low-amplitude signals and have a normal pulse frequency within the range of 0.5 to 4 Hz.
“PPG signals are low-amplitude signals and have a normal pulse frequency within the range of 0.5 to 4 Hz [195,196].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [195] Wong, 2005, A near-infrared heart rate measurement IC with very low cutoff frequency using current steering technique, IEEE Trans. Circuits Syst. I Regul. Pap. source
Not checked in the free run.
- cites [196] Moraes, J.L., Rocha, M.X., Vasconcelos, G.G., Vasconcelos Filho, J.E., de Albuquerque, V.H.C., and Alexandria, A.R. (2018). Advances in photopletysmography signal analysis for biomedical applications. Sensors, 18. source
Not checked in the free run.
May combine two claims.
PremisecitationWith the change in transmural pressure, there is a related compliance change in the affected vessel, and the vessel reaches maximum compliance when the transmural pressure is zero.
“With the change in transmural pressure, there is also a related compliance change in the affected vessel. The vessel reaches maximum compliance when the transmural pressure is zero [205].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [205] Forster, 1986, Oscillometric determination of diastolic, mean and systolic blood pressure—A numerical model, J. Biomech. Eng. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
InterpretationreasoningWith a low bandwidth of around 5 Hz, PPG signals can be extracted from high frequency noise using a low-pass filter and from low frequency noise using a high-pass filter.
“With a low bandwidth of around 5 Hz, PPG signals can be extracted out from high frequency noise using a low-pass filter and from low frequency noise using a high-pass filter (e.g., to reject 60 Hz room lights and to reject DC sunlight).”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationWang et al.'s ear-worn sensor method was effective in shielding from noise due to ambient light, as the DC level of the PPG signal was less than 2 nA when the LEDs were switched off.
“This method was found to be effective in shielding from noise due to ambient light, as the DC level of the PPG signal was considerably low (measured to be less than 2 nA) when the LEDs were switched off.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [200] Wang, 2007, Multichannel reflective PPG earpiece sensor with passive motion cancellation, IEEE Trans. Biomed. Circuits Syst. source
Not checked in the free run.
May combine two claims.
InterpretationcitationSeveral characteristics of the PPG waveform are affected by changes in contact force, including the AC amplitude, the DC offset amplitude, the ratio of AC/DC amplitudes, and the normalized pulse area.
“Several characteristics of the PPG waveform are affected by changes in contact force, including; the AC amplitude, the DC offset amplitude, the ratio of AC/DC amplitudes, and the normalized pulse area.”
Quote in the paperSubstantial claim
The paper relies on this without support.
The findings of the cited studies on specific PPG devices and populations are applicable to PPG-based continuous cardiovascular monitoring devices in general.
“Systems aiming to derive such complex parameters do not always account for realistic sources of noise, as testing is performed within controlled parameter spaces.”
The paper relies on this without support.
The Fitzpatrick scale is a valid proxy for skin pigmentation relevant to PPG light-tissue interactions.
“The most common way to characterize skin tone is via the Fitzpatrick Scale [37].”
The paper relies on this without support.
The physiological changes associated with obesity measured in the cited studies are present in the target population of PPG users.
“The variation of BMI across individuals is thus a potential source of variation for PPG measurements.”
The paper relies on this without support.
Gender is binary.
“While the wide spectrum of ages can make it a difficult parameter to incorporate into models, gender is binary, making it easier to incorporate.”
The paper relies on this without support.
PWV is highly dependent on vascular stiffness and is observable through PPG.
“PWV is highly dependent on vascular stiffness and is observable through PPG.”
The paper relies on this without support.
Pulse pressure is dependent on both arterial stiffness and cardiac output.
“Pulse pressure is dependent on both arterial stiffness and cardiac output.”
The paper relies on this without support.
Motion artifacts are picked up by PPG sensors and cause fluctuations within the collected signal.
“Within all settings, motion artifacts are picked up by the sensor and cause fluctuations within the collected signal.”
The paper relies on this without support.
Ambient light intensity is generally multiple orders of magnitude larger than the pulsatile (AC) component of the PPG, which can lead to signal saturation.
“The ambient light intensity can be a zero frequency (e.g., DC), such as sunlight, or at a variable frequency, as from room lights (e.g., 60 Hz in the US) and is generally multiple orders of magnitude larger than the pulsatile (AC) component of the PPG, which can lead to signal saturation.”
The paper relies on this without support.
Only a limited number of RCIM PPG devices have FDA approval.
“However, only a limited number of RCIM PPG devices have FDA approval.”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Technological limitations are primarily sensor-related (LED wavelength, light intensity, placement), while algorithmic limitations stem from biased datasets, poor calibration, and underrepresentation of darker skin in training cohorts.
“Technological limitations are primarily sensor-related - such as LED wavelength, light intensity, and placement - while algorithmic limitations stem from biased datasets, poor calibration, and underrepresentation of darker skin in training cohorts.”
From Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring (2025) · cites this paper as [28]
Section 2.1 (Skin Tone; page unavailable) links PPG accuracy to LED wavelength and source–detector placement, and reports greater error for darker skin tones. The Introduction also acknowledges hardware and software sources of noise. However, the cited passages do not establish that technological limitations are primarily sensor-related or that algorithmic limitations stem from biased training datasets, poor calibration, or underrepresentation of darker skin in training cohorts; the study with fewer dark-skinned participants is described as a device validation, not a training cohort. (LLM, conf=0.86)
Experimental — accuracy not yet measuredStatistically reliable patterns tied to phenotype, particularly adiposity, and to a lesser extent skin pigmentation, were evident and align with optical and physiological mechanisms described in prior work.
“Nonetheless, statistically reliable patterns tied to phenotype, particularly adiposity, and, to a lesser extent, skin pigmentation, were evident and align with the optical and physiological mechanisms described in prior work [13,43].”
From PPG-Based Heart Rate Accuracy in Hispanic Adults with Fitzpatrick III–V Skin Tones: An Evaluation of Body Composition and Skin-Tone Effects (2026) · cites this paper as [43]
Melanin attenuates the wavelength of incident light and limits penetration to subcutaneous tissue because it is highly light-absorbing.
“Melanin attenuates the wavelength of the incident light and limits the penetration to the subcutaneous tissue because is high light-absorbing [80,101].”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [101]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Pulse wave velocity is said to both decrease and increase with age
MinorThe text says pulse wave velocity (PWV) decreases with age, but later says it increases with age. Table 3 also lists an increase.
“PWV, which has been shown to decrease as a function of age.”
“an increase in pulse wave velocity with age is related to a diminished dicrotic notch.”
Hartmann study's mean age conflicts with its stated age range
MinorThe same 36-person Hartmann study is described as having ages from 33 to 58 and a mean age of 33. If 33 is the minimum and at least one participant is 58, the mean must be greater than 33-even when rounded to the nearest year.
“this work featured 36 individuals ranging in age from only 33 to 58.”
“Hartmann et al. performed a similar study in 36 healthy subjects (12 male, 24 female, mean age of 33)”
Photodiode absorption is said to increase while absorbed light decreases
MinorThe text says the absorption coefficient increases with temperature, then concludes that the detectors absorb less light at higher temperatures. Those statements point in opposite directions under the stated explanation.
“For silicon photodiodes, the absorption coefficient increases with temperature”
“the detectors will absorb less light at higher temperatures”
Cites
1 retracted
Uçar, M. K., Bozkurt, M. R., Bilgin, C., & Polat, K. (2018) · Neural Computing and Applications · doi:10.1007/s00521-016-2365-x · Cited by 106 · see citing works
RetractedHeart Disease and Stroke Statistics—2023 Update: A Report From the American Heart Association
Tsao, C. W., Aday, A. W., Almarzooq, Z. I., Anderson, C. A. M., Arora, P., Avery, C. L., Baker-Smith, C. M., Beaton, A. Z., Boehme, A. K., Buxton, A. E., et al. (2023) · Circulation · doi:10.1161/cir.0000000000001123 · Cited by 8,300 · see citing works
General Cardiovascular Risk Profile for Use in Primary Care: The Framingham Heart Study
D’Agostino, R. B., Vasan, R. S., Pencina, M. J., Wolf, P. A., Cobain, M., Massaro, J. M., & Kannel, W. B. (2008) · Circulation · doi:10.1161/circulationaha.107.699579 · Cited by 7,491 · see citing works
Heart Disease and Stroke Statistics—2011 Update: A Report From the American Heart Association
Roger, V. L., Go, A. S., Lloyd-Jones, D. M., Adams, R. J., Berry, J. D., Brown, T. M., Carnethon, M. R., Dai, S., de Simone, G., Ford, E. S., et al. (2011) · Circulation · doi:10.1161/cir.0b013e3182009701 · Cited by 6,149 · see citing works
Rehospitalizations among Patients in the Medicare Fee-for-Service Program
Jencks, S. F., Williams, M. V., & Coleman, E. A. (2009) · New England Journal of Medicine · doi:10.1056/nejmsa0803563 · Cited by 5,164 · see citing works
Impact of Atrial Fibrillation on the Risk of Death: The Framingham Heart Study
Benjamin, E. J., Wolf, P. A., D’Agostino, R. B., Silbershatz, H., Kannel, W. B., & Levy, D. (1998) · Circulation · doi:10.1161/01.cir.98.10.946 · Cited by 4,575 · see citing works
The validity and practicality of sun-reactive skin types I through VI
Fitzpatrick, T. B. (1988) · Archives of Dermatology · doi:10.1001/archderm.1988.01670060015008 · Cited by 4,339 · see citing works
Optical properties of biological tissues: A review
Jacques, S. L. (2013) · Physics in Medicine and Biology · doi:10.1088/0031-9155/58/11/r37 · Cited by 4,008 · see citing works
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967 · see citing works
Poirier, P., Giles, T. D., Bray, G. A., Hong, Y., Stern, J. S., Pi-Sunyer, F. X., & Eckel, R. H. (2006) · Circulation · doi:10.1161/circulationaha.106.171016 · Cited by 3,165 · see citing works
Absorption spectrum (380–700 nm) of pure water II Integrating cavity measurements
Pope, R. M., & Fry, E. S. (1997) · Applied Optics · doi:10.1364/ao.36.008710 · Cited by 2,303 · see citing works
Body Mass Index: Obesity, BMI, and Health A Critical Review
Nuttall, F. Q. (2015) · Nutrition Today · doi:10.1097/nt.0000000000000092 · Cited by 2,284 · see citing works
D'Orazio, J., Jarrett, S., Amaro-Ortiz, A., & Scott, T. (2013) · International Journal of Molecular Sciences · doi:10.3390/ijms140612222 · Cited by 1,984 · see citing works
Robust pulse rate from chrominance-based rPPG
de Haan, G., & Jeanne, V. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2266196 · Cited by 1,378 · see citing works
Gender differences in the regulation of blood pressure
Reckelhoff, J. F. (2001) · Hypertension · doi:10.1161/01.hyp.37.5.1199 · Cited by 1,268 · see citing works
Algorithmic principles of remote PPG
Wang, W., den Brinker, A. C., Stuijk, S., & de Haan, G. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2609282 · Cited by 1,234 · see citing works
On the Analysis of Fingertip Photoplethysmogram Signals
Elgendi, M. (2012) · Current Cardiology Reviews · doi:10.2174/157340312801215782 · Cited by 1,206 · see citing works
Ghamari, M. (2018) · International Journal of Biosensors & Bioelectronics · doi:10.15406/ijbsbe.2018.04.00125 · Cited by 945 · see citing works
Toward ubiquitous blood pressure monitoring via pulse transit time: Theory and practice
Mukkamala, R., Hahn, J.-O., Inan, O. T., Mestha, L. K., Kim, C.-S., Toreyin, H., & Kyal, S. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2441951 · Cited by 935 · see citing works
Wearable photoplethysmographic sensors—past and present
Tamura, T., Maeda, Y., Sekine, M., & Yoshida, M. (2014) · Electronics · doi:10.3390/electronics3020282 · Cited by 912 · see citing works
Sandby-Møller, J., Poulsen, T., & Wulf, H. C. (2003) · Acta Dermato-Venereologica · doi:10.1080/00015550310015419 · Cited by 737 · see citing works
Investigating sources of inaccuracy in wearable optical heart rate sensors
Bent, B., Goldstein, B. A., Kibbe, W. A., & Dunn, J. P. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0226-6 · Cited by 673 · cited 4× · see citing works
“published an experimental analysis of error observed in optical heart rate sensors manufactured by Apple (Apple Watch 4), Fitbit (Fitbit Charge 2), Garmin (Garmin Vivosmart 3), Xiaomi (Xiaomi Miband 3), Empatica (Empatica E4), and Biovotion (Biovation Everion) as a function of skin tone [46].”
