25-Hydroxyvitamin D: A Difficult Analyte

Open access
Carter, Graham D · 2012 · Clinical Chemistry
Cited by 135

The DOI matches its Crossref record.

Abstract

In their article in the current issue of Clinical Chemistry, Farrell et al. (1) describe the latest in a number of studies (2, 3) highlighting the method-related variability in serum 25-hydroxyvitamin D (25-OHD)2 results. There is common agreement that 25-OHD is a “difficult” analyte. That quality is generally ascribed to its hydrophobic nature, its existence in several different molecular forms, its tight binding to vitamin D– binding protein (VDBP) and, until recently, the absence of reference materials or a reference measurement procedure (RMP) against which assays could be standardized. This last problem was mitigated to some extent with the introduction in 2009 of the NIST Standard Reference Materials (SRM 972 and SRM 2972) and the acceptance of the NIST and University of Ghent liquid chromatography–tandem mass spectrometry (LC-MS/MS) assays as RMPs (4). Unfortunately, 3 of the 4 SRM 972 reference materials are either spiked with exogenous metabolites or diluted with equine serum, characteristics that render these reference materials unsuitable for many immunoassays (5, 6). The history of 25-OHD methodology could serve as a case study of the consequences of transferring a rigorous but labor-intensive method from the unhurried atmosphere of the research laboratory to the bustle of a routine clinical laboratory having to meet tight deadlines. The pioneering competitive protein-binding (CPB) method of Haddad and Chyu (7) involved solvent extraction and chromatography, with the results of every sample corrected for procedural losses. Subsequent attempts at simplifying a CPB method by omitting the chromatography stage proved unsuccessful (8), and an automated version (Nichols Advantage) introduced in 2004 was withdrawn in 2006. The successive abandonment of sample extraction, chromatography, and correction for procedural losses in immunoassays has undoubtedly contributed to the inconsistencies reported by Farrell et al., as have differences in assay standardization. Nevertheless, results submitted to the international Vitamin D External Quality Assessment Scheme (DEQAS) have shown a gradual reduction in interlaboratory imprecision (CV) in recent years—from >30% in 1995 to 15% in 2011. This reduction may be due to the increasing dominance of automated immunoassays, in particular the DiaSorin LIAISON Total method and, more recently, greater use of the NIST SRMs in standardization. Whatever the explanation, it appears that there is currently reasonable agreement, with 1 or 2 notable exceptions, between most 25-OHD methods (9). Nevertheless, the fact that a change of method can affect the continuity of results has implications for the monitoring of patients on vitamin D supplements and for long-term epidemiologic studies. Even within the same method, the consistency of results can be compromised by subtle lot-to-lot changes in reagent kits or more overtly by a change of antibody or in a reformulation of reagents. With the introduction of the RMPs and the expected release in 2012 of a new generation of human serum–based SRMs, assay standardization should become less of an issue, and attention is likely to focus on the reliability of individual results, a far more intractable problem. The use of summary statistics in method comparisons, particularly the correlation coefficient, and linear regression equations can disguise a marked variance in the results given by 2 methods for the same sample. Method differences are best revealed by the use of Bland–Altman plots, which allow discrepancies between results for individual samples to be seen at a glance. These apparently random errors undermine the confidence in the veracity of all 25-OHD results and suggest that some methods are subject to interference from other components of the sample matrix. The study reported by Heijboer et al. (10) (also in the current issue) not only identifies a potential culprit (VDBP) but quantifies its effect. The automated 25-OHD methods, which have necessarily abandoned solvent extraction or protein precipitation to free 25-OHD from binding proteins, are likely to be more susceptible to variations in VDBP concentration. Manufacturers have been coy about disclosing details of the separation step used in these assays, but most appear to involve the denaturation of VDBP via a pH change. Whatever the detail, there is an assumption—with little or no published evidence—that 100% of 25-OHD is displaced from VDBP or, if not, that the percentage displaced in different samples is consistent. The data presented in the report of Heijboer et al. suggest this is not the case. Four of the 5 nonextraction immunoassays compared with the reference LC-MS/MS method had a highly significant correlation of bias to the VDBP concentration. Moreover, the scatter of results indicates that the bias was far from consistent; this finding offers one explanation for the arbitrary nature of method differences with individual samples. A recent and surprising twist in t

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1373/clinchem.2011.180562

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 135

Metadata

Authors
Carter, G. D.
Venue
Clinical Chemistry
Published
2012-03-01
Volume
58
Issue
3
Pages
486-488
ISSN
  • 0009-9147 (ISSN-L)
  • 1530-8561
OpenAlex
W1977406374
Cited by
135
References
0
FWCI
13.40
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: 25-Hydroxyvitamin D: A Difficult Analyte. https://app.cochat.ai/papers/10.1373/clinchem.2011.180562
The paper
Carter, G. D. (2012). 25-Hydroxyvitamin D: A Difficult Analyte. Clinical Chemistry. https://doi.org/10.1373/clinchem.2011.180562

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.