Accuracy Evaluation of a Novel Reusable Patch Pump Prototype

Open access
Mende, Jochen, Baumstark, Annette, Waldenmaier, Delia, Amirouche, Amin, Barraud, Antoine, Fridez, Pierre, Haug, Cornelia, Freckmann, Guido·2023·Journal of Diabetes Science and Technology

Abstract

A prototype of an insulin patch pump that operates with a double pump mechanism was tested. Accuracy of bolus delivery of 0.2 U and 1.0 U, and a basal rate of 1.0 U/h were evaluated with a microgravimetric method. In addition, occlusion detection time at basal rates of 0.1 U/h and 1.0 U/h was assessed. Mean deviation from target was lower than 2% for both bolus sizes. Regarding basal rate accuracy, mean deviation over 72 hours was lower than 1%. Occlusion detection occurred in less than 30 minutes with both basal rates. Our study results suggest that the tested pump prototype provides delivery accuracy and occlusion detection that is similar or even better compared to all commercially available pumps tested with a similar experimental approach.

AI snapshot

Written from the paper’s full text.

TL;DR

This bench study evaluated a novel reusable patch pump prototype (Sigi) using microgravimetric methods. It found high bolus and basal delivery accuracy and faster occlusion detection than previously tested commercial pumps.

Key findings

  • Bolus accuracy: mean deviations from target were +1.8% ± 2.5% for 0.2 U boluses and +0.3% ± 2.4% for 1.0 U boluses; 89.3% and 95.3% of boluses were within ±5%, respectively, and all boluses were within ±15%.
  • Basal accuracy at 1.0 U/h over 72 hours: mean total deviation was −0.3% ± 0.3%; 97.7% of 1-hour windows were within ±5% and 99.8% within ±15% of target.
  • Occlusion detection time: mean 29 minutes at 0.1 U/h (range 28-30 minutes) and 10 minutes at 1.0 U/h (range 8-18 minutes).
  • The authors conclude the prototype’s delivery accuracy was similar or better than commercially available pumps tested in comparable settings, with distinctly shorter occlusion detection times.

Methods

  • Bench microgravimetric study using 3 individual prototype devices; no human testing and no ethical approval required; testing was based on IEC 60601-2-24 and previously published procedures.
  • Bolus accuracy: 0.2 U and 1.0 U boluses, 25 consecutive boluses per run, 6 runs each, with weight gain recorded 6 minutes after delivery start. Basal accuracy: 1.0 U/h over 72 hours, weight recorded every 2.5 minutes, repeated 6 times. Insulin aspart (NovoRapid PumpCart) was used with a density of 1.005 g/ml and a 30 cm fluoropolymer tube replacing the original cannula.
  • Occlusion detection time: original cannula occluded with a clamp at basal rates of 0.1 U/h and 1.0 U/h, repeated 6 times per rate; mean, minimum, and maximum times were calculated.

Limitations

  • The paper states comparability to previous studies is limited by differences in test setup, including a smaller number of individual pumps used and fewer repetitions performed.
  • The paper states this was the first accuracy evaluation of the pump, which is still under development.
  • Evident from the text: the study was bench-only, used only 3 prototype devices and 6 runs per condition, and did not assess clinical outcomes or in-human performance.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1177/19322968221097997

Cites

7 of 7 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Claims the paper makes

8 of 9 cited sources checked · 8 supported · 0 flagged

  • Interpretationcitation
    § Discussion · p. 4

    Deviations from target for the prototype were similar or lower compared to other patch pumps currently available on the market tested in a similar experimental setting.

    “However, the results of this investigation showed that deviations from target were similar or lower compared to other patch pumps currently available on the market and that have been tested in a similar experimental setting. 7,8”
    Quote in the paperSubstantial claim2 of 2 sources checked
    • cites [7] 7. Ziegler R, Oliver N, Waldenmaier D, Mende J, Haug C, Freckmann G. Evaluation of the accuracy of current tubeless pumps for continuous subcutaneous insulin infusion. Diabetes Technol Ther . 2021;23(5):350-357. source
    • cites [8] 8. Baumstark A, Mende J, Uchiyama J, Haug C, Freckmann G. Description of a novel patch pump for insulin delivery and comparative accuracy evaluation. J Diabetes Sci Technol . 2021:19322968211000441. <!-- image --> <!-- image --> <!-- image --> source
    • Rests on an assumption: Despite differences in test setup, the results from previous studies of other pumps are comparable enough to support the conclusion that the prototype's deviations and ODT are similar or better.
  • Interpretationown data
    § Abstract · p. 1

    The prototype's delivery accuracy is similar or even better than that of all commercially available pumps tested in a comparable experimental setting.

