Composition and morphology of wear debris in failed uncemented total hip replacement
Abstract
Interfacial membranes collected at revision from 11 failed uncemented Ti-alloy total hip replacements were examined. Particles in the membranes were characterised by electron microscopy, microchemical spectroscopy and particle size analysis. Most were polyethylene and had a mean size of 0.53 micron +/- 0.3. They were similar to the particles seen in the base resin used in the manufacture of the acetabular implants. Relatively few titanium particles were seen. Fragments of bone, stainless steel and silicate were found in small amounts. Most of the polyethylene particles were too small to be seen by light microscopy. Electron microscopy and spectroscopic techniques are required to provide an accurate description of this debris.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Wear byproducts from total joint replacements include particulate polymeric, ceramic, and metallic debris and metallic ions, which cause pro-inflammatory responses that can persist for the life of the implant.
“The byproducts of wear include particulate polymeric, ceramic and metallic debris (mostly in the nano, submicron, and micron size range) and in the case of metal-on-metal implants, metallic ions [143]. These byproducts are well known to cause pro-inflammatory responses which can persist for the life of the implant [144,145].”
From Macrophage polarization: an opportunity for improved outcomes in biomaterials and regenerative medicine (2012) · cites this paper as [143]
Cited by
Sorted by citations
Emergence of Nanoplastic in the Environment and Possible Impact on Human Health
Lehner, R., Weder, C., Petri-Fink, A., & Rothen-Rutishauser, B. (2019) · Environmental Science & Technology · doi:10.1021/acs.est.8b05512 · Cited by 1,242 · see citing works
Brown, B. N., Ratner, B. D., Goodman, S. B., Amar, S., & Badylak, S. F. (2012) · Biomaterials · doi:10.1016/j.biomaterials.2012.02.034 · Cited by 926 · see citing works
“submicron, and micron size range) and in the case of metal-onmetal implants, metallic ions [143].”
BackgroundImpact of Microplastics and Nanoplastics on Human Health
Yee, M. S.-L., Hii, L., Looi, C.-K., Lim, W., Wong, S. F., Kok, Y.-Y., Tan, B.-K., Wong, C.-Y., & Leong, C. (2021) · Nanomaterials · doi:10.3390/nano11020496 · Cited by 1,077 · see citing works
“A study of failed titanium alloy total hip arthroplasty cases found that a majority of wear debris in the interfacial membranes consisted of polyethylene particles with an average diameter of 530 nm [124].”
BackgroundBiological reactions to wear debris in total joint replacement
Ingham, E., & Fisher, J. (2000) · Proceedings of the Institution of mechanical engineers. Part H, journal of engineering in medicine · doi:10.1243/0954411001535219 · Cited by 753
The role of macrophages in osteolysis of total joint replacement
Ingham, E., & Fisher, J. (2005) · Biomaterials · doi:10.1016/j.biomaterials.2004.04.035 · Cited by 667 · see citing works
Abu-Amer, Y., Darwech, I., & Clohisy, J. (2007) · Arthritis Research & Therapy · doi:10.1186/ar2170 · Cited by 556 · see citing works
“types of cells has also been reported [12,21-24].”
BackgroundStructure-property correlations in a combinatorial library of degradable biomaterials
Brocchini, S., James, K., Tangpasuthadol, V., & Kohn, J. (1998) · Journal of Biomedical Materials Research · doi:10.1002/(sici)1097-4636(199810)42:1 · Cited by 188 · see citing works
The potential impact of nano- and microplastics on human health: Understanding human health risks.
Winiarska, E., Jutel, M., & Zemelka‐Wiącek, M. (2024) · Environmental Research · doi:10.1016/j.envres.2024.118535 · Cited by 484 · see citing works
Isolation of predominantly submicron-sized UHMWPE wear particles from periprosthetic tissues.