BackgroundMethodInfluentialThe influence of age and sex on skin thickness, skin collagen and density
SHUSTER, S., BLACK, M. M., & McVITIE, E. (1975) · British Journal of Dermatology · doi:10.1111/j.1365-2133.1975.tb05113.x · Cited by 640 · see citing works
Shcherbina, A., Mattsson, C., Waggott, D., Salisbury, H., Christle, J., Hastie, T., Wheeler, M., & Ashley, E. (2017) · Journal of Personalized Medicine · doi:10.3390/jpm7020003 · Cited by 625 · see citing works
The use of photoplethysmography for assessing hypertension
Elgendi, M., Fletcher, R., Liang, Y., Howard, N., Lovell, N. H., Abbott, D., Lim, K., & Ward, R. (2019) · npj Digital Medicine · doi:10.1038/s41746-019-0136-7 · Cited by 625 · see citing works
Takazawa, K., Tanaka, N., Fujita, M., Matsuoka, O., Saiki, T., Aikawa, M., Tamura, S., & Ibukiyama, C. (1998) · Hypertension · doi:10.1161/01.hyp.32.2.365 · Cited by 589 · see citing works
Antelmi, I., De Paula, R. S., Shinzato, A. R., Peres, C. A., Mansur, A. J., & Grupi, C. J. (2004) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2003.09.065 · Cited by 555 · see citing works
Optical absorption of hemoglobin
Prahl, S. A. (1999) · Medical Entomology and Zoology · Cited by 552 · see citing works
Characteristics of the Aging Skin
Farage, M. A., Miller, K. W., Elsner, P., & Maibach, H. I. (2013) · Advances in Wound Care · doi:10.1089/wound.2011.0356 · Cited by 535 · see citing works
Kochanek, K. D., Murphy, S. L., Xu, J., & Arias, E. (2019) · PubMed · Cited by 454 · see citing works
de Jongh, R. T., Serné, E. H., IJzerman, R. G., de Vries, G., & Stehouwer, C. D. A. (2004) · Circulation · doi:10.1161/01.cir.0000129772.26647.6f · Cited by 407 · see citing works
Motion Artifact Reduction in Photoplethysmography Using Independent Component Analysis
Kim, B., & Yoo, S. K. (2006) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.869784 · Cited by 400 · see citing works
Zhang, Z. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2406332 · Cited by 388 · see citing works
Charlton, P. H., Bonnici, T., Tarassenko, L., Clifton, D. A., Beale, R., & Watkinson, P. J. (2016) · Physiological Measurement · doi:10.1088/0967-3334/37/4/610 · Cited by 366 · see citing works
Spontaneous rhythms in physiological control systems
HYNDMAN, B. W., KITNEY, R. I., & SAYERS, B. M. (1971) · Nature · doi:10.1038/233339a0 · Cited by 361 · see citing works
Skin thickness measurement by pulsed ultrasound: its reproducibility, validation and variability
Tan, C. Y., Statham, B. N., Marks, R., & Payne, P. A. (1982) · British Journal of Dermatology · doi:10.1111/j.1365-2133.1982.tb14702.x · Cited by 346 · see citing works
A Novel Approach for Motion Artifact Reduction in PPG Signals Based on AS-LMS Adaptive Filter
Ram, M. R., Madhav, K. V., Krishna, E. H., Komalla, N. R., & Reddy, K. A. (2011) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2011.2175832 · Cited by 345 · see citing works
Burton, A. C. (1939) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1939.127.3.437 · Cited by 333 · see citing works
Monte-Moreno, E. (2011) · Artificial Intelligence in Medicine · doi:10.1016/j.artmed.2011.05.001 · Cited by 311 · see citing works
Pheomelanin as well as Eumelanin Is Present in Human Epidermis
Thody, A. J., Higgins, E. M., Wakamatsu, K., Ito, S., Burchill, S. A., & Marks, J. M. (1991) · Journal of Investigative Dermatology · doi:10.1111/1523-1747.ep12480680 · Cited by 311 · see citing works
Advances in Photopletysmography Signal Analysis for Biomedical Applications
Moraes, J. L., Rocha, M. X., Vasconcelos, G. G., Vasconcelos Filho, J. E., De Albuquerque, V. H. C., & Alexandria, A. R. (2018) · Sensors · doi:10.3390/s18061894 · Cited by 292 · see citing works
Alian, A. A., & Shelley, K. H. (2014) · Best Practice & Research Clinical Anaesthesiology · doi:10.1016/j.bpa.2014.08.006 · Cited by 289 · see citing works
Gibney, M. A., Arce, C. H., Byron, K. J., & Hirsch, L. J. (2010) · Current Medical Research and Opinion · doi:10.1185/03007995.2010.481203 · Cited by 283 · see citing works
Stahl, S. E., An, H.-S., Dinkel, D. M., Noble, J. M., & Lee, J.-M. (2016) · BMJ Open Sport & Exercise Medicine · doi:10.1136/bmjsem-2015-000106 · Cited by 276 · see citing works
Oltulu, P., İnce, B., Kökbudak, N., Fındık, S., & Kılınç, F. (2018) · Turkish Journal of Plastic Surgery · doi:10.4103/tjps.tjps_2_17 · Cited by 276 · see citing works
The thickness of the epidermis
Whitton, J. T., & Everall, J. D. (1973) · British Journal of Dermatology · doi:10.1111/j.1365-2133.1973.tb03007.x · Cited by 271 · see citing works
Two-Stage Approach for Detection and Reduction of Motion Artifacts in Photoplethysmographic Data
Krishnan, R., Natarajan, B., & Warren, S. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2009.2039568 · Cited by 263 · see citing works
Current progress of photoplethysmography and SPO2 for health monitoring
TAMURA, T. (2019) · Biomedical Engineering Letters · doi:10.1007/s13534-019-00097-w · Cited by 260 · see citing works
Noninvasive pulse oximetry utilizing skin reflectance photoplethysmography
Mendelson, Y., & Ochs, B. D. (1988) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.7286 · Cited by 251 · see citing works
Dooley, E. E., Golaszewski, N. M., & Bartholomew, J. B. (2017) · JMIR mHealth and uHealth · doi:10.2196/mhealth.7043 · Cited by 242 · see citing works
Teng, X. F., & Zhang, Y. (2004) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2003.1280811 · Cited by 241 · see citing works
Laurent, A., Mistretta, F., Bottigioli, D., Dahel, K., Goujon, C., Nicolas, J. F., Hennino, A., & Laurent, P. E. (2007) · Vaccine · doi:10.1016/j.vaccine.2007.05.046 · Cited by 235 · see citing works
Artifact-resistant power-efficient design of finger-ring plethysmographic sensors
Rhee, S., Yang, B.-H., & Asada, H. H. (2001) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.930904 · Cited by 235 · see citing works
Wilkoff, B. L., Auricchio, A., Brugada, J., Cowie, M., Ellenbogen, K. A., Gillis, A. M., Hayes, D. L., Howlett, J. G., Kautzner, J., Love, C. J., et al. (2008) · Europace · doi:10.1093/europace/eun122 · Cited by 231 · see citing works
A Motion-Tolerant Adaptive Algorithm for Wearable Photoplethysmographic Biosensors
Yousefi, R., Nourani, M., Ostadabbas, S., & Panahi, I. M. S. (2013) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2013.2264358 · Cited by 229 · see citing works
Photoplethysmographic derivation of respiratory rate: A review of relevant physiology
Meredith, D. J., Clifton, D., Charlton, P., Brooks, J., Pugh, C. W., & Tarassenko, L. (2012) · Journal of Medical Engineering & Technology · doi:10.3109/03091902.2011.638965 · Cited by 226 · see citing works
A new, short-recorded photoplethysmogram dataset for blood pressure monitoring in China
Liang, Y., Chen, Z., Liu, G., & Elgendi, M. (2018) · Scientific Data · doi:10.1038/sdata.2018.20 · Cited by 215 · see citing works
Maeda, Y., Sekine, M., & Tamura, T. (2011) · Journal of Medical Systems · doi:10.1007/s10916-010-9505-0 · Cited by 203 · see citing works
Allen, J., & Murray, A. (2003) · Physiological Measurement · doi:10.1088/0967-3334/24/2/306 · Cited by 202 · see citing works
Lee, J., Matsumura, K., Yamakoshi, K.-I., Rolfe, P., Tanaka, S., & Yamakoshi, T. (2013) · 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2013.6609852 · Cited by 202 · see citing works
<i>In Vivo</i> Studies of the Evolution of Physical Properties of the Human Skin with Age
Leveque, J. L., Corcuff, P., Rigal, J. D., & Agache, P. (1984) · International Journal of Dermatology · doi:10.1111/j.1365-4362.1984.tb04061.x · Cited by 199 · see citing works
Gender differences in skin: A review of the literature
Dao, H., & Kazin, R. A. (2007) · Gender Medicine · doi:10.1016/s1550-8579(07)80061-1 · Cited by 199 · see citing works
Gender differences in large artery stiffness pre-and post puberty
Ahimastos, A. A., Formosa, M., Dart, A. M., & Kingwell, B. A. (2003) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2003-030722 · Cited by 196 · see citing works
Effect of dermal thickness, tissue composition, and body site on skin biomechanical properties
Smalls, L. K., Wickett, R. R., & Visscher, M. O. (2006) · Skin Research and Technology · doi:10.1111/j.0909-725x.2006.00135.x · Cited by 192 · see citing works
Predicting Arterial Stiffness From the Digital Volume Pulse Waveform
Alty, S. R., Angarita-Jaimes, N., Millasseau, S., & Chowienczyk, P. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.897805 · Cited by 192 · see citing works
Skin changes in the obese patient
Hirt, P. A., Castillo, D. E., Yosipovitch, G., & Keri, J. E. (2019) · Journal of the American Academy of Dermatology · doi:10.1016/j.jaad.2018.12.070 · Cited by 182 · see citing works
Age-related changes in peripheral pulse timing characteristics at the ears, fingers and toes
Allen, J., & Murray, A. (2002) · Journal of Human Hypertension · doi:10.1038/sj.jhh.1001478 · Cited by 179 · see citing works
Portable out-of-hospital electrocardiography: A review of current technologies
Bansal, A., & Joshi, R. (2018) · Journal of Arrhythmia · doi:10.1002/joa3.12035 · Cited by 174 · cited 1× · see citing works
“The blood pressure cuff, ECG patch, and Holter monitor are also not often used for long-term remote monitoring [24].”
