Short- and long-term reproducibility of peripheral superficial vein depth and diameter measurements using ultrasound imaging
The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.
Abstract
Background: Ultrasound imaging is used for diagnosis, treatment, and blood vessel visualization during venous catheter placement. However, various physiological factors (e.g., body temperature and exercise) influence vein diameters, which are expected to exhibit daily or diurnal fluctuations. Therefore, this study aimed to determine the intraday (short-term) and interday (long-term) reproducibility of repeated measurements of the depth and diameter of peripheral superficial veins. Methods: Twenty-three healthy young women (mean age, 21.7 ± 0.8 years) participated in the study to examine the short- and long-term reproducibility of the depth and diameter of the cutaneous vein in the left elbow fossa acquired by ultrasound imaging. Short-term measurement intervals were 10 s, and the probe was released from the skin for each acquisition, which was repeated five consecutive times. Long-term measurements were performed at the same time on the next day following the same procedure. The acquired images were analyzed for vein depth and diameter using ImageJ software. The intraclass correlation coefficient (ICC) was calculated to determine the short- and long-term reproducibility of the measurements. The relationship between the venous depth and venous diameter intra-individual variation was analyzed, as well as the influence of body composition (body fat and muscle mass) on the venous diameter and depth. Results: For vein depth measurements, the short- and long-term ICCs were 0.94–0.96 and 0.88, respectively. For the vein diameter, the short- and long-term ICCs were 0.94–0.97 and 0.67, respectively. The short-term ICCs for both vein depth and diameter exceeded 0.9, indicating that the ultrasound vascular measurement was sufficiently reliable. However, long-term reproducibility was slightly lower, especially for the vein diameter. No correlation was found between the intra-individual variation of the vein diameter and vein depth. Although the vein diameter and body fat mass uncorrelated, the vein depth and body fat mass significantly correlated (r = 0.675, 95% confidence interval = 0.281–0.830). Conclusions: The long-term reproducibility of vein diameters was somewhat lower than that of the short-term reproducibility. This could be attributed to fluctuations in the physiological state of the participant rather than to the instability of the measurement. Therefore, ultrasound measurement of the peripheral superficial vein is sufficiently reliable.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Repeated measurements can show biological variation that exceeds inter-observer measurement error [ 15 ], underscoring that real-time measurement is more useful than relying on age/size charts alone.”
From Ultrasound vessel sizing before cannulation: a measurement-first approach for anesthesiologists (2026) · cites this paper as [15]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
van Loon, F. H. J., Buise, M. P., Claassen, J. J. F., Dierick-van Daele, A. T. M., & Bouwman, A. R. A. (2018) · British Journal of Anaesthesia · doi:10.1016/j.bja.2018.04.047 · Cited by 195
van Loon, F. H. J., van Hooff, L. W. E., de Boer, H. D., Koopman, S. S. H. A., Buise, M. P., Korsten, H. H. M., Dierick-van Daele, A. T. M., & Bouwman, A. R. A. (2019) · Journal of Clinical Medicine · doi:10.3390/jcm8020144 · Cited by 157
Research works printed but not linked: 32
- Lee L, DeCara JM. Point-of-care ultrasound. Curr Cardiol Rep. 2020;22:149. · DOI
- Mariani PJ, Setla JA. Palliative ultrasound for home care hospice patients. Acad Emerg Med. 2010;17:293–6. · doi:10.1111/j.1553-2712.2009.00678.x
- Steinwandel U, Gibson NP, Rippey JC, Towell A, Rosman J. Use of ultrasound by registered nurses-a systematic literature review. J Ren Care. 2017;43:132–42. · doi:10.1111/jorc.12191
