A Novel Algorithm for Movement Artifact Removal in ECG Signals Acquired from Wearable Systems Applied to Horses
Open accessThe DOI matches its Crossref record; 72 of the 72 references it cites were found in a registry or an index.
Abstract
This study reports on a novel method to detect and reduce the contribution of movement artifact (MA) in electrocardiogram (ECG) recordings gathered from horses in free movement conditions. We propose a model that integrates cardiovascular and movement information to estimate the MA contribution. Specifically, ECG and physical activity are continuously acquired from seven horses through a wearable system. Such a system employs completely integrated textile electrodes to monitor ECG and is also equipped with a triaxial accelerometer for movement monitoring. In the literature, the most used technique to remove movement artifacts, when noise bandwidth overlaps the primary source bandwidth, is the adaptive filter. In this study we propose a new algorithm, hereinafter called Stationary Wavelet Movement Artifact Reduction (SWMAR), where the Stationary Wavelet Transform (SWT) decomposition algorithm is employed to identify and remove movement artifacts from ECG signals in horses. A comparative analysis with the Normalized Least Mean Square Adaptive Filter technique (NLMSAF) is performed as well. Results achieved on seven hours of recordings showed a reduction greater than 40% of MA percentage (between before- and after- the application of the proposed algorithm). Moreover, the comparative analysis with the NLMSAF, applied to the same ECG recordings, showed a greater reduction of MA percentage in favour of SWMAR with a statistical significant difference (p-value < 0.0.5).
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures














Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“By using additional sensors, such as triaxial accelerometers and gyroscopes, to synchronously monitor the movement, many methods have been proposed to detect and suppress the motion artifacts superimposed on MCG (ECG) signals, such as the adaptive filters (46) and the stationary wavelet transform (47, 48).”
From A movable unshielded magnetocardiography system (2023) · cites this paper as [47]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
A theory for multiresolution signal decomposition: the wavelet representation
Mallat, S. G. (1989) · IEEE Transactions on Pattern Analysis and Machine Intelligence · doi:10.1109/34.192463 · Cited by 21,372 · cited 1×
“[67], the signal was decomposed into various resolution levels.”
BackgroundGoldberger, A. L., Amaral, L. A. N., Glass, L., Hausdorff, J. M., Ivanov, P. C., Mark, R. G., Mietus, J. E., Moody, G. B., Peng, C.-K., & Stanley, H. E. (2000) · Circulation · doi:10.1161/01.cir.101.23.e215 · Cited by 14,809
Adaptive noise cancelling: Principles and applications
Widrow, B., Glover, J. R., McCool, J. M., Kaunitz, J., Williams, C. S., Hearn, R. H., Zeidler, J. R., Eugene Dong, Jr., & Goodlin, R. C. (1975) · Proceedings of the IEEE · doi:10.1109/proc.1975.10036 · Cited by 3,939 · cited 1×
“It uses a “primary” input containing the corrupted signal (signal with artifacts) and a “reference” input containing artifacts correlated in some unknown way with the primary noise [30].”
MethodComparing methods of measurement: why plotting difference against standard method is misleading
Bland, J. M., & Altman, D. G. (1995) · The Lancet · doi:10.1016/s0140-6736(95)91748-9 · Cited by 2,343
Medical Instrumentation-Application and Design
Webster, J. G. (1978) · Journal of Clinical Engineering · doi:10.1097/00004669-197807000-00017 · Cited by 1,847
Robust Statistics: Theory and Methods
Maronna, R. A., Martin, R. D., & Yohai, V. J. (2006) · Wiley Series in Probability and Statistics · doi:10.1002/0470010940 · Cited by 1,655
A Wavelet-Based ECG Delineator: Evaluation on Standard Databases
Martinez, J. P., Almeida, R., Olmos, S., Rocha, A. P., & Laguna, P. (2004) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2003.821031 · Cited by 1,605 · cited 1×
“In order to test the performance of the QRS detection algorithm proposed here, 19 ECG records (Fs = 250Hz) gathered from the human QT database [61, 62] provided by PhysioNet [63] (PN-DB) were used.”