Campbell, P., Ma, S., Yeom, B., Mckellop, H., Schmalzried, T., & Amstutz, H. (1995) · Journal of Biomedical Materials Research · doi:10.1002/jbm.820290118 · Cited by 378
Current Concepts Review - Wear in Total Hip and Knee Replacements*
Schmalzried, T., & Callaghan, J. (1999) · doi:10.2106/00004623-199901000-00016 · Cited by 341
The Biology of Aseptic Osteolysis
Holt, G., Murnaghan, C., Reilly, J. H., & Meek, R. M. D. (2007) · Clinical Orthopaedics and Related Research · doi:10.1097/blo.0b013e31804b4147 · Cited by 319 · see citing works
“1 m to several millimeters.(13,76,77,80,120) Macrophages appear most sensitive to submicron-sized particles, with giant cells forming around larger (larger than 10 m) particles (Fig 1).”
BackgroundConcentration- and composition-dependent effects of metal ions on human MG-63 osteoblasts.
Hallab, N., Vermes, C., Messina, C., Roebuck, K., Glant, T., & Jacobs, J. (2002) · Journal of Biomedical Materials Research · doi:10.1002/jbm.10106 · Cited by 282 · see citing works
Self-citationHuman monocyte response to particulate biomaterials generated in vivo and in vitro
Shanbhag, A. S., Jacobs, J., Black, J., Galante, J., & Glant, T. (1995) · Journal of Orthopaedic Research · doi:10.1002/jor.1100130520 · Cited by 272 · see citing works
Self-citationShanbhag, A. S., Hasselman, C., & Rubash, H. (1997) · Clinical Orthopaedics and Related Research · doi:10.1097/00003086-199711000-00005 · Cited by 271
Self-citationWear Debris in Total Joint Replacements
Jacobs, J., Shanbhag, A. S., Glant, T., Black, J., & Galante, J. (1994) · Journal of the American Academy of Orthopaedic Surgeons · doi:10.5435/00124635-199407000-00004 · Cited by 267 · see citing works
Self-citationA unified theory of wear for ultra-high molecular weight polyethylene in multi-directional sliding
Wang, A. (2001) · doi:10.1016/s0043-1648(00)00522-6 · Cited by 242
Vermes, C., Chandrasekaran, R., Jacobs, J., Galante, J., Roebuck, K., & Glant, T. (2001) · Journal of Bone and Joint Surgery. American volume · doi:10.2106/00004623-200102000-00007 · Cited by 238
Self-citationTHE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS*
Cannon, D., Sim, F., Zuckerman, J., & Pritchard, D. (2000) · doi:10.2106/00004623-197153060-00024 · Cited by 235
Wear in total hip and knee replacements.
Schmalzried, T., & Callaghan, J. (2000) · Journal of Bone and Joint Surgery. American volume · doi:10.2106/00004623-200012000-00020 · Cited by 229
Bi, Y., Seabold, J. M., Kaar, S. G., Ragab, A. A., Goldberg, V. M., Anderson, J. M., & Greenfield, E. M. (2001) · Journal of Bone and Mineral Research · doi:10.1359/jbmr.2001.16.11.2082 · Cited by 227 · see citing works
Tipper, J., Ingham, E., Hailey, J. L., Besong, A. A., Fisher, J., Wroblewski, B., & Stone, M. (2000) · Journal of materials science. Materials in medicine · doi:10.1023/a:1008901302646 · Cited by 215
Smith, S., Dowson, D., & Goldsmith, A. (2001) · Proceedings of the Institution of mechanical engineers. Part H, journal of engineering in medicine · doi:10.1243/0954411011533724 · Cited by 213
Lee, S.-H., Brennan, F., Jacobs, J., Urban, R., Ragasa, D., & Glant, T. (1997) · Journal of Orthopaedic Research · doi:10.1002/jor.1100150107 · Cited by 208 · see citing works
Self-citationKobayashi, A., Freeman, M., Bonfield, W., Kadoya, Y., Yamac, T., Al-Saffar, N., Scott, G., & Revell, P. (1997) · Journal of Bone and Joint Surgery-british Volume · doi:10.1302/0301-620x.79b5.7602 · Cited by 204
Biological reaction to debris in relation to joint prostheses
