Bone Mineral and Body Composition Measurements: Cross-Calibration of Pencil-Beam and Fan-Beam Dual-Energy X-Ray Absorptiometers
Abstract
Pencil-beam dual-energy X-ray absorptiometers (DXA) are being replaced with instruments that rely solely on fan-beam technology. However, information has been lacking regarding the translation of bone mineral and body composition data between the two devices. We have compared total body scans using pencil-beam (Hologic QDR-2000W) and fan-beam (Hologic QDR-4500A) instruments for 33 children (ages 3–18 years) and 14 adults. Bone mineral content (BMC), bone mineral density (BMD), fat, lean, and body fatness (%fat) values were highly correlated (r2 = 0.984–0.998) between the two DXA instruments. The mean differences between the paired measurements were: ΔBMC = 7.5 ± 73.6 g, ΔBMD = 0.0074 ± 0.0252 g/cm2, Δlean = 1.05 ± 1.8 kg, Δfat = −0.77 ± 1.7 kg, and Δ%fat = −0.94% ± 2.5%. The BMC and BMD values were not statistically different, whereas the differences for the body composition values were significant (p < 0.02–0.005). Regression equations are provided for conversion of bone and body composition data between pencil-beam and fan-beam values for the whole body. To test the performance of these equations for a second group (23 subjects), predicted values were compared with the measured data obtained using the fan-beam instrument. The mean differences were −1.0% to 1.4%, except for body fat mass, where the difference was 6.4%. For cross-sectional studies, the two DXA technologies can be considered equivalent after using the translational equations provided. For longitudinal studies in which small changes in body composition for the individual are to be detected, we recommend that the same DXA instrument be used whenever possible. For example, transition from a pencil-beam to a fan-beam instrument could, in extreme cases, result in differences as large as 19% for the estimate of body fat mass.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Differences in intra- and inter-manufacturer DXA estimates are small.
“Studies have shown that the differences in intraand inter-manufacturer estimates are small (11,14 -16).”
From Comparison of Body Composition Methods in Obese African‐American Women (2006) · cites this paper as [11]
DXA estimates of %BF are accurate, but values can differ by DXA instrumentation.
“Estimates of %BF are also accurate (11)(12)(13), but values can differ based on the DXA instrumentation utilized (e.g., Lunar, Hologic, Norland).”
From Comparison of Body Composition Methods in Obese African‐American Women (2006) · cites this paper as [11]
DXA provides accurate estimates of bone mineral content and bone mineral density.
“The DXA method has been shown to provide accurate estimates of bone mineral content and bone mineral density (11).”
From Comparison of Body Composition Methods in Obese African‐American Women (2006) · cites this paper as [11]
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965 · see citing works
Beckett, A. H., Rowland, M., & TURNER, P. (1965) · The Lancet · doi:10.1016/s0140-6736(65)91037-8 · Cited by 16,917 · see citing works
Mazess, R., Barden, H. S., Bisek, J., & Hanson, J. (1990) · American Journal of Clinical Nutrition · doi:10.1093/ajcn/51.6.1106 · Cited by 1,465 · see citing works
Heymsfield, S. B., Wang, J., Heshka, S., Kehayias, J. J., & Pierson, R. N. (1989) · American Journal of Clinical Nutrition · doi:10.1093/ajcn/49.6.1283 · Cited by 277 · see citing works
Komiya, S., Mitsuzono, R., & Ube, M. (1995) · 体力科學 · doi:10.7600/jspfsm1949.44.211 · Cited by 256 · see citing works
Dual-Energy X-Ray Absorptiometry
Lohman, T. G., & Chen, Z. (2005) · Human Body Composition · doi:10.5040/9781492596950.ch-005 · Cited by 219 · see citing works
JENSEN, M. D., Kanaley, J. A., Roust, L. R., O’Brien, P. C., BRAUN, J. S., DUNN, W. L., & Wahner, H. W. (1993) · Mayo Clinic Proceedings · doi:10.1016/s0025-6196(12)60695-8 · Cited by 213 · see citing works
(1994) · PubMed · Cited by 165 · see citing works
Tothill, P., Avenell, A., & Reid, D. M. (1994) · British Journal of Radiology · doi:10.1259/0007-1285-67-804-1210 · Cited by 133 · see citing works
Abrahamsen, B., Gram, J., Hansen, T. B., et al. (1995) · Bone · doi:10.1016/8756-3282(94)00054-9 · Cited by 117 · see citing works
Tothill, P., Hannan, W. J., Cowen, S. J., & Freeman, C. P. L. (1997) · Journal of Bone and Mineral Research · doi:10.1359/jbmr.1997.12.11.1908 · Cited by 109 · see citing works
Dual X-ray absorptiometry: a comparison between fan beam and pencil beam scans
Blake, G. M., Parker, J. C., Buxton, F. M. A., et al. (1993) · British Journal of Radiology · doi:10.1259/0007-1285-66-790-902 · Cited by 95 · see citing works
Cross-calibration of DXA equipment: Upgrading from a hologic QDR 1000/W to a QDR 2000
Faulkner, K. G., Gl�er, C., ESTILO, M. J., et al. (1993) · Calcified Tissue International · doi:10.1007/bf00308312 · Cited by 81 · see citing works
Comparison of body composition and bone mineral measurements from two DXA instruments in young men
Modlesky, C., Lewis, R., Yetman, K. A., Rose, B. S., Rosskopf, L. B., Snow, T., & Sparling, P. B. (1996) · American Journal of Clinical Nutrition · doi:10.1093/ajcn/64.5.669 · Cited by 70 · see citing works
Ellis, K. J., Shypailo, R. J., Hergenroeder, A. C., Pérez, M. T., & Abrams, S. A. (1996) · Journal of Bone and Mineral Research · doi:10.1002/jbmr.5650110616 · Cited by 64 · see citing works
Self-citationComparison of two Hologic DXA systems (QDR 1000 and QDR 4500/A).
