Controlling a motorized orthosis to follow elbow volitional movement: tests with individuals with pathological tremor

Open access
Herrnstadt, Gil, McKeown, Martin J., Menon, Carlo · 2019 · Journal of NeuroEngineering and Rehabilitation

Abstract

BACKGROUND: There is a need for alternative treatment options for tremor patients who do not respond well to medications or surgery, either due to side effects or poor efficacy, or that are excluded from surgery. The study aims to evaluate feasibility of a voluntary-driven, speed-controlled tremor rejection approach with individuals with pathological tremor. The suppression approach was investigated using a robotic orthosis for suppression of elbow tremor. Importantly, the study emphasizes the performance in relation to the voluntary motion. METHODS: Nine participants with either Essential Tremor (ET) or Parkinson's disease (PD) were recruited and tested off medication. The participants performed computerized pursuit tracking tasks following a sinusoid and a random target, both with and without the suppressive orthosis. The impact of the Tremor Suppression Orthosis (TSO) at the tremor and voluntary frequencies was determined by the relative power change calculated from the Power Spectral Density (PSD). Voluntary motion was, in addition, assessed by position and velocity tracking errors. RESULTS: The suppressive orthosis resulted in a 94.4% mean power reduction of the tremor (p < 0.001) - a substantial improvement over reports in the literature. As for the impact to the voluntary motion, paired difference tests revealed no statistical effect of the TSO on the relative power change (p = 0.346) and velocity tracking error (p = 0.283). A marginal effect was observed for the position tracking error (p = 0.05). The interaction torque with the robotic orthosis was small (0.62 Nm) when compared to the maximum voluntary torque that can be exerted by adult individuals at the elbow joint. CONCLUSIONS: Two key contributions of this work are first, a recently proposed approach is evaluated with individuals with tremor demonstrating high levels of tremor suppression; second, the impact of the approach to the voluntary motion is analyzed comprehensively, showing limited inhibition. This study also seeks to address a gap in studies with individuals with tremor where the impact of engineering solutions on voluntary motion is unreported. This study demonstrates feasibility of the wearable technology as an effective treatment that removes tremor with limited impediment to intentional motion. The goal for such wearable technology is to help individuals with pathological tremor regain independence in activities affected by the tremor condition. Further investigations are needed to validate the technology.

Overview

CoChat Summary

## TL;DR

This preliminary study tested a voluntary-driven, speed-controlled tremor suppression orthosis on 9 participants with Essential Tremor or Parkinson’s disease and found a mean 94.4% tremor power reduction. The device produced limited interference with voluntary movement, with no significant effect on voluntary power change or velocity tracking error and only a marginal effect on position error.

## Key findings

  • In 9 off-medication participants, the Tremor Suppression Orthosis (TSO) reduced tremor power by a mean of 94.37% (SD 7.27%), with p < 0.001 versus no suppression.
  • Tremor reduction differed by task: random targets showed −96.96% (SD 2.73), slow targets −96.76% (SD 2.28), sinusoid targets −91.78% (SD 9.34), and fast targets −91.97% (SD 9.57); random and slow motions had better suppression (p = 0.037 and p = 0.015, respectively).
  • For voluntary motion, there was no statistically significant difference between TSO and measurement orthosis (MO) for relative power change (p = 0.346) or velocity tracking error (p = 0.283); position tracking error was marginally significant (p = 0.05).
  • Mean voluntary metrics were close between devices: power change 4.32% (SD 16.37) for MO vs 10.96% (SD 22.91) for TSO; position RMSE 7.3° (SD 1.9) vs 8.2° (SD 1.9); velocity RMSE 0.39 rad/s (SD 0.04) vs 0.41 rad/s (SD 0.07).

## Methods

  • Design: preliminary within-subject feasibility study comparing a Tremor Suppression Orthosis (TSO) against a Measurement Orthosis (MO) during elbow pursuit-tracking tasks.
  • Sample/data: 9 participants with Essential Tremor (ET) or Parkinson’s disease (PD), tested off medication; each session lasted about 2 h, and participants performed sinusoid and pseudo-random target-following tasks under slow and fast conditions.
  • Analysis: tremor suppression was quantified from PSD-based relative power change in the 2–10 Hz band; voluntary motion was assessed by position and velocity RMSE and by power change in the 0–2 Hz band; statistics included Wilcoxon Signed Rank tests, paired t-tests, and two-way repeated-measures ANOVA.