Revell, P. A., Al-Saffar, N., & Kobayashi, A. (1997) · Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine · doi:10.1243/0954411971534304 · Cited by 195 · see citing works
Revell, P. A. (2008) · Journal of The Royal Society Interface · doi:10.1098/rsif.2008.0142 · Cited by 198 · see citing works
Spinal Implant Debris-Induced Osteolysis
Hallab, N., Cunningham, B., & Jacobs, J. (2003) · Spine · doi:10.1097/00007632-200310151-00006 · Cited by 197
Self-citationOsteolysis in Association with a Total Hip Arthroplasty with Ceramic Bearing Surfaces*
Yoon, T., Rowe, S., Jung, S., Seon, K. J., & Maloney, W. (1998) · Journal of Bone and Joint Surgery. American volume · doi:10.2106/00004623-199810000-00007 · Cited by 179
Gutwein, L., & Webster, T. (2004) · Biomaterials · doi:10.1016/j.biomaterials.2003.10.090 · Cited by 177
The pathology of total joint arthroplasty
Bauer, T. W., & Schils, J. (1999) · Skeletal Radiology · doi:10.1007/s002560050552 · Cited by 172 · see citing works
“Several laboratories have developed methods to isolate and characterize the debris [23, 28, 29], and have reported the size, tissue concentration, and composition of”
MethodCellular mediators secreted by interfacial membranes obtained at revision total hip arthroplasty.
Shanbhag, A. S., Jacobs, J. J., Black, J., Galante, J., & Glant, T. (1995) · Journal of Arthroplasty · doi:10.1016/s0883-5403(05)80152-4 · Cited by 167 · see citing works
Self-citationSaikko, V., Ahlroos, T., Calonius, O., & Keränen, J. (2001) · Biomaterials · doi:10.1016/s0142-9612(00)00306-9 · Cited by 166
Suppression of Osteoblast Function by Titanium Particles*†
Yao, J., Cs-Szabó, G., Jacobs, J., Kuettner, K., & Glant, T. (1997) · Journal of Bone and Joint Surgery. American volume · doi:10.2106/00004623-199701000-00011 · Cited by 166 · see citing works
Self-citationMetal-on-metal hip resurfacing arthroplasty: A review of periprosthetic biological reactions
Mabilleau, G., Kwon, Y.-M., Pandit, H., Murray, D., & Sabokbar, A. (2008) · Acta Orthopaedica · doi:10.1080/17453670810016795 · Cited by 163
Polyethylene wear in total hip and knee arthroplasties.
Lewis, G. (1997) · Journal of Biomedical Materials Research · doi:10.1002/(sici)1097-4636(199721)38:1 · Cited by 163 · see citing works
The effects of lubricant composition on in vitro wear testing of polymeric acetabular components.
Wang, A., Essner, A., & Schmidig, G. (2004) · Journal of Biomedical Materials Research. Part B - Applied biomaterials · doi:10.1002/jbm.b.10077 · Cited by 161
New experience with alumina-on-alumina ceramic bearings for total hip arthroplasty.
D'Antonio, J., Capello, W., Manley, M., & Bierbaum, B. (2002) · Journal of Arthroplasty · doi:10.1054/arth.2002.32183 · Cited by 161
“The actual biologic event is caused by macrophage-induced resorption of bone at the prosthesis–bone interface secondary to the presence of particulate polyethylene debris [15,16].”
BackgroundCoCrMo Metal-on-Metal Hip Replacements
Liao, Y., Hoffman, E. E., Wimmer, M., Fischer, A., Jacobs, J., & Marks, L. (2013) · Physical Chemistry, Chemical Physics - PCCP · doi:10.1039/c2cp42968c · Cited by 159
Self-citationFern�ndez, E., Gil, F. J., Best, S. M., Ginebra, M., Driessens, F. C. M., & Planell, J. A. (1998) · Journal of Biomedical Materials Research · doi:10.1002/(sici)1097-4636(19980915)41:4 · Cited by 105 · see citing works
Moreno, J. L., MIKHAILENKO, I. A., Tondravi, M. M., & Keegan, A. D. (2007) · Journal of Leukocyte Biology · doi:10.1189/jlb.0107058 · Cited by 150 · see citing works
“In agreement with this data, others have reported that the size of the cell correlates with the size of the internalized particle [39–41].”