    “Our study results suggest that the tested pump prototype provides delivery accuracy and occlusion detection that is similar or even better compared to all commercially available pumps tested with a similar experimental approach.”
    • Also stated in: § Conclusions · p. 4 · plain
      “This first study's results showed that the tested insulin patch pump prototype provides a delivery accuracy that is similar or even better compared to all commercially available pumps that have been tested in a comparable experimental setting and that ODT was distinctly shorter.”
    Quote in the paperSubstantial claim2 of 2 sources checked
    • cites [7] 7. Ziegler R, Oliver N, Waldenmaier D, Mende J, Haug C, Freckmann G. Evaluation of the accuracy of current tubeless pumps for continuous subcutaneous insulin infusion. Diabetes Technol Ther . 2021;23(5):350-357. source
    • cites [8] 8. Baumstark A, Mende J, Uchiyama J, Haug C, Freckmann G. Description of a novel patch pump for insulin delivery and comparative accuracy evaluation. J Diabetes Sci Technol . 2021:19322968211000441. <!-- image --> <!-- image --> <!-- image --> source
    • Rests on an assumption: Despite differences in test setup, the results from previous studies of other pumps are comparable enough to support the conclusion that the prototype's deviations and ODT are similar or better.
  • Findingown data
    § Results · p. 4

    For 1.0 U boluses, 95.3% were within ±5% of target.

    “The percentage of boluses within ± 5% of the target was 89.3% for the 0.2 U bolus and 95.3% for the 1 U bolus (Figure 4).”
    Quote in the paperSubstantial claim
  • Findingown data
    § Results · p. 4

    For 0.2 U boluses, 89.3% were within ±5% of target.

    “The percentage of boluses within ± 5% of the target was 89.3% for the 0.2 U bolus and 95.3% for the 1 U bolus (Figure 4).”
    Quote in the paperSubstantial claim
  • Findingown data
    § Results · p. 4

    For basal rate 1.0 U/h over 72 hours, 97.7% of 1-hour windows were within ±5% of target.

    “Mean total deviation from target over 72 hours was -0.3% ± 0.3% with 99.8% of 1-hour windows within ± 15% from target and 97.7% of 1-hour windows within ± 5% from target.”
    Quote in the paperSubstantial claim
  • May combine two claims. 6
  • Classified as probably not a substantial claim. 4
  • May combine two claims.

    Interpretationcitation
    § Abstract · p. 1

    The prototype's occlusion detection time was similar or better (distinctly shorter) than that of commercially available pumps tested in comparable settings.

    “Our study results suggest that the tested pump prototype provides delivery accuracy and occlusion detection that is similar or even better compared to all commercially available pumps tested with a similar experimental approach.”
    • Also stated in: § Conclusions · p. 4 · plain
      “This first study's results showed that the tested insulin patch pump prototype provides a delivery accuracy that is similar or even better compared to all commercially available pumps that have been tested in a comparable experimental setting and that ODT was distinctly shorter.”
    • Also stated in: § Discussion · p. 4 · boosted
      “In this bench test, occlusion alarms, which are an important safety feature of insulin pumps, were triggered noticeably quicker than for any insulin pump tested so far. 4,7”
    Quote in the paperSubstantial claim2 of 2 sources checked
    • cites [4] 4. Freckmann G, Kamecke U, Waldenmaier D, Haug C, Ziegler R. occlusion detection time in insulin pumps at two different basal rates. J Diabetes Sci Technol . 2018;12(3):608-613. source
    • cites [7] 7. Ziegler R, Oliver N, Waldenmaier D, Mende J, Haug C, Freckmann G. Evaluation of the accuracy of current tubeless pumps for continuous subcutaneous insulin infusion. Diabetes Technol Ther . 2021;23(5):350-357. source
    • Rests on an assumption: Despite differences in test setup, the results from previous studies of other pumps are comparable enough to support the conclusion that the prototype's deviations and ODT are similar or better.
  • Classified as probably not a substantial claim. May combine two claims.

    Study design and conductcitation
    § Methods · p. 2

    Experimental testing was based on IEC 60601-2-24 and on procedures published previously.

    “Experimental testing was based on IEC 60601-2-24 and on procedures published previously. 3-5”
    Quote in the paperSubstantial claim2 of 3 sources checked
    • cites [3] 3. IEC 60601-2-24:2012 Medical electrical equipment-Part 2-24: Particular requirements for the basic safety and essential performance of infusion pumps and controllers [German version EN 60601-2-24:2015].

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

    • cites [4] 4. Freckmann G, Kamecke U, Waldenmaier D, Haug C, Ziegler R. occlusion detection time in insulin pumps at two different basal rates. J Diabetes Sci Technol . 2018;12(3):608-613. source
    • cites [5] 5. Kamecke U, Waldenmaier D, Haug C, Ziegler R, Freckmann G. Establishing methods to determine clinically relevant bolus and basal rate delivery accuracy of insulin pumps. J Diabetes Sci Technol . 2019;13(1):60-67. source
  • May combine two claims.

    Findingown data
    § Results · p. 4

    Minimum and maximum occlusion detection times at basal rate 0.1 U/h were 28 and 30 minutes.