- Totenhofer R, Luck L, Wilkes L. Point of care ultrasound use by registered nurses and nurse practitioners in clinical practice: an integrative review. Collegian. 2021;28:456–63. · DOI
- Suzuki M, Iguchi Y, Igawa Y, Yoshida M, Sanada H, Miyazaki H, Homma Y. Ultrasound-assisted prompted voiding for management of urinary incontinence of nursing home residents: efficacy and feasibility. Int J Urol. 2016;23:786–90. · doi:10.1111/iju.13156
- Matsumoto M, Yoshida M, Yabunaka K, Nakagami G, Miura Y, Fujimaki S, et al. Safety and efficacy of a defecation care algorithm based on ultrasonographic bowel observation in Japanese home-care settings: a single-case, multiple-baseline study. Geriatr Gerontol Int. 2020;20:187–94. · DOI
- Lan Y, Li N, Song Q, Zhang M-b, Luo Y-k, Zhang Y. Correlation and agreement between superb micro-vascular imaging and contrast-enhanced ultrasound for assessing radiofrequency ablation treatment of thyroid nodules: a preliminary study. Image and clinical analysis of common carotid web: a case report. BMC Med Imaging. 2021;21:175. · DOI
- Schlunk F, Kuthe J, Harmel P, Audebert H, Hanning U, Bohner G, et al. Volumetric accuracy of different imaging modalities in acute intracerebral hemorrhage. BMC Med Imaging. 2022;22:9. · doi:10.1186/s12880-022-00735-3
- Bahl A, Pandurangadu AV, Tucker J, Bagan M. A randomized controlled of ultrasound for nurse-performed IV placement in difficult access ED patients. Am J Emerg Med. 2016;34:1950–4. · DOI
- Duran-Gehring P, Bryant L, Reynolds JA, Aldridge P, Kalynych CJ, Guirgis FW. Ultrasound-guided peripheral intravenous catheter training results in physician-level success for emergency department technicians. J Ultrasound Med. 2016;35:2343–52. · DOI · Paper page
- Yalçınlı S, Akarca FK, Can Ö, Şener A, Akbinar C. Factors affecting the first-attempt success rate of intravenous cannulation in older people. J Clin Nurs. 2019;28:2206–13. · DOI · Paper page
- Witting MD, Schenkel SM, Lawner BJ, Euerle BD. Effects of vein width and depth on ultrasound-guided peripheral intravenous success rates. J Emerg Med. 2010;39:70–5. · DOI
- Panebianco NL, Fredette JM, Szyld D, Sagalyn EB, Pines JM, Dean AJ. What you see (sonographically) is what you get: vein and patient characteristics associated with successful ultrasound-guided peripheral intravenous placement in patients with difficult access. Acad Emerg Med. 2009;16:1298–303. · DOI · Paper page
- Yamagami Y, Tomita K, Tsujimoto T, Inoue T. Tourniquet application after local forearm warming to improve venodilation for peripheral intravenous cannulation in young and middle-aged adults: a single-blind prospective randomized controlled trial. Int J Nurs Stud. 2017;72:1–7. · DOI
- Tokizawa Y, Tsujimoto T, Inoue T. Duration of venodilation for peripheral intravenous cannulation, as induced by a thermal stimulus on the forearm. Biol Res Nurs. 2017;19:206–12. · DOI
- Yamagami Y, Tsujimoto T, Inoue T. How long should local warming for venodilation be used for peripheral intravenous cannulation? A prospective observational study. Int J Nurs Stud. 2018;79:52–7. · DOI
- Sharp R, Childs J, Bulmer AC, Esterman A. The effect of oral hydration and localised heat on peripheral vein diameter and depth: a randomised controlled trial. Appl Nurs Res. 2018;42:83–8. · DOI
- Yamagami Y, Inoue T. Patient position affects venodilation for peripheral intravenous cannulation. Biol Res Nurs. 2020;22:226–33. · DOI
- Aellig WH. Clinical pharmacology, physiology and pathophysiology of superficial veins-1. Br J Clin Pharmacol. 1994;38:181–96. · DOI
- Gignon L, Roger C, Bastide S, Alonso S, Zieleskiewicz L, Quintard H, et al. Influence of diaphragmatic motion on inferior Vena cava diameter respiratory variations in healthy volunteers. Anesthesiology. 2016;124:1338–46. · DOI