ResultBearing surface options for total hip replacement in young patients.
Heisel, C., Silva, M., & Schmalzried, T. (2003) · Instructional Course Lectures · doi:10.2106/00004623-200307000-00026 · Cited by 155
Metal versus polyethylene wear particles in total hip replacements. A review.
Doorn*, P., Campbell, P., & Amstutz, H. (1996) · Clinical Orthopaedics and Related Research · doi:10.1097/00003086-199608001-00018 · Cited by 154 · see citing works
Opalchenova, G., Dyulgerova, E., & Petrov, O. E. (1996) · Journal of Biomedical Materials Research · doi:10.1002/(sici)1097-4636(199606)31:2 · Cited by 4 · see citing works
Kaufman, A. M., Alabre, C. I., Rubash, H. E., & Shanbhag, A. S. (2007) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.31467 · Cited by 146 · see citing works
Self-citationBiologic and tribologic considerations of alternative bearing surfaces.
Campbell, P., Shen, F.-W., & Mckellop, H. (2004) · Clinical Orthopaedics and Related Research · doi:10.1097/00003086-200401000-00017 · Cited by 139
Yang, S.-Y., Ren, W., Park, Y., Sieving, A. L., Hsu, S., Nasser, S., & Wooley, P. (2002) · Biomaterials · doi:10.1016/s0142-9612(02)00032-7 · Cited by 137
Turell, M., Wang, A.-G., & Bellare, A. (2003) · doi:10.1016/s0043-1648(03)00357-0 · Cited by 131
“Wear debris collected from periprosthetic tissue retrieval studies showed that the majority of debris particles were micron or submicron is size and that particles could generally be categorized as fibrous or particulate in shape [10–14].”
BackgroundUrban, R., Jacobs, J., Sumner, D. R., Peters, C., Voss, F., & Galante, J. (1996) · Journal of Bone and Joint Surgery. American volume · doi:10.2106/00004623-199607000-00012 · Cited by 126
Self-citationKhan, I., Smith, N., Jones, E., Finch, D., & Cameron, R. (2005) · Biomaterials · doi:10.1016/j.biomaterials.2004.02.064 · Cited by 122
Yao, J., Glant, T. T., Lark, M., Mikecz, K., Jacobs, J. J., Hutchinson, N., Hoerrner, L. A., Kuettner, K., & Galante, J. O. (1995) · Journal of Bone and Mineral Research · doi:10.1002/jbmr.5650100920 · Cited by 121
Self-citationPathogenesis of bone loss after total hip arthroplasty.
Rubash, H., Sinha, R., Shanbhag, A. S., & Kim, S.-Y. (1998) · The Orthopedic cllinics of North America · doi:10.1016/s0030-5898(05)70316-3 · Cited by 120
Self-citationIn vivo corrosion of cobalt-chromium and titanium wear particles.
Shahgaldi, B., Heatley, F. W., Dewar, A., & Corrin, B. (1995) · Journal of Bone and Joint Surgery-british Volume · doi:10.1302/0301-620x.77b6.7593115 · Cited by 120
Comparison of in vitro with in vivo characteristics of wear particles from metal-metal hip implants.
Catelas, I., Medley, J., Campbell, P., Huk, O., & Bobyn, J. (2004) · Journal of Biomedical Materials Research. Part B - Applied biomaterials · doi:10.1002/jbm.b.20036 · Cited by 119
Schwartz, Z., Lohmann, C. H., Vocke, A. K., Sylvia, V. L., Cochran, D. L., Dean, D. D., & Boyan, B. D. (2001) · Journal of Biomedical Materials Research · doi:10.1002/1097-4636(20010905)56:3 · Cited by 83 · see citing works
Self-citationBiologic Effects of Cobalt Chrome in Cell and Animal Models
Howie, D. W., Rogers, S. D., McGee, M. A., & Haynes, D. R. (1996) · Clinical Orthopaedics and Related Research · doi:10.1097/00003086-199608001-00019 · Cited by 113 · see citing works
Effects of soluble metals on human peri-implant cells.
