Cardiovascular magnetic resonance physics for clinicians: part I
Open accessThe DOI matches its Crossref record; 22 of the 22 references it cites were found in a registry or an index.
Abstract
There are many excellent specialised texts and articles that describe the physical principles of cardiovascular magnetic resonance (CMR) techniques. There are also many texts written with the clinician in mind that provide an understandable, more general introduction to the basic physical principles of magnetic resonance (MR) techniques and applications. There are however very few texts or articles that attempt to provide a basic MR physics introduction that is tailored for clinicians using CMR in their daily practice. This is the first of two reviews that are intended to cover the essential aspects of CMR physics in a way that is understandable and relevant to this group. It begins by explaining the basic physical principles of MR, including a description of the main components of an MR imaging system and the three types of magnetic field that they generate. The origin and method of production of the MR signal in biological systems are explained, focusing in particular on the two tissue magnetisation relaxation properties (T1 and T2) that give rise to signal differences from tissues, showing how they can be exploited to generate image contrast for tissue characterisation. The method most commonly used to localise and encode MR signal echoes to form a cross sectional image is described, introducing the concept of k-space and showing how the MR signal data stored within it relates to properties within the reconstructed image. Before describing the CMR acquisition methods in detail, the basic spin echo and gradient pulse sequences are introduced, identifying the key parameters that influence image contrast, including appearances in the presence of flowing blood, resolution and image acquisition time. The main derivatives of these two pulse sequences used for cardiac imaging are then described in more detail. Two of the key requirements for CMR are the need for data acquisition first to be to be synchronised with the subject's ECG and to be fast enough for the subject to be able to hold their breath. Methods of ECG synchronisation using both triggering and retrospective gating approaches, and accelerated data acquisition using turbo or fast spin echo and gradient echo pulse sequences are therefore outlined in some detail. It is shown how double inversion black blood preparation combined with turbo or fast spin echo pulse sequences acquisition is used to achieve high quality anatomical imaging. For functional cardiac imaging using cine gradient echo pulse sequences two derivatives of the gradient echo pulse sequence; spoiled gradient echo and balanced steady state free precession (bSSFP) are compared. In each case key relevant imaging parameters and vendor-specific terms are defined and explained.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
RARE imaging: A fast imaging method for clinical MR
Hennig, J., Nauerth, A., & Friedburg, H. (1986) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910030602 · Cited by 2,081 · cited 1×
“The turbo or fast spin echo pulse sequence [20,21] generates multiple echoes by applying multiple 180° pulses after the initial 90° pulse (Figure 22).”
Cardiovascular T2-star (T2*) magnetic resonance for the early diagnosis of myocardial iron overload
Anderson, L. (2001) · European Heart Journal · doi:10.1053/euhj.2001.2822 · Cited by 1,682 · cited 1×
“For T2*-weighted gradient echo, the image contrast is strongly influenced by the presence of magnetic susceptibility effects and can be used to detect the presence of iron, for example where there is haemorrhage or iron loading of tissue [14].”
Principles and applications of balanced SSFP techniques
Scheffler, K., & Lehnhardt, S. (2003) · European Radiology · doi:10.1007/s00330-003-1957-x · Cited by 714 · cited 2×
“After a number of repetitions this gives rise to a steady state condition where the transverse magnetisation from two or three successive repetition periods combine to give a much greater signal [15,16].”
"Black blood" T2-weighted inversion-recovery MR imaging of the heart.
Simonetti, O. P., Finn, J. P., White, R. D., Laub, G., & Henry, D. A. (1996) · Radiology · doi:10.1148/radiology.199.1.8633172 · Cited by 560 · cited 2×
“In order to improve the effectiveness of black blood imaging it is common to use a black blood magnetisation preparation scheme in combination with the spin echo pulse sequence [9,22].”
