Initial experience with a SiPM-based PET/CT scanner: influence of acquisition time on image quality
Open accessAbstract
BACKGROUND: A newly introduced PET/CT scanner (Discovery Meaningful Insights-DMI, GE Healthcare) includes the silicon photomultiplier (SiPM) with time-of-flight (TOF) technology first used in the GE SIGNA PET/MRI. In this study, we investigated the impact of various acquisition times on image quality using this SiPM-based PET/CT. METHODS: We reviewed data from 58 participants with cancer who were scanned using the DMI PET/CT scanner. The administered dosages ranged 295.3-429.9 MBq (mean ± SD 356.3 ± 37.4) and imaging started at 71-142 min (mean ± SD 101.41 ± 17.52) after administration of the radiopharmaceutical. The patients' BMI ranged 19.79-46.16 (mean ± SD 26.55 ± 5.53). We retrospectively reconstructed the raw TOF data at 30, 60, 90, and 120 s/bed and at the standard image acquisition time per clinical protocol (180 or 210 s/bed depending on BMI). Each reconstruction was reviewed blindly by two nuclear medicine physicians and scored 1-5 (1-poor, 5-excellent quality). The liver signal-to-noise ratio (SNR) was used as a quantitative measure of image quality. RESULTS: The average scores ± SD of the readers were 2.61 ± 0.83, 3.70 ± 0.92, 4.36 ± 0.82, 4.82 ± 0.39, and 4.91 ± 0.91 for the 30, 60, 90, and 120 s/bed and at standard acquisition time, respectively. Inter-reader agreement on image quality assessment was good, with a weighted kappa of 0.80 (95% CI 0.72-0.81). In the evaluation of the effects of time per bed acquisition on semi-quantitative measurements, we found that the only time point significantly different from the standard time were 30 and 60 s (both with P < 0.001). The effects of dose and BMI were not statistically significant (P = 0.195 and 0.098, respectively). There was a significant positive effect of time on SNR (P < 0.001), as well as a significant negative effect of weight (P < 0.001). CONCLUSIONS: Our results suggest that despite significant delays from injection to imaging (due to comparison with standard PET/CT) compared to standard clinical operations and even in a population with average BMI > 25, images can be acquired as fast as 90 s/bed using the SiPM PET/CT and still result in very good image quality (average score > 4).
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Systematic SUVmax increases of >5-10% have not typically been observed at acquisition times per bed position of >30-60 s, at least in SiPM-equipped systems and/or patients with BMI <25 kg/m2.
“However, systematic SUVmax increases of >5–10% have not typically been observed at acquisition times per bed position of >30–60s, at least during investigations of SiPM-equipped systems and/or patients with BMI <25 kg/m2 [79,81,82].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [81]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
The citation lists haven’t been fetched yet
References 21 · Cited by 74
Research works printed but not linked: 19
- [1]Ter-Pogossian MM Phelps ME Hoffman EJ Mullani NA A positron-emission transaxial tomograph for nuclear imaging (PETT) Radiology 1975 114 89 98 10.1148/114.1.89 1208874 · DOI
- [2]Phelps ME Hoffman EJ Mullani NA Ter-Pogossian MM Application of annihilation coincidence detection to transaxial reconstruction tomography J Nucl Med 1975 16 210 224 1113170
- [3]Gambhir SS Molecular imaging of cancer with positron emission tomography Nat Rev Cancer 2002 2 683 693 10.1038/nrc882 12209157 · DOI
- [4]Beyer T Townsend DW Brun T Kinahan PE Charron M Roddy R A combined PET/CT scanner for clinical oncology J Nucl Med 2000 41 1369 1379 10945530
- [5]Miles KA. PET-CT in oncology: making the most of CT. Cancer Imaging. 2008;8 Spec no A:S87-S93. 10.1102/1470-7330.2008.9015. PMC2582506 18852084 · DOI
- [6]Kadrmas DJ Casey ME Conti M Jakoby BW Lois C Townsend DW Impact of time-of-flight on PET tumor detection J Nucl Med 2009 50 1315 1323 10.2967/jnumed.109.063016 19617317 PMC2786272 · DOI · Paper page
