A Novel Algorithm for Movement Artifact Removal in ECG Signals Acquired from Wearable Systems Applied to Horses
Open accessThe DOI matches its Crossref record; 72 of the 72 references it cites were found in a registry or an index.
Abstract
This study reports on a novel method to detect and reduce the contribution of movement artifact (MA) in electrocardiogram (ECG) recordings gathered from horses in free movement conditions. We propose a model that integrates cardiovascular and movement information to estimate the MA contribution. Specifically, ECG and physical activity are continuously acquired from seven horses through a wearable system. Such a system employs completely integrated textile electrodes to monitor ECG and is also equipped with a triaxial accelerometer for movement monitoring. In the literature, the most used technique to remove movement artifacts, when noise bandwidth overlaps the primary source bandwidth, is the adaptive filter. In this study we propose a new algorithm, hereinafter called Stationary Wavelet Movement Artifact Reduction (SWMAR), where the Stationary Wavelet Transform (SWT) decomposition algorithm is employed to identify and remove movement artifacts from ECG signals in horses. A comparative analysis with the Normalized Least Mean Square Adaptive Filter technique (NLMSAF) is performed as well. Results achieved on seven hours of recordings showed a reduction greater than 40% of MA percentage (between before- and after- the application of the proposed algorithm). Moreover, the comparative analysis with the NLMSAF, applied to the same ECG recordings, showed a greater reduction of MA percentage in favour of SWMAR with a statistical significant difference (p-value < 0.0.5).
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures














Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“By using additional sensors, such as triaxial accelerometers and gyroscopes, to synchronously monitor the movement, many methods have been proposed to detect and suppress the motion artifacts superimposed on MCG (ECG) signals, such as the adaptive filters (46) and the stationary wavelet transform (47, 48).”
From A movable unshielded magnetocardiography system (2023) · cites this paper as [47]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
A theory for multiresolution signal decomposition: the wavelet representation
Mallat, S. G. (1989) · IEEE Transactions on Pattern Analysis and Machine Intelligence · doi:10.1109/34.192463 · Cited by 21,372 · cited 1×
“[67], the signal was decomposed into various resolution levels.”
BackgroundGoldberger, A. L., Amaral, L. A. N., Glass, L., Hausdorff, J. M., Ivanov, P. C., Mark, R. G., Mietus, J. E., Moody, G. B., Peng, C.-K., & Stanley, H. E. (2000) · Circulation · doi:10.1161/01.cir.101.23.e215 · Cited by 14,809
Adaptive noise cancelling: Principles and applications
Widrow, B., Glover, J. R., McCool, J. M., Kaunitz, J., Williams, C. S., Hearn, R. H., Zeidler, J. R., Eugene Dong, Jr., & Goodlin, R. C. (1975) · Proceedings of the IEEE · doi:10.1109/proc.1975.10036 · Cited by 3,939 · cited 1×
“It uses a “primary” input containing the corrupted signal (signal with artifacts) and a “reference” input containing artifacts correlated in some unknown way with the primary noise [30].”
MethodComparing methods of measurement: why plotting difference against standard method is misleading
Bland, J. M., & Altman, D. G. (1995) · The Lancet · doi:10.1016/s0140-6736(95)91748-9 · Cited by 2,343
Medical Instrumentation-Application and Design
Webster, J. G. (1978) · Journal of Clinical Engineering · doi:10.1097/00004669-197807000-00017 · Cited by 1,847
Robust Statistics: Theory and Methods
Maronna, R. A., Martin, R. D., & Yohai, V. J. (2006) · Wiley Series in Probability and Statistics · doi:10.1002/0470010940 · Cited by 1,655
A Wavelet-Based ECG Delineator: Evaluation on Standard Databases
Martinez, J. P., Almeida, R., Olmos, S., Rocha, A. P., & Laguna, P. (2004) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2003.821031 · Cited by 1,605 · cited 1×
“In order to test the performance of the QRS detection algorithm proposed here, 19 ECG records (Fs = 250Hz) gathered from the human QT database [61, 62] provided by PhysioNet [63] (PN-DB) were used.”