MethodApplications of adaptive filtering to ECG analysis: noise cancellation and arrhythmia detection
Thakor, N. V., & Zhu, Y.-S. (1991) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.83591 · Cited by 885 · cited 2×
“Successive implementations were then improved with the introduction of recursivity [31].”
BackgroundMethodA Real-Time Algorithm for Signal Analysis with the Help of the Wavelet Transform
Holschneider, M., Kronland-Martinet, R., Morlet, J., & Tchamitchian, P. (1989) · Inverse Problems and Theoretical Imaging · doi:10.1007/978-3-642-97177-8_28 · Cited by 633
Laguna, P., Mark, R. G., Goldberg, A., & Moody, G. B. (2002) · Computers in Cardiology 1997 · doi:10.1109/cic.1997.648140 · Cited by 606 · cited 1×
“In order to test the performance of the QRS detection algorithm proposed here, 19 ECG records (Fs = 250Hz) gathered from the human QT database [61, 62] provided by PhysioNet [63] (PN-DB) were used.”
MethodAutomatic Detection of Wave Boundaries in Multilead ECG Signals: Validation with the CSE Database
Laguna, P., Jané, R., & Caminal, P. (1994) · Computers and Biomedical Research · doi:10.1006/cbmr.1994.1006 · Cited by 490
A Review of Fetal ECG Signal Processing Issues and Promising Directions
Sameni (2010) · The Open Pacing, Electrophysiology & Therapy Journal · doi:10.2174/1876536x01003010004 · Cited by 410
A theory of multirate filter banks
Vetterli, M. (1987) · IEEE Transactions on Acoustics, Speech, and Signal Processing · doi:10.1109/tassp.1987.1165137 · Cited by 410
Time-invariant orthonormal wavelet representations
Pesquet, J.-C., Krim, H., & Carfantan, H. (1996) · IEEE Transactions on Signal Processing · doi:10.1109/78.533717 · Cited by 407 · cited 1×
“In this work we propose a novel algorithm based on the stationary wavelet transform (SWT) [45] method for robustly estimating MAs for reliable removal.”
MethodHolschneider, M. (1995) · doi:10.1093/oso/9780198534815.001.0001 · Cited by 381
Causes of poor performance of horses during training, racing, or showing: 348 cases (1992–1996)
Martin, B. B., Reef, V. B., Parente, E. J., & Sage, A. D. (2000) · Journal of the American Veterinary Medical Association · doi:10.2460/javma.2000.216.554 · Cited by 294 · cited 1×
“It has been suggested that both sopraventricular and ventricular premature depolarisation during physical activity can be considered a clue of poor performance in the horses [52].”
BackgroundECG Recording on a Bed During Sleep Without Direct Skin-Contact
Lim, Y. G., Kim, K. K., & Park, K. S. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2006.889194 · Cited by 237
Minimizing Electrode Motion Artifact by Skin Abrasion
Tam, H., & Webster, J. G. (1977) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1977.326117 · Cited by 220 · cited 1×
“Moreover, textile electrodes offer a greater dynamical adaptation to anatomical profiles [24].”
BackgroundRecommendations for Management of Equine Athletes with Cardiovascular Abnormalities
Reef, V. B., Bonagura, J., Buhl, R., McGurrin, M. K. J., Schwarzwald, C. C., van Loon, G., & Young, L. E. (2014) · Journal of Veterinary Internal Medicine · doi:10.1111/jvim.12340 · Cited by 214 · cited 2×
“, glue, adhesive bendage, etc) and the portability achieved by miniaturized holter ECG acquisition platforms, the ECG-related artifacts are even more important in equine sports [47, 53].”
BackgroundMultichannel Electrocardiogram Decomposition Using Periodic Component Analysis
Sameni, R., Jutten, C., & Shamsollahi, M. B. (2008) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2008.919714 · Cited by 207
Arrhythmia ECG Noise Reduction by Ensemble Empirical Mode Decomposition
Chang, K.-M. (2010) · Sensors · doi:10.3390/s100606063 · Cited by 188
Reducing Motion Artifacts and Interference in Biopotential Recording
Webster, J. G. (1984) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1984.325244 · Cited by 181 · cited 1×
“specifically, MAs are mostly generated by skin stretching [19] that is recognized to be the most common source of MAs [20].”
