The OSCAR-IB Consensus Criteria for Retinal OCT Quality Assessment
Open accessAbstract
BACKGROUND: Retinal optical coherence tomography (OCT) is an imaging biomarker for neurodegeneration in multiple sclerosis (MS). In order to become validated as an outcome measure in multicenter studies, reliable quality control (QC) criteria with high inter-rater agreement are required. METHODS/PRINCIPAL FINDINGS: A prospective multicentre study on developing consensus QC criteria for retinal OCT in MS: (1) a literature review on OCT QC criteria; (2) application of these QC criteria to a training set of 101 retinal OCT scans from patients with MS; (3) kappa statistics for inter-rater agreement; (4) identification reasons for inter-rater disagreement; (5) development of new consensus QC criteria; (6) testing of the new QC criteria on the training set and (7) prospective validation on a new set of 159 OCT scans from patients with MS. The inter-rater agreement for acceptable scans among OCT readers (n = 3) was moderate (kappa 0·45) based on the non-validated QC criteria which were entirely based on the ophthalmological literature. A new set of QC criteria was developed based on recognition of: (O) obvious problems, (S) poor signal strength, (C) centration of scan, (A) algorithm failure, (R) retinal pathology other than MS related, (I) illumination and (B) beam placement. Adhering to these OSCAR-IB QC criteria increased the inter-rater agreement to kappa from moderate to substantial (0.61 training set and 0.61 prospective validation). CONCLUSIONS: This study presents the first validated consensus QC criteria for retinal OCT reading in MS. The high inter-rater agreement suggests the OSCAR-IB QC criteria to be considered in the context of multicentre studies and trials in MS.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Tewarie and colleagues developed quality control guidelines for OCT in multiple sclerosis known as the OSCAR-IB criteria.
“Tewarie and colleagues developed quality control guidelines for OCT in MS -known as the OSCAR-IB criteria. 92”
From The use of optical coherence tomography and visual evoked potentials in the 2024 McDonald diagnostic criteria for multiple sclerosis (2025) · cites this paper as [92]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
The Measurement of Observer Agreement for Categorical Data
Landis, J. R., & Koch, G. G. (1977) · Biometrics · doi:10.2307/2529310 · Cited by 81,148 · see citing works
Statistical Methods for Rates and Proportions
Fleiss, J. L., Levin, B., & Paik, M. C. (2003) · Wiley Series in Probability and Statistics · doi:10.1002/0471445428 · Cited by 18,646 · see citing works
Flagged first, then by citations
Research works printed but not linked: 14
- [1]Frohman E Costello F Zivadinov R Stuve O Conger A 2006 Optical coherence tomography in multiple sclerosis. Lancet Neurol 5 853 863 16987732
- [2]Petzold A de Boer J Schippling S Vermersch P Kardon R 2010 Optical coherence tomography in multiple sclerosis: a systematic review and meta-analysis. Lancet Neurol 9 921 932 20723847
- [3]Green AJ McQuaid S Hauser SL Allen IV Lyness R 2010 Ocular pathology in multiple sclerosis: retinal atrophy and inflammation irrespective of disease duration. Brain 133 1591 1601 20410146 PMC2877904
- [4]Jindahra P Hedges TH Mendoza-Santiesteban CE Plant GT 2010 Optical coherence tomography of the retina: applications in neurology. Curr Opin Neurol 23 16 23 20009925
- [5]Barkhof F Calabresi PA Miller DH Reingold SC 2009 Imaging outcomes for neuroprotection and repair in multiple sclerosis trials. Nat Rev Neurol 5 256 266 19488083
- [6]Domalpally A Danis RP Zhang B Myers D Kruse CN 2009 Quality issues in interpretation of optical coherence tomograms in macular diseases. Retina 29 775 781 19373128
- [9]Vizzeri G Bowd C Medeiros FA Weinreb RN Zangwill LM 2009 Effect of Signal Strength and Improper Alignment on the Variability of Stratus Optical Coherence Tomography Retinal Nerve Fiber Layer Thickness Measurements. American Journal of Ophthalmology 148 245 255 PMC2848173 19427621
- [10]Yim Lui Cheung C Kai Shun Leung C Lin D Pang CP Shun Chiu Lam D 2008 Relationship between Retinal Nerve Fiber Layer Measurement and Signal Strength in Optical Coherence Tomography. Ophthalmology 115 1347 1351 18294689
- [11]Huang J Liu X Wu Z Sadda S 2011 Image quality affects macular and retinal nerve fiber layer thickness measurements on fourier-domain optical coherence tomography. Ophthalmic Surg Lasers Imaging 42 216 21 21449533
- [12]Kanamori A Nakamura M Escano MFT Seya R Maeda H 2003 Evaluation of the glaucomatous damage on retinal nerve fiber layer thickness measured by optical coherence tomography. American Journal of Ophthalmology 135 513 520 12654369
