The effect of femoral head diameter upon lubrication and wear of metal-on-metal total hip replacements
The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
It has been found that a remarkable reduction in the wear of metal-on-metal hip joints can be achieved by simply increasing the diameter of the joint. A tribological evaluation of metal-on-metal joints of 16, 22,225, 28 and 36 mm diameter was conducted in 25 per cent bovine serum using a hip joint simulator. The joints were subject to dynamic motion and loading cycles simulating walking for both lubrication and wear studies. For each size of joint in the lubrication study, an electrical resistivity technique was used to detect the extent of surface separation through a complete walking cycle. Wear of each size of joint was measured gravimetrically in wear tests of at least 2 × 10 6 cycles duration. Joints of 16 and 22.225mm diameter showed no surface separation in the lubrication study. This suggested that wear would be proportional to the sliding distance and hence joint size in this boundary lubrication regime. A 28 mm diameter joint showed only limited evidence of surface separation suggesting that these joints were operating in a mixed lubrication regime. A 36 mm diameter joint showed surface separation for considerable parts of each walking cycle and hence evidence of the formation of a protective lubricating film. Wear testing of 16 and 22.225mm diameter metal-on-metal joints gave mean wear rates of 4.85 and 6.30mm 3/10 6 cycles respectively. The ratio of these wear rates, 0.77, is approximately the same as the joint diameters ratio, 16/22.225 or 0.72, as expected from simple wear theory for dry or boundary lubrication conditions. No bedding-in was observed with these smaller diameter joints. For the 28 mm diameter joint, from 0 to 2 × 10 6 cycles, the mean wear rate was 1.62 mm 3/10 6 cycles as the joints bedded-in. Following bedding-in, from 2.0 × 10 6 to 4.7 × 10 6 cycles, the wear rate was 0.54mm 3/10 6 cycles. As reported previously by Goldsmith in 2000 [1], the mean steady state wear rate of the 36 mm diameter joints was lower than those of all the other diameters at 0.07 mm 3/10 6 cycles. For a range of joints of various diameters, subjected to identical test conditions, mean wear rates differed by almost two orders of magnitude. This study has demonstrated that the application of sound tribological principles to prosthetic design can reduce the wear of metal-on-metal joints, using currently available materials, to a negligible level.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Composition and morphology of wear debris in failed uncemented total hip replacement
Shanbhag, A., Jacobs, J., Glant, T., Gilbert, J., Black, J., & Galante, J. (1994) · The Journal of Bone and Joint Surgery. British volume · doi:10.1302/0301-620x.76b1.8300684 · Cited by 387
Cited by
Smith, A. J., Dieppe, P., Vernon, K., Porter, M., & Blom, A. W. (2012) · The Lancet · doi:10.1016/s0140-6736(12)60353-5 · Cited by 568
Metadata
- Authors
- Smith, S. L., Dowson, D., & Goldsmith, A. A. J.
- Venue
- Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Publisher
- SAGE Publications
- Published
- 2001-02-01
- Volume
- 215
- Issue
- 2
- Pages
- 161-170
- ISSN
- 0954-4119 (ISSN-L)
- 2041-3033
- OpenAlex
- W2083636405
- Cited by
- 198
- References
- 1
- FWCI
- 8.07
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: The effect of femoral head diameter upon lubrication and wear of metal-on-metal total hip replacements. https://app.cochat.ai/papers/10.1243/0954411011533724
- The paper
- Smith, S. L., Dowson, D., & Goldsmith, A. A. J. (2001). The effect of femoral head diameter upon lubrication and wear of metal-on-metal total hip replacements. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. https://doi.org/10.1243/0954411011533724
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.