MethodApplications of adaptive filtering to ECG analysis: noise cancellation and arrhythmia detection
Thakor, N. V., & Zhu, Y.-S. (1991) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.83591 · Cited by 885 · cited 2×
“Successive implementations were then improved with the introduction of recursivity [31].”
BackgroundMethodA Real-Time Algorithm for Signal Analysis with the Help of the Wavelet Transform
Holschneider, M., Kronland-Martinet, R., Morlet, J., & Tchamitchian, P. (1989) · Inverse Problems and Theoretical Imaging · doi:10.1007/978-3-642-97177-8_28 · Cited by 633
Research works printed but not linked: 2
- 23. Thakor N, Webster J. The origin of skin potential and its variations. In: Proc. Ann. Conf. Eng. Biol. Med. vol. 20; 1978. p. 212.
- 49. Petersen E, et al. Prevalence of arrhythmias during and immediately after racing in Standardbred trotters–is there an association between arrhythmias and myocardial hypertrophy. In: Proceedings of the BEVA Congress, Liverpool, United Kingdom; 2008. p. 138.
Cited by
Toward a new generation of smart skins
Someya, T., & Amagai, M. (2019) · Nature Biotechnology · doi:10.1038/s41587-019-0079-1 · Cited by 474
Wearable Smart Textiles for Long-Term Electrocardiography Monitoring—A Review
Nigusse, A. B., Mengistie, D. A., Malengier, B., Tseghai, G. B., & Langenhove, L. V. (2021) · Sensors · doi:10.3390/s21124174 · Cited by 184
A movable unshielded magnetocardiography system
Xiao, W., Sun, C., Shen, L., Feng, Y., Liu, M., Wu, Y., Liu, X., Wu, T., Peng, X., & Guo, H. (2023) · Science Advances · doi:10.1126/sciadv.adg1746 · Cited by 116
“…sensors, such as triaxial accelerometers and gyroscopes, to synchronously monitor the movement, many methods have been proposed to detect and suppress the motion artifacts superimposed on MCG (ECG) signals, such as the adaptive filters ( 46 ) and the stationary wavelet transform ( 47 , 48 ).”
Appathurai, A., Carol, J. J., Raja, C., Kumar, S. N., Daniel, A. V., Malar, A. J. G., Fred, A. L., & Krishnamoorthy, S. (2019) · Measurement · doi:10.1016/j.measurement.2019.02.040 · Cited by 94
Steenkiste, G. V., van Loon, G., & Crevecoeur, G. (2020) · Scientific Reports · doi:10.1038/s41598-019-57025-2 · Cited by 64
Redaelli, V., Luzi, F., Mazzola, S., Bariffi, G. D., Zappaterra, M., Costa, L. N., & Padalino, B. (2019) · Animals · doi:10.3390/ani9030084 · Cited by 54
“Further studies are needed to confirm our findings, testing the animals during maximal exercise and competitions, monitoring HR, HRV, and other physiological parameters, such as respiratory rate and rectal temperature, continuously and in real time using new technological tools such as wearable systems applied to the…”
MethodPlant Electrical Signal Classification Based on Waveform Similarity
Chen, Y., Zhao, D., Wang, Z., Wang, Z., Tang, G., & Huang, L. (2016) · Algorithms · doi:10.3390/a9040070 · Cited by 56
“The common artifacts reduction methods in Electrocardiograph (ECG) and Electroencephalograph (EEG) includes wavelet transform [49,50], digital filter [51–53], independent component analysis (ICA) [54], artificial neural networks [55,56], empirical mode decomposition (EMD) [57], morphological filter [58,59], and Hilber…”
MethodRemoval of Motion Artifacts in Capacitive Electrocardiogram Acquisition: A Review
Sirtoli, V. G., Liamini, M., Lins, L. T., Lessard-Tremblay, M., Cowan, G. E. R., Zednik, R. J., & Gagnon, G. (2023) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2023.3270661 · Cited by 58
“The main learning algorithms used for MA removal are: LMS [68], [75], Normalized LMS (NLMS) [70], [76], [84], Error Normalized LMS (ENLMS) [87] and Recursive Least Square (RLS) [74], [88].”
