Accuracy and feasibility of point-of-care and continuous blood glucose analysis in critically ill ICU patients
Open accessAbstract
Introduction: To obtain strict glucose regulation, an accurate and feasible bedside glucometry method is essential. We evaluated three different types of point-of-care glucometry in seriously ill intensive care unit (ICU) patients. The study was performed as a single-centre, prospective, observational study in a 12-bed medical ICU of a university hospital. Methods: Patients with an expected ICU stay of more than 48 hours were included. Because the reference laboratory delivers glucose values after approximately 30 to 60 minutes, which is too slow to use in a glucose regulation protocol and for calibration of the subcutaneous continuous glucose monitoring system (CGMS) (CGMS System Gold), we first validated the ICU-based blood gas/glucose analyser ABL715 (part 1 of the study). Subsequently, part 2 was performed: after inserting (and calibrating) the subcutaneous CGMS, heparinised arterial blood samples were drawn from an arterial line every 6 hours and analysed on both the Precision PCx point-of-care meter using test strips and on the blood gas/glucose analyser ABL715. CGMS glucose data were downloaded after 24 to 72 hours. The results of the paired measurements were analysed as a scatter plot by the method of Bland and Altman and were expressed as a correlation coefficient. Results: Part 1: Four hundred and twenty-four blood samples were drawn from 45 critically ill ICU patients. The ICU-based blood gas/glucose analyser ABL715 provided a good estimate of conventional laboratory glucose assessment: the correlation coefficient was 0.95. In the Clarke error grid, 96.8% of the paired measurements were in the clinically acceptable zones A and B. Part 2: One hundred sixty-five paired samples were drawn from 19 ICU patients. The Precision PCx point-of-care meter showed a correlation coefficient of 0.89. Ninety-eight point seven percent of measurements were within zones A and B. The correlation coefficient for the subcutaneous CGMS System Gold was 0.89. One hundred percent of measurements were within zones A and B. Conclusion: The ICU-based blood glucose analyser ABL715 is a rapid and accurate alternative for laboratory glucose determination and can serve as a standard for ICU blood glucose measurements. The Precision PCx is a good alternative, but feasibility may be limited because of the blood sample handling. The subcutaneous CGMS System Gold is promising, but real-time glucose level reporting is necessary before it can be of clinical use in the ICU. When implementing a glucose-insulin algorithm in patient care or research, one should realise that the absolute glucose level may differ systematically among various measuring methods, influencing targeted glucose levels.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Intensive insulin therapy in the medical ICU
Van den Berghe, G., Wilmer, A., Hermans, G., Meersseman, W., Wouters, P. J., Milants, I., Van Wijngaerden, E., Bobbaers, H., & Bouillon, R. (2006) · New England Journal of Medicine · doi:10.1056/nejmoa052521 · Cited by 3,639
Hyperglycemia: an independent marker of in-hospital mortality in patients with undiagnosed diabetes
Umpierrez, G. E., Isaacs, S. D., Bazargan, N., You, X., Thaler, L. M., & Kitabchi, A. E. (2002) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jcem.87.3.8341 · Cited by 2,195
Capes, S. E., Hunt, D., Malmberg, K., & Gerstein, H. C. (2000) · The Lancet · doi:10.1016/s0140-6736(99)08415-9 · Cited by 2,187
Cited by
Point-of-care testing (POCT): Current techniques and future perspectives
Luppa, P. B., Müller, C., Schlichtiger, A., & Schlebusch, H. (2011) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2011.01.019 · Cited by 660
Jacobi, J., Bircher, N., Krinsley, J., Agus, M., Braithwaite, S. S., Deutschman, C., Freire, A. X., Geehan, D., Kohl, B., Nasraway, S. A., et al. (2012) · Critical Care Medicine · doi:10.1097/ccm.0b013e3182653269 · Cited by 571
Krinsley, J. S., Egi, M., Kiss, A., Devendra, A. N., Schuetz, P., Maurer, P. M., Schultz, M. J., van Hooijdonk, R. T., Kiyoshi, M., Mackenzie, I. M., et al. (2013) · Critical Care · doi:10.1186/cc12547 · Cited by 369
Perioperative Glycemic Control: An Evidence-based Review
Lipshutz, A. K. M., Gropper, M. A., Warner, D. S., & Warner, M. A. (2009) · Anesthesiology · doi:10.1097/aln.0b013e3181948a80 · Cited by 259
Continuous glucose monitoring: an Endocrine Society Clinical Practice Guideline
Klonoff, D. C., Buckingham, B., Christiansen, J. S., Montori, V. M., Tamborlane, W. V., Vigersky, R. A., & Wolpert, H. (2011) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2010-2756 · Cited by 211
Metadata
- Authors
- Corstjens, A. M., Ligtenberg, J. J., van der Horst, I. C., Spanjersberg, R., Lind, J. S., Tulleken, J. E., Meertens, J. H., & Zijlstra, J. G.
- Venue
- Critical Care
- Published
- 2006-09-18
- Volume
- 10
- Issue
- 5
- Pages
- R135-R135
- ISSN
- 1364-8535 (ISSN-L)
- OpenAlex
- W2113662426
- Cited by
- 163
- References
- 4
- FWCI
- 6.86
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Accuracy and feasibility of point-of-care and continuous blood glucose analysis in critically ill ICU patients. https://app.cochat.ai/papers/10.1186/cc5048
- The paper
- Corstjens, A. M., Ligtenberg, J. J., van der Horst, I. C., Spanjersberg, R., Lind, J. S., Tulleken, J. E., Meertens, J. H., & Zijlstra, J. G. (2006). Accuracy and feasibility of point-of-care and continuous blood glucose analysis in critically ill ICU patients. Critical Care. https://doi.org/10.1186/cc5048
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.