CoChat

Automated Assay for the Determination of Methylmalonic Acid, Total Homocysteine, and Related Amino Acids in Human Serum or Plasma by Means of Methylchloroformate Derivatization and Gas Chromatography–Mass Spectrometry

Windelberg, Amrei, Årseth, Ove, Kvalheim, Gry, Ueland, Per Magne · 2005 · Clinical Chemistry
Cited by 154

The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.

Abstract

Background: The combined measurement of methylmalonic acid (MMA) and total homocysteine (tHcy) in serum or plasma is useful in diagnosing and distinguishing between cobalamin and folate deficiencies. We developed and validated an isotope-dilution gas chromatography–mass spectrometry (GC-MS) method with automated sample workup for the determination of MMA, tHcy, and the related amino acids Met, total cysteine (tCys), Ser, and Gly in serum or plasma. Methods: Serum or plasma samples (100 μL) were treated with a reductant (dithioerythritol), deproteinized with ethanol, and derivatized and extracted in a single step by the addition of methylchloroformate and toluene. All liquid handling was performed in 96-well (1 mL) microtiter plates by a robotic workstation. The N(S)-methoxycarbonyl ethyl ester derivatives were analyzed by GC-MS in the selected-ion monitoring mode. Results: Detection limits (signal-to-noise ratio, 5:1) were between 0.03 μmol/L (MMA) and 10 μmol/L (Ser, tCys). The assay was linear to 100 μmol/L for MMA and tHcy and to 1000 μmol/L for Met, tCys, Ser, and Gly. The within-day CVs ranged from 0.7% to 3.6% (n = 20), and the between-day CVs from 2.1% to 8.1% (n = 20). The recovery was between 79% and 99% for the different analytes. Conclusion: This assay combines a simple and automated sample preparation with selective and sensitive GC-MS analysis and is well suited for the combined measurement of MMA, tHcy, and the related amino acids.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1373/clinchem.2005.053835

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 1 references

Cited by

See all 154

Metadata

Authors
Windelberg, A., Årseth, O., Kvalheim, G., & Ueland, P. M.
Venue
Clinical Chemistry
Published
2005-11-01
Volume
51
Issue
11
Pages
2103-2109
ISSN
  • 0009-9147 (ISSN-L)
  • 1530-8561
OpenAlex
W2149340056
Cited by
154
References
1
FWCI
4.69
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Automated Assay for the Determination of Methylmalonic Acid, Total Homocysteine, and Related Amino Acids in Human Serum or Plasma by Means of Methylchloroformate Derivatization and Gas Chromatography–Mass Spectrometry. https://app.cochat.ai/papers/10.1373/clinchem.2005.053835
The paper
Windelberg, A., Årseth, O., Kvalheim, G., & Ueland, P. M. (2005). Automated Assay for the Determination of Methylmalonic Acid, Total Homocysteine, and Related Amino Acids in Human Serum or Plasma by Means of Methylchloroformate Derivatization and Gas Chromatography–Mass Spectrometry. Clinical Chemistry. https://doi.org/10.1373/clinchem.2005.053835

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.