High Resolution T1ρ Mapping of In Vivo Human Knee Cartilage at 7T
Open accessAbstract
PURPOSE: Spin lattice relaxation time in rotating frame (T1ρ) mapping of human knee cartilage has shown promise in detecting biochemical changes during osteoarthritis. Due to higher field strength, MRI at 7T has advantages in term of SNR compared to clinical MR scanners and this can be used to increase in image resolution. Objective of current study was to evaluate the feasibility of high resolution T1ρ mapping of in vivo human knee cartilage at 7T MR scanner. MATERIALS AND METHODS: In this study we have used a T1ρ prepared GRE pulse sequence for obtaining high resolution (in plan resolution = 0.2 mm2) T1ρ MRI of human knee cartilage at 7T. The effect of a global and localized reference frequency and reference voltage setting on B0, B1 and T1ρ maps in cartilage was evaluated. Test-retest reliability results of T1ρ values from asymptomatic subjects as well as T1ρ maps from abnormal cartilage of two human subjects are presented. These results are compared with T1ρ MRI data obtained from 3T. RESULTS: Our approach enabled acquisition of 3D-T1ρ data within allowed SAR limits at 7T. SNR of cartilage on T1ρ weighted images was greater than 90. Off-resonance effects present in the cartilage B0, B1 and T1ρ maps obtained using global shim and reference frequency and voltage setting, were reduced by the proposed localized reference frequency and voltage setting. T1ρ values of cartilage obtained with the localized approach were reproducible. Abnormal knee cartilage showed elevated T1ρ values in affected regions. T1ρ values at 7T were significantly lower (p<0.05) compared to those obtained at 3T. CONCLUSION: In summary, by using proposed localized frequency and voltage setting approach, high-resolution 3D-T1ρ maps of in vivo human knee cartilage can be obtained in clinically acceptable scan times (<30 min) and SAR constraints, which provides the ability to characterize cartilage molecular integrity.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
With higher field strength T1ρ decreases.
“and it is well known fact that with higher field strength T1ρ decreases (Singh et al., 2014).”
From T1rho MRI and CSF biomarkers in diagnosis of Alzheimer's disease (2015) · cites this paper as Singh et al., 2014
The Results section (‘T1ρ Analysis at 3T vs 7T’) reports significantly lower T1ρ values at 7T than at 3T in human knee cartilage (n = 6). That supports the direction of the comparison, but the paper does not establish an unqualified rule that T1ρ decreases with higher field strength across tissues or measurement conditions; the Discussion also notes that the spin-lock power used could have contributed to the lower values. (LLM, conf=0.86)
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
No references could be resolved yet.
Research works printed but not linked: 27
- [1]Cohen ZA, McCarthy DM, Kwak SD, Legrand P, Fogarasi F, et al (1999) Knee cartilage topography, thickness, and contact areas from MRI: in-vitro calibration and in-vivo measurements. Osteoarthritis Cartilage 7: 95– 109. 10367018
- [2]Saarakkala S, Julkunen P, Kiviranta P, Makitalo J, Jurvelin JS, et al (2010) Depth-wise progression of osteoarthritis in human articular cartilage: investigation of composition, structure and biomechanics. Osteoarthritis Cartilage 18: 73– 81. 19733642
- [3]Blanco FJ, Guitian R, Vazquez-Martul E, de Toro FJ, Galdo F (1998) Osteoarthritis chondrocytes die by apoptosis. A possible pathway for osteoarthritis pathology. Arthritis Rheum 41: 284– 289. 9485086
- [4]Young AA, McLennan S, Smith MM, Smith SM, Cake MA, et al (2006) Proteoglycan 4 downregulation in a sheep meniscectomy model of early osteoarthritis. Arthritis Res Ther 8: R41. 16469119 PMC1526595
- [5]Pakin SK, Cavalcanti C, La Rocca R, Schweitzer ME, Regatte RR (2006) Ultra-high-field MRI of knee joint at 7.0 T: preliminary experience. Acad Radiol 13: 1135– 1142. 16935725
- [6]Chang G, Wiggins GC, Xia D, Lattanzi R, Madelin G, et al (2012) Comparison of a 28-channel receive array coil and quadrature volume coil for morphologic imaging and T2 mapping of knee cartilage at 7T. J Magn Reson Imaging 35: 441– 448. 22095723 PMC3265647
- [7]Regatte RR, Schweitzer ME (2007) Ultra-high-field MRI of the musculoskeletal system at 7.0 T. J Magn Reson Imaging 25: 262– 269. 17260399
- [8]Redfield AG (1955) Nuclear Magnetic Resonance Saturation and Rotary Saturation in Solids. Physical Review 98: 13.