BackgroundRemoving artifacts from electrocardiographic signals using independent components analysis
Barros, A. K., Mansour, A., & Ohnishi, N. (1998) · Neurocomputing · doi:10.1016/s0925-2312(98)00056-3 · Cited by 180
An Adaptive Kalman Filter for ECG Signal Enhancement
Vullings, R., de Vries, B., & Bergmans, J. W. M. (2011) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2010.2099229 · Cited by 174
Fetal Electrocardiogram Enhancement by Time-Sequenced Adaptive Filtering
Ferrara, E. R., & Widraw, B. (1982) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1982.324973 · Cited by 172
Electrocardiogram signals de-noising using lifting-based discrete wavelet transform
Erçelebi, E. (2004) · Computers in Biology and Medicine · doi:10.1016/s0010-4825(03)00090-8 · Cited by 161 · cited 1×
“ECG morphology is strongly affected by movement artifacts (MAs), which must be therefore effectively removed [7, 8].”
BackgroundArtifact removal from electroencephalograms using a hybrid BSS-SVM algorithm
Shoker, L., Sanei, S., & Chambers, J. (2005) · IEEE Signal Processing Letters · doi:10.1109/lsp.2005.855539 · Cited by 160
Lee, J., McManus, D. D., Merchant, S., & Chon, K. H. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2011.2175729 · Cited by 159 · cited 1×
“Moreover, other techniques such as principal component (PCA) and independent component (ICA) analysis [39–42] as well as Empirical Mode Decomposition (EMD) [43, 44] have been explored, but nevertheless the MA issue needs to be still addressed.”
MethodLanatà, A., Valenza, G., Greco, A., Gentili, C., Bartolozzi, R., Bucchi, F., Frendo, F., & Scilingo, E. P. (2014) · IEEE Transactions on Intelligent Transportation Systems · doi:10.1109/tits.2014.2365681 · Cited by 156
Self-citationChawla, M. P. S. (2010) · Applied Soft Computing · doi:10.1016/j.asoc.2010.08.001 · Cited by 143
Reducing skin potential motion artefact by skin abrasion
Burbank, D. P., & Webster, J. G. (1978) · Medical & Biological Engineering & Computing · doi:10.1007/bf02442929 · Cited by 130
Errors in the computerized electrocardiogram interpretation of cardiac rhythm
Shah, A. P., & Rubin, S. A. (2007) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2007.03.008 · Cited by 129 · cited 1×
“In addition, computerised ECG interpretation programs have a high rate of misclassification, especially in non-sinus rhythm events [73].”
BackgroundThe origin of skin-stretch-caused motion artifacts under electrodes
Talhouet, H. D., & Webster, J. G. (1996) · Physiological Measurement · doi:10.1088/0967-3334/17/2/003 · Cited by 124
Adaptive reduction of motion artifact in the electrocardiogram
Tong, D. A., Bartels, K. A., & Honeyager, K. S. (2002) · Proceedings of the Second Joint 24th Annual Conference and the Annual Fall Meeting of the Biomedical Engineering Society] [Engineering in Medicine and Biology · doi:10.1109/iembs.2002.1106451 · Cited by 122 · cited 1×
“In fact, the information obtained from accelerometer and magneto-resistive sensors were recently used [37, 38].”
MethodLanatà, A., Scilingo, E. P., Nardini, E., Loriga, G., Paradiso, R., & De-Rossi, D. (2010) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2009.2037614 · Cited by 111
Self-citationAdaptive noise cancelling of motion artifact in stress ECG signals using accelerometer
Raya, M. A. D., & Sison, L. G. (2002) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2002.1106637 · Cited by 100 · cited 1×
“In fact, the information obtained from accelerometer and magneto-resistive sensors were recently used [37, 38].”
MethodSurvey of cardiac arrhythmias during submaximal and maximal exercise in Thoroughbred racehorses
Ryan, N., Marr, C. M., & McGladdery, A. J. (2005) · Equine Veterinary Journal · doi:10.2746/0425164054530713 · Cited by 98
Comparison of methods for adaptive removal of motion artifact
Hamilton, P. S., Curley, M., Aimi, R., & Sae-Hau, C. (2002) · Computers in cardiology · doi:10.1109/cic.2000.898537 · Cited by 93 · cited 1×
“In some cases, the need to measure the abovementioned parameters led to the development of novel hardware prototypes [1].”
