Systematic evaluation of three different commercial software solutions for automatic segmentation for adaptive therapy in head-and-neck, prostate and pleural cancer
Open accessThe DOI matches its Crossref record; 14 of the 14 references it cites were found in a registry or an index.
Abstract
Purpose: To validate, in the context of adaptive radiotherapy, three commercial software solutions for atlas-based segmentation. Methods and materials: Fifteen patients, five for each group, with cancer of the Head&Neck, pleura, and prostate were enrolled in the study. In addition to the treatment planning CT (pCT) images, one replanning CT (rCT) image set was acquired for each patient during the RT course. Three experienced physicians outlined on the pCT and rCT all the volumes of interest (VOIs). We used three software solutions (VelocityAI 2.6.2 (V), MIM 5.1.1 (M) by MIMVista and ABAS 2.0 (A) by CMS-Elekta) to generate the automatic contouring on the repeated CT. All the VOIs obtained with automatic contouring (AC) were successively corrected manually. We recorded the time needed for: 1) ex novo ROIs definition on rCT; 2) generation of AC by the three software solutions; 3) manual correction of AC. To compare the quality of the volumes obtained automatically by the software and manually corrected with those drawn from scratch on rCT, we used the following indexes: overlap coefficient (DICE), sensitivity, inclusiveness index, difference in volume, and displacement differences on three axes (x, y, z) from the isocenter. Results: The time saved by the three software solutions for all the sites, compared to the manual contouring from scratch, is statistically significant and similar for all the three software solutions. The time saved for each site are as follows: about an hour for Head&Neck, about 40 minutes for prostate, and about 20 minutes for mesothelioma. The best DICE similarity coefficient index was obtained with the manual correction for: A (contours for prostate), A and M (contours for H&N), and M (contours for mesothelioma). Conclusions: From a clinical point of view, the automated contouring workflow was shown to be significantly shorter than the manual contouring process, even though manual correction of the VOIs is always needed.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Measures of the Amount of Ecologic Association Between Species
Dice, L. R. (1945) · Ecology · doi:10.2307/1932409 · Cited by 11,874 · cited 1×
“Since its introduction, DICE similarity coefficient (DSC) index [13] has been widely used in the evaluation of deformable image registration results.”
MethodMohler, J., Bahnson, R. R., Boston, B., Busby, J. E., D'Amico, A., Eastham, J. A., Enke, C. A., George, D., Horwitz, E. M., Huben, R. P., et al. (2010) · Journal of the National Comprehensive Cancer Network · doi:10.6004/jnccn.2010.0012 · Cited by 897
Markou, K. D., Nikolaou, A. C., & Themelis, C. (2011) · Cited by 757
The 2010 NCCN Clinical Practice Guidelines in Oncology on Prostate Cancer
Mohler, J. L. (2010) · Journal of the National Comprehensive Cancer Network · doi:10.6004/jnccn.2010.0010 · Cited by 614
Atlas-Based Auto-segmentation of Head and Neck CT Images
Han, X., Hoogeman, M. S., Levendag, P. C., Hibbard, L. S., Teguh, D. N., Voet, P., Cowen, A. C., & Wolf, T. K. (2008) · Lecture Notes in Computer Science · doi:10.1007/978-3-540-85990-1_52 · Cited by 195
Teguh, D. N., Levendag, P. C., Voet, P. W. J., Al-Mamgani, A., Han, X., Wolf, T. K., Hibbard, L. S., Nowak, P., Akhiat, H., Dirkx, M. L. P., et al. (2011) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2010.07.009 · Cited by 194
Zhang, T., Chi, Y., Meldolesi, E., & Yan, D. (2007) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2007.01.038 · Cited by 177 · cited 1×
“As underlined in the literature [18], mapping planning contours to daily diagnostic CT images, instead of daily MVCT or kV CT, would facilitate adaptive replanning.”
BackgroundTracking the dose distribution in radiation therapy by accounting for variable anatomy
Schaly, B., Kempe, J., Bauman, G. S., Battista, J. J., & Dyk, J. V. (2004) · Physics in Medicine and Biology · doi:10.1088/0031-9155/49/5/010 · Cited by 146 · cited 2×
“Discussion The need to replan and adapt treatment for internal anatomy variations due to tumor shrinkage and shape deformation [1,3] has increased over the years in order to make better use of highly conformal treatment techniques.”