    “Minimum and maximum ODT were 28 minutes and 30 minutes at a BR of 0.1 U/h and 8 minutes and 18 minutes at a BR of 1.0 U/h, respectively.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data
    § Results · p. 4

    All delivered boluses (0.2 U and 1.0 U) were within ±15% of target.

    “All boluses were within ± 15% from target.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data
    § Results · p. 4

    Minimum and maximum occlusion detection times at basal rate 1.0 U/h were 8 and 18 minutes.

    “Minimum and maximum ODT were 28 minutes and 30 minutes at a BR of 0.1 U/h and 8 minutes and 18 minutes at a BR of 1.0 U/h, respectively.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Premisenone given
    § Corresponding Author: · p. 1

    The lever actuations synchronized with the advancements of the primary mechanism together shall ensure precise insulin delivery.

    “These lever actuations are synchronized with the advancements of the primary mechanism, which together shall ensure precise insulin delivery.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Study design and conductnone given
    § Methods · p. 3

    Room temperature and humidity were within a controlled range.

    “Room temperature and humidity were within a controlled range.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationown data
    § Discussion · p. 4

    The prototype showed a high level of accuracy regarding the tested basal rate and both tested bolus sizes.

    “The prototype showed a high level of accuracy regarding the tested BR and both tested boluses sizes.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S4 (based).

    Premisereasoning
    § Introduction · p. 1

    Associating the primary piston pump and secondary micro-delivery mechanism shall ensure both short- and long-term pumping accuracy and precision.

    “Associating the 2 technologies shall ensure both short- and long-term pumping accuracy and precision.”
    Quote in the paperNot a substantial claim
    • Rests on an assumption: Piston pumps are very accurate on average delivered volume but may lack precision on each pumping cycle due to stick-slip of the piston inside the cartridge.
    • The authors state this as an assumption: Micro-pumps are very precise on each delivery cycle, but if not calibrated correctly, will cumulate error and result in less accuracy of the volume injected on the long run.
9 of 9
  • The paper relies on this without support.

    Piston pumps are very accurate on average delivered volume but may lack precision on each pumping cycle due to stick-slip of the piston inside the cartridge.

    “The reason for this combination is that, from experience piston pumps are very accurate on the average delivered volume but may lack precision on each pumping cycle due to the stickslip of the piston inside the cartridge.”
  • The authors state this as an assumption.

    Micro-pumps are very precise on each delivery cycle, but if not calibrated correctly, will cumulate error and result in less accuracy of the volume injected on the long run.

    “On the other hand, micro-pumps are assumed to be very precise on each delivery cycle, but, if not calibrated correctly, will cumulate the error and thus result in less accuracy of the volume injected on the long run.”
  • The paper relies on this without support.

    Despite differences in test setup, the results from previous studies of other pumps are comparable enough to support the conclusion that the prototype's deviations and ODT are similar or better.

    “Comparability to data of previous studies is limited because of differences in the test setup (eg, a smaller number of individual pumps used and fewer repetitions performed).”
    “However, the results of this investigation showed that deviations from target were similar or lower compared to other patch pumps currently available on the market and that have been tested in a similar experimental setting. 7,8”
  • The paper relies on this without support.

    This was the first accuracy evaluation study with this insulin pump prototype.

    “This was the first accuracy evaluation study with this insulin pump which is currently under development.”
4 of 4

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

Each contradiction between two passages listed here was confirmed by an independent check.

  • Abstract says all occlusions were detected in less than 30 minutes

    Minor

    The abstract says detection occurred in less than 30 minutes at both basal rates, but the reported maximum at 0.1 U/h was 30 minutes.

1 of 1

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Mende, J., Baumstark, A., Waldenmaier, D., Amirouche, A., Barraud, A., Fridez, P., Haug, C., & Freckmann, G.
Affiliations
  • Universität Ulm (DE)
  • AMF Medical SA, Ecublens, Switzerland
Corresponding author
Mende, Jochen
Venue
Journal of Diabetes Science and Technology
Published
2022-05-20
Volume
17
Issue
6
Pages
1644-1648
ISSN
1932-2968 (ISSN-L)
OpenAlex
W4281257499
Cited by
1
References
7
Citations by year
2024: 1
FWCI
0.15
Type
journal article
Language
en
Open access
Locations
  • Journal of Diabetes Science and Technology (publishedVersion)
  • PubMed (publishedVersion)
  • PubMed Central (submittedVersion)
Funders
AMF Medical SA
Topics
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
  • Physical Sciences › Engineering › Biomedical Engineering › Intravenous Infusion Technology and Safety
  • Life Sciences › Pharmacology, Toxicology and Pharmaceutics › Pharmaceutical Science › Advancements in Transdermal Drug Delivery
Keywords
occlusion detection, insulin delivery, pump prototype
MeSH terms
Humans, Hypoglycemic Agents, Insulin, Insulin Infusion Systems, Transdermal Patch, Insulin, Regular, Human

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-13.

Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-13.

Cited sources are checked by gpt-6-sol.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review