- [13]Otani T Yamaguchi Y Kishi S 2011 Improved visualization of henle fiber layer by changing the measurement beam angle on optical coherence tomography. Retina 31 497 501 21102368
- [14]Balk L de Vries–Knoppert WAEJ Petzold A 2011 A simple sign for recognizing off–center OCT measurement beam placement in the context of multicentre studies. Retina [in press] PMC3493550 23144857
- [15]Talman LS Bisker ER Long DA Sackel DJ Galetta KM 2010 Longitudinal study of vision and retinal nerve fiber layer thickness in multiple sclerosis. Ann Neurol 67 749 760 20517936 PMC2901775
- [16]Costello F 2011 Evaluating the Use of Optical Coherence Tomography in Optic Neuritis. Multiple sclerosis int 2011 2011 148394 PMC3196333 22096626
Cited by
Sorted by citations
Vidal-Jordana, A., Rovira, A., Calderon, W., Arrambide, G., Castilló, J., Moncho, D., Rahnama, K., Collorone, S., Toosy, A. T., Ciccarelli, O., et al. (2024) · Neurology · doi:10.1212/wnl.0000000000207805 · Cited by 88 · see citing works
Saidha, S., Green, A. J., Leocani, L., Vidal-Jordana, A., Kenney, R. C., Bsteh, G., Outteryck, O., Thompson, A., Montalban, X., Coetzee, T., et al. (2025) · The Lancet Neurology · doi:10.1016/s1474-4422(25)00275-3 · Cited by 66 · see citing works
Self-citationOertel, F. C., Krämer, J., Motamedi, S., Keihani, A., Zimmermann, H. G., Dimitriou, N. G., Condor-Montes, S., Bereuter, C., Cordano, C., Abdelhak, A., et al. (2023) · Neurology Neuroimmunology & Neuroinflammation · doi:10.1212/nxi.0000000000200092 · Cited by 33 · see citing works
Self-citationSocioeconomic disparity is associated with faster retinal neurodegeneration in multiple sclerosis
Vasileiou, E. S., Filippatou, A. G., Pimentel Maldonado, D., Kalaitzidis, G., Ehrhardt, H., Lambe, J., Saidha, S., Sotirchos, E. S., Mowry, E. M., Calabresi, P. A., et al. (2021) · Brain · doi:10.1093/brain/awab342 · Cited by 32 · see citing works
Wicki, C. A., Hanson, J. V. M., & Schippling, S. (2018) · Current Opinion in Neurology · doi:10.1097/wco.0000000000000604 · Cited by 16 · see citing works
Self-citation
- Authors
- Tewarie, P., Balk, L., Costello, F., Green, A., Martin, R., Schippling, S., & Petzold, A.
- Affiliations
- Amsterdam UMC Location Vrije Universiteit Amsterdam (NL)
- University of Calgary (CA)
- University of California, San Francisco (US)
- University Medical Center Hamburg-Eppendorf (DE)
- UCL Queen Square Institute of Neurology (GB)
- Corresponding author
- Petzold, Axel
- Venue
- PLoS ONE
- Publisher
- Public Library of Science (PLoS)
- Published
- 2012-04-19
- Volume
- 7
- Issue
- 4
- Pages
- e34823
- ISSN
- 1932-6203 (ISSN-L)
- PMC
- PMC3334941
- OpenAlex
- W2075584671
- Cited by
- 627
- References
- 2
- Citations by year
- 2026: 66 · 2025: 83 · 2024: 61 · 2023: 57 · 2022: 59 · 2021: 55 · 2020: 51 · 2019: 42 · 2018: 36 · 2017: 27 · 2016: 23 · 2015: 28 · 2014: 22 · 2013: 12 · 2012: 5
- FWCI
- 15.48
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Locations
- PLoS ONE (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Amsterdam UMC (VU Amsterdam) - Institutional Repository (publishedVersion, other-oa)
- VU Research Portal (publishedVersion, other-oa)
- Zurich Open Repository and Archive (University of Zurich) (acceptedVersion, cc-by)
- Digital Academic REpository of VU University Amsterdam (Vrije Universiteit Amsterdam) (submittedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- UCL Discovery (University College London) (submittedVersion)
- Pure Amsterdam UMC (submittedVersion, other-oa)
- Zurich Open Repository and Archive (University of Zurich) (submittedVersion, cc-by)
- Data Archiving and Networked Services (DANS) (submittedVersion, other-oa)
- Zurich Open Repository and Archive (University of Zurich)
- Funders
- Vanderbilt University Medical Center
- Gemeinnützige Hertie-Stiftung
- Topics
- Health Sciences › Medicine › Pathology and Forensic Medicine › Multiple Sclerosis Research Studies
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Retinal Imaging and Analysis
- Health Sciences › Medicine › Ophthalmology › Glaucoma and retinal disorders
- Keywords
- retinal optical coherence tomography, multiple sclerosis, interobserver agreement, consensus criteria, neurodegeneration biomarkers, outcome measure validation, signal strength
- MeSH terms
- Evaluation Studies as Topic, Humans, Multiple Sclerosis, Optic Neuritis, Prospective Studies, Quality Assurance, Health Care, Quality Control, Observer Variation, Consensus, Tomography, Optical Coherence
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review