Automated Measurement of Salivary Cortisol

Open access
van Aken, Maarten O, Romijn, Johannes A, Miltenburg, Johannes A, Lentjes, Eef G W M · 2003 · Clinical Chemistry
Cited by 93

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Abstract

A new serum cortisol assay was introduced on the Elecsys (Roche), a random access analyzer, with reportedly good performance in the low (nmol/L) concentration range (preliminary data provided by the manufacturer) and with low cross-reactivity with cortisone [0.3% at 2.7 μmol/L cortisone (package insert)]. This prompted us to evaluate the performance of this new assay for the measurement of salivary cortisol. For this study, saliva samples were collected with a Salivette® (Sarstedt), with an insert containing a sterile polyester swab for collection of the saliva, yielding a clear and particle-free sample. The salivettes were used according to the instructions provided by the manufacturer. Samples collected this way are stable at room temperature for at least a week and, therefore, offer the opportunity to collect samples at home (1)(2). Salivettes containing saliva were centrifuged at 2000g for 10 min, and the filtrates were stored frozen (−20 °C). Before analysis, the samples were thawed, mixed, and placed on the Elecsys analyzer without pretreatment. The Roche cortisol assay is a competitive electrochemiluminescence immunoassay (ECLIA) that uses a sheep polyclonal antibody. Endogenous cortisol contained in the sample is liberated from the binding proteins by danazol and, subsequently, competes with a cortisol derivative (a cortisol–peptide–Tris bipyridyl ruthenium complex) for the binding sites on the biotinylated antibody. After the addition of streptavidin-coated paramagnetic particles, the biotin on the antibody can bind to the streptavidin of the microparticle and form a complex. This complex is then captured on the surface of the magnetic electrode. Electrical stimulation of the ruthenium complex induces chemiluminescent emission, which is measured by a photomultiplier. The assay was calibrated against Enzymun-Test-Cortisol, which in turn was calibrated via isotope dilution-mass spectrometry. The cortisol assay was used as instructed by the manufacturer without modifications. The sample volume for the assay was 20 μL. Processing time is 18 min. The assay linearity of the cortisol determinations in saliva was studied according to the NCCLS EP-6 protocol, in a saliva sample with a low cortisol concentration with added cortisol (Sigma Chemical Co) at a concentration of 240 nmol/L (3). This sample was diluted with the untreated sample. The correlation between added and measured cortisol was linear with slope = 0.99, intercept = −2.9 nmol/L, r2 = 0.998. A precision profile was established based on saliva samples with or without cortisol additions. Each sample was aliquoted 10 times and stored frozen. An analysis of the samples was performed on 10 different days over a time span of 4 months (Fig. 1 ). The functional detection limit (20% interassay CV) as determined from these measurements is 2.0 nmol/L. The performance of the Elecsys cortisol assay in the low concentration range renders it suitable for the measurement of cortisol concentrations in saliva. Precision profile of the Elecsys salivary cortisol assay. Each point is the mean of 10 measurements. Interassay precision was evaluated according to the NCCLS EP-5 protocol (4). Two pools of saliva with different cortisol concentration were aliquoted, and cortisol was measured on 20 separate days, twice in each sample. Interassay CV was 12% at 11 nmol/L and 5.0% at 50 nmol/L. The somewhat higher CVs, compared with those found in the former experiment, reflected a shift in the concentrations after a calibration. Cross-reactivity by other steroids was investigated by adding increasing amounts of the interfering compound to a saliva sample with a cortisol concentration of 5 nmol/L. Apparent cortisol concentration at 10, 100, and 1000 nmol/L of the cross-reactant cortisone and 11-deoxycortisol never exceeded, respectively, 0.6% and 1.8% of the added amount. For 6β-hydroxycortisol, however, there was an increasing interference, i.e., apparent cortisol at 10 nmol/L 30%, at 100 nmol/L 40%, and at 1000 nmol/L 50% of the added amount. For 21-deoxycortisol at a concentration of 10, 100, and 1000 nmol/L, apparent cortisol concentrations were 44%, 24%, and 14%, respectively. Importantly, interference by cortisone, an abundant steroid metabolite in saliva, is negligible. By contrast, the presence of 6β-hydroxycortisol or 21-deoxycortisol in saliva has not been reported. These steroids are easily excreted in urine (6β-hydroxycortisol) or are present only at low concentrations in serum, which makes it unlikely that the concentrations in saliva will interfere with the cortisol assay. We compared the Elecsys cortisol assay with an in-house RIA (5) in salivary samples. Correlation of salivary cortisol measured by ECLIA (Elecsys) vs RIA was: Elecsys = 0.92 × RIA − 1.6 nmol/L [n = 34; r = 0.84; SD of slope = 0.1; SD intercept = 1.9; Sy · x = 4.3 nmol/L (Deming regression analysis)]. Reference intervals were estimated from an unselected group of healthy individuals, 26 male an

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1373/49.8.1408

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Cited by

See all 93

Metadata

Authors
van Aken, M. O., Romijn, J. A., Miltenburg, J. A., & Lentjes, E. G. W. M.
Venue
Clinical Chemistry
Published
2003-08-01
Volume
49
Issue
8
Pages
1408-1409
ISSN
  • 0009-9147 (ISSN-L)
  • 1530-8561
OpenAlex
W2099145455
Cited by
93
References
2
FWCI
1.27
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.