Barthe, N., Braillon, P. M., Ducassou, D., & Basse‐Cathalinat, B. (1997) · British Journal of Radiology · doi:10.1259/bjr.70.835.9245885 · Cited by 60 · see citing works
A Preliminary Evaluation of the Lunar Expert-XL for Bone Densitometry and Vertebral Morphometry
Lang, T., Takada, M., Gee, R., et al. (1997) · Journal of Bone and Mineral Research · doi:10.1359/jbmr.1997.12.1.136 · Cited by 54 · see citing works
Correcting the Magnification Error of Fan Beam Densitometers
Griffiths, M. R., Noakes, K. A., & POCOCK, N. A. (1997) · Journal of Bone and Mineral Research · doi:10.1359/jbmr.1997.12.1.119 · Cited by 54 · see citing works
Body composition by DEXA in older adults: accuracy and influence of scan mode
Clasey, J. L., HARTMAN, M. L., Kanaley, J. A., Wideman, L., Teates, C. D., Bouchard, C., & WELTMAN, A. L. (1997) · Medicine & Science in Sports & Exercise · doi:10.1097/00005768-199704000-00020 · Cited by 50 · see citing works
Dual-energy X-ray absorptiometry results differ between machines
Paton, N. I., Macallan, D. C., Jebb, S. A., Pazianas, M., & Griffin, G. E. (1995) · The Lancet · doi:10.1016/s0140-6736(95)92737-9 · Cited by 44 · see citing works
Dual energy X-ray absorptiometry: inter-machine variability.
Tataranni, P. A., Pettitt, D. J., & Ravussin, É. (1996) · PubMed · Cited by 39 · see citing works
Pierson, R. N., Wang, J., Thornton, J. C., et al. (1995) · Calcified Tissue International · doi:10.1007/bf00296337 · Cited by 25 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Visser, M., Kritchevsky, S. B., Goodpaster, B. H., Newman, A. B., Nevitt, M., Stamm, E., & Harris, T. B. (2002) · Journal of the American Geriatrics Society · doi:10.1046/j.1532-5415.2002.50217.x · Cited by 845 · see citing works
Schoeller, D. A., Tylavsky, F. A., Baer, D. J., Chumlea, W. C., Earthman, C. P., Fuerst, T., Harris, T. B., Heymsfield, S. B., Horlick, M., Lohman, T. G., et al. (2005) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/81.5.1018 · Cited by 253 · see citing works
Cribb, P. J., Williams, A. D., Carey, M. F., & Hayes, A. (2006) · International Journal of Sport Nutrition and Exercise Metabolism · doi:10.1123/ijsnem.16.5.494 · Cited by 177 · see citing works
Gillette-Guyonnet, S. (2003) · Mechanisms of Ageing and Development · doi:10.1016/s0047-6374(02)00199-9 · Cited by 32 · see citing works
Comparison of Body Composition Methods in Obese African‐American Women
Newton, R. L., Alfonso, A., York‐Crowe, E., Walden, H., White, M. A., Ryan, D., & Williamson, D. A. (2006) · Obesity · doi:10.1038/oby.2006.55 · Cited by 27 · see citing works
- Authors
- Ellis, K. J., & Shypailo, R. J.
- Affiliations
- Baylor College of Medicine (US)
- Corresponding author
- Ellis, Kenneth J.
- ORCID
- Shypailo, Roman J.
- Venue
- Journal of Bone and Mineral Research
- Publisher
- Oxford University Press (OUP)
- Published
- 1998-10-01
- Volume
- 13
- Issue
- 10
- Pages
- 1613-1618
- ISSN
- 0884-0431 (ISSN-L)
- 1523-4681
- OpenAlex
- W2070650287
- Cited by
- 108
- References
- 22
- Citations by year
- 2024: 1 · 2023: 1 · 2021: 2 · 2020: 1 · 2019: 4 · 2018: 1 · 2017: 1 · 2016: 1 · 2015: 2 · 2014: 2 · 2013: 4 · 2012: 6
- FWCI
- 2.25
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- academic.oup.com/journals/pages/open_access/funder_policies/chorus/standard_publication_model (vor)
- Locations
- Journal of Bone and Mineral Research (publishedVersion)
- PubMed (publishedVersion)
- Funders
- U.S. Department of Agriculture
- Agricultural Research Service
- Topics
- Health Sciences › Medicine › Physiology › Body Composition Measurement Techniques
- Health Sciences › Medicine › Orthopedics and Sports Medicine › Bone health and osteoporosis research
- Social Sciences › Arts and Humanities › Archeology › Forensic Anthropology and Bioarchaeology Studies
- Keywords
- dual-energy X-ray absorptiometry, cross-calibration, bone mineral content, bone mineral density, body composition, fat mass, lean body mass, children, adults, conversion equations
- MeSH terms
- Adult, Body Composition, Bone and Bones, Calibration, Child, Child, Preschool, Humans, Middle Aged, Minerals, Absorptiometry, Photon
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review