Cine MR Angiography of the Heart with Segmented True Fast Imaging with Steady-State Precession
Carr, J. C., Simonetti, O., Bundy, J., Li, D., Pereles, S., & Finn, J. P. (2001) · Radiology · doi:10.1148/radiology.219.3.r01jn44828 · Cited by 468 · cited 1×
“5 Tesla the balanced SSFP gradient echo sequence is used for most functional imaging applications and volumetric assessments due to the high intrinsic contrast it achieves between blood and myocardium throughout the cardiac cycle [30-33].”
Rapid NMR imaging of dynamic processes using the FLASII technique
Frahm, J., Haase, A., & Matthaei, D. (1986) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910030217 · Cited by 381 · cited 1×
“The use of a low flip angle is important for spoiled gradient echo techniques [12,13] as it allows the TR to be reduced to much lower values than are possible for spin echo techniques (Figure 19).”
Barkhausen, J., Ruehm, S. G., Goyen, M., Buck, T., Laub, G., & Debatin, J. F. (2001) · Radiology · doi:10.1148/radiology.219.1.r01ap12264 · Cited by 380 · cited 1×
“5 Tesla the balanced SSFP gradient echo sequence is used for most functional imaging applications and volumetric assessments due to the high intrinsic contrast it achieves between blood and myocardium throughout the cardiac cycle [30-33].”
Cardiovascular Magnetic Resonance
Pennell, D. J. (2010) · Circulation · doi:10.1161/circulationaha.108.811547 · Cited by 338
Cardiac SSFP imaging at 3 Tesla
Schär, M., Kozerke, S., Fischer, S. E., & Boesiger, P. (2004) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20024 · Cited by 307
Spoiling of transverse magnetization in steady‐state sequences
Zur, Y., Wood, M. L., & Neuringer, L. J. (1991) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910210210 · Cited by 306 · cited 1×
“gradient at the end of each TR period, or by using a technique known as rf spoiling [11] so that its contribution to subsequent TR periods is suppressed.”
Cited by
Magnetic nanoparticles in nanomedicine: a review of recent advances
Wu, K., Su, D., Liu, J., Saha, R., & Wang, J.-P. (2019) · Nanotechnology · doi:10.1088/1361-6528/ab4241 · Cited by 509
T1 Mapping in Characterizing Myocardial Disease: A Comprehensive Review
Puntmann, V. O., Peker, E., Chandrashekhar, Y., & Nagel, E. (2016) · Circulation Research · doi:10.1161/circresaha.116.307974 · Cited by 337
Magnetic Nanomaterials as Contrast Agents for MRI
Caspani, S., Magalhães, R., Araújo, J. P., & Sousa, C. T. (2020) · Materials · doi:10.3390/ma13112586 · Cited by 229
“This parameter is defined as the time required for Mz to recover to approximately 63% of its equilibrium value after the application of an RF pulse, as represented in Figure 1 [13].”
BackgroundCardiovascular magnetic resonance artefacts
Ferreira, P. F., Gatehouse, P. D., Mohiaddin, R. H., & Firmin, D. N. (2013) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-15-41 · Cited by 191
“Also, alternatives to these sections are available elsewhere such as the excellent physics articles from Ridgway and Biglands in JCMR 2010 and 2012 [4,5].”
BackgroundRole of Cardiovascular Imaging in the Follow-Up of Patients with Fontan Circulation
Moscatelli, S., Borrelli, N., Sabatino, J., Leo, I., Avesani, M., Montanaro, C., & Di Salvo, G. (2022) · Children · doi:10.3390/children9121875 · Cited by 21
“CMR can overcome echocardiography’s limitations of suboptimal acoustic windows, operator-dependent measurements, and non-classical cardiac anatomy [10,28,29].”
Background
Metadata
- Authors
- Ridgway, J. P.
- Venue
- Journal of Cardiovascular Magnetic Resonance
- Publisher
- Elsevier BV
- Published
- 2010-10
- Volume
- 12
- Issue
- 1
- Pages
- 71
- ISSN
- 1097-6647 (ISSN-L)
- OpenAlex
- W4232468181
- Cited by
- 218
- References
- 22
- FWCI
- 4.58
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.