BackgroundHartmann, V., Liu, H., Chen, F., Qiu, Q., Hughes, S., & Zheng, D. (2019) · Frontiers in Physiology · doi:10.3389/fphys.2019.00198 · Cited by 173 · see citing works
The effect of contacting force on photoplethysmographic signals
Teng, X. F., & Zhang, Y. T. (2004) · Physiological Measurement · doi:10.1088/0967-3334/25/5/020 · Cited by 171 · see citing works
Signal quality measures for pulse oximetry through waveform morphology analysis
Sukor, J. A., Redmond, S. J., & Lovell, N. H. (2011) · Physiological Measurement · doi:10.1088/0967-3334/32/3/008 · Cited by 166 · see citing works
Potential errors in pulse oximetry: I. Pulse oximeter evaluation*
RALSTON, A. C., WEBB, R. K., & RUNCIMAN, W. B. (1991) · Anaesthesia · doi:10.1111/j.1365-2044.1991.tb09410.x · Cited by 166 · see citing works
Basal perfusion of the cutaneous microcirculation: Measurements as a function of anatomic position
Tur, E., Tur, M., Maibach, H. I., & Guy, R. H. (1983) · Journal of Investigative Dermatology · doi:10.1111/1523-1747.ep12522619 · Cited by 163 · see citing works
Influence of skin type and wavelength on light wave reflectance
Fallow, B. A., Tarumi, T., & Tanaka, H. (2013) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-013-9436-7 · Cited by 162 · see citing works
Bousefsaf, F., Maaoui, C., & Pruski, A. (2013) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2013.05.010 · Cited by 161 · see citing works
Increased skin capillary density in treated essential hypertensive patients
DEBBABI, H., UZAN, L., MOURAD, J., SAFAR, M., LEVY, B., & TIBIRICA, E. (2006) · American Journal of Hypertension · doi:10.1016/j.amjhyper.2005.10.021 · Cited by 156 · see citing works
Lee, B., Han, J., Baek, H. J., Shin, J. H., Park, K. S., & Yi, W. J. (2010) · Physiological Measurement · doi:10.1088/0967-3334/31/12/003 · Cited by 155 · see citing works
Khan, E., Al Hossain, F., Uddin, S. Z., Alam, S. K., & Hasan, M. K. (2016) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2466075 · Cited by 153 · see citing works
Lee, J., Kim, M., Park, H.-K., & Kim, I. Y. (2020) · Sensors · doi:10.3390/s20051493 · Cited by 153 · see citing works
Validation of Biofeedback Wearables for Photoplethysmographic Heart Rate Tracking.
(2016) · PubMed Central · Cited by 150 · see citing works
The effects of obesity on lung volumes and oxygenation
Littleton, S. W., & Tulaimat, A. (2017) · Respiratory Medicine · doi:10.1016/j.rmed.2017.01.004 · Cited by 150 · see citing works
Everything you need to know about NOx
Baukal, C. E. (2005) · Metal Finishing · doi:10.1016/s0026-0576(05)80816-5 · Cited by 140 · see citing works
Motion Artifact Reduction for Wrist-Worn Photoplethysmograph Sensors Based on Different Wavelengths
Zhang, Y., Song, S., Vullings, R., Biswas, D., Simões-Capela, N., van Helleputte, N., van Hoof, C., & Groenendaal, W. (2019) · Sensors · doi:10.3390/s19030673 · Cited by 140 · see citing works
Limiting racial disparities and bias for wearable devices in health science research
Colvonen, P. J., DeYoung, P. N., Bosompra, N.-O. A., & Owens, R. L. (2020) · Sleep · doi:10.1093/sleep/zsaa159 · Cited by 137 · see citing works
Proctor, D. N., Beck, K. C., Shen, P. H., Eickhoff, T. J., Halliwill, J. R., & Joyner, M. J. (1998) · Journal of Applied Physiology · doi:10.1152/jappl.1998.84.2.599 · Cited by 135 · see citing works
von Wowern, E., Östling, G., Nilsson, P. M., & Olofsson, P. (2015) · PLOS ONE · doi:10.1371/journal.pone.0135659 · Cited by 134 · see citing works
Hypertension Assessment via ECG and PPG Signals: An Evaluation Using MIMIC Database
Liang, Y., Chen, Z., Ward, R. K., & Elgendi, M. (2018) · Diagnostics · doi:10.3390/diagnostics8030065 · Cited by 132 · see citing works
Yamakoshi, K., Shimazu, H., Shibata, M., & Kamiya, A. (1982) · Medical & Biological Engineering & Computing · doi:10.1007/bf02442797 · Cited by 129 · see citing works
Derraik, J. G. B., Rademaker, M., Cutfield, W. S., Pinto, T. E., Tregurtha, S., Faherty, A., Peart, J. M., Drury, P. L., & Hofman, P. L. (2014) · PLoS ONE · doi:10.1371/journal.pone.0086637 · Cited by 129 · see citing works
Motin, M. A., Karmakar, C. K., & Palaniswami, M. S. (2017) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2017.2679108 · Cited by 129 · see citing works
The analysis of PPG morphology: Investigating the effects of aging on arterial compliance
Yousef, Q., Reaz, M. B. I., & Ali, M. A. M. (2012) · Measurement Science Review · doi:10.2478/v10048-012-0036-3 · Cited by 126 · see citing works
Photoplethysmogram signal quality estimation using repeated Gaussian filters and cross-correlation
Karlen, W., Kobayashi, K., Ansermino, J. M., & Dumont, G. A. (2012) · Physiological Measurement · doi:10.1088/0967-3334/33/10/1617 · Cited by 123 · see citing works
Statistical approach for the detection of motion/noise artifacts in Photoplethysmogram
Selvaraj, N., Mendelson, Y., Shelley, K. H., Silverman, D. G., & Chon, K. H. (2011) · 2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2011.6091232 · Cited by 119 · see citing works
Rees, J., & Robertson, K. (2010) · Acta Dermato Venereologica · doi:10.2340/00015555-0875 · Cited by 117 · see citing works
Multichannel reflective PPG earpiece sensor with passive motion cancellation
Wang, L., Lo, B. P., & Yang, G.-Z. (2007) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2007.910900 · Cited by 114 · see citing works
Zhang, Y., Liu, B., & Zhang, Z. (2015) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2015.05.006 · Cited by 109 · see citing works
Neural network for photoplethysmographic respiratory rate monitoring
Johansson, A. (2003) · Medical & Biological Engineering & Computing · doi:10.1007/bf02348427 · Cited by 109 · see citing works
Role of Biological Sex in Normal Cardiac Function and in its Disease Outcome – A Review
Prabhavathi, K. (2014) · JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH · doi:10.7860/jcdr/2014/9635.4771 · Cited by 108 · see citing works
Toward Generating More Diagnostic Features from Photoplethysmogram Waveforms
Elgendi, M., Liang, Y., & Ward, R. K. (2018) · Diseases · doi:10.3390/diseases6010020 · Cited by 107 · see citing works
Oscillometric determination of diastolic, mean and systolic blood pressure—A numerical model
Forster, F. K., & Turney, D. (1986) · Journal of Biomechanical Engineering · doi:10.1115/1.3138629 · Cited by 104 · see citing works
Chan, K. W., & Zhang, Y. (2003) · Proceedings of IEEE Sensors/Proceedings of IEEE Sensors ... · doi:10.1109/icsens.2002.1037314 · Cited by 104 · see citing works
THE EFFECT OF BLOOD PRESSURE ALTERATION ON THE PULSE WAVE VELOCITY
Nye, E. R. (1964) · Heart · doi:10.1136/hrt.26.2.261 · Cited by 103 · see citing works
Burton, A. C., & Taylor, R. M. (1940) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1940.129.3.565 · Cited by 103 · see citing works
Carter, S. A., & TATE, R. B. (1996) · Journal of Vascular Surgery · doi:10.1016/s0741-5214(96)70101-5 · Cited by 102 · see citing works
Optical properties of blood in motion
Lindberg, L.-G., & Oberg, P. A. (1993) · Optical Engineering · doi:10.1117/12.60688 · Cited by 102 · see citing works
Development of real-time motion artifact reduction algorithm for a wearable photoplethysmography
Han, H., Kim, M.-J., & Kim, J. (2007) · 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2007.4352596 · Cited by 101 · see citing works
Addison, P. S., Watson, J. N., Mestek, M. L., & Mecca, R. S. (2012) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-011-9332-y · Cited by 97 · see citing works
Olsen, L. O., Takiwaki, H., & Serup, J. (1995) · Skin Research and Technology · doi:10.1111/j.1600-0846.1995.tb00021.x · Cited by 96 · see citing works
Adaptive pulse segmentation and artifact detection in photoplethysmography for mobile applications
Karlen, W., Ansermino, J. M., & Dumont, G. (2012) · 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/embc.2012.6346628 · Cited by 94 · see citing works
Theoretical study on the effect of sensor contact force on pulse transit time
Teng, X.-F., & Zhang, Y.-T. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.900815 · Cited by 94 · see citing works
The influence of gender and age on the thickness and echo-density of skin
Firooz, A., Rajabi‐Estarabadi, A., Zartab, H., Pazhohi, N., Fanian, F., & Janani, L. (2017) · Skin Research and Technology · doi:10.1111/srt.12294 · Cited by 92 · see citing works
Shelley, K. H., Tamai, D., Jablonka, D., Gesquiere, M., Stout, R. G., & Silverman, D. G. (2005) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000145063.01043.4b · Cited by 91 · see citing works