MethodGhaleb, F. A. A., Kamat, M. B., Salleh, M., Rohani, M. F., & Razak, S. B. A. (2018) · PLoS ONE · doi:10.1371/journal.pone.0207176 · Cited by 50
“Thus, QRS detection algorithm is used as a pre-step before removing the motion artefact in some studies such as in [24].”
MethodA Wearable System for the Evaluation of the Human-Horse Interaction: A Preliminary Study
Guidi, A., Lanatà, A., Baragli, P., Valenza, G., & Scilingo, E. P. (2016) · Electronics · doi:10.3390/electronics5040063 · Cited by 45
Self-citationMotion artifact removal based on periodical property for ECG monitoring with wearable systems
Zou, C., Qin, Y., Sun, C., Li, W., & Chen, W. (2017) · Pervasive and Mobile Computing · doi:10.1016/j.pmcj.2017.06.026 · Cited by 41
“For example, a novel method to detect and reduce the motion artifacts in ECG recordings gathered from horses in free movement conditions was proposed in [15], in which the technique for removing the motion artifacts is adaptive filtering method based on the Stationary Wavelet Transform (SWT) decomposition algorithm, a…”
MethodNardelli, M., Lanatà, A., Valenza, G., Felici, M., Baragli, P., & Scilingo, E. P. (2019) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2019.101666 · Cited by 32
Self-citationCritical review of electrocardiography measurement systems and technology
Kalra, A. M., Lowe, A., & Al‐Jumaily, A. M. (2018) · Measurement Science and Technology · doi:10.1088/1361-6501/aaf2b7 · Cited by 35
“Approaches to motion artefact reduction include modifica tion of materials involved in the skin electrode interface and by implementing models and algorithms for reducing motion artefact contribution [78], as discussed in the following sec tion.”
MethodLanata, A., Nardelli, M., Valenza, G., Baragli, P., DrAniello, B., Alterisio, A., Scandurra, A., Semin, G. R., & Scilingo, E. P. (2018) · 2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2018.8512327 · Cited by 27
Self-citationQuantitative heartbeat coupling measures in human-horse interaction
Lanatà, A., Guidi, A., Valenza, G., Baragli, P., & Scilingo, E. P. (2016) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2016.7591286 · Cited by 29
“More precisely, R-peaks related to the human ECG signals were detected by means of the well known Pan-Tompkins method [28], while the method proposed in [27] was used to detect the R-peaks”
MethodSelf-citationValidation of smart textile electrodes for electrocardiogram monitoring in free-moving horses
Guidi, A., Lanatà, A., Valenza, G., Scilingo, E. P., & Baragli, P. (2016) · Journal of Veterinary Behavior · doi:10.1016/j.jveb.2016.10.001 · Cited by 24
Self-citationClark‐Price, S. C., Lascola, K. M., Carter, J. E., da Cunha, A. F., Donaldson, L. L., Doherty, T. J., Martin‐Flores, M., Hofmeister, E. H., Keating, S. C. J., Mama, K. R., et al. (2017) · American Journal of Veterinary Research · doi:10.2460/ajvr.78.6.668 · Cited by 24
Baragli, P., Vitale, V., Sighieri, C., Lanatà, A., Palagi, E., & Reddon, A. R. (2017) · Scientific Reports · doi:10.1038/s41598-017-16729-z · Cited by 24
“After a semiautomatic procedure for visual inspection and editing 22 , the ECG signal was then offline processed using proprietary software where the Stationary Wavelet Transform decomposition algorithm was applied to identify and remove movement artefacts from ECG traces in horses 29 .”
Self-citationArtifact elimination in ECG signal using wavelet transform
Nguyen, T.-N., Nguyen, T.-H., & Ngo, V.-T. (2020) · TELKOMNIKA (Telecommunication Computing Electronics and Control) · doi:10.12928/telkomnika.v18i2.14403 · Cited by 19
Felici, M., Nardelli, M., Lanatà, A., Sgorbini, M., Scilingo, E. P., & Baragli, P. (2020) · Equine Veterinary Journal · doi:10.1111/evj.13296 · Cited by 16
“This results in changes in the ECG during exercise, which can often lead to misinterpretation or misleadingly normal ECG waves [5].”