## Limitations

  • The paper states that this is a preliminary study and that further investigations/larger populations are needed to validate the technology.
  • The authors note that the study did not evaluate the passive effect of the TSO’s mass and inertia on tremor suppression or voluntary motion, and that future work should test the TSO in off mode to assess that contribution.
  • The authors also identify possible order effects, fatigue, and training effects because the TSO was tested after the MO; they suggest randomizing device order in future studies.
  • Evident from the text: the sample was small (n = 9) and included only 2 PD participants, so subgroup conclusions are limited.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1186/s12984-019-0484-1

Cites

59 of 59 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Figures

Fig. 1 Orthoses and Computer Interface. a TSO components and TSO donned. b MO components and MO donned. c Pursuit target graphical interface seen by the study participants. Cursors moved vertically. The outlined circle indicates the position of the target on the screen and the filled orange circle indicates the participant’s elbow cursor position. Fully extended elbow position was considered as the zero angle and corresponded to the cursor located at the bottom of the graphical interface window
Fig. 1 Orthoses and Computer Interface. a TSO components and TSO donned. b MO components and MO donned. c Pursuit target graphical interface seen by the study participants. Cursors moved vertically. The outlined circle indicates the position of the target on the screen and the filled orange circle indicates the participant’s elbow cursor position. Fully extended elbow position was considered as the zero angle and corresponded to the cursor located at the bottom of the graphical interface windowCC BY · PubMed Central
Fig. 2 Tremor motion related measures. a Tremor power change and b MO tremor fundumental frequency (from PSD) when performing the pursuit tracking tasks. The dotted horizontal lines in a and b indicate the participants’ mean. The scatter plots are composed of four unique markers associated with the four motion cases of the pursuit tracking experiment, namely sine slow (ss), sine fast (sf), random slow (rs) and random fast (rf)
Fig. 2 Tremor motion related measures. a Tremor power change and b MO tremor fundumental frequency (from PSD) when performing the pursuit tracking tasks. The dotted horizontal lines in a and b indicate the participants’ mean. The scatter plots are composed of four unique markers associated with the four motion cases of the pursuit tracking experiment, namely sine slow (ss), sine fast (sf), random slow (rs) and random fast (rf)CC BY · PubMed Central
Fig. 3 Voluntary motion related measures. a Voluntary component power changes for the MO (PCvMO) and the TSO (PCvTSO). The horizontal dashed line indicates zero value. b Position pursuit tracking RMSE for the fMO (epfMO) and TSO (epTSO). c Velocity pursuit tracking RMSE for the fMO (evfMO) and TSO (evTSO). d Voluntary interaction torque (same as Fe in Fig. 4) root mean square for the TSO. The scatter plots are composed of four unique markers associated with the four motion cases of the pursuit tracking experiment, namely sine slow (ss), sine fast (sf), random slow (rs) and random fast (rf)
Fig. 3 Voluntary motion related measures. a Voluntary component power changes for the MO (PCvMO) and the TSO (PCvTSO). The horizontal dashed line indicates zero value. b Position pursuit tracking RMSE for the fMO (epfMO) and TSO (epTSO). c Velocity pursuit tracking RMSE for the fMO (evfMO) and TSO (evTSO). d Voluntary interaction torque (same as Fe in Fig. 4) root mean square for the TSO. The scatter plots are composed of four unique markers associated with the four motion cases of the pursuit tracking experiment, namely sine slow (ss), sine fast (sf), random slow (rs) and random fast (rf)CC BY · PubMed Central
Fig. 4 Participant T06 sine and pseudo-random target tasks. a Slow sinusoid target motion b Slow pseudo-random target motion. For both subfigures a and b, the top two subplots show the position and velocity associated to the target T, the TSO and the MO. The velocity plot also shows the fMO velocity. The pseudo-random position plot also shows the successive random targets (solid green square wave). The bottom two subplots refer to the TSO interaction torque and the PSD obtained for both the TSO and MO. The measured torque Fm represents the combined voluntary and tremor components. The estimated torque Fe represents the voluntary component as a result of online filtering
Fig. 4 Participant T06 sine and pseudo-random target tasks. a Slow sinusoid target motion b Slow pseudo-random target motion. For both subfigures a and b, the top two subplots show the position and velocity associated to the target T, the TSO and the MO. The velocity plot also shows the fMO velocity. The pseudo-random position plot also shows the successive random targets (solid green square wave). The bottom two subplots refer to the TSO interaction torque and the PSD obtained for both the TSO and MO. The measured torque Fm represents the combined voluntary and tremor components. The estimated torque Fe represents the voluntary component as a result of online filteringCC BY · PubMed Central
Fig. 5 Performance Metrics Correlations Matrix. Each correlation is composed of 36 data points (nine participants performing the four motion cases)
Fig. 5 Performance Metrics Correlations Matrix. Each correlation is composed of 36 data points (nine participants performing the four motion cases)CC BY · PubMed Central