A Flexible, Low Noise Reflective PPG Sensor Platform for Ear-Worn Heart Rate Monitoring
Patterson, J. A. C., McIlwraith, D. C., & Yang, G.-Z. (2009) · 2009 Sixth International Workshop on Wearable and Implantable Body Sensor Networks · doi:10.1109/bsn.2009.16 · Cited by 91 · see citing works
High frequency ultrasound to assess skin thickness in healthy adults
Van Mulder, T. J. S., de Koeijer, M., Theeten, H., Willems, D., Van Damme, P., Demolder, M., De Meyer, G., Beyers, K. C. L., & Vankerckhoven, V. (2017) · Vaccine · doi:10.1016/j.vaccine.2016.07.039 · Cited by 89 · see citing works
Nilsson, L., Goscinski, T., Kalman, S., Lindberg, L. ‐., & Johansson, A. (2007) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.2007.01375.x · Cited by 88 · see citing works
Allen, J., Frame, J. R., & Murray, A. (2002) · Physiological Measurement · doi:10.1088/0967-3334/23/2/312 · Cited by 86 · see citing works
Chatterjee, S., & Kyriacou, P. A. (2019) · Sensors · doi:10.3390/s19040789 · Cited by 86 · see citing works
The absence of vasoconstrictor reflexes in the forehead circulation. Effects of cold
Hertzman, A. B., & Roth, L. W. (1942) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1942.136.4.692 · Cited by 85 · see citing works
Motion Artifact Reduction in Photoplethysmographic Signals using Singular Value Decomposition
Reddy, K. A., & Kumar, V. J. (2007) · Conference proceedings - IEEE Instrumentation/Measurement Technology Conference · doi:10.1109/imtc.2007.379467 · Cited by 83 · see citing works
Chong, J. W., Dao, D. K., Salehizadeh, S. M. A., McManus, D. D., Darling, C. E., Chon, K. H., & Mendelson, Y. (2014) · Annals of Biomedical Engineering · doi:10.1007/s10439-014-1080-y · Cited by 79 · see citing works
Motion artifact cancellation in wearable photoplethysmography using gyroscope
Lee, H., Chung, H., & Lee, J. (2019) · IEEE Sensors Journal · doi:10.1109/jsen.2018.2879970 · Cited by 79 · see citing works
Joannides, R., Costentin, A., Iacob, M., Compagnon, P., Lahary, A., & Thuillez, C. (2002) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.00209.2001 · Cited by 75 · see citing works
Signal recovery in imaging photoplethysmography
Holton, B. D., Mannapperuma, K., Lesniewski, P. J., & Thomas, J. C. (2013) · Physiological Measurement · doi:10.1088/0967-3334/34/11/1499 · Cited by 73 · see citing works
Nabeel, P. M., Karthik, S., Joseph, J., & Sivaprakasam, M. (2018) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2018.2800539 · Cited by 72 · see citing works
Rajala, S., Lindholm, H., & Taipalus, T. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aac7ac · Cited by 70 · see citing works
Boonya-ananta, T., Rodriguez, A. J., Ajmal, A., Du Le, V. N., Hansen, A. K., Hutcheson, J. D., & Ramella-Roman, J. C. (2021) · Scientific Reports · doi:10.1038/s41598-021-82124-4 · Cited by 70 · see citing works
Self-citationFrancischetti, E. A., Tibirica, E., da Silva, E. G., Rodrigues, E., Celoria, B. M., & de Abreu, V. G. (2011) · Microvascular Research · doi:10.1016/j.mvr.2011.01.002 · Cited by 69 · see citing works
Simultaneous monitoring of ballistocardiogram and photoplethysmogram using a camera
Shao, D., Tsow, F., Liu, C., Yang, Y., & Tao, N. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2585109 · Cited by 69 · see citing works
Dehkordi, P., Garde, A., Molavi, B., Ansermino, J. M., & Dumont, G. A. (2018) · Frontiers in Physiology · doi:10.3389/fphys.2018.00948 · Cited by 66 · see citing works
Nowara, E. M., McDuff, D., & Veeraraghavan, A. (2020) · 2020 IEEE/CVF Conference on Computer Vision and Pattern Recognition Workshops (CVPRW) · doi:10.1109/cvprw50498.2020.00150 · Cited by 66 · see citing works
Microvascular dysfunction in healthy insulin-sensitive overweight individuals
Czernichow, S., Greenfield, J. R., Galan, P., Bastard, J.-P., Charnaux, N., Samaras, K., Safar, M. E., Blacher, J., Hercberg, S., & Levy, B. I. (2010) · Journal of Hypertension · doi:10.1097/hjh.0b013e328333d1fc · Cited by 64 · see citing works
Wong, A., Pun, K.-P., Zhang, Y.-T., & Hung, K. (2005) · IEEE Transactions on Circuits and Systems I: Regular Papers · doi:10.1109/tcsi.2005.857767 · Cited by 63 · see citing works
Gibbs, P., & Asada, H. H. (2005) · Proceedings of the ... American Control Conference/Proceedings of the American Control Conference · doi:10.1109/acc.2005.1470193 · Cited by 62 · see citing works
MRI and CT in the diagnosis of coronary artery disease: indications and applications
Nikolaou, K., Alkadhi, H., Bamberg, F., Leschka, S., & Wintersperger, B. J. (2010) · Insights into Imaging · doi:10.1007/s13244-010-0049-0 · Cited by 60 · see citing works
Validation of a Photoplethysmographic Heart Rate Monitor: Polar OH1
Hermand, E., Cassirame, J., Ennequin, G., & Hue, O. (2019) · International Journal of Sports Medicine · doi:10.1055/a-0875-4033 · Cited by 59 · see citing works
New aging index using signal features of both photoplethysmograms and acceleration plethysmograms
Ahn, J. M. (2017) · Healthcare Informatics Research · doi:10.4258/hir.2017.23.1.53 · Cited by 58 · see citing works
Applications of photoelectric plethysmography in peripheral vascular disease
Hertzman, A. B., & Dillon, J. B. (1940) · American Heart Journal · doi:10.1016/s0002-8703(40)90534-8 · Cited by 57 · see citing works
Gyroscope vs. accelerometer measurements of motion from wrist PPG during physical exercise
Casson, A. J., Vazquez Galvez, A., & Jarchi, D. (2016) · ICT Express · doi:10.1016/j.icte.2016.11.003 · Cited by 57 · see citing works
Nilsson, L., Johansson, A., & Kalman, S. (2003) · Physiological Measurement · doi:10.1088/0967-3334/24/4/009 · Cited by 55 · see citing works
The effect of applied sensor contact force on pulse transit time
Teng, X. F., & Zhang, Y. T. (2006) · Physiological Measurement · doi:10.1088/0967-3334/27/8/002 · Cited by 53 · see citing works
Value of Quantitative Photoplethysmography for Functional Vascular Diagnostics
Schultz-Ehrenburg, U., & Blažek, V. (2001) · Skin Pharmacology and Physiology · doi:10.1159/000056362 · Cited by 53 · see citing works
Measuring venous oxygenation using the photoplethysmograph waveform
Walton, Z. D., Kyriacou, P. A., Silverman, D. G., & Shelley, K. H. (2010) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-010-9248-y · Cited by 53 · see citing works
The detection of peripheral venous pulsation using the pulse oximeter as a plethysmograph
Shelley, K. H., Dickstein, M., & Shulman, S. M. (1993) · Journal of Clinical Monitoring · doi:10.1007/bf02886699 · Cited by 48 · see citing works
Evaluation of the Possible Use of PPG Waveform Features Measured at Low Sampling Rate
Fujita, D., & Suzuki, A. (2019) · IEEE Access · doi:10.1109/access.2019.2914498 · Cited by 48 · see citing works
Effect of probe contact pressure on the photoplethysmographic assessment of conduit artery stiffness
Grabovskis, A., Marcinkevics, Z., Rubins, U., & Kviesis-Kipge, E. (2013) · Journal of Biomedical Optics · doi:10.1117/1.jbo.18.2.027004 · Cited by 47 · see citing works
Comparison of respiratory-induced variations in photoplethysmographic signals
Li, J., Jin, J., Chen, X., Sun, W., & Guo, P. (2010) · Physiological Measurement · doi:10.1088/0967-3334/31/3/009 · Cited by 47 · see citing works
Hsiu, H., Hsu, C.-L., & Wu, T.-L. (2011) · Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine · doi:10.1177/0954411910396288 · Cited by 46 · see citing works
Photoplethysmography: Part 1 Comparison with laser Doppler flowmetry
Lindberg, L. G., Tamura, T., & Öberg, P. Å. (1991) · Medical & Biological Engineering & Computing · doi:10.1007/bf02446294 · Cited by 46 · see citing works
Krishnan, R., Natarajan, B., & Warren, S. (2008) · 2008 IEEE International Conference on Acoustics, Speech and Signal Processing · doi:10.1109/icassp.2008.4517684 · Cited by 45 · see citing works
Askarian, B., Jung, K., & Chong, J. W. (2019) · Sensors · doi:10.3390/s19132846 · Cited by 45 · see citing works
Investigating the effects of temperature on photoplethysmography
Khan, M., Pretty, C. G., Amies, A. C., Elliott, R., Shaw, G. M., & Chase, J. G. (2015) · IFAC-PapersOnLine · doi:10.1016/j.ifacol.2015.10.166 · Cited by 41 · see citing works
Design, development and clinical validation of a wrist-based optical heart rate monitor
Preejith, S. P., Alex, A., Joseph, J., & Sivaprakasam, M. (2016) · 2016 IEEE International Symposium on Medical Measurements and Applications (MeMeA) · doi:10.1109/memea.2016.7533786 · Cited by 41 · see citing works