BackgroundMilanesi, M., Martini, N., Vanello, N., Positano, V., Santarelli, M. F., & Landini, L. (2007) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-007-0293-8 · Cited by 90
Cardiac arrhythmias in clinically healthy showjumping horses
Buhl, R., Meldgaard, C. W., & BARBESGAARD, L. (2010) · Equine Veterinary Journal · doi:10.1111/j.2042-3306.2010.00185.x · Cited by 81 · cited 1×
“When manual inspection occurs, the corrupted ECG segments are usually completely removed, thus leading to a great loss of signal and related information [58].”
BackgroundMotion Artifact Removal In Ecg Signals Using Multi-Resolution Thresholding
Muma, M., Strasser, F., & Zoubir, A. M. (2012) · Zenodo (CERN European Organization for Nuclear Research) · doi:10.5281/zenodo.42801 · Cited by 73 · cited 1×
“Moreover, it overcomes many limitations due to empirical thresholds proposed by previous research [46].”
BackgroundCPR artifact removal from human ECG using optimal multichannel filtering
Aase, S. O., Eftestøl, T., Husøy, J. H., Sunde, K., & Steen, P. A. (2000) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.880095 · Cited by 73
(2010) · Elsevier eBooks · doi:10.1016/c2009-0-48166-2 · Cited by 66 · cited 2×
“, glue, adhesive bendage, etc) and the portability achieved by miniaturized holter ECG acquisition platforms, the ECG-related artifacts are even more important in equine sports [47, 53].”
BackgroundA Multimodal Transducer for Cardiopulmonary Activity Monitoring in Emergency
Lanatà, A., Scilingo, E. P., & Rossi, D. D. (2009) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2009.2024414 · Cited by 61
Self-citationMovement-induced potentials in surface electrodes
Ödman, S., & Öberg, P. Å. (1982) · Medical & Biological Engineering & Computing · doi:10.1007/bf02441351 · Cited by 57 · cited 1×
“specifically, MAs are mostly generated by skin stretching [19] that is recognized to be the most common source of MAs [20].”
BackgroundElectrocardiography in horses – part 1: how to make a good recording
Verheyen, T., Decloedt, A., Clercq, D. D., Deprez, P. R., Sys, S. U., & van Loon, G. (2010) · Vlaams Diergeneeskundig Tijdschrift · doi:10.21825/vdt.87463 · Cited by 57 · cited 3×
“This work underlines that quantity and quality of body movements is a fundamental variable that needs to be taken into account for the implementation of algorithms addressing the movement artifact issue [56].”
BackgroundMethodInfluentialMotion artifact suppression: A review of post-processing techniques
Hedley, M., & Yan, H. (1992) · Magnetic Resonance Imaging · doi:10.1016/0730-725x(92)90014-q · Cited by 55
Liu, Y., & PECHT, M. G. (2006) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2006.260006 · Cited by 54 · cited 1×
“Furthermore, skin stretch was also estimated by using non-invasive optical sensors incorporated into ECG electrodes [36].”
