Can unmanned aerial systems (drones) be used for the routine transport of chemistry, hematology, and coagulation laboratory specimens?
Open accessAbstract
BACKGROUND: Unmanned Aerial Systems (UAS or drones) could potentially be used for the routine transport of small goods such as diagnostic clinical laboratory specimens. To the best of our knowledge, there is no published study of the impact of UAS transportation on laboratory tests. METHODS: Three paired samples were obtained from each one of 56 adult volunteers in a single phlebotomy event (336 samples total): two tubes each for chemistry, hematology, and coagulation testing respectively. 168 samples were driven to the flight field and held stationary. The other 168 samples were flown in the UAS for a range of times, from 6 to 38 minutes. After the flight, 33 of the most common chemistry, hematology, and coagulation tests were performed. Statistical methods as well as performance criteria from four distinct clinical, academic, and regulatory bodies were used to evaluate the results. RESULTS: Results from flown and stationary sample pairs were similar for all 33 analytes. Bias and intercepts were <10% and <13% respectively for all analytes. Bland-Altman comparisons showed a mean difference of 3.2% for Glucose and <1% for other analytes. Only bicarbonate did not meet the strictest (Royal College of Pathologists of Australasia Quality Assurance Program) performance criteria. This was due to poor precision rather than bias. There were no systematic differences between laboratory-derived (analytic) CV's and the CV's of our flown versus terrestrial sample pairs however CV's from the sample pairs tended to be slightly higher than analytic CV's. The overall concordance, based on clinical stratification (normal versus abnormal), was 97%. Length of flight had no impact on the results. CONCLUSIONS: Transportation of laboratory specimens via small UASs does not affect the accuracy of routine chemistry, hematology, and coagulation tests results from selfsame samples. However it results in slightly poorer precision for some analytes.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Some evidence from high-income countries suggests that drones can reduce delivery times and improve the quality of blood components.
“Some evidence from high-income countries suggests that drones can reduce delivery times 14 and improve the quality of blood components.15,16”
From Effect of unmanned aerial vehicle (drone) delivery on blood product delivery time and wastage in Rwanda: a retrospective, cross-sectional study and time series analysis (2022) · cites this paper as [14]
Citation 14 is attached to delivery times, so this source is assessed for the claim that drones improve the quality of blood components. The Conclusions (page unavailable) report no systematic differences in test results between flown and terrestrial diagnostic specimens, but slightly worse precision in the flown samples. The study does not report an improvement in blood-component quality. (LLM, conf=0.83)
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measuredA study in the USA found that drones resulted in a median time saving of 16 min for automated external defibrillator delivery.
“For example, a study in the USA found that drones resulted in a median time saving of 16 min for automated external defibrillator delivery.14”
From Effect of unmanned aerial vehicle (drone) delivery on blood product delivery time and wastage in Rwanda: a retrospective, cross-sectional study and time series analysis (2022) · cites this paper as [14]
The Methods > Study Design and Results > Flight Time passages concern laboratory specimens and the effect of flight duration on test results. Searches for automated external defibrillators, AED delivery, and a 16-minute median saving found no passage addressing the claim; page numbers are unavailable in our copy. (LLM, conf=0.78)
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Arcury, T. A., Gesler, W. M., Preisser, J. S., Sherman, J., Spencer, J., & Perin, J. (2005) · Health Services Research · doi:10.1111/j.1475-6773.2005.00346.x · Cited by 594 · see citing works
Tanser, F. C., Gijsbertsen, B., & Herbst, K. (2006) · Social Science & Medicine · doi:10.1016/j.socscimed.2006.01.015 · Cited by 350 · see citing works
Significance Tests for Coefficients of Variation and Variability Profiles
Sokal, R. R., & Braumann, C. A. (1980) · Systematic Biology · doi:10.1093/sysbio/29.1.50 · Cited by 306 · see citing works
Gerbig, C., Lin, J. C., Wofsy, S. C., Daube, B. C., Andrews, A. E., Stephens, B. B., Bakwin, P. S., & Grainger, C. A. (2003) · Journal of Geophysical Research Atmospheres · doi:10.1029/2002jd003018 · Cited by 184 · see citing works
Medically Useful Criteria for Analytic Performance of Laboratory Tests