Hallab, N., Anderson, S., Caicedo, M., Brasher, A., Mikecz, K., & Jacobs, J. (2005) · Journal of Biomedical Materials Research. Part A · doi:10.1002/jbm.a.30345 · Cited by 114
Self-citationCurrent concepts in osteolysis
Ollivere, B. J., Wimhurst, J. A., Clark, I. M., & Donell, S. T. (2012) · Journal of Bone and Joint Surgery - British Volume · doi:10.1302/0301-620x.94b1.28047 · Cited by 113 · see citing works
Caicedo, M. S., Samelko, L., McAllister, K., Jacobs, J. J., & Hallab, N. J. (2013) · Journal of Orthopaedic Research® · doi:10.1002/jor.22411 · Cited by 118 · see citing works
Self-citationIn vitro human monocyte response to wear particles of titanium alloy containing vanadium or niobium.
Rogers, S. D., Howie, D., Graves, S., Pearcy, M., & Haynes, D. (1997) · Journal of Bone and Joint Surgery-british Volume · doi:10.1302/0301-620x.79b2.0790311 · Cited by 113
“In our study the wear particles of each material were of comparable size to those found in vivo.(28,30-33) Studies have shown that tissues retrieved during revision surgery for aseptic loosening have had an average of 818 g titanium per gram of tissue.”
ResultThe role of fibroblasts and fibroblast‐derived factors in periprosthetic osteolysis
Koreny, T., Tunyogi-Csapó, M., Gál, I., Vermes, C., Jacobs, J. J., & Glant, T. T. (2006) · Arthritis & Rheumatism · doi:10.1002/art.22134 · Cited by 112 · see citing works
Self-citationPolyethylene and titanium particles induce osteolysis by similar, lymphocyte‐independent, mechanisms
Taki, N., Tatro, J. M., Nalepka, J. L., Togawa, D., Goldberg, V. M., Rimnac, C., & Greenfield, E. (2005) · Journal of Orthopaedic Research · doi:10.1016/j.orthres.2004.08.023 · Cited by 108
Rader, C., Sterner, T., Jakob, F., Schütze, N., & Eulert, J. (1999) · Journal of Arthroplasty · doi:10.1016/s0883-5403(99)90035-9 · Cited by 107 · see citing works
Schmalzried, T., Campbell, P., Schmitt, A., Brown, I., & Amstutz, H. (1997) · Journal of Biomedical Materials Research · doi:10.1002/(sici)1097-4636(199723)38:3 · Cited by 107 · see citing works
Roy, T., Choudhury, D., Ghosh, S., Mamat, A. B., & Pingguan-Murphy, B. (2015) · doi:10.1016/j.ceramint.2014.08.123 · Cited by 107
“their use [6-9].”
Wound dressing with sustained anti-microbial capability
Loke, W., Lau, S.-K., Yong, L. L., Khor, E., & Chow, K. S. (2000) · Journal of Biomedical Materials Research · doi:10.1002/(sici)1097-4636(2000)53:1 · Cited by 175 · see citing works
Self-citationMacrophage response to cross-linked and conventional UHMWPE.
Sethi, R., Neavyn, M., Rubash, H., & Shanbhag, A. S. (2003) · Biomaterials · doi:10.1016/s0142-9612(03)00056-5 · Cited by 106 · see citing works
“Since the predominant wear debris identified in periimplant tissues is ultra-high molecular weight polyethylene (UHMWPE) from the acetabular liner, significant efforts have been directed at improving the quality of the articulating polymer [31–33].”
