Systematisation of spatial uncertainties for comparison between a MR and a CT-based radiotherapy workflow for prostate treatments
Open accessThe DOI matches its Crossref record; 18 of the 18 references it cites were found in a registry or an index.
Abstract
BACKGROUND: In the present work we compared the spatial uncertainties associated with a MR-based workflow for external radiotherapy of prostate cancer to a standard CT-based workflow. The MR-based workflow relies on target definition and patient positioning based on MR imaging. A solution for patient transport between the MR scanner and the treatment units has been developed. For the CT-based workflow, the target is defined on a MR series but then transferred to a CT study through image registration before treatment planning, and a patient positioning using portal imaging and fiducial markers. METHODS: An "open bore" 1.5T MRI scanner, Siemens Espree, has been installed in the radiotherapy department in near proximity to a treatment unit to enable patient transport between the two installations, and hence use the MRI for patient positioning. The spatial uncertainty caused by the transport was added to the uncertainty originating from the target definition process, estimated through a review of the scientific literature. The uncertainty in the CT-based workflow was estimated through a literature review. RESULTS: The systematic uncertainties, affecting all treatment fractions, are reduced from 3-4 mm (1Sd) with a CT based workflow to 2-3 mm with a MR based workflow. The main contributing factor to this improvement is the exclusion of registration between MR and CT in the planning phase of the treatment. CONCLUSION: Treatment planning directly on MR images reduce the spatial uncertainty for prostate treatments.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Multimodality image registration by maximization of mutual information
Maes, F., Collignon, A., Vandermeulen, D., Marchal, G., & Suetens, P. (1997) · IEEE Transactions on Medical Imaging · doi:10.1109/42.563664 · Cited by 4,547 · cited 1×
“The trend is, however, to use mutual information (MI) registration based directly on the patient anatomy [27,28].”
Multi-modal volume registration by maximization of mutual information
Wells, W. M., Viola, P., Atsumi, H., Nakajima, S., & Kikinis, R. (1996) · Medical Image Analysis · doi:10.1016/s1361-8415(01)80004-9 · Cited by 1,959 · cited 1×
“The trend is, however, to use mutual information (MI) registration based directly on the patient anatomy [27,28].”
van Herk, M., Remeijer, P., Rasch, C., & Lebesque, J. V. (2000) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/s0360-3016(00)00518-6 · Cited by 1,531 · cited 1×
“[16,17].”
Schenck, J. F. (1996) · Medical Physics · doi:10.1118/1.597854 · Cited by 1,522 · cited 1×
“air pockets in the rectal cavity, can generate susceptibility-generated field changes up to ± 10 ppm [26].”
Jovicich, J., Czanner, S., Greve, D., Haley, E., van der Kouwe, A., Gollub, R., Kennedy, D., Schmitt, F., Brown, G., MacFall, J., et al. (2006) · NeuroImage · doi:10.1016/j.neuroimage.2005.09.046 · Cited by 1,265 · cited 1×
“Even though this device specific corrections only correct for intrinsic gradient non-linearity connected to a specific type of scanner/gradient coil, it has been shown that this kind of correction yields a spatial accuracy better than 2% [23,25], which is sufficient when region of interest in the patient is close to t…”
Definition of the prostate in CT and MRI: a multi-observer study
Rasch, C., Barillot, I., Remeijer, P., Touw, A., van Herk, M., & Lebesque, J. V. (1999) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/s0360-3016(98)00351-4 · Cited by 565 · cited 1×
“Uncertainty in prostate delineation Rasch et al [ 18 ] has from a study with 18 patient analysed by 3 physicians reported an uncertainty, in the prostate delineation on axial MR study, of 2 mm at the base of seminal vesicles and up 2.8 mm in the prostate apex.”
InfluentialLagendijk, J. J. W., Raaymakers, B. W., Raaijmakers, A. J. E., Overweg, J., Brown, K. J., Kerkhof, E. M., van der Put, R. W., Hårdemark, B., van Vulpen, M., & van der Heide, U. A. (2008) · Radiotherapy and Oncology · doi:10.1016/j.radonc.2007.10.034 · Cited by 511
Moseley, D. J., White, E. A., Wiltshire, K. L., Rosewall, T., Sharpe, M. B., Siewerdsen, J. H., Bissonnette, J., Gospodarowicz, M. K., Warde, P. R., Catton, C., et al. (2007) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2006.10.039 · Cited by 304 · cited 1×
“Literature indicates that a manual registration typically results in uncertainty of around 0.7 mm in the HF and RL direction, and 1.4 mm in the AP direction [37, 38 ].”
InfluentialTarget Definition in Prostate, Head, and Neck
Rasch, C., Steenbakkers, R., & van Herk, M. (2005) · Seminars in Radiation Oncology · doi:10.1016/j.semradonc.2005.01.005 · Cited by 210 · cited 1×
“Background MR images are well suited for target delineation, not only for the prostate [1], but also for many other tumours, such as brain lesions [2,3] and head and neck tumours [4,5], which explains the growing interest for MR in radiotherapy [6-12].”