- [9]Akella SV, Regatte RR, Wheaton AJ, Borthakur A, Reddy R (2004) Reduction of residual dipolar interaction in cartilage by spin-lock technique. Magn Reson Med 52: 1103– 1109. 15508163
- [10]Wheaton AJ, Dodge GR, Elliott DM, Nicoll SB, Reddy R (2005) Quantification of cartilage biomechanical and biochemical properties via T1rho magnetic resonance imaging. Magn Reson Med 54: 1087– 1093. 16200568
- [11]Wang N, Xia Y (2012) Orientational dependent sensitivities of T2 and T1rho towards trypsin degradation and Gd-DTPA2- presence in bovine nasal cartilage. MAGMA 25: 297– 304. 22071581 PMC5682561
- [12]Borthakur A, Wheaton AJ, Gougoutas AJ, Akella SV, Regatte RR, et al (2004) In vivo measurement of T1rho dispersion in the human brain at 1.5 tesla. J Magn Reson Imaging 19: 403– 409. 15065163
- [13]Regatte RR, Akella SV, Wheaton AJ, Lech G, Borthakur A, et al (2004) 3D-T1rho-relaxation mapping of articular cartilage: in vivo assessment of early degenerative changes in symptomatic osteoarthritic subjects. Acad Radiol 11: 741– 749. 15217591
- [14]Wheaton AJ, Borthakur A, Kneeland JB, Regatte RR, Akella SV, et al (2004) In vivo quantification of T1rho using a multislice spin-lock pulse sequence. Magn Reson Med 52: 1453– 1458. 15562469
- [15]Pakin SK, Schweitzer ME, Regatte RR (2006) 3D-T1rho quantitation of patellar cartilage at 3.0 T. J Magn Reson Imaging 24: 1357– 1363. 17058202
- [16]Li X, Benjamin Ma C, Link TM, Castillo DD, Blumenkrantz G, et al (2007) In vivo T(1rho) and T(2) mapping of articular cartilage in osteoarthritis of the knee using 3T MRI. Osteoarthritis Cartilage 15: 789– 797. 17307365 PMC2040334
- [17]Witschey WR 2nd, Borthakur A, Elliott MA, Mellon E, Niyogi S, et al (2007) Artifacts in T1 rho-weighted imaging: compensation for B(1) and B(0) field imperfections. J Magn Reson 186: 75– 85. 17291799 PMC1995435
- [18]Witschey WR, Borthakur A, Elliott MA, Mellon E, Niyogi S, et al (2007) Compensation for spin-lock artifacts using an off-resonance rotary echo in T1rhooff-weighted imaging. Magn Reson Med 57: 2– 7. 17191245 PMC2877388
- [19]Witschey WR, Borthakur A, Elliott MA, Fenty M, Sochor MA, et al (2008) T1rho-prepared balanced gradient echo for rapid 3D T1rho MRI. J Magn Reson Imaging 28: 744– 754. 18777535 PMC2593793
- [20]Witschey WR, Borthakur A, Elliott MA, Magland J, McArdle EL, et al (2009) Spin-locked balanced steady-state free-precession (slSSFP). Magn Reson Med 62: 993– 1001. 19672947 PMC2859035
- [21]Witschey WR, Borthakur A, Fenty M, Kneeland BJ, Lonner JH, et al (2010) T1rho MRI quantification of arthroscopically confirmed cartilage degeneration. Magn Reson Med 63: 1376– 1382. 20432308 PMC2933515
- [22]Goto H, Iwama Y, Fujii M, Aoyama N, Kubo S, et al (2012) A preliminary study of the T1rho values of normal knee cartilage using 3T-MRI. Eur J Radiol 81: e796– 803. 22525597
- [23]Borthakur A, Wheaton A, Charagundla SR, Shapiro EM, Regatte RR, et al (2003) Three-dimensional T1rho-weighted MRI at 1.5 Tesla. J Magn Reson Imaging 17: 730– 736. 12766904