- [7]Schlemmer HP Pichler BJ Schmand M Burbar Z Michel C Ladebeck R Simultaneous MR/PET imaging of the human brain: feasibility study Radiology 2008 248 1028 1035 10.1148/radiol.2483071927 18710991 · DOI
- [8]Shao Y Cherry SR Farahani K Meadors K Siegel S Silverman RW Simultaneous PET and MR imaging Phys Med Biol 1997 42 1965 1970 10.1088/0031-9155/42/10/010 9364592 · DOI
- [10]Fraum TJ Fowler KJ McConathy J PET/MRI: emerging clinical applications in oncology Acad Radiol 2016 23 220 236 10.1016/j.acra.2015.09.008 26521689 · DOI
- [11]Roncali E Cherry SR Application of silicon photomultipliers to positron emission tomography Ann Biomed Eng 2011 39 1358 1377 10.1007/s10439-011-0266-9 21321792 PMC3069330 · DOI
- [12]Ullah MN Pratiwi E Cheon J Choi H Yeom JY Instrumentation for time-of-flight positron emission tomography Nucl Med Mol Imaging 2016 50 112 122 10.1007/s13139-016-0401-5 27275359 PMC4870467 · DOI
- [13]Grant AM Lee BJ Chang CM Levin CS Simultaneous PET/MR imaging with a radio frequency-penetrable PET insert Med Phys 2017 44 112 120 10.1002/mp.12031 28102949 PMC5372382 · DOI
- [14]Seifert S van der Lei G van Dam HT Schaart DR First characterization of a digital SiPM based time-of-flight PET detector with 1 mm spatial resolution Phys Med Biol 2013 58 3061 3074 10.1088/0031-9155/58/9/3061 23587636 · DOI
- [15]Iagaru A Mittra E Minamimoto R Jamali M Levin C Quon A Simultaneous whole-body time-of-flight 18F-FDG PET/MRI: a pilot study comparing SUVmax with PET/CT and assessment of MR image quality Clin Nucl Med 2015 40 1 8 10.1097/RLU.0000000000000611 25489952 PMC6319383 · DOI
- [16]Minamimoto R Levin C Jamali M Holley D Barkhodari A Zaharchuk G Improvements in PET image quality in time of flight (TOF) simultaneous PET/MRI Mol Imaging Biol 2016 18 776 781 10.1007/s11307-016-0939-8 26884058 · DOI · Paper page
- [17]Baratto L Park SY Hatami N Davidzon G Srinivas S Gambhir SS (18) F-FDG silicon photomultiplier PET/CT: a pilot study comparing semi-quantitative measurements with standard PET/CT PLoS One 2017 12 e0178936 10.1371/journal.pone.0178936 28582472 PMC5459477 · DOI
- [18]McDermott GM Chowdhury FU Scarsbrook AF Evaluation of noise equivalent count parameters as indicators of adult whole-body FDG-PET image quality Ann Nucl Med 2013 27 855 861 10.1007/s12149-013-0760-2 23925895 · DOI
- [19]Fukukita H Senda M Terauchi T Suzuki K Daisaki H Matsumoto K Japanese guideline for the oncology FDG-PET/CT data acquisition protocol: synopsis of version 1.0 Ann Nucl Med 2010 24 325 334 10.1007/s12149-010-0377-7 20401547 · DOI
- [20]Graham MM Badawi RD Wahl RL Variations in PET/CT methodology for oncologic imaging at U.S. academic medical centers: an imaging response assessment team survey J Nucl Med 2011 52 311 317 10.2967/jnumed.109.074104 21233185 PMC3889016 · DOI
- Authors
- Sonni, I., Baratto, L., Park, S., Hatami, N., Srinivas, S., Davidzon, G., Gambhir, S. S., & Iagaru, A.
- Affiliations
- Lawrence Berkeley National Laboratory (US)
- Stanford Medicine (US)
- Corresponding author
- Iagaru, Andrei
- ORCID
- Venue
- EJNMMI Physics
- Published
- 2018-04-18
- Volume
- 5
- Issue
- 1
- Pages
- 9-9
- ISSN
- 2197-7364 (ISSN-L)
- PMC
- PMC5904089
- OpenAlex
- W2801241905
- Cited by
- 74
- References
- 21
- Citations by year
- 2026: 2 · 2025: 6 · 2024: 7 · 2023: 9 · 2022: 9 · 2021: 18 · 2020: 12 · 2019: 8 · 2017: 3
- FWCI
- 6.56
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- EJNMMI Physics (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- OSTI OAI (U.S. Department of Energy Office of Scientific and Technical Information) (submittedVersion)
- Funders
- GE Healthcare
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Cardiac Imaging and Diagnostics
- Physical Sciences › Physics and Astronomy › Radiation › Radiation Detection and Scintillator Technologies
- Keywords
- PET/CT, silicon photomultipliers, time-of-flight PET, acquisition time, image quality, image reconstruction, cancer patients, radiation dose, body mass index, reader agreement, semi-quantitative measurement
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review