Skendzel, L. P., Barnett, R. N., & Platt, R. B. (1985) · American Journal of Clinical Pathology · doi:10.1093/ajcp/83.2.200 · Cited by 131 · see citing works
General Principles of Specimen Collection and Transport
Wilson, M. L. (1996) · Clinical Infectious Diseases · doi:10.1093/clinids/22.5.766 · Cited by 91 · see citing works
Westgard, J. O. (1992) · Clinical Chemistry · doi:10.1093/clinchem/38.7.1226 · Cited by 85 · see citing works
Pneumatic tube system induced haemolysis: assessing sample type susceptibility to haemolysis
Sodi, R., Darn, S. M., & Stott, A. (2004) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1258/000456304323019631 · Cited by 67 · see citing works
The Good, the Bad, and the Unknown: Quality of Clinical Laboratories in Kampala, Uganda
Elbireer, A., Jackson, J. B., Sendagire, H., Opio, A. A., Bagenda, D. S., & Amukele, T. K. (2013) · PLoS ONE · doi:10.1371/journal.pone.0064661 · Cited by 61 · see citing works
Self-citationAirborne observations of spatial and temporal variability of tropospheric carbon dioxide
Anderson, B. E., Gregory, G., Collins, J., Sachse, G., Conway, T., & Whiting, G. (1996) · Oceanographic Literature Review · doi:10.1029/95jd00413 · Cited by 55 · see citing works
The adiabatic lapse‐rate for dry and saturated air
Brunt, D. (1933) · Quarterly Journal of the Royal Meteorological Society · doi:10.1002/qj.49705925204 · Cited by 53 · see citing works
Keshgegian, A. A., & Bull, G. E. (1992) · American Journal of Clinical Pathology · doi:10.1093/ajcp/97.4.535 · Cited by 52 · see citing works
Jones, G. R. D., Sikaris, K. A., & Gill, J. (2012) · Europe PMC (PubMed Central) · Cited by 51 · see citing works
Evaluation of a Computer-directed Pneumatic-tube System for Pneumatic Transport of Blood Specimens
Weaver, D. K., Miller, D., Leventhal, E. A., & Tropeano, V. (1978) · American Journal of Clinical Pathology · doi:10.1093/ajcp/70.3.400 · Cited by 40 · see citing works
Implementation of a Pneumatic-tube System for Transport of Blood Specimens
Poznanski, W. J., Smith, F. M., & Bodley, F. (1978) · American Journal of Clinical Pathology · doi:10.1093/ajcp/70.2.291 · Cited by 34 · see citing works
A rationale for parsimonious laboratory term mapping by frequency.
Vreeman, D. J., Finnell, J. T., & Overhage, J. M. (2007) · PubMed Central · Cited by 22 · see citing works
Amukele, T. K., Schroeder, L. F., Jackson, J. B., & Elbireer, A. (2014) · American Journal of Clinical Pathology · doi:10.1309/ajcpcya54dwzqpqt · Cited by 16 · see citing works
Self-citationEvaluation of a Pneumatic-tube System for Delivery of Blood Specimens to the Blood Bank
Bruner, K. W., & Kissling, C. W. (1980) · American Journal of Clinical Pathology · doi:10.1093/ajcp/73.4.593 · Cited by 14 · see citing works
Ladenson, J. H., Scott, M. G., Klarkowski, D. B., & Seyoum, M. (2003) · Clinical Chemistry · doi:10.1373/49.1.162 · Cited by 9 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Drone transportation of blood products
Amukele, T., Ness, P. M., Tobian, A. A. R., Boyd, J., & Street, J. (2017) · Transfusion · doi:10.1111/trf.13900 · Cited by 194 · see citing works
Self-citationNisingizwe, M. P., Ndishimye, P., Swaibu, K., Nshimiyimana, L., Karame, P., Dushimiyimana, V., Musabyimana, J. P., Musanabaganwa, C., Nsanzimana, S., & Law, M. R. (2022) · The Lancet Global Health · doi:10.1016/s2214-109x(22)00048-1 · Cited by 175 · see citing works
- Authors
- Amukele, T. K., Sokoll, L. J., Pepper, D., Howard, D. P., & Street, J.
- Affiliations
- Johns Hopkins University (US)
- University of Michigan (US)
- NextGen Aeronautics (United States) (US)
- Corresponding author
- Amukele, Timothy K.
- Venue
- PLOS ONE
- Publisher
- Public Library of Science (PLoS)
- Published
- 2015-07-29
- Volume
- 10
- Issue
- 7
- Pages
- e0134020
- ISSN
- 1932-6203 (ISSN-L)
- PMC
- PMC4519103
- OpenAlex
- W1631786717
- Cited by
- 139
- References
- 19
- Citations by year
- 2026: 10 · 2025: 11 · 2024: 11 · 2023: 11 · 2022: 22 · 2021: 14 · 2020: 15 · 2019: 16 · 2018: 8 · 2017: 12 · 2016: 9
- FWCI
- 7.12
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- PLoS ONE (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Figshare (submittedVersion, cc-by)
- Topics
- Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Engineering › Aerospace Engineering › UAV Applications and Optimization
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Clinical Biochemistry › Bacterial Identification and Susceptibility Testing
- Keywords
- unmanned aerial systems, unmanned aerial vehicles, point-of-care testing, precision, bicarbonate, glucose
- MeSH terms
- Adult, Aircraft, Blood Coagulation Tests, Feasibility Studies, Female, Hematologic Tests, Humans, Male, Specimen Handling, Time Factors, Phlebotomy, Clinical Chemistry Tests
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review