BackgroundSelf-citationRahbek, O., Overgaard, S., Lind, M., Bendix, K., Bünger, C., & Søballe, K. (2001) · Journal of Bone and Joint Surgery-british Volume · doi:10.1302/0301-620x.83b3.10667 · Cited by 104
Besong, A. A., Tipper, J. L., Ingham, E., Stone, M. H., Wroblewski, B., & Fisher, J. (1998) · Journal of Bone and Joint Surgery-british Volume · doi:10.1302/0301-620x.80b2.0800340 · Cited by 97
Methodology for Long-Term Wear Testing of Total Knee Replacements
Walker, P., Blunn, G., Perry, J., Bell, C. J., Sathasivam, S., Andriacchi, T., Paul, J. P., Haider, H., & Campbell, P. (2000) · Clinical Orthopaedics and Related Research · doi:10.1097/00003086-200003000-00032 · Cited by 93
Characterization and Comparison of Wear Debris from Failed Total Hip Implants of Different Types*
Hirakawa, K., Bauer, T., Stulberg, B., Wilde, A. H., & Secic, M. (1996) · Journal of Bone and Joint Surgery. American volume · doi:10.2106/00004623-199608000-00014 · Cited by 93
The role of adsorbed endotoxin in particle‐induced stimulation of cytokine release
Cho, D. R., Shanbhag, A. S., Hong, C., Baran, G., & Goldring, S. (2002) · Journal of Orthopaedic Research · doi:10.1016/s0736-0266(01)00179-6 · Cited by 92 · see citing works
Self-citationIllgen, R., Forsythe, T. M., Pike, J., Laurent, M., & Blanchard, C. (2008) · Journal of Arthroplasty · doi:10.1016/j.arth.2007.05.043 · Cited by 90
“2) [21,28,29].”
BackgroundRotating Hinge Total Knee Arthroplasty in Severely Affected Knees
Westrich, G., Mollano, A. V., Sculco, T., Buly, R., Laskin, R., & Windsor, R. (2000) · Clinical Orthopaedics and Related Research · doi:10.1097/00003086-200010000-00023 · Cited by 88
Macrophage phagocytosis of polyethylene particulate in vitro.
Voronov, I., Santerre, J., Hinek, A., Callahan, J., Sandhu, J., & Boynton, E. (1998) · Journal of Biomedical Materials Research · doi:10.1002/(sici)1097-4636(199801)39:1 · Cited by 86 · see citing works
IN VITRO HUMAN MONOCYTE RESPONSE TO WEAR PARTICLES OF TITANIUM ALLOY CONTAINING VANADIUM OR NIOBIUM
Rogers, S. D., Howie, D., Graves, S., Pearcy, M., & Haynes, D. (1997) · doi:10.1302/0301-620x.79b2.7192 · Cited by 85
Polyethylene and metal wear particles: characteristics and biological effects
Catelas, I., Wimmer, M., & Utzschneider, S. (2011) · Seminars in Immunopathology · doi:10.1007/s00281-011-0242-3 · Cited by 84
“4 μm (1 μm=10 m), but with some larger plateletlike particles up to 250 μm as well as fibrils or shreds [25, 28, 31, 39].”
BackgroundDean, D., Schwartz, Z., Liu, Y., Blanchard, C., Agrawal, C., Mabrey, J., Sylvia, V., Lohmann, C., & Boyan, B. (1999) · Journal of Bone and Joint Surgery. American volume · doi:10.2106/00004623-199904000-00002 · Cited by 82 · see citing works
Standardized analysis of UHMWPE wear particles from failed total joint arthroplasties.