BackgroundAutomatic Segmentation of Whole Breast Using Atlas Approach and Deformable Image Registration
Reed, V. K., Woodward, W. A., Zhang, L., Strom, E. A., Perkins, G. H., Tereffe, W., Oh, J., Yu, T. K., Bedrosian, I., Whitman, G. J., et al. (2008) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2008.07.001 · Cited by 121
Dosimetric Evaluation of Automatic Segmentation for Adaptive IMRT for Head-and-Neck Cancer
Tsuji, S. Y., Hwang, A., WEINBERG, V. K., Yom, S. S., Quivey, J. M., & Xia, P. (2010) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2009.06.012 · Cited by 116 · cited 3×
“In this context, a fast, robust, and automatic regionof-interest (ROI) delineation method is needed and has led to a growing array of automatic contouring (AC) software [5-7].”
BackgroundVoet, P. W. J., Dirkx, M. L. P., Teguh, D. N., Hoogeman, M. S., Levendag, P. C., & Heijmen, B. J. M. (2011) · Radiotherapy and Oncology · doi:10.1016/j.radonc.2010.11.017 · Cited by 115 · cited 2×
“[6] found underdosages in the PTV of up to 11 Gy even for DSC coefficients of 0.”
Backgroundvan Dam, I. E., van Sörnsen de Koste, J. R., Hanna, G. G., Muirhead, R., Slotman, B. J., & Senan, S. (2010) · Radiotherapy and Oncology · doi:10.1016/j.radonc.2010.05.003 · Cited by 69 · cited 1×
“VelocityAI uses the basis-spline (B-spline) method [12] for deformable registrations.”
MethodA qualitative and a quantitative analysis of an auto-segmentation module for prostate cancer
Huyskens, D. P., Maingon, P., Vanuytsel, L., Remouchamps, V. M., Roques, T. W., Dubray, B. M., Haas, B., Kunz, P., Coradi, T., Bühlman, R., et al. (2008) · Radiotherapy and Oncology · doi:10.1016/j.radonc.2008.08.007 · Cited by 68 · cited 1×
“In this context, a fast, robust, and automatic regionof-interest (ROI) delineation method is needed and has led to a growing array of automatic contouring (AC) software [5-7].”
BackgroundTechnology assessment of automated atlas based segmentation in prostate bed contouring
Hwee, J., Louie, A. V., Gaede, S., Bauman, G. S., D'Souza, D. P., Sexton, T., Lock, M. I., Ahmad, B., & Rodrigues, G. (2011) · Radiation Oncology · doi:10.1186/1748-717x-6-110 · Cited by 58 · cited 1×
“The use of atlas-based tools to delineate OARs for cancer sites including H&N [11,14], breast [15], endometrium [16] and prostate [17] have shown to reduce volume delineation variability and the total time required to contour.”
Cited by
Segmentation of organs‐at‐risks in head and neck CT images using convolutional neural networks
Ibragimov, B., & Xing, L. (2017) · Medical Physics · doi:10.1002/mp.12045 · Cited by 502
“0 software systems demonstrated better performance of 81%, 83%, and 84% DSC, respectively, on larynx segmentation, and 63%, 66%, and 65% DSC, respectively, on pharynx segmentation.(45) Clinicians demonstrated an agreement of 60%(48) and 83%(40) DSC for larynx segmentation.”
InfluentialVision 20/20: Perspectives on automated image segmentation for radiotherapy
Sharp, G., Fritscher, K. D., Pekar, V., Peroni, M., Shusharina, N., Veeraraghavan, H., & Yang, J. (2014) · Medical Physics · doi:10.1118/1.4871620 · Cited by 375
Nikolov, S., Blackwell, S., Zverovitch, A., Mendes, R., Livne, M., De Fauw, J., Patel, Y., Meyer, C., Askham, H., Romera-Paredes, B., et al. (2021) · Journal of Medical Internet Research · doi:10.2196/26151 · Cited by 366
Wong, J., Fong, A., McVicar, N., Smith, S., Giambattista, J., Wells, D., Kolbeck, C., Giambattista, J., Gondara, L., & Alexander, A. (2020) · Radiotherapy and Oncology · doi:10.1016/j.radonc.2019.10.019 · Cited by 246
Autosegmentation for thoracic radiation treatment planning: A grand challenge at AAPM 2017
Yang, J., Veeraraghavan, H., Armato, S. G., Farahani, K., Kirby, J. S., Kalpathy‐Kramer, J., van Elmpt, W., Dekker, A., Han, X., Feng, X., et al. (2018) · Medical Physics · doi:10.1002/mp.13141 · Cited by 240
Metadata
- Authors
- La Macchia, M., Fellin, F., Amichetti, M., Cianchetti, M., Gianolini, S., Paola, V., Lomax, A. J., & Widesott, L.
- Venue
- Radiation Oncology
- Published
- 2012-09-18
- Volume
- 7
- Issue
- 1
- Pages
- 160-160
- ISSN
- 1748-717X (ISSN-L)
- OpenAlex
- W2108057069
- Cited by
- 172
- References
- 14
- FWCI
- 6.54
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.