BackgroundSelf-citationViolence Detection From ECG Signals: A Preliminary Study
Ferdinando, H., Ye, L., Han, T., Zhang, Z., Sun, G., Huuki, T., Sepp, T., & Alasaarela, E. (2017) · Journal of Pattern Recognition Research · doi:10.13176/11.790 · Cited by 14
Design and Implementation of Wearable Dynamic Electrocardiograph Real-Time Monitoring Terminal
Gong, Z., & Ding, Y. (2019) · IEEE Access · doi:10.1109/access.2019.2958992 · Cited by 14
“signal could be effectively reduced [8].”
The role of nonlinear coupling in Human-Horse Interaction: A preliminary study
Lanatà, A., Guidi, A., Valenza, G., Baragli, P., & Scilingo, E. P. (2017) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2017.8037075 · Cited by 13
“Concerning human ECG traces, R-peaks were found through the well-known PanTompkins algorithm [22], while, concerning equine ECG traces, the method proposed in [23] by our group was used.”
MethodSelf-citationVitale, V., Viu, J., Armengou, L., Ríos, J., & José‐Cunilleras, E. L. V. (2020) · American Journal of Veterinary Research · doi:10.2460/ajvr.81.2.147 · Cited by 10
Investigating Textile-Based Electrodes for ECG Monitoring in Veterinary Clinical Practice
Nigusse, A. B., Malengier, B., Mengistie, D. A., Maru, A., & Langenhove, L. V. (2022) · Autex Research Journal · doi:10.2478/aut-2022-0027 · Cited by 9
“Electrocardiography ( ECG ) is the graphical representation of the electrical activity of the heart that contains essential infor - mation concerning the heart condition [ 1,2 ] .”
BackgroundAlternative measurement systems for recording cardiac activity in animals: a pilot study
Kahankova, R. V., Kolařík, J., Brablík, J., Barnova, K., Simkova, I., & Martínek, R. (2022) · Animal Biotelemetry · doi:10.1186/s40317-022-00286-y · Cited by 10
“The studies [19, 83, 84] presented a wearable ECG monitoring system for the capture of cardiac activity in horses.”
BackgroundReservoir sedimentation and flushing
White, W. R. (1990) · Cited by 7
Aarts, R. M., Siegers, E. W., Bragança, F. M. S., & van Weeren, P. R. (2025) · Equine Veterinary Journal · doi:10.1111/evj.70092 · Cited by 14
“27,28 Data can be collected and stored in Holter-mode (e.g., on the device using a SD-card) and transmitted later to a laptop or computer via Bluetooth ® , or collected in the telemetric mode to transmit data to a laptop or computer in real time.”
Vezzosi, T., Vitale, V., Sgorbini, M., Tognetti, R., & Bonelli, F. (2019) · Journal of Equine Veterinary Science · doi:10.1016/j.jevs.2019.06.001 · Cited by 6
Hopenfeld, B. (2022) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2022.01.07.475456 · Cited by 6
Development of 3D-Formed Textile-Based Electrodes with Flexible Interconnect Ribbon
Veske, P., Steenkiste, G. V., Thienpondt, S., Cools, J., Pauw, H. D., & Bossuyt, F. (2025) · Sensors · doi:10.3390/s25020414 · Cited by 5
Ghaleb, F. A. A., Kamat, M. B., Salleh, M., Rohani, M. F., & Hadji, S. E. (2017) · Lecture notes on data engineering and communications technologies · doi:10.1007/978-3-319-59427-9_13 · Cited by 5
Blood pressure wave propagation—a multisensor setup for cerebral autoregulation studies
Zienkiewicz, A., Favre, M., Ferdinando, H., Iring-Sanchez, S., Serrador, J. M., & Myllylä, T. (2021) · Physiological Measurement · doi:10.1088/1361-6579/ac3629 · Cited by 4
“…thewidths of these blocks depend on the heartbeat, HR variability (HRV)was used to determine the maximumwidth of each block that contains S1 and S2 using robust statistics, which utilised themedian and median absolute deviation (mad), as suggested by Lanata et al (2015): width median HRV mad HRV1.”