The ageing cardiovascular system
Bolton, E., & Rajkumar, C. (2011) · Reviews in Clinical Gerontology · doi:10.1017/s0959259810000389 · Cited by 40 · see citing works
The effect of local mild cold exposure on pulse transit time
Zhang, X.-Y., & Zhang, Y.-T. (2006) · Physiological Measurement · doi:10.1088/0967-3334/27/7/008 · Cited by 40 · see citing works
Reducing motion artifacts in photoplethysmograms by using relative sensor motion: Phantom study
Wijshoff, R. W. C. G. R., Mischi, M., Veen, J., & Van Der Lee, A. M. (2012) · Journal of Biomedical Optics · doi:10.1117/1.jbo.17.11.117007 · Cited by 40 · see citing works
Zong, C., & Jafari, R. (2015) · 2015 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2015.7320268 · Cited by 39 · see citing works
The vasomotor components in the vascular reactions in the finger to cold
Hertzman, A. B., & Roth, L. W. (1942) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1942.136.4.669 · Cited by 39 · see citing works
Effects of skin thickness and skinfold compressibility on skinfold thickness measurement
Martin, A. D., Drinkwater, D. T., Clarys, J. P., Daniel, M., & Ross, W. D. (1992) · American Journal of Human Biology · doi:10.1002/ajhb.1310040404 · Cited by 38 · see citing works
Central venous pulsations associated with a falsely low oxygen saturation measured by pulse oximetry
Sami, H. M., Kleinman, B. S., & Lonchyna, V. A. (1991) · Journal of Clinical Monitoring · doi:10.1007/bf01619351 · Cited by 36 · see citing works
Hahn, M., Hahn, C., Jünger, M., Steins, A., Zuder, D., Klyscz, T., Büchtemann, A., Rassner, G., & Blazek, V. (1999) · Microvascular Research · doi:10.1006/mvre.1998.2130 · Cited by 33 · see citing works
Couceiro, R., Carvalho, P., Paiva, R. P., Henriques, J., & Muehlsteff, J. (2012) · 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/embc.2012.6346497 · Cited by 33 · see citing works
Effects of skin surface temperature on photoplethysmograph
Jeong, I. C., Yoon, H., Kang, H., & Yeom, H. (2014) · Journal of Healthcare Engineering · doi:10.1260/2040-2295.5.4.429 · Cited by 33 · see citing works
Kapur, V. K., Wilsdon, A. G., Au, D., Avdalovic, M., Enright, P., Fan, V. S., Hansel, N. N., Heckbert, S. R., Jiang, R., Krishnan, J. A., et al. (2013) · Respiratory Care · doi:10.4187/respcare.02008 · Cited by 33 · see citing works
Spectral analysis of resting variability of some circulatory parameters in man.
Peňáz, J., Honzíková, N., & Fiser, B. (1978) · PubMed · Cited by 32 · see citing works
Optical Fibre-Based Pulse Oximetry Sensor with Contact Force Detection
Liu, C., Correia, R., Ballaji, H., Korposh, S., Hayes-Gill, B., & Morgan, S. (2018) · Sensors · doi:10.3390/s18113632 · Cited by 32 · see citing works
Suzuki, S., & Oguri, K. (2009) · 2009 Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2009.5332505 · Cited by 31 · see citing works
Dutt, D. N., & Shruthi, S. (2015) · International Conference on Communications and Signal Processing (ICCSP) · doi:10.1109/iccsp.2015.7322766 · Cited by 30 · see citing works
Analysing the effects of cold, normal, and warm digits on transmittance pulse oximetry
Khan, M., Pretty, C. G., Amies, A. C., Elliott, R. B., Chiew, Y. S., Shaw, G. M., & Chase, J. G. (2016) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2015.12.006 · Cited by 30 · see citing works
Vibrotactile and thermal perception and its relation to finger skin thickness
Lundström, R., Dahlqvist, H., Hagberg, M., & Nilsson, T. (2018) · Clinical Neurophysiology Practice · doi:10.1016/j.cnp.2018.01.001 · Cited by 30 · see citing works
Imaging Photoplethysmography: What are the Best Locations on the Face to Estimate Heart Rate?
Fallet, S., Moser, V., Braun, F., & Vesin, J.-M. (2016) · 2016 Computing in Cardiology Conference (CinC) · doi:10.22489/cinc.2016.098-236 · Cited by 28 · see citing works
Peripheral Blood Flow Regulation in Human Obesity and Metabolic Syndrome
Limberg, J. K., Morgan, B. J., & Schrage, W. G. (2016) · Exercise and Sport Sciences Reviews · doi:10.1249/jes.0000000000000083 · Cited by 27 · see citing works
A contact-force regulated photoplethysmography (PPG) platform
Sim, J. K., Ahn, B., & Doh, I. (2018) · AIP Advances · doi:10.1063/1.5020914 · Cited by 27 · see citing works
Age and gender do not influence the ability to detect respiration by photoplethysmography
Nilsson, L., Goscinski, T., Johansson, A., Lindberg, L.-G., & Kalman, S. (2006) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-006-9050-z · Cited by 26 · see citing works
Rauh, R. D., Posfay, A., & Mück-Weymann, M. (2003) · Clinical Physiology and Functional Imaging · doi:10.1046/j.1475-0961.2003.00516.x · Cited by 26 · see citing works
Excessive weight favours skin physiology-up to a point: Another expression of the obesity paradox
Monteiro Rodrigues, L. M., Palma, L., Santos, O., Almeida, M. A., Bujan, J., & Tavares, L. (2017) · Skin Pharmacology and Physiology · doi:10.1159/000464338 · Cited by 25 · see citing works
Sex-Related Differences in Photoplethysmography Signals Measured From Finger and Toe
Dehghanojamahalleh, S., & Kaya, M. (2019) · IEEE Journal of Translational Engineering in Health and Medicine · doi:10.1109/jtehm.2019.2938506 · Cited by 25 · see citing works
Addison, P. S., Jacquel, D., Foo, D. M. H., & Borg, U. R. (2018) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-017-0076-1 · Cited by 24 · see citing works
Skin Collagen and Thickness in Simple Obesity
Black, M. M., Bottoms, E., & Shuster, S. (1971) · BMJ · doi:10.1136/bmj.4.5780.149 · Cited by 23 · see citing works
Non-invasive Studies on Age Related Parameters Using a Blood Volume Pulse Sensor
Jayasree, V. K., Sandhya, T., & Radhakrishnan, P. (2008) · Measurement Science Review · doi:10.2478/v10048-008-0020-0 · Cited by 23 · see citing works
Kim, J., Lee, T., Kim, J.-H., & Ko, H. (2015) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2015.7319867 · Cited by 22 · see citing works
A System for Continuous Estimating and Monitoring Cardiac Output via Arterial Waveform Analysis.
Vakily, A., Parsaei, H., Movahhedi, M. M., & Sahmeddini, M. A. (2017) · PubMed Central · Cited by 22 · see citing works
Lee, H., Chung, H., Kim, J.-W., & Lee, J. (2019) · IEEE Sensors Journal · doi:10.1109/jsen.2019.2894640 · Cited by 22 · see citing works
Perfusion Changes at the Forehead Measured by Photoplethysmography during a Head-Down Tilt Protocol
Ysehak Abay, T., Shafqat, K., & Kyriacou, P. A. (2019) · Biosensors · doi:10.3390/bios9020071 · Cited by 22 · see citing works
Guajardo, I., Ayer, A., Johnson, A. D., Ganz, P., Mills, C., Donovan, C., Scherzer, R., Shah, S. J., Peralta, C. A., & Dubin, R. F. (2018) · Hemodialysis International · doi:10.1111/hdi.12544 · Cited by 21 · see citing works
Critical care trainees call for pulse oximetry reform
Hidalgo, D. C., Olusanya, O., & Harlan, E. (2021) · The Lancet Respiratory Medicine · doi:10.1016/s2213-2600(21)00102-8 · Cited by 21 · see citing works
Chin, L.-C., Huang, T.-Y., Yu, C.-L., Wu, C.-H., Hsu, C.-C., & Yu, H.-S. (1999) · Atherosclerosis · doi:10.1016/s0021-9150(99)00135-5 · Cited by 20 · see citing works
A 0.5-Hz High-Pass Cutoff Dual-Loop Transimpedance Amplifier for Wearable NIR Sensing Device
Wong, A., Leung, K. N., Pun, K., & Zhang, Y. (2010) · IEEE Transactions on Circuits & Systems II Express Briefs · doi:10.1109/tcsii.2010.2048401 · Cited by 20 · see citing works
Experimental validation of dual PPG local pulse wave velocity probe
Nabeel, P. M., Karthik, S., Joseph, J., & Sivaprakasam, M. (2017) · 2017 IEEE International Symposium on Medical Measurements and Applications (MeMeA) · doi:10.1109/memea.2017.7985911 · Cited by 20 · see citing works
A novel sensor for wrist based optical heart rate monitor
Mohapatra, P., Preejith, S. P., & Sivaprakasam, M. (2017) · 2017 IEEE International Instrumentation and Measurement Technology Conference (I2MTC) · doi:10.1109/i2mtc.2017.7969842 · Cited by 19 · see citing works
The reactions of the digital artery and minute pad arteries to local cold
Hertzman, A. B., & Roth, L. W. (1942) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1942.136.4.680 · Cited by 17 · see citing works
Phuong, C., & Maibach, H. I. (2017) · Textbook of Aging Skin · doi:10.1007/978-3-662-47398-6_93 · Cited by 16 · see citing works
Vascular Distensibility and Functions of the Arterial and Venous Systems
(2012) · Elsevier eBooks · doi:10.1016/b978-1-4160-5451-1.00015-3 · Cited by 15 · see citing works
Exploring the anatomy and physiology of ageing. Part 1--The cardiovascular system.