MethodECG Enhancement by Adaptive Cancellation of Electrosurgical Interference
Yelderman, M., Widrow, B. C., Cioffi, J. M., Hesler, E., & Leddy, J. A. (1983) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1983.325039 · Cited by 53
Haaksma, J., Brouwer, J., van den Berg, M. P., Dijk, W. A., Dassen, W., Crijns, H., Mulder, L. J. M., & Mulder, G. (1995) · Computing in cardiology · Cited by 52
An ECG-on-Chip for Wearable Cardiac Monitoring Devices
Deepu, C. J., Xu, X., Zou, X., Yao, L., & Lian, Y. (2010) · IEEE International Workshop on Electronic Design, Test and Applications (DELTA) · doi:10.1109/delta.2010.43 · Cited by 49
Multichannel Techniques for Motion Artifacts Removal from Electrocardiographic Signals
Milanesi, M., Martini, N., Vanello, N., Positano, V., Santarelli, M. F., Paradiso, R., Rossi, D. D., & Landini, L. (2006) · Conference proceedings · doi:10.1109/iembs.2006.260464 · Cited by 46
Cottin, F., Barrey, É., Lopes, P., & Billat, V. L. (2006) · Equine Veterinary Journal · doi:10.1111/j.2042-3306.2006.tb05540.x · Cited by 43
On the deconvolution analysis of electrodermal activity in bipolar patients
Greco, A., Lanata, A., Valenza, G., Rota, G., Vanello, N., & Scilingo, E. P. (2012) · 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/embc.2012.6347529 · Cited by 42
Self-citationThe Optimal De-noising Algorithm for ECG Using Stationary Wavelet Transform
Li, S., & Lin, J. (2009) · doi:10.1109/csie.2009.999 · Cited by 42 · cited 2×
“Here, the Haar wavelet Eq (16) was chosen because it has the advantage of being numerically faster, theoretically simpler, physically easier to interpret and exactly reversible without any edge effects: cðtÞ ¼ 1; 0 t < 1=2 1; 1=2 t < 1 0; otherwise ð16Þ 8>>< >>: In this work, a decomposition level equal to 5, as sugge…”
MethodVitale, V., Balocchi, R., Varanini, M., Sgorbini, M., Macerata, A., Sighieri, C., & Baragli, P. (2013) · Journal of Veterinary Behavior · doi:10.1016/j.jveb.2013.02.003 · Cited by 39 · cited 2×
“Moreover, since restriction of movement in horses dramatically increases their sympathetic activity, it is therefore recommended to be avoided [54].”
BackgroundSelf-citationEffect of adaptive motion-artifact reduction on QRS detection.
Hamilton, P. S., Curley, M., & Aimi, R. (2000) · PubMed · Cited by 39
Skolidis, G., Clayton, R. H., & Sanguinetti, G. (2008) · CiteSeer X (The Pennsylvania State University) · Cited by 38
Diseases of the heart and vessels
Young, L. E., & van Loon, G. (2013) · Elsevier eBooks · doi:10.1016/b978-0-7020-4771-8.00032-6 · Cited by 32
Zito, D., Pepe, D., Neri, B., Zito, F., Rossi, D. D., & Lanatà, A. (2008) · International Journal of Telemedicine and Applications · doi:10.1155/2008/328597 · Cited by 31
Adaptive noise removal in the ECG using the Block LMS algorithm
Rahman, M. Z. U., Shaik, R. A., & Reddy, D. V. R. K. (2009) · IEEE International Conference on Adaptive Science & Technology · doi:10.1109/icastech.2009.5409698 · Cited by 31
Trachsel, D. S., Bitschnau, C., Waldern, N. M., Weishaupt, M. A., & Schwarzwald, C. C. (2010) · Equine Veterinary Journal · doi:10.1111/j.2042-3306.2010.00172.x · Cited by 30 · cited 2×
“Moreover, possible mistakes or disagreement could also be due to miscounting events in a large number of events [57].”
BackgroundAdaptive Removal of Motion Artifact
Hamilton, P. S., & Curley, M. (2002) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.1997.754531 · Cited by 30 · cited 1×
“Other studies have introduced stretch sensors to measure MAs effect [33, 34].”
BackgroundExperimental study: brachial motion artifact reduction in the ECG
Luo, S., & Tompkins, W. J. (2002) · Computers in cardiology · doi:10.1109/cic.1995.482564 · Cited by 20 · cited 1×
“Other researchers proposed to measure the voltage difference from two adjacent electrodes to estimate a noise reference signal [35].”
BackgroundComputerized ECG recording in horses during a standardized exercise test
Scheffer, C. J. W., & van Oldruitenborgh-Oosterbaan, M. M. S. (1996) · Veterinary Quarterly · doi:10.1080/01652176.1996.9694601 · Cited by 18 · cited 1×
“The problems related to MAs in the equine ECG become more relevant during exercise [55].”
BackgroundRemoval of artifacts in ECG using Empirical mode decomposition
Anapagamini, S. A., & Rajavel, R. (2013) · doi:10.1109/iccsp.2013.6577061 · Cited by 11 · cited 1×
“ECG morphology is strongly affected by movement artifacts (MAs), which must be therefore effectively removed [7, 8].”