Prostate volume contouring: A 3D analysis of segmentation using 3DTRUS, CT, and MR
Smith, W. L., Lewis, C., Bauman, G., Rodrigues, G., D’Souza, D., Ash, R., Ho, D., Venkatesan, V., Downey, D., & Fenster, A. (2007) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2006.11.027 · Cited by 209 · cited 1×
“[19] reported a radial uncertainty of 0.”
Nichol, A. M., Brock, K. K., Lockwood, G. A., Moseley, D. J., Rosewall, T., Warde, P. R., Catton, C. N., & Jaffray, D. A. (2007) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2006.08.021 · Cited by 176 · cited 1×
“An investigation by Nichol et al [39] indicates that a systematic deformation of the prostate during radiotherapy leads to drift in the relation between the centre of mass for the markers and centre of mass for the contoured prostate.”
MRI-based treatment planning for radiotherapy: Dosimetric verification for prostate IMRT
Chen, L., Price, R. A., Wang, L., Li, J., Qin, L., McNeeley, S., Ma, C.-M. C., Freedman, G. M., & Pollack, A. (2004) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2004.05.068 · Cited by 172 · cited 1×
“Background MR images are well suited for target delineation, not only for the prostate [1], but also for many other tumours, such as brain lesions [2,3] and head and neck tumours [4,5], which explains the growing interest for MR in radiotherapy [6-12].”
Beltran, C. J., Herman, M., & Davis, B. J. (2007) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2007.08.040 · Cited by 155 · cited 1×
“[42].”
MR imaging and MR spectroscopic imaging in the pre-treatment evaluation of prostate cancer
Hricak, H. (2005) · British Journal of Radiology · doi:10.1259/bjr/11253478 · Cited by 138
Roberson, P. L., McLaughlin, P. W., Narayana, V., Troyer, S., Hixson, G. V., & Kessler, M. L. (2005) · Medical Physics · doi:10.1118/1.1851920 · Cited by 93 · cited 2×
“A few studies have been performed evaluating the accuracy and precision of MI registration for CT and MR studies of the prostate; the registration uncertainty has been reported to be around 2 mm [29,31].”
Residual set-up errors and margins in on-line image-guided prostate localization in radiotherapy
Poulsen, P. R., Muren, L. P., & Høyer, M. (2007) · Radiotherapy and Oncology · doi:10.1016/j.radonc.2007.08.006 · Cited by 48 · cited 2×
“Paulsen et al [34] observed a systematic discrepancy of almost 1 mm when comparing 2 different imaging modalities at 2 different accelerators.”
The use of mutual information in registration of CT and MRI datasets post permanent implant
McLaughlin, P. W., Narayana, V., Kessler, M. L., McShan, D. L., Troyer, S., Marsh, L. H., Hixson, G. V., & Roberson, P. L. (2004) · Brachytherapy · doi:10.1016/j.brachy.2004.06.001 · Cited by 47 · cited 2×
“A few studies have been performed evaluating the accuracy and precision of MI registration for CT and MR studies of the prostate; the registration uncertainty has been reported to be around 2 mm [29,31].”
Chen, L., Nguyen, T.-B., Jones, É., Chen, Z., Luo, W., Wang, L., Price, R., Pollack, A. S., & Ma, C. (2007) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2007.02.033 · Cited by 43 · cited 1×
“Background MR images are well suited for target delineation, not only for the prostate [1], but also for many other tumours, such as brain lesions [2,3] and head and neck tumours [4,5], which explains the growing interest for MR in radiotherapy [6-12].”
Cited by
A review of substitute CT generation for MRI-only radiation therapy
Edmund, J. M., & Nyholm, T. (2017) · Radiation Oncology · doi:10.1186/s13014-016-0747-y · Cited by 378
Self-citationJohansson, A., Karlsson, M., & Nyholm, T. (2011) · Medical Physics · doi:10.1118/1.3578928 · Cited by 324
Self-citationJohnstone, E., Wyatt, J. J., Henry, A. M., Short, S. C., Sebag-Montefiore, D., Murray, L., Kelly, C. G., McCallum, H. M., & Speight, R. (2018) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2017.08.043 · Cited by 317
A review on medical imaging synthesis using deep learning and its clinical applications
Wang, T., Lei, Y., Fu, Y., Wynne, J. F., Curran, W. J., Liu, T., & Yang, X. (2021) · Journal of Applied Clinical Medical Physics · doi:10.1002/acm2.13121 · Cited by 279
Maspero, M., Savenije, M. H. F., Dinkla, A. M., Seevinck, P. R., Intven, M. P. W., Jurgenliemk-Schulz, I. M., Kerkmeijer, L. G. W., & van den Berg, C. A. T. (2018) · Physics in Medicine & Biology · doi:10.1088/1361-6560/aada6d · Cited by 253
Metadata
- Authors
- Nyholm, T., Nyberg, M., Karlsson, M. G., & Karlsson, M.
- Venue
- Radiation Oncology
- Published
- 2009-11-17
- Volume
- 4
- Issue
- 1
- Pages
- 54-54
- ISSN
- 1748-717X (ISSN-L)
- OpenAlex
- W2157025461
- Cited by
- 165
- References
- 18
- FWCI
- 2.87
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.