Knight, J., & Nigam, Y. (2008) · PubMed · Cited by 14 · see citing works
Influence of motion artifacts on photoplethysmographic signals for measuring pulse rates
Lee, Y., Han, H., & Kim, J. (2008) · International Conference on Control, Automation and Systems (ICCAS) · doi:10.1109/iccas.2008.4694635 · Cited by 12 · see citing works
Photoplethysmography system for blood pulsation detection in unloaded artery conditions
Grabovskis, A., Marcinkevics, Z., Rubenis, O., Rubins, U., & Lusa, V. (2012) · SPIE Proceedings · doi:10.1117/12.922649 · Cited by 11 · see citing works
Anatomy and physiology of ageing 1: the cardiovascular system
Knight, J. (2017) · Cronfa (Swansea University) · Cited by 11 · see citing works
Altintas, A. A., Aust, M. C., Krämer, R., Vogt, P. M., & Altintas, M. A. (2016) · Journal of Biomedical Optics · doi:10.1117/1.jbo.21.3.036009 · Cited by 10 · see citing works
Photoplethysmography signal analysis to assess obesity, age group and hypertension
Ferdinando, H., Huotari, M., & Myllyla, T. (2019) · 2019 41st Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2019.8857570 · Cited by 9 · see citing works
Iacopi, E., Riitano, N., Dini, V., Berta, R. D., Pieruzzi, L., Janowska, A., Anselmino, M., Piaggesi, A., & Romanelli, M. (2020) · Diabetes, Metabolic Syndrome and Obesity: Targets and Therapy · doi:10.2147/dmso.s229042 · Cited by 9 · see citing works
Nijland, R., Jongsma, H. W., Verbruggen, I. M., & Nijhuis, J. G. (1996) · Journal of Clinical Monitoring · doi:10.1007/bf00857643 · Cited by 8 · see citing works
Flagged first, then by citations
2 of the 212 references couldn’t be matched to a known paper.
Research works printed but not linked: 3
- Hasan, 2015, A New First-Derivative Related Index to Assess Pulse Wave Transit Time from a Photoplethysmographic Waveform: Age Dependence and Agreement with Normative Data, Int. J. Bioinform. Biomed. Eng., 1, 276
- Goepfert, L., Petersen, C.L., Ansermino, J.M., and Gorges, M. (2018, January 10–13). Evaluation of a Combined PPG/MPG Sensor in Healthy Volunteers. Proceedings of the Anesthesia and Analgesia, Miami, FL, USA.
- Sazonov, E. (2014). Wearable Sensors: Fundamentals, Implementation and Applications, Elsevier.
Cited by
Sorted by citations
Wearable Pressure Sensors for Pulse Wave Monitoring
Meng, K., Xiao, X., Wei, W., Chen, G., Nashalian, A., Shen, S., Xiao, X., & Chen, J. (2022) · Advanced Materials · doi:10.1002/adma.202109357 · Cited by 883 · see citing works
Min, S., Kim, D. H., Joe, D. J., Kim, B. W., Jung, Y.-H., Lee, J. H., Lee, B., Doh, I., An, J., Youn, Y., et al. (2023) · Advanced Materials · doi:10.1002/adma.202301627 · Cited by 285 · see citing works
de Zambotti, M., Goldstein, C. A., Cook, J. D., Menghini, L., Altini, M., Cheng, P., & Robillard, R. (2023) · SLEEP · doi:10.1093/sleep/zsad325 · Cited by 262 · see citing works
Quigley, K. S., Gianaros, P. J., Norman, G. J., Jennings, J. R., Berntson, G. G., & de Geus, E. J. C. (2024) · Psychophysiology · doi:10.1111/psyp.14604 · Cited by 253 · see citing works
The 2023 wearable photoplethysmography roadmap
Charlton, P. H., Allen, J., Bailón, R., Baker, S., Behar, J. A., Chen, F., Clifford, G. D., Clifton, D. A., Davies, H. J., Ding, C., et al. (2023) · Physiological Measurement · doi:10.1088/1361-6579/acead2 · Cited by 210 · see citing works
Self-citationLiu, G., Lv, Z., Batool, S. A., Li, M.-Z., Zhao, P., Guo, L., Wang, Y., Zhou, Y., & Han, S. (2023) · Small · doi:10.1002/smll.202207879 · Cited by 191 · see citing works
A Review of Wearable Multi-Wavelength Photoplethysmography
Ray, D., Collins, T., Woolley, S., & Ponnapalli, P. (2023) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2021.3121476 · Cited by 155 · see citing works
Cabanas, A. M., Fuentes-Guajardo, M., Latorre, K., León, D., & Martín-Escudero, P. (2022) · Sensors · doi:10.3390/s22093402 · Cited by 124 · see citing works
A review of the effect of skin pigmentation on pulse oximeter accuracy
Al-Halawani, R., Charlton, P. H., Qassem, M., & Kyriacou, P. A. (2023) · Physiological Measurement · doi:10.1088/1361-6579/acd51a · Cited by 119 · see citing works
Coutrot, M., Dudoignon, E., Joachim, J., Gayat, É., Vallée, F., & Dépret, F. (2021) · Anaesthesia Critical Care & Pain Medicine · doi:10.1016/j.accpm.2021.100964 · Cited by 119 · see citing works
Scardulla, F., Cosoli, G., Spinsante, S., Poli, A., Iadarola, G., Pernice, R., Busacca, A., Pasta, S., Scalise, L., & D’Acquisto, L. (2023) · Measurement · doi:10.1016/j.measurement.2023.113150 · Cited by 120 · see citing works
Face2PPG: An Unsupervised Pipeline for Blood Volume Pulse Extraction From Faces
Casado, C. Á., & López, M. B. (2023) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2023.3307942 · Cited by 109 · see citing works
Continuous cuffless blood pressure monitoring with a wearable ring bioimpedance device
Sel, K., Osman, D., Huerta, N., Edgar, A., Pettigrew, R. I., & Jafari, R. (2023) · npj Digital Medicine · doi:10.1038/s41746-023-00796-w · Cited by 100 · see citing works
Sel, K., Mohammadi, A., Pettigrew, R. I., & Jafari, R. (2023) · npj Digital Medicine · doi:10.1038/s41746-023-00853-4 · Cited by 95 · see citing works
Roos, L. G., & Slavich, G. M. (2023) · Brain Behavior and Immunity · doi:10.1016/j.bbi.2023.08.008 · Cited by 100 · see citing works
Consumer Wearable Health and Fitness Technology in Cardiovascular Medicine
Petek, B. J., Al‐Alusi, M. A., Moulson, N., Grant, A. J., Besson, C., Guseh, J. S., Wasfy, M. M., Grémeaux, V., Churchill, T. W., & Baggish, A. L. (2023) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2023.04.054 · Cited by 106 · see citing works
Data quality evaluation in wearable monitoring
Böttcher, S., Vieluf, S., Bruno, E., Joseph, B., Epitashvili, N., Biondi, A., Zabler, N., Glasstetter, M., Dümpelmann, M., Van Laerhoven, K., et al. (2022) · Scientific Reports · doi:10.1038/s41598-022-25949-x · Cited by 91 · see citing works
Zhou, Z.-B., Cui, T., Li, D., Jian, J., Li, Z., Ji, S., Li, X., Xu, J., Liu, H., Yang, Y., et al. (2023) · Materials · doi:10.3390/ma16062133 · Cited by 81 · see citing works
Validating cuffless continuous blood pressure monitoring devices
Hu, J., Martin, G., Iyengar, S., Kovell, L. C., Plante, T. B., van Helmond, N., Dart, R. A., Brady, T. M., Turkson‐Ocran, R., & Juraschek, S. P. (2023) · Cardiovascular Digital Health Journal · doi:10.1016/j.cvdhj.2023.01.001 · Cited by 74 · see citing works
Heart rate variability: Evaluating a potential biomarker of anxiety disorders
Tomasi, J., Zai, C. C. H., Pouget, J. G., Tiwari, A. K., & Kennedy, J. L. (2023) · Psychophysiology · doi:10.1111/psyp.14481 · Cited by 66 · see citing works
Decoding tissue biomechanics using conformable electronic devices
Yoon, H., Kim, J., Sadat, D., Barrett, A., Ko, S. H., & Dağdeviren, C. (2024) · Nature Reviews Materials · doi:10.1038/s41578-024-00729-3 · Cited by 63 · see citing works
Peripheral perfusion index of pulse oximetry in adult patients: a narrative review
Sun, X., He, H., Xu, M., & Long, Y. (2024) · European journal of medical research · doi:10.1186/s40001-024-02048-3 · Cited by 55 · see citing works
Six‐arm Stellat Dendritic‐PbS Flexible Infrared Photodetector for Intelligent Healthcare Monitoring
Zhang, T., Ling, C., Wang, X., Feng, B., Cao, M., Xue, X., Xue, Q., Zhang, J., Zhu, L., Wang, C., et al. (2022) · Advanced Materials Technologies · doi:10.1002/admt.202200250 · Cited by 54 · see citing works
Blood pressure measurement using only a smartphone
Frey, L., Menon, C., & Elgendi, M. (2022) · npj Digital Medicine · doi:10.1038/s41746-022-00629-2 · Cited by 55 · see citing works
Cuffless Blood Pressure in clinical practice: challenges, opportunities and current limits.