Mabrey, J., Afsar-Keshmiri, A., Engh, G., Sychterz, C. J., Wirth, M., Rockwood, C. A., & Agrawal, C. (2002) · Journal of Biomedical Materials Research · doi:10.1002/jbm.10302 · Cited by 82
Carmody, E., Schwarz, E. M., Puzas, J., Rosier, R., & O'Keefe, R. J. (2002) · Arthritis & Rheumatism · doi:10.1002/art.10227 · Cited by 80 · see citing works
Sawano, H., Warisawa, S., & Ishihara, S. (2009) · doi:10.1016/j.precisioneng.2009.01.005 · Cited by 79
Howling, G. I., Sakoda, H., Antonarulrajah, A., Marrs, H., Stewart, T., Appleyard, S., Rand, B., Fisher, J., & Ingham, E. (2003) · Journal of Biomedical Materials Research. Part B - Applied biomaterials · doi:10.1002/jbm.b.10068 · Cited by 79
Hip simulator wear testing according to the newly introduced standard ISO 14242
Kaddick, C., & Wimmer, M. A. (2001) · Proceedings of the Institution of mechanical engineers. Part H, journal of engineering in medicine · doi:10.1243/0954411011536019 · Cited by 79
Lerouge, S., Huk, O., Yahia, L., Witvoet, J., & Sedel, L. (1997) · Journal of Bone and Joint Surgery-british Volume · doi:10.1302/0301-620x.79b1.0790135 · Cited by 78
“Study of the pseudomembranes from such cases has shown that polyethylene particles generated by friction at both articular and non-articular interfaces are the most frequent component of this debris.(1,2) They are found in considerable concentration in the periprosthetic tissues, and are associated with intense cellul…”
BackgroundGibon, E., Ma, T., Ren, P.-G., Fritton, K., Biswal, S., Yao, Z., Smith, L. R., & Goodman, S. B. (2011) · Journal of Orthopaedic Research · doi:10.1002/jor.21548 · Cited by 77 · see citing works
Lange, T., Schilling, A., Peters, F., Haag, F., Morlock, M., Rueger, J., & Amling, M. (2009) · Biomaterials · doi:10.1016/j.biomaterials.2009.06.023 · Cited by 73 · see citing works
Fritz, E. A., Glant, T., Vermes, C., Jacobs, J., & Roebuck, K. (2006) · Journal of Biomedical Materials Research. Part A · doi:10.1002/jbm.a.30609 · Cited by 73
Self-citationNinomiya, J. T., Struve, J. A., Stelloh, C. T., Toth, J. M., & Crosby, K. E. (2001) · Journal of Orthopaedic Research · doi:10.1016/s0736-0266(00)00061-9 · Cited by 73
Hallab, N., McAllister, K., Brady, M. F., & Jarman-Smith, M. (2012) · Journal of Biomedical Materials Research. Part B - Applied biomaterials · doi:10.1002/jbm.b.31974 · Cited by 71 · see citing works
Wear evaluation of cobalt-chromium alloy for use in a metal-on-metal hip prosthesis.
John, K. R. S., Zardiackas, L., & Poggie, R. (2004) · Journal of Biomedical Materials Research. Part B - Applied biomaterials · doi:10.1002/jbm.b.10053 · Cited by 71
Richards, L., Brown, C., Stone, M., Fisher, J., Ingham, E., & Tipper, J. (2008) · Journal of Bone and Joint Surgery-british Volume · doi:10.1302/0301-620x.90b8.20737 · Cited by 70
“However, this study also had the benefit of using FEG-SEM, and was able to characterise the lower end of the size distribution of UHMWPE particles, i.e. those in the size range 0.1 μm.”