MethodKang, Y., Choi, S., Koo, C., & Joung, Y.-H. (2022) · Sensors · doi:10.3390/s22218388 · Cited by 6
Rein tension and heart rate variability in horses: an experiment on experience
Galotti, A., Eisersiö, M., Yngvesson, J., Lanatà, A., Maglieri, V., Palagi, E., & Baragli, P. (2025) · Journal of Animal Science · doi:10.1093/jas/skaf146 · Cited by 2
Self-citationHai, B. H. (2023) · Mathematical Modelling and Engineering Problems · doi:10.18280/mmep.100317 · Cited by 2
“ECG signals are frequently affected by various interference components, including electromechanical components, electroencephalograms, and white noise, which cause abnormal variations in the recorded ECG signal's frequency and amplitude [1].”
BackgroundThe impact of skin preparation method on electrocardiogram quality in horses.
McCrae, P., Spong, H., Mahnam, A., Bashura, Y., & Pearson, W. (2024) · PubMed Central · Cited by 2
Optimized Cuffless Blood Pressure Measurement Using ECG, PPG and Linear RegressionPreprint
Boschi, F., Figini, V., Pagana, G., & Visintin, M. (2023) · Preprints.org · doi:10.20944/preprints202312.0273.v1 · Cited by 1
Nardelli, M., Lanatà, A., Valenza, G., Sgorbini, M., Baragli, P., & Scilingo, E. P. (2018) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2018.8512373 · Cited by 1
“Of note, during strenuous exercise the ECG signal becomes more difficult to be critically evaluated because of high heart rates and big number of movement artifacts [9]–[11].”
BackgroundSelf-citationDhara, S. K., Bhanja, N., & Khampariya, P. (2023) · Intelligent Decision Technologies · doi:10.3233/idt-220130 · Cited by 1
“Artifact Removal Phase (ARP) [26]: In ARP, the suggested AF detection model utilized artifact removal as a preprocessing method, where the raw signal data ES colle z is given as the input for removing the artifacts.”
Enhancing the Reliability of Resting ECGs via Deep Learning–Driven Motion Artifact DetectionPreprint
Nanayakkara, S., & Meegama, R. G. N. (2025) · medRxiv · doi:10.1101/2025.08.26.25334522
“Motion artifacts, which result from involuntary muscle movements and skin stretching, can distort ECG signals and complicate interpretation [5,6].”
Face to FACE® – Investigating horses' perception of facial expressions performed by an android robot
Baragli, P., Frassineti, L., Felici, M., Ricci-Bonot, C., Galotti, A., Sgorbini, M., Palagi, E., Cominelli, L., Scilingo, E. P., Scopa, C., et al. (2026) · Applied Animal Behaviour Science · doi:10.1016/j.applanim.2026.106998
Self-citation
Metadata
- Authors
- Lanata, A., Guidi, A., Baragli, P., Valenza, G., & Scilingo, E. P.
- Affiliations
- University of Pisa (IT)
- Venue
- PLOS ONE
- Publisher
- Public Library of Science (PLoS)
- Published
- 2015-10-20
- Volume
- 10
- Issue
- 10
- Pages
- e0140783
- ISSN
- 1932-6203 (ISSN-L)
- PMC
- PMC4618928
- OpenAlex
- W2100756879
- Cited by
- 42
- References
- 72
- Citations by year
- 2026: 1 · 2025: 4 · 2024: 1 · 2023: 5 · 2022: 4 · 2021: 2 · 2020: 4 · 2019: 6 · 2018: 4 · 2017: 6 · 2016: 4
- FWCI
- 3.40
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- PLoS ONE (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Florence Research (University of Florence) (submittedVersion, other-oa)
- RePEc: Research Papers in Economics (submittedVersion)
- CINECA IRIS Institutial research information system (University of Pisa) (submittedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Keywords
- ECG denoising, horses, wearable ECG monitoring, stationary wavelet transform, adaptive filtering, triaxial accelerometer, textile electrodes, equine electrocardiography
- MeSH terms
- Algorithms, Animals, Electrocardiography, Horses, Movement, Signal Processing, Computer-Assisted, Artifacts, Wavelet Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.