Performance of a new hand-held device for exhaled nitric oxide measurement in adults and children
Open accessThe DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
Abstract
Background: Exhaled nitric oxide (NO) measurement has been shown to be a valuable tool in the management of patients with asthma. Up to now, most measurements have been done with stationary, chemiluminescence-based NO analysers, which are not suitable for the primary health care setting. A hand-held NO analyser which simplifies the measurement would be of value both in specialized and primary health care. In this study, the performance of a new electrochemical hand-held device for exhaled NO measurements (NIOX MINO) was compared with a standard stationary chemiluminescence unit (NIOX). Methods: A total of 71 subjects (6–60 years; 36 males), both healthy controls and atopic patients with and without asthma were included. The mean of three approved exhalations (50 ml/s) in each device, and the first approved measurement in the hand-held device, were compared with regard to NO readings (Bland-Altman plots), measurement feasibility (success rate with 6 attempts) and repeatability (intrasubject SD). Results: Success rate was high (≥ 84%) in both devices for both adults and children. The subjects represented a FE NO range of 8–147 parts per billion (ppb). When comparing the mean of three measurements (n = 61), the median of the intrasubject difference in exhaled NO for the two devices was -1.2 ppb; thus generally the hand-held device gave slightly higher readings. The Bland-Altman plot shows that the 95% limits of agreement were -9.8 and 8.0 ppb. The intrasubject median difference between the NIOX and the first approved measurement in the NIOX MINO was -2.0 ppb, and limits of agreement were -13.2 and 10.2 ppb. The median repeatability for NIOX and NIOX MINO were 1.1 and 1.2 ppb, respectively. Conclusion: The hand-held device (NIOX MINO) and the stationary system (NIOX) are in clinically acceptable agreement both when the mean of three measurements and the first approved measurement (NIOX MINO) is used. The hand-held device shows good repeatability, and it can be used successfully on adults and most children. The new hand-held device will enable the introduction of exhaled NO measurements into the primary health care.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
(2005) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.200406-710st · Cited by 3,203
Increased amount of nitric oxide in exhaled air of asthmatics
Alving, K., Weitzberg, E., & Lundberg, J. (1993) · European Respiratory Journal · doi:10.1183/09031936.93.06091368 · Cited by 1,232 · cited 1×
“Background Since the original reports of the presence of nitric oxide (NO) in exhaled breath of mammals including humans [1], and of the increased levels in subjects with asthma [2], there has been a rapidly increasing interest in the measurement of exhaled NO.”
BackgroundSelf-citationEndogenous nitric oxide is present in the exhaled air of rabbits, guinea pigs and humans
Gustafsson, L. E., Leone, A. M., Persson, M. G., Wiklund, N. P., & Moncada, S. (1991) · Biochemical and Biophysical Research Communications · doi:10.1016/0006-291x(91)91268-h · Cited by 964
Cited by
Menzies, D., Nair, A., Williamson, P. A., Schembri, S., Al-Khairalla, M. Z. H., Barnes, M., Fardon, T. C., McFarlane, L., Magee, G. J., & Lipworth, B. J. (2006) · JAMA · doi:10.1001/jama.296.14.1742 · Cited by 278
Portable Exhaled Nitric Oxide Measurement
Menzies, D., Nair, A., & Lipworth, B. J. (2007) · Chest · doi:10.1378/chest.06-1335 · Cited by 177
Coggan, A. R., Leibowitz, J. L., Kadkhodayan, A., Thomas, D. P., Ramamurthy, S., Spearie, C. A., Waller, S., Farmer, M., & Peterson, L. R. (2015) · Nitric Oxide · doi:10.1016/j.niox.2014.08.014 · Cited by 159
Metadata
- Authors
- Alving, K., Janson, C., & Nordvall, L.
- Venue
- Respiratory Research
- Published
- 2006-04-20
- Volume
- 7
- Issue
- 1
- Pages
- 67-67
- ISSN
- 1465-9921 (ISSN-L)
- 1465-993X
- OpenAlex
- W1878613384
- Cited by
- 191
- References
- 3
- FWCI
- 10.00
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.