Verification

Claims the paper makes

13 Valid · 4 Not valid · 16 Not checked · 69 Own results and interpretations

26 more claims

Create a free account to see all 102 claims and every source verdict.

See all claims — free account
  • Probably not a substantial claim. 46
  • It may state two separate claims as one. 39
  • The quote only partly states this claim. 2

Create a free account to see these claims.

Citations

Cites

Show all 60 references

Research works printed but not linked: 2

  1. Herrnstadt G, Menon C. Voluntary-driven elbow orthosis with speed-controlled tremor suppression. Front Bioeng Biotechnol. 2016;4:1–10. · DOI · Paper page
  2. Roberts MJ. Signals and systems - analysis using transform methods and MATLAB, 2nd_edition, 1st ed. Boston: McGraw-Hill; 2004.

Cited by

See all 31

Metadata

Authors
Herrnstadt, G., McKeown, M. J., & Menon, C.
Affiliations
  • Simon Fraser University (CA)
  • University of British Columbia (CA)
Corresponding author
Menon, Carlo
Venue
Journal of NeuroEngineering and Rehabilitation
Published
2019-02-01
Volume
16
Issue
1
Pages
23-23
ISSN
1743-0003 (ISSN-L)
OpenAlex
W2937390004
Cited by
30
References
60
Citations by year
2026: 1 · 2025: 3 · 2024: 8 · 2023: 5 · 2022: 4 · 2021: 5 · 2020: 2 · 2019: 2
FWCI
1.14
Type
journal article
Language
en
Open access
Licence
  • cc-by
  • creativecommons.org/licenses/by/4.0/ (tdm)
Locations
  • Journal of NeuroEngineering and Rehabilitation (publishedVersion, cc-by)
  • Europe PMC (PubMed Central) (publishedVersion, other-oa)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • cIRcle (University of British Columbia) (submittedVersion, cc-by)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
  • Summit (Simon Fraser University) (submittedVersion, cc-by)
  • Open Collections
Funders
  • Natural Sciences and Engineering Research Council of Canada
  • Canadian Institutes of Health Research
  • Michael Smith Foundation for Health Research
  • International Essential Tremor Foundation
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Prosthetics and Rehabilitation Robotics
  • Life Sciences › Neuroscience › Cognitive Neuroscience › Motor Control and Adaptation
  • Health Sciences › Medicine › Clinical Neurology › Deep Brain Stimulation and Movement Disorders
Keywords
pathological tremor, tremor suppression, robotic orthosis, essential tremor, Parkinson's disease, voluntary movement, pursuit tracking, wearable devices, speed control, tracking error, power spectral density
MeSH terms
Aged, Aged, 80 and over, Algorithms, Biomechanical Phenomena, Elbow, Female, Humans, Male, Middle Aged, Movement, Orthotic Devices, Parkinson Disease, Prosthesis Design, Psychomotor Performance, Robotics, Torque, Essential Tremor

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

The claim verdicts this report counts are experimental: their accuracy is not yet measured.

A language model drew the claims from the paper on 2026-09-30, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).

A second model labelled each claim on 2026-09-30 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).

Cited sources are checked with a language model’s help (gpt-6-sol).

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Controlling a motorized orthosis to follow elbow volitional movement: tests with individuals with pathological tremor. https://app.cochat.ai/papers/10.1186/s12984-019-0484-1
The paper
Herrnstadt, G., McKeown, M. J., & Menon, C. (2019). Controlling a motorized orthosis to follow elbow volitional movement: tests with individuals with pathological tremor. Journal of NeuroEngineering and Rehabilitation. https://doi.org/10.1186/s12984-019-0484-1

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.