Influence of Thyroid Hormone Autoantibodies on 7 Thyroid Hormone Assays
The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
Spurious results of thyroid function tests (TFT) can be recognized when they do not reflect the clinical status of the patient or are not internally consistent [e.g., increased FT4 with nonsuppressed thyroid-stimulating hormone (TSH)]. Potential causes of spurious TFT results include nonspecific binding of endogenous circulating factors, such as heterophilic antibodies, with assay reagents(1), the presence of albumin variants found in familial dysalbuminemic hyperthyroxinaemia(2), and thyroid hormone autoantibodies (THAA)(3). We describe a patient who had discordant TFTs due to circulating THAA and report differences in TFT results obtained on a variety of automated immunoassay platforms. A 19-year-old man presented with tiredness spanning a 12-month period following a prolonged viral illness. A family history of hypothyroidism prompted a request for TFTs. These showed increased FT4 of 60 pmol/L (reference interval 10–25 pmol/L) and FT3 of 8.1 pmol/L (reference interval 2.5–6.5 pmol/L), but an nonsuppressed TSH of 0.9 mU/L (reference interval 0.3–5.5 mU/L). The tests were performed on an Immulite 2500 (Siemens Medical Solutions Diagnostics). The patient was clinically euthyroid. Use of residual samples was in accordance with the Human Tissue Act of 2004 and the guidelines of the Royal College of Pathologists. The terms of the Declaration of Helsinki were also fully observed. Test results for heterophilic antibodies using heterophile blocking tubes (Scantibodies) were found to be negative. A screening test for familial dysalbuminemic hyperthyroxinaemia(4) was also negative. Incubation of the patient’s serum with radiolabeled T4 tracer in the presence of 8-anilino-1-naphthalene sulfonic acid, followed by precipitation with polyethylene glycol(5), indicated the presence of T4 autoantibodies as the cause of the erroneously high FT4 result. A similar screen for T3 autoantibodies(5), using a similar method but with radiolabeled T3, was negative for this patient. Serum FT4 measured by equilibrium dialysis (EDFT4) was within the reference interval (Table 1 ). Although the general effects of THAA on assay methods have been previously reviewed(3), the specific effects on currently used assays methods are not precisely known. We compared our TFT results for this patient with results obtained with different analytical platforms: Advia Centaur (Siemens Medical Solutions Diagnostics), Elecsys E170 (Roche Diagnostics), TOSOH AIA 1800 (Tosoh Bioscience), Architect c800 (Abbott Limited Diagnostics), UniCel DxI 800 (Beckman Coulter UK), and Wallac Delfia (PerkinElmer UK). TFT Measurements by 7 Different Immunoassays and Equilibrium Dialysis.1 Patient results outside reference intervals are in bold type. NA indicates not measured. TFT Measurements by 7 Different Immunoassays and Equilibrium Dialysis.1 Patient results outside reference intervals are in bold type. NA indicates not measured. Measurements of TSH obtained with the different systems did not show marked assay-dependent variation (Table 1 ). The Immulite 2500 was the only assay that gave an increased FT3 result of 9.1 pmol/L (reference interval 2.5–6.5 pmol/L). Although no T3 autoantibodies were detected, endogenous antibodies to T4 may possibly cross-react with the T3 label used in the Immulite assay. FT4 results varied from within the reference range with Wallac Delfia to >200% of the upper reference limit with the Advia Centaur, suggesting that observed differences in FT4 results are likely due to differences in assay design. One-step procedures showed the greatest positive interference; high results were observed with both the Advia Centaur and Immulite 2500. The TOSOH AIA 1800, another 1-step assay, appeared less affected, a finding that may be attributable to a different tracer used for signal detection. Both the Advia Centaur and Immulite 2500 use chemiluminescent substrates, whereas the TOSOH uses a fluorogenic tracer. The different molecular sizes of the labels may influence whether they are recognized by the THAA. Two-step methods appear not to be susceptible to THAA interference, because the procedure ensures that there is no contact between serum components and analog tracer(3). This characteristic was evident in the Wallac Delfia, which gave results comparable to the EDFT4. Conversely, the other 2-step assays, Abbott Architect and Beckman DXI 800, gave borderline low results, which may have been related to the nature of the tracer used in these assays. Both the Abbott and Beckman assays use T3 acridinium–labeled tracer in the second step of the reaction, when the tracer binds to unbound sites of the capture antibody. Any THAA left after the washing step will bind only the analyte and not the tracer, and hence increase the signal and decrease the FT4 concentration detected. The Elecsys E170 is a competitive 1-step assay that differs in that it involves 2 incubations steps, but without a washing step in between. First the serum and a ruthenium-labeled capt
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Antibody interference in thyroid assays: a potential for clinical misinformation
Després, N., & Grant, A. M. (1998) · Clinical Chemistry · doi:10.1093/clinchem/44.3.440 · Cited by 262
Cited by
Interferences With Thyroid Function Immunoassays: Clinical Implications and Detection Algorithm
Favresse, J., Burlacu, M.-C., Maiter, D., & Gruson, D. (2018) · Endocrine Reviews · doi:10.1210/er.2018-00119 · Cited by 258
Analytical error and interference in immunoassay: minimizing risk
Sturgeon, C. M., & Viljoen, A. (2011) · Annals of Clinical Biochemistry: International Journal of Laboratory Medicine · doi:10.1258/acb.2011.011073 · Cited by 228
What should be done when thyroid function tests do not make sense?: Atypical thyroid function tests
Gurnell, M., Halsall, D. J., & Chatterjee, V. K. (2011) · Clinical Endocrinology · doi:10.1111/j.1365-2265.2011.04023.x · Cited by 88
Diagnostic procedure in suspected Graves’ disease
Goichot, B., Leenhardt, L., Massart, C., Raverot, V., Tramalloni, J., & Iraqi, H. (2018) · Annales d Endocrinologie · doi:10.1016/j.ando.2018.08.002 · Cited by 32
Metadata
- Authors
- Zouwail, S. A., O’Toole, A. M., Clark, P. M. S., & Begley, J. P.
- Venue
- Clinical Chemistry
- Publisher
- Oxford University Press (OUP)
- Published
- 2008-05-01
- Volume
- 54
- Issue
- 5
- Pages
- 927-928
- ISSN
- 0009-9147 (ISSN-L)
- 1530-8561
- OpenAlex
- W2032008088
- Cited by
- 36
- References
- 1
- FWCI
- 1.09
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.