Henry, B., Merz, M., Hoang, H., Abdulkarim, G., Wosik, J., & Schoettker, P. (2024) · Blood Pressure · doi:10.1080/08037051.2024.2304190 · Cited by 54 · see citing works
Artificial Intelligence for the Detection and Treatment of Atrial Fibrillation
Harmon, D., Sehrawat, O., Maanja, M., Wight, J., & Noseworthy, P. A. (2023) · Arrhythmia & Electrophysiology Review · doi:10.15420/aer.2022.31 · Cited by 48 · see citing works
Yao, P., Xue, N., Yin, S., You, C., Guo, Y., Shi, Y., Liu, T., Yao, L., Zhou, J., Sun, J., et al. (2022) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2022.3167059 · Cited by 49 · see citing works
Automated loss of pulse detection on a consumer smartwatch
Shah, K., Wang, A., Chen, Y., Munjal, J., Chhabra, S., Stange, A., Wei, E., Phan, T., Giest, T., Hawkins, B., et al. (2025) · Nature · doi:10.1038/s41586-025-08810-9 · Cited by 44 · see citing works
Artificial Intelligence in Atrial Fibrillation: From Early Detection to Precision Therapy
Karakasis, P., Theofilis, P., Sagris, Μ., Pamporis, K., Stachteas, P., Sidiropoulos, G., Vlachakis, P. K., Patoulias, D., Antoniadis, A. P., & Fragakis, N. (2025) · Journal of Clinical Medicine · doi:10.3390/jcm14082627 · Cited by 47 · see citing works
Photoplethysmography—new applications for an old technology: a sleep technology review
Ryals, S. M., Chiang, A. A.-P., Schutte-Rodin, S., Chandrakantan, A., Verma, N., Holfinger, S. J., Abbasi-Feinberg, F., Bandyopadhyay, A., Baron, K. G., Bhargava, S., et al. (2022) · Journal of Clinical Sleep Medicine · doi:10.5664/jcsm.10300 · Cited by 43 · see citing works
Are Activity Wrist-Worn Devices Accurate for Determining Heart Rate during Intense Exercise?
Martín-Escudero, P., Cabanas, A. M., Dotor-Castilla, M. L., Galindo-Canales, M., Miguel‐Tobal, F., Fernández, C., Fuentes, M. E., & Giannetti, R. (2023) · Bioengineering · doi:10.3390/bioengineering10020254 · Cited by 40 · see citing works
Rodrigues, E., Lima, D., Barbosa, P. E. E. S., Gonzaga, K. D. A., Guerra, R. O., Pimentel, M. M., Barbosa, H. A., & Maciel, Á. C. C. (2022) · Sensors · doi:10.3390/s22052001 · Cited by 38 · see citing works
Skin colour affects the accuracy of medical oxygen sensors
Keller, M. D., Harrison-Smith, B., Patil, C., & Arefin, M. S. (2022) · Nature · doi:10.1038/d41586-022-03161-1 · Cited by 38 · see citing works
Singh, S., Bennett, M. R., Chen, C., Shin, S., Ghanbari, H., & Nelson, B. W. (2024) · Journal of Medical Internet Research · doi:10.2196/62769 · Cited by 36 · see citing works
“Researchers have identified several factors that influence the accuracy of PPG-based measurements, including skin temperature, sensor contact pressure, skin thickness, and hydration level [5,6].”
An In-Ear PPG-Based Blood Glucose Monitor: A Proof-of-Concept Study
Hammour, G., & Mandic, D. P. (2023) · Sensors · doi:10.3390/s23063319 · Cited by 39 · see citing works
Al-Halawani, R., Qassem, M., & Kyriacou, P. A. (2024) · Scientific Reports · doi:10.1038/s41598-024-58435-7 · Cited by 36 · see citing works
A Survey of Photoplethysmography and Imaging Photoplethysmography Quality Assessment Methods
Desquins, T., Bousefsaf, F., Pruski, A., & Maaoui, C. (2022) · Applied Sciences · doi:10.3390/app12199582 · Cited by 33 · see citing works
Contactless Arterial Blood Pressure Waveform Monitoring with mmWave Radar
Hu, Q., Zhang, Q., Lü, H., Wu, S., Zhou, Y., Huang, Q., Chen, H., Chen, Y.-C., & Zhao, N. (2024) · Proceedings of the ACM on Interactive Mobile Wearable and Ubiquitous Technologies · doi:10.1145/3699781 · Cited by 32 · see citing works
Performance evaluation of a wrist-worn reflectance pulse oximeter during sleep
Jung, H., Kim, D., Lee, W. Y., Seo, H., Seo, J.-W., Choi, J., & Joo, E. Y. (2022) · Sleep Health · doi:10.1016/j.sleh.2022.04.003 · Cited by 31 · see citing works
Wearable PZT Piezoelectric Sensor Device for Accurate Arterial Pressure Pulse Waveform Measurement
Li, M.-Y., Aoyama, J., Inayoshi, K., & Zhang, H. (2025) · Advanced Electronic Materials · doi:10.1002/aelm.202400852 · Cited by 31 · see citing works
Babu, A., Ranpariya, S., Sinha, D. K., Chatterjee, A., & Mandal, D. (2023) · Advanced Materials Technologies · doi:10.1002/admt.202202021 · Cited by 30 · see citing works
Advances in PPG Sensors Data Acquisition With Light-to-Digital Converters: A Review
Chettri, N., Aprile, A., Bonizzoni, E., & Malcovati, P. (2024) · IEEE Sensors Journal · doi:10.1109/jsen.2024.3420170 · Cited by 29 · see citing works
Cohen, J. B., Byfield, R. L., Hardy, S. T., Juraschek, S. P., Miller, N. H., Mukkamala, R., Picone, D. S., Thiele, R. H., Yang, E., & Brady, T. M. (2025) · Hypertension · doi:10.1161/hyp.0000000000000254 · Cited by 29 · see citing works
Hu, X., Sgherza, T. R., Nothrup, J. B., Fresco, D. M., Naragon‐Gainey, K., & Bylsma, L. M. (2024) · Behavior Research Methods · doi:10.3758/s13428-024-02387-3 · Cited by 29 · see citing works
PPG and Bioimpedance-Based Wearable Applications in Heart Rate Monitoring—A Comprehensive Review
Lapsa, D., Janeliukštis, R., Metshein, M., & Selavo, L. (2024) · Applied Sciences · doi:10.3390/app14177451 · Cited by 30 · see citing works
Koch, M., Matzke, I., Huhn, S., Sié, A., Boudo, V., Compaoré, G., Maggioni, M. A., Bunker, A., Bärnighausen, T., Dambach, P., et al. (2023) · JMIR mhealth and uhealth · doi:10.2196/46980 · Cited by 24 · see citing works
Guarducci, S., Jayousi, S., Caputo, S., & Mucchi, L. (2025) · Sensors · doi:10.3390/s25020556 · Cited by 29 · see citing works
Sato, S., Hiratsuka, T., Hasegawa, K., Watanabe, K., Obara, Y., Kariya, N., Shinba, T., & Matsui, T. (2023) · Sensors · doi:10.3390/s23083867 · Cited by 25 · see citing works
Advances in Photoplethysmography for Personalized Cardiovascular Monitoring
Kim, S., Xiao, X., & Chen, J. (2022) · Biosensors · doi:10.3390/bios12100863 · Cited by 25 · see citing works
Gajda, R., Gajda, J., Czuba, M., Knechtle, B., & Drygas, W. (2023) · Sports Medicine · doi:10.1007/s40279-023-01948-4 · Cited by 24 · see citing works
Matzke, I., Huhn, S., Koch, M., Maggioni, M. A., Munga, S., Muma, J. O., Odhiambo, C. O., Kwaro, D., Obor, D., Bärnighausen, T., et al. (2024) · JMIR mhealth and uhealth · doi:10.2196/54669 · Cited by 21 · see citing works
Khalili, M., Lingawi, S., Hutton, J., Fordyce, C. B., Christenson, J. M., Shadgan, B., Grunau, B. E., & Kuo, C. (2024) · Scientific Reports · doi:10.1038/s41598-024-74117-w · Cited by 25 · see citing works
Hermans, F., Arents, E., Blondeel, A., Janssens, W., Cardinaels, N., Calders, P., Troosters, T., Derom, E., & Demeyer, H. (2024) · JMIR mhealth and uhealth · doi:10.2196/56027 · Cited by 23 · see citing works
Heartbeat Detection in Gyrocardiography Signals without Concurrent ECG Tracings
Parlato, S. J., Centracchio, J., Esposito, D., Bifulco, P., & Andreozzi, E. (2023) · Sensors · doi:10.3390/s23136200 · Cited by 23 · see citing works
Ho, M. Y., Pham, H. M., Saeed, A., & Ma, D. (2025) · Scientific Data · doi:10.1038/s41597-025-04453-7 · Cited by 25 · see citing works
Photoplethysmography based atrial fibrillation detection: a continually growing field
Ding, C., Xiao, R., Wang, W., Holdsworth, E., & Hu, X. (2024) · Physiological Measurement · doi:10.1088/1361-6579/ad37ee · Cited by 25 · see citing works
Oxygen Saturation Behavior by Pulse Oximetry in Female Athletes: Breaking Myths
Martín-Escudero, P., Cabanas, A. M., Fuentes, M. E., & Galindo-Canales, M. (2021) · Biosensors · doi:10.3390/bios11100391 · Cited by 25 · see citing works
Tourist Experiences Recommender System Based on Emotion Recognition with Wearable Data
Granados, L. H. S., Mendoza-Moreno, J. F., Chantre-Astaiza, Á., Muñoz-Organero, M., & Ramírez-González, G. (2021) · Sensors · doi:10.3390/s21237854 · Cited by 22 · see citing works
External factors that affect the photoplethysmography waveforms
Pi, I., Pi, I., & Wu, W. (2021) · SN Applied Sciences · doi:10.1007/s42452-021-04906-9 · Cited by 21 · see citing works
Sanches, C. A., Silva, G. A., Librantz, A. F. H., Sampaio, L. M. M., & Belan, P. A. (2023) · Journal of Medical Internet Research · doi:10.2196/47112 · Cited by 23 · see citing works
Sarkar, S., Ghosh, A., & Kalra, J. (2023) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2023.107558 · Cited by 21 · see citing works
Establishing best practices in photoplethysmography signal acquisition and processing
Charlton, P. H., Pilt, K., & Kyriacou, P. A. (2022) · Physiological Measurement · doi:10.1088/1361-6579/ac6cc4 · Cited by 23 · see citing works
Validity of heart rate measurements in wrist-based monitors across skin tones during exercise
Hung, S. H., Serwa, K., Rosenthal, G., & Eng, J. J. (2025) · PLOS ONE · doi:10.1371/journal.pone.0318724 · Cited by 21 · see citing works
Quality Assessment and Morphological Analysis of Photoplethysmography in Daily Life