- Shibasaki S, Kishino T, Yokoyama T, Sunahara S, Harashima K, Nakajima S, et al. Sonographic detection of physiological lower leg oedema in the late afternoon in healthy young women. Clin Physiol Funct Imaging. 2020;40:381–4. · doi:10.1111/cpf.12654
- Sharp R, Gordon A, Mikocka-Walus A, Childs J, Grech C, Cummings M, Esterman A. Vein measurement by peripherally inserted central catheter nurses using ultrasound: a reliability study. JAVA J Assoc Vasc Access. 2013;18:234–8. · doi:10.1016/j.java.2013.08.001
- Steinwandel U, Gibson N, Towell A, Rippey JJR, Rosman J. Can a renal nurse assess fluid status using ultrasound on the inferior vena cava? A cross-sectional interrater study. Hemodial Int. 2018;22:261–9. · doi:10.1111/hdi.12606
- Fields JM, Lee PA, Jenq KY, Mark DG, Panebianco NL, Dean AJ. The interrater reliability of inferior Vena cava ultrasound by bedside clinician sonographers in emergency department patients. Acad Emerg Med. 2011;18:98–101. · DOI
- Mesin L, Giovinazzo T, D’Alessandro S, Roatta S, Raviolo A, Chiacchiarini F, et al. Improved repeatability of the estimation of pulsatility of inferior Vena cava. Ultrasound Med Biol. 2019;45:2830–43. · DOI
- Schindelin J, Arganda-Carreras I, Frise E, Kaynig V, Longair M, Pietzsch T, et al. Fiji: an open-source platform for biological-image analysis. Nat Methods. 2012;9:676–82. · DOI · Paper page
- Kobayashi H. Inter- and intra-individual variations of heart rate variability in Japanese males. J Physiol Anthropol. 2007;26:173–7. · DOI
- Sharp R, Cummings M, Childs J, Fielder A, Mikocka-Walus A, Grech C, Esterman A. Measurement of vein diameter for peripherally inserted central catheter (PICC) insertion: an observational study. J Infus Nurs. 2015;38:351–7. · DOI
- Kim IS, Kang SS, Park JH, Hong SJ, Shin KM, Yoon YJ, Kim MS. Impact of sex, age and BMI on depth and diameter of the infraclavicular axillary vein when measured by ultrasonography. Eur J Anaesthesiol. 2011;28:346–50. · doi:10.1097/eja.0b013e3283416674
- Ma Q, Shi X, Ji J, Chen L, Tian Y, Hao J, et al. The diagnostic accuracy of inferior vena cava respiratory variation in predicting volume responsiveness in patients under different breathing status following abdominal surgery. BMC Anesthesiol. 2022;22:1–12. · DOI
- Lim KJ, Lee JM, Byon HJ, Kim HS, Kim CS, Lee SK, et al. The effect of full expiration on the position and size of the subclavian vein in spontaneously breathing adults. Anesth Analg. 2013;117:109–13. · doi:10.1213/ane.0b013e31826257f4
- Folino A, Benzo M, Pasquero P, Laguzzi A, Mesin L, Messere A, et al. Vena cava responsiveness to controlled isovolumetric respiratory efforts. J Ultrasound Med. 2017;36:2113–23. · DOI
Cited by
Matsumoto, M., Tateishi, A., Kobayashi, H., & Hashiguchi, N. (2025) · PLOS One · doi:10.1371/journal.pone.0323367 · Cited by 5
“Prior studies have shown that vein depth was positively correlated with fat mass [23].”
Self-citationUltrasound vessel sizing before cannulation: a measurement-first approach for anesthesiologists
Linares, L. A. R., & Lopes, S. C. (2026) · Journal of Anesthesia, Analgesia and Critical Care · doi:10.1186/s44158-026-00399-3
Metadata
- Authors
- Miharu, M., Nobuko, H., & Hiromitsu, K.
- Venue
- BMC Medical Imaging
- Published
- 2022-12-02
- Volume
- 22
- Issue
- 1
- ISSN
- 1471-2342
- Cited by
- 8
- References
- 2
- Language
- en
- Licence
- creativecommons.org/licenses/by/4.0 (tdm)
- creativecommons.org/licenses/by/4.0 (vor)
- Funders
- Japan Society for the Promotion of Science (JP 20K19023)
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Short- and long-term reproducibility of peripheral superficial vein depth and diameter measurements using ultrasound imaging. https://app.cochat.ai/papers/10.1186/s12880-022-00945-9
- The paper
- Miharu, M., Nobuko, H., & Hiromitsu, K. (2022). Short- and long-term reproducibility of peripheral superficial vein depth and diameter measurements using ultrasound imaging. BMC Medical Imaging. https://doi.org/10.1186/s12880-022-00945-9
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.