Dosing Accuracy of Insulin Aspart FlexPens after Transport through the Pneumatic Tube System
Open accessThe DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
Purpose: The purpose of this study was to evaluate whether transporting insulin aspart FlexPens via a pneumatic tube system affects the dosing accuracy of the pens. Methods: A total of 115 Novo Nordisk FlexPens containing insulin aspart were randomly assigned to be transported via a pneumatic tube system (n = 92) or to serve as the control (n = 23). Each pen was then randomized to 10 international unit (IU) doses (n = 25) or 30 IU doses (n = 67), providing 600 and 603 doses, respectively, for the pneumatic tube group. The control group also received random assignment to 10 IU doses (n = 6) or 30 IU doses (n = 17), providing 144 and 153 doses, respectively. Each dose was expelled using manufacturer instructions. Weights were recorded, corrected for specific gravity, and evaluated based on acceptable International Organization for Standardization (ISO) dosing limits. Results: In the group of pens transported through the pneumatic tube system, none of the 600 doses of 10 IU (0.0%; 95% CI, 0.0 to 0.6) and none of the 603 doses of 30 IU (0.0%; 95% CI, 0.0 to 0.6) fell outside of the range of acceptable weights. Correspondingly, in the control group, none of the 144 doses at 10 IU (0.0%; 95% CI, 0.0 to 2.5) and none of the 153 doses at 30 IU (0.0%; 95% CI, 0.0 to 2.4) were outside of acceptable ISO limits. Conclusion: Transportation via pneumatic tube system does not appear to compromise dosing accuracy. Hospital pharmacies may rely on the pneumatic tube system for timely and accurate transport of insulin aspart FlexPens.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Korytkowski, M., Bell, D., Jacobsen, C., Suwannasari, R., & FlexPen® Study Team (2003) · Clinical Therapeutics · doi:10.1016/s0149-2918(03)80337-5 · Cited by 267
Cited by
(2019) · doi:10.1089/dia.2018.0376 · Cited by 38
“005 g/mL for insulin aspart.(10) For each individual bolus, deviation from target (1 or 10 U, respectively) was calculated.”
MethodKamecke, U., Waldenmaier, D., Haug, C., Ziegler, R., & Freckmann, G. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818788349 · Cited by 25
(2020) · doi:10.1111/pedi.12993 · Cited by 23
Accuracy Evaluation of a Novel Reusable Patch Pump Prototype
Mende, J., Baumstark, A., Waldenmaier, D., Amirouche, A., Barraud, A., Fridez, P., Haug, C., & Freckmann, G. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221097997 · Cited by 1
Metadata
- Authors
- Ward, L. G., Heckman, M. G., Warren, A. I., & Tran, K.
- Venue
- Hospital Pharmacy
- Publisher
- SAGE Publications
- Published
- 2013-01
- Volume
- 48
- Issue
- 1
- Pages
- 33-38
- ISSN
- 0018-5787 (ISSN-L)
- 1945-1253
- OpenAlex
- W1999197986
- Cited by
- 12
- References
- 1
- FWCI
- 0.52
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Dosing Accuracy of Insulin Aspart FlexPens after Transport through the Pneumatic Tube System. https://app.cochat.ai/papers/10.1310/hpj4801-33
- The paper
- Ward, L. G., Heckman, M. G., Warren, A. I., & Tran, K. (2013). Dosing Accuracy of Insulin Aspart FlexPens after Transport through the Pneumatic Tube System. Hospital Pharmacy. https://doi.org/10.1310/hpj4801-33
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.