The IFCC Reference Measurement System for HbA1c: A 6-Year Progress Report
Abstract
Background: The IFCC Reference Measurement System for hemoglobin (Hb)A1c (IFCC-RM) has been developed within the framework of metrologic traceability and is embedded in a network of 14 reference laboratories. This paper describes the outcome of 12 intercomparison studies (periodic evaluations to control essential elements of the IFCC-RM). Methods: Each study included: unknown samples (to test individual network laboratories); known samples (controls); recently manufactured calibrators (to check calculated assigned value); stored calibrators (to test stability) and a calibration-set (to calibrate the IFCC-RM). The unknown samples are measured by use of the IFCC-RM and the designated comparison methods [DCMs; the National Glycohemoglobin Standardization Program (NGSP) in the US, Japanese Diabetes Society/Japanese Society for Clinical Chemistry (JDS/JSCC) in Japan, and Mono-S in Sweden] are used to investigate the stability of the Master Equation (ME), the relationship between IFCC-RM and DCMs. Results: A total of 105 IFCC-RM data sets were evaluated: 95 were approved, 5 were not, and for 5 no data were submitted. Trend analysis of the MEs, expressed as change in percentage HbA1c per year, revealed 0.000% (NGSP, not significant), −0.030%, (JDS/JSCC; significant) and −0.016% (Mono-S; not significant). Evaluation of long-term performance revealed no systematic change over time; 2 laboratories showed significant bias, 1 poor reproducibility. The mean HbA1c determined by laboratories performing mass spectrometry (MS) was the same as the mean determined by laboratories using capillary electrophoresis (CE), but the reproducibility at laboratories using CE was better. One batch of new calibrators was not approved. All stored calibrators were stable. Conclusion: A sound reference system is in place to ensure continuity and stability of the analytical anchor for HbA1c.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Hoelzel, W., Weykamp, C., Jeppsson, J.-O., Miedema, K., Barr, J. R., Goodall, I., Hoshino, T., John, W. G., Kobold, U., Little, R., et al. (2004) · Clinical Chemistry · doi:10.1373/clinchem.2003.024802 · Cited by 690
Self-citationApproved IFCC reference method for the measurement of HbA1c in human blood
Jeppsson, J.-O., Kobold, U., Barr, J., Finke, A., Hoelzel, W., Hoshino, T., Miedema, K., Mosca, A., Mauri, P., Paroni, R., et al. (2002) · Clinical Chemistry and Laboratory Medicine · doi:10.1515/cclm.2002.016 · Cited by 655
Self-citation
Cited by
International Expert Committee report on the role of the A1c assay in the diagnosis of diabetes
The International Expert Committee (2009) · Diabetes Care · doi:10.2337/dc09-9033 · Cited by 3,185
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Kirkman, M. S., Lernmark, A., Metzger, B. E., & Nathan, D. M. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2010.161596 · Cited by 1,360
Self-citationSacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Kirkman, M. S., Lernmark, A., Metzger, B. E., & Nathan, D. M. (2011) · Diabetes Care · doi:10.2337/dc11-9998 · Cited by 718
Self-citationLittle, R. R., Rohlfing, C. L., & Sacks, D. B. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2010.148841 · Cited by 365
Self-citationSacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Diabetes Care · doi:10.2337/dci23-0036 · Cited by 351
“The IFCC offers manufacturers calibrators and controls as well as a monitoring program (221).”
Self-citation
Metadata
- Authors
- Weykamp, C., John, W. G., Mosca, A., Hoshino, T., Little, R., Jeppsson, J.-O., Goodall, I., Miedema, K., Myers, G., Reinauer, H., Sacks, D. B., Slingerland, R., & Siebelder, C.
- Venue
- Clinical Chemistry
- Publisher
- Oxford University Press (OUP)
- Published
- 2008-02-01
- Volume
- 54
- Issue
- 2
- Pages
- 240-248
- ISSN
- 0009-9147 (ISSN-L)
- 1530-8561
- OpenAlex
- W2157828221
- Cited by
- 206
- References
- 2
- FWCI
- 12.47
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: The IFCC Reference Measurement System for HbA1c: A 6-Year Progress Report. https://app.cochat.ai/papers/10.1373/clinchem.2007.097402
- The paper
- Weykamp, C., John, W. G., Mosca, A., Hoshino, T., Little, R., Jeppsson, J.-O., Goodall, I., Miedema, K., Myers, G., Reinauer, H., Sacks, D. B., Slingerland, R., & Siebelder, C. (2008). The IFCC Reference Measurement System for HbA1c: A 6-Year Progress Report. Clinical Chemistry. https://doi.org/10.1373/clinchem.2007.097402
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.