MethodInfluentialMurine model of prosthesis failure for the long‐term study of aseptic loosening
Yang, S.-Y., Yu, H.-Y., Gong, W., Wu, B., Mayton, L., Costello, R. E., & Wooley, P. (2007) · Journal of Orthopaedic Research · doi:10.1002/jor.20342 · Cited by 70 · see citing works
“Numerous polymeric andmetallic wear debris particles are usually present in periprosthetic tissue, many of them engulfed by or in close contact with macrophages and other inflammatory cells.(29,30) Cytokines such as interleukin-1 (IL-1b), tumor necrosis factor alpha (TNFa), interleukin-6 (IL-6), and prostaglandin E2 h…”
BackgroundCoathup, M., Blackburn, J., Goodship, A., Cunningham, J., Smith, T., & Blunn, G. (2005) · Biomaterials · doi:10.1016/j.biomaterials.2004.10.020 · Cited by 70
Osteoblast and Chrondrocyte Proliferation in the Presence of Alumina And Titania Nanoparticles
Gutwein, L., & Webster, T. (2002) · doi:10.1023/a:1019920105207 · Cited by 70
Campbell, P., Doom, P., Dorey, F. J., & Amstutz, H. (1996) · Proceedings of the Institution of mechanical engineers. Part H, journal of engineering in medicine · doi:10.1243/pime_proc_1996_210_409_02 · Cited by 70
NUMBER OF POLYETHYLENE PARTICLES AND OSTEOLYSIS IN TOTAL JOINT REPLACEMENTS
Kobayashi, A., Freeman, M., Bonfield, W., Kadoya, Y., Yamac, T., Al-Saffar, N., Scott, G., & Revell, P. (1997) · doi:10.1302/0301-620x.79b5.0790844 · Cited by 69
Garrigues, G. E., Cho, D. R., Rubash, H. E., Goldring, S. R., Herndon, J. H., & Shanbhag, A. S. (2004) · Biomaterials · doi:10.1016/j.biomaterials.2004.06.034 · Cited by 58 · see citing works
Self-citationVermes, C., Glant, T., Hallab, N., Fritz, E. A., Roebuck, K., & Jacobs, J. (2001) · Journal of Arthroplasty · doi:10.1054/arth.2001.28719 · Cited by 68 · see citing works
Self-citationHailey, J. L., Ingham, E., Stone, M. H., Wroblewski, B., & Fisher, J. (1996) · Proceedings of the Institution of mechanical engineers. Part H, journal of engineering in medicine · doi:10.1243/pime_proc_1996_210_385_02 · Cited by 65
Positive cytokine production in failed metal-on-metal total hip replacements
Campbell, P., Wang, M. Y., Amstutz, H., & Goodman, S. (2002) · Acta Orthopaedica Scandinavica · doi:10.1080/000164702321022767 · Cited by 62
“Most of the metal wear particles analyzed by Doorn et al. are below this range, but most polyethylene particles are within it (Shanbhag et al. 1994).”
BackgroundDeng, Z.-T., Wang, Z.-H., Jin, J., Wang, Y., Bao, N.-R., Gao, Q., & Zhao, J.-N. (2017) · Acta Biomaterialia · doi:10.1016/j.actbio.2016.11.051 · Cited by 61 · see citing works
“studies have suggested that biomaterial wear particles such as UHMWPE (ultra-high molecular weight 407 polyethylene), titanium, Cobalt Chromium, PMMA (polymethylmethacrylate) particles were generated 408 during the daily functions of the prosthetic joint, which are believed playing a critical role in the 409 pathogene…”
Background
- Authors
- Shanbhag, A., Jacobs, J., Glant, T., Gilbert, J., Black, J., & Galante, J.
- Affiliations
- Rush University (US)
- Venue
- The Journal of Bone and Joint Surgery. British volume
- Published
- 1994-01
- Volume
- 76-B
- Issue
- 1
- Pages
- 60-67
- ISSN
- 0301-620X (ISSN-L)
- 2044-5377
- OpenAlex
- W1914858881
- Cited by
- 387
- References
- 0
- Citations by year
- 2026: 1 · 2025: 4 · 2024: 3 · 2023: 4 · 2022: 3 · 2021: 9 · 2020: 3 · 2019: 3 · 2018: 7 · 2017: 5 · 2016: 5 · 2015: 4 · 2014: 12 · 2013: 11 · 2012: 9
- FWCI
- 26.08
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Journal of Bone and Joint Surgery - British Volume (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Surgery › Orthopaedic implants and arthroplasty
- Health Sciences › Medicine › Surgery › Orthopedic Infections and Treatments
- Physical Sciences › Materials Science › Materials Chemistry › Titanium Alloys Microstructure and Properties
- Keywords
- wear debris, cementless total hip arthroplasty, revision hip arthroplasty, polyethylene wear particles, titanium alloys, interfacial membrane, particle size, implant failure, bone fragments, stainless steel particles, silicate particles
- MeSH terms
- Aged, Cementation, Female, Hip Joint, Hip Prosthesis, Humans, Male, Microscopy, Electron, Scanning, Middle Aged, Particle Size, Polyethylenes, Prosthesis Failure, Reoperation, Titanium, Spectroscopy, Fourier Transform Infrared
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review