Moscato, S., Palmerini, L., Palumbo, P., & Chiari, L. (2022) · Frontiers in Digital Health · doi:10.3389/fdgth.2022.912353 · Cited by 20 · see citing works
Støve, M. P., & Hansen, E. C. K. (2022) · Journal of Sports Sciences · doi:10.1080/02640414.2023.2180160 · Cited by 19 · see citing works
Heartbeat detector from ECG and PPG signals based on wavelet transform and upper envelopes
Merino-Monge, M., Castro-García, J. A., Lebrato-Vázquez, C., Gómez, I., & Molina-Cantero, A. J. (2023) · Physical and Engineering Sciences in Medicine · doi:10.1007/s13246-023-01235-6 · Cited by 19 · see citing works
Fabric‐Based Ultrasonic Sensor with Integrated Piezoelectric Composite for Blood Pressure Monitoring
Li, B., Wang, T., Luo, D., Peng, P., Wang, Y., Liu, L., Wang, H., & Liu, H. (2023) · Advanced Materials Technologies · doi:10.1002/admt.202201814 · Cited by 19 · see citing works
Liu, J., Hu, S., Wang, Y., Hu, Q., Wang, D., & Yang, C. (2023) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2023.3331975 · Cited by 19 · see citing works
Hellqvist, H., Karlsson, M. K., Hoffman, J., Kahan, T., & Spaak, J. (2024) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2024.1350726 · Cited by 19 · see citing works
Predicting Blood Pressures for Pregnant Women by PPG and Personalized Deep Learning
Nguyen, D. H., Chao, P. C.-P., Yan, H. F., Tu, T.-Y., Cheng, C.-H., & Phan, T.-P. (2024) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2024.3386707 · Cited by 19 · see citing works
Kimball, J. P., Inan, O. T., Convertino, V. A., Cardin, S., & Sawka, M. N. (2022) · Sensors · doi:10.3390/s22020442 · Cited by 19 · see citing works
Reali, P., Lolatto, R., Coelli, S., Tartaglia, G., & Bianchi, A. M. (2022) · Sensors · doi:10.3390/s22041428 · Cited by 18 · see citing works
Wu, Z., Li, X., Feng, Z., Wan, C., Ya, L., Li, T., Yang, Q., Liu, X., Ren, M., Li, J., et al. (2023) · Advanced Healthcare Materials · doi:10.1002/adhm.202202629 · Cited by 19 · see citing works
Sinichi, M., Gevonden, M. J., & Krabbendam, L. (2025) · Psychophysiology · doi:10.1111/psyp.70004 · Cited by 22 · see citing works
Cardiodiagnostics Based on Photoplethysmographic Signals
Georgieva-Tsaneva, G. N., Gospodinova, E. P., & Cheshmedzhiev, K. (2022) · Diagnostics · doi:10.3390/diagnostics12020412 · Cited by 17 · see citing works
Tissue mimicking materials and finger phantom design for pulse oximetry
Rodriguez, A., Vasudevan, S., Farahmand, M., Weininger, S., Vogt, W. C., Scully, C. G., Ramella‐Roman, J. C., & Pfefer, T. J. (2024) · Biomedical Optics Express · doi:10.1364/boe.518967 · Cited by 17 · see citing works
Self-citationHoang, T. T., Cunio, A. M., Zhao, S., Nguyen, T., Peng, S., Liaw, S., Barber, T. J., Zhang, J., Farajikhah, S., Dehghani, F., et al. (2024) · Advanced Sensor Research · doi:10.1002/adsr.202400039 · Cited by 17 · see citing works
Yaşar, M. N., Sica, M., O’Flynn, B., Tedesco, S. A., & Menolotto, M. (2024) · Scientific Data · doi:10.1038/s41597-024-03254-8 · Cited by 17 · see citing works
Chang, S.-C., Wang, C., Cheng, K., & Lee, G. (2025) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2025.137949 · Cited by 16 · see citing works
Kantrowitz, A. B., Ben‐David, K., Morris, M., Wittels, H. L., Wishon, M. J., McDonald, S. M., Renaghan, E., Feigenbaum, L. A., & Wittels, S. H. (2025) · Frontiers in Physiology · doi:10.3389/fphys.2025.1630032 · Cited by 20 · see citing works
Wu, M., Liu, I.-F., Tzeng, Y.-H., & Wang, L. (2022) · Physiological Measurement · doi:10.1088/1361-6579/ac7d91 · Cited by 16 · see citing works
Balaraman, K., & Pushpanathan, A. C. S. (2023) · Journal of Machine and Computing · doi:10.53759/7669/jmc202303030 · Cited by 16 · see citing works
Lv, Y., Cen, G., Li, W., Zhao, C., & Mai, W. (2023) · Science China Materials · doi:10.1007/s40843-023-2613-5 · Cited by 16 · see citing works
Monnink, S. H. J., van Vliet, M., Kuiper, M. J., Constandse, J. C., Hoftijzer, D., Muller, M., & Ronner, E. (2024) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-024-01229-z · Cited by 16 · see citing works
Ling, C., Rong, C., Men, B., Wang, J., Sun, J., Zhang, T., Zhang, L., Guo, T., Zhou, P., & Liu, W. (2025) · Advanced Healthcare Materials · doi:10.1002/adhm.202402507 · Cited by 15 · see citing works
EarPPG: Securing Your Identity with Your Ears
Choi, S., Yim, J., Jin, Y., Gao, Y., Li, J., & Jin, Z. (2023) · ACM International Conference on Intelligent User Interfaces (IUI) · doi:10.1145/3581641.3584070 · Cited by 15 · see citing works
Multiwavelength tissue-mimicking phantoms with tunable vessel pulsation
Jenne, S., & Zappe, H. P. (2023) · Journal of Biomedical Optics · doi:10.1117/1.jbo.28.4.045003 · Cited by 15 · see citing works
Lønfeldt, N. N., Olesen, K. V., Das, S., Mora‐Jensen, A. C., Pagsberg, A. K., & Clemmensen, L. K. H. (2023) · Frontiers in Psychiatry · doi:10.3389/fpsyt.2023.1231024 · Cited by 15 · see citing works
Mather, J. D., Hayes, L. D., Mair, J. L., & Sculthorpe, N. F. (2024) · Frontiers in Digital Health · doi:10.3389/fdgth.2024.1326511 · Cited by 17 · see citing works
Radar-Based Heart Rate Sensing on the Smart Glasses
Huang, W., Rajbhandary, P. L., Shiu, S., Ho, J. S., Zhu, J. B., Wilson, B., & Ye, G. (2024) · IEEE Microwave and Wireless Technology Letters · doi:10.1109/lmwt.2024.3391660 · Cited by 15 · see citing works
Wearable Blood Pressure Monitoring Devices: Understanding Heterogeneity in Design and Evaluation
Cheung, M. Y., Sabharwal, A., Coté, G. L., & Veeraraghavan, A. (2024) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2024.3434344 · Cited by 16 · see citing works
Self-citationAl-Halawani, R., Qassem, M., & Kyriacou, P. A. (2024) · Journal of Biomedical Optics · doi:10.1117/1.jbo.29.s3.s33305 · Cited by 17 · see citing works
Insights into vascular physiology from sleep photoplethysmography
Yılmaz, G., Ong, J. L., Ling, L. H., & Chee, M. W. L. (2023) · SLEEP · doi:10.1093/sleep/zsad172 · Cited by 18 · see citing works
Akther, M., Quezada, A. K., Hossain, M. A., Salas, J. I., Alam, M. M., & Uddin, M. J. (2025) · Nano Trends · doi:10.1016/j.nwnano.2025.100165 · Cited by 14 · see citing works
Liu, H., Pan, F., Lei, X., Hui, J., Gong, R., Feng, J., & Zheng, D. (2023) · Frontiers in Physiology · doi:10.3389/fphys.2023.1085871 · Cited by 14 · see citing works
Zhang, Y., Qiu, S., Du, K., Wu, S. P., Xiang, T., Zheng, K., Liu, Z., Chen, H., Ji, N., Wang, F., et al. (2026) · Nano-Micro Letters · doi:10.1007/s40820-025-02003-9 · Cited by 16 · see citing works
Davydov, R. V., Zaitceva, A. Y., Davydov, V. V., Isakova, D. D., & Mazing, M. S. (2023) · Journal of Personalized Medicine · doi:10.3390/jpm13030443 · Cited by 13 · see citing works
Quantification of the Phenomena Affecting Reflective Arterial Photoplethysmography
Rovas, G., Bikia, V., & Stergiopulos, N. (2023) · Bioengineering · doi:10.3390/bioengineering10040460 · Cited by 14 · see citing works
Hesar, M. E., Seyedsadrkhani, N. S., Khan, D., Naghashian, A., Piekarski, M., Gall, H. J., SCHERMULY, R. T., Ghofrani, H. A., & Ingebrandt, S. (2023) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2023.115693 · Cited by 13 · see citing works
Kanniainen, M., Pukkila, T., Kuisma, J., Molkkari, M., Lajunen, K., & Räsänen, E. (2023) · Frontiers in Physiology · doi:10.3389/fphys.2023.1299104 · Cited by 14 · see citing works
- Authors
- Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L.
- Affiliations
- Texas A&M University (US)
- Florida International University (US)
- Department of Biomedical Engineering, Florida International University, Miami, FL 33174, USA
- Texas A&M Engineering Experiment Station (US)
- Corresponding author
- McShane, Michael J.
- Coté, Gerard L.
- Venue
- Biosensors
- Publisher
- MDPI AG
- Published
- 2021-04-16
- Volume
- 11
- Issue
- 4
- Pages
- 126
- ISSN
- 2079-6374 (ISSN-L)
- PMC
- PMC8073123
- OpenAlex
- W3155015885
- Cited by
- 348
- References
- 212
- Citations by year
- 2026: 80 · 2025: 74 · 2024: 70 · 2023: 74 · 2022: 40 · 2021: 10
- FWCI
- 4.09
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Biosensors (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Florida International University Digital Commons (Florida International University) (submittedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- National Science Foundation (1648451)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Keywords
- photoplethysmography, wearable health monitoring, cardiovascular monitoring, motion artifacts, ambient light interference, skin tone, body temperature, heart rate variability, blood pressure estimation, respiration, venous pulsation
- MeSH terms
- Wearable Electronic Devices, Blood Pressure, Heart Rate, Humans, Monitoring, Physiologic, Respiration, Signal Processing, Computer-Assisted, Artifacts, Photoplethysmography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review