- [24]Kim M, Gillen J, Landman BA, Zhou J, van Zijl PC (2009) Water saturation shift referencing (WASSR) for chemical exchange saturation transfer (CEST) experiments. Magn Reson Med 61: 1441– 1450. 19358232 PMC2860191
- [25]Singh A, Haris M, Cai K, Kassey VB, Kogan F, et al (2012) Chemical exchange saturation transfer magnetic resonance imaging of human knee cartilage at 3T and 7T. Magn Reson Med 68: 588– 594. 22213239 PMC4067761
- [26]Schneider CA, Rasband WS, Eliceiri KW (2012) NIH Image to ImageJ: 25 years of image analysis. Nat Methods 9: 671– 675. 22930834 PMC5554542
- [27]Firbank MJ, Coulthard A, Harrison RM, Williams ED (1999) A comparison of two methods for measuring the signal to noise ratio on MR images. Phys Med Biol 44: N261– 264. 10616158
Cited by
Sorted by citations
MacKay, J. W., Low, S. B. L., Smith, T. O., Toms, A. P., McCaskie, A. W., & Gilbert, F. J. (2018) · Osteoarthritis and Cartilage · doi:10.1016/j.joca.2017.11.018 · Cited by 148 · see citing works
T1rho MRI and CSF biomarkers in diagnosis of Alzheimer's disease
Haris, M., Yadav, S. K., Rizwan, A., Singh, A., Cai, K., Kaura, D., Wang, E., Davatzikos, C., Trojanowski, J. Q., Melhem, E. R., et al. (2015) · NeuroImage: Clinical · doi:10.1016/j.nicl.2015.02.016 · Cited by 33 · see citing works
Self-citation
- Authors
- Singh, A., Haris, M., Cai, K., Kogan, F., Hariharan, H., & Reddy, R.
- Affiliations
- Indian Institute of Technology Delhi (IN)
- Sidra Medical and Research Center (QA)
- University of Illinois Chicago (US)
- University of Pennsylvania (US)
- Corresponding author
- Singh, Anup
- Venue
- PLoS ONE
- Publisher
- Public Library of Science (PLoS)
- Published
- 2014-05-15
- Volume
- 9
- Issue
- 5
- Pages
- e97486
- ISSN
- 1932-6203 (ISSN-L)
- PMC
- PMC4022681
- OpenAlex
- W2006429895
- Cited by
- 48
- References
- 0
- Citations by year
- 2025: 1 · 2024: 4 · 2022: 3 · 2021: 4 · 2020: 4 · 2019: 5 · 2018: 6 · 2017: 3 · 2016: 10 · 2015: 7 · 2014: 1
- FWCI
- 5.48
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Locations
- PLoS ONE (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, other-oa)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- National Institutes of Health
- National Institute of Arthritis and Musculoskeletal and Skin Diseases (R01-AR45404, R01AR045404, NIH-R01AR45404)
- National Institute of Biomedical Imaging and Bioengineering (P41EB015893, 41EB015893)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
- Health Sciences › Medicine › Rheumatology › Osteoarthritis Treatment and Mechanisms
- Physical Sciences › Physics and Astronomy › Nuclear and High Energy Physics › NMR spectroscopy and applications
- Keywords
- T1rho mapping, knee cartilage, 7T MRI, spin-lattice relaxation, off-resonance effects, test-retest reliability, osteoarthritis, signal-to-noise ratio, specific absorption rate
- MeSH terms
- Adult, Cartilage, Articular, Humans, Image Processing, Computer-Assisted, Knee, Knee Joint, Magnetic Resonance Imaging, Middle Aged, Reproducibility of Results, Osteoarthritis, Knee, Imaging, Three-Dimensional, Young Adult, Signal-To-Noise Ratio
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review