BackgroundUsing an injection signal to reduce motion artifacts in capacitive ECG measurements
Serteyn, A., Vullings, R., Meftah, M., & Bergmans, J. W. M. (2013) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2013.6610620 · Cited by 10
Wu, Y., & Rangayyan, R. M. (2007) · Conference proceedings - Canadian Conference on Electrical and Computer Engineering · doi:10.1109/ccece.2007.220 · Cited by 10
Evaluation of clinical and electrocardiographic changes during the euthanasia of horses
Buhl, R., Andersen, L. O. F., Karlshøj, M., & Kanters, J. K. (2013) · The Veterinary Journal · doi:10.1016/j.tvjl.2012.11.016 · Cited by 9
Valvular regurgitations in the horse: The importance of an exercise ECG
Buhl, R. (2009) · The Veterinary Journal · doi:10.1016/j.tvjl.2009.06.018 · Cited by 4
Research works printed but not linked: 2
- 23. Thakor N, Webster J. The origin of skin potential and its variations. In: Proc. Ann. Conf. Eng. Biol. Med. vol. 20; 1978. p. 212.
- 49. Petersen E, et al. Prevalence of arrhythmias during and immediately after racing in Standardbred trotters–is there an association between arrhythmias and myocardial hypertrophy. In: Proceedings of the BEVA Congress, Liverpool, United Kingdom; 2008. p. 138.
Cited by
Toward a new generation of smart skins
Someya, T., & Amagai, M. (2019) · Nature Biotechnology · doi:10.1038/s41587-019-0079-1 · Cited by 474
Wearable Smart Textiles for Long-Term Electrocardiography Monitoring—A Review
Nigusse, A. B., Mengistie, D. A., Malengier, B., Tseghai, G. B., & Langenhove, L. V. (2021) · Sensors · doi:10.3390/s21124174 · Cited by 184
A movable unshielded magnetocardiography system
Xiao, W., Sun, C., Shen, L., Feng, Y., Liu, M., Wu, Y., Liu, X., Wu, T., Peng, X., & Guo, H. (2023) · Science Advances · doi:10.1126/sciadv.adg1746 · Cited by 116
“…sensors, such as triaxial accelerometers and gyroscopes, to synchronously monitor the movement, many methods have been proposed to detect and suppress the motion artifacts superimposed on MCG (ECG) signals, such as the adaptive filters ( 46 ) and the stationary wavelet transform ( 47 , 48 ).”
Appathurai, A., Carol, J. J., Raja, C., Kumar, S. N., Daniel, A. V., Malar, A. J. G., Fred, A. L., & Krishnamoorthy, S. (2019) · Measurement · doi:10.1016/j.measurement.2019.02.040 · Cited by 94
Steenkiste, G. V., van Loon, G., & Crevecoeur, G. (2020) · Scientific Reports · doi:10.1038/s41598-019-57025-2 · Cited by 64
Metadata
- Authors
- Lanata, A., Guidi, A., Baragli, P., Valenza, G., & Scilingo, E. P.
- Affiliations
- University of Pisa (IT)
- Venue
- PLOS ONE
- Publisher
- Public Library of Science (PLoS)
- Published
- 2015-10-20
- Volume
- 10
- Issue
- 10
- Pages
- e0140783
- ISSN
- 1932-6203 (ISSN-L)
- PMC
- PMC4618928
- OpenAlex
- W2100756879
- Cited by
- 42
- References
- 72
- Citations by year
- 2026: 1 · 2025: 4 · 2024: 1 · 2023: 5 · 2022: 4 · 2021: 2 · 2020: 4 · 2019: 6 · 2018: 4 · 2017: 6 · 2016: 4
- FWCI
- 3.40
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- PLoS ONE (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Florence Research (University of Florence) (submittedVersion, other-oa)
- RePEc: Research Papers in Economics (submittedVersion)
- CINECA IRIS Institutial research information system (University of Pisa) (submittedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Keywords
- ECG denoising, horses, wearable ECG monitoring, stationary wavelet transform, adaptive filtering, triaxial accelerometer, textile electrodes, equine electrocardiography
- MeSH terms
- Algorithms, Animals, Electrocardiography, Horses, Movement, Signal Processing, Computer-Assisted, Artifacts, Wavelet Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.