Comparison of Several Point-of-Care Testing (POCT) Glucometers with an Established Laboratory Procedure for the Diagnosis of Type 2 Diabetes Using the Discordance Rate. A New Statistical Approach
The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.
Abstract
The applicability of point-of-care testing (POCT) glucometers for monitoring blood glucose concentrations has been demonstrated. However, their use in diagnosing type 2 diabetes is still debated. Therefore, a new statistical procedure for estimating discordance rates (DRs) was applied in comparing a well-established laboratory method (Ebio) with another laboratory method (Cobas Integra 700) and with several POCT glucometers (Accu-Chek, Accutrend, Elite, HemoCue, Omni) in detecting glucose intolerance states. All procedures led to parallel glucose concentration patterns in capillary blood, venous plasma, and venous blood during oral glucose tolerance tests. However, the mean concentrations differed more or less. The Ebio and Integra results agreed within a maximal deviation of 3%. In blood samples, the HemoCue and Accutrend results were closest to the laboratory procedures (Ebio and Integra) and the highest differences were obtained with the Elite. Comparing whole blood values with those obtained in the aqueous blood compartment (Omni), even greater differences were observed. When all procedures were referred to the same glucose standard, the Ebio, Integra, Accutrend, and Omni results remained almost unchanged, whereas the Elite "moved" toward the Ebio results, and the Accu-Chek results toward the Omni results. Thus, traceability to an aqueous standard was observed with the Ebio, Integra, Accutrend, and Elite in all three sample systems. The Accu-Chek was only traceable in the presence of albumin, and HemoCue was not traceable at all. The clinical relevance of the differences observed between Ebio and POCT glucometers was tested by comparing the relative number of discordant classifications. The highest DRs were observed in the fasting state. They were higher in capillary blood than in the other sample systems. The DRs were found higher with POCT glucometers than with the other established laboratory procedure (Integra). Thus, at least in the fasting state, all POCT glucometers were less reliable than the established laboratory procedures and above the chosen criteria of clinical acceptability (DR < or = 5%). After transforming all glucometer results with a regression function (bias correction), the DRs were less than 5% if compared with the Ebio procedure in all sample systems. In conclusion, the WHO recommendation not to use POCT glucometers for diagnosing type 2 diabetes must be supported. However, after proper recalibration, the tested systems were acceptable. Therefore, manufacturers should reconsider their calibration procedure. Those POCT procedures should be preferred that can be referred to aqueous glucose solutions.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Boyd, J. C., & Bruns, D. E. (2001) · Clinical Chemistry · doi:10.1093/clinchem/47.2.209 · Cited by 227
Analytical performance characteristics should be judged against objective quality specifications
Fraser, C. G., & Petersen, P. H. (1999) · Clinical Chemistry · doi:10.1093/clinchem/45.3.321 · Cited by 194 · cited 2×
“Another approach was developed for comparison studies based on clinical decisions (12), which is a class 1 procedure according to Fraser and Hyltoft Petersen (10).”
BackgroundMethod
Cited by
Can Serum β-Hydroxybutyrate Be Used to Diagnose Diabetic Ketoacidosis?
Sheikh-Ali, M., Karon, B. S., Basu, A., Kudva, Y. C., Muller, L. A., Xu, J., Schwenk, W. F., & Miles, J. M. (2008) · Diabetes Care · doi:10.2337/dc07-1683 · Cited by 159
Assessing the quality of glucose monitor studies: a critical evaluation of published reports
Mahoney, J., & Ellison, J. (2007) · Clinical Chemistry · doi:10.1373/clinchem.2006.083493 · Cited by 48
“The reference lists of the selected articles were also reviewed and personal files were hand searched for additional reports.”
BackgroundEvaluation of hematocrit interference with MyStar extra and seven competitive devices
Demircik, F., Ramljak, S., Hermanns, I., Pfützner, A., & Pfützner, A. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814565790 · Cited by 22
Metadata
- Authors
- Püntmann, I., Wosniok, W., & Haeckel, R.
- Venue
- Clinical Chemistry and Laboratory Medicine
- Publisher
- Walter de Gruyter GmbH
- Published
- 2003-01-17
- Volume
- 41
- Issue
- 6
- Pages
- 809-20
- ISSN
- 1434-6621 (ISSN-L)
- OpenAlex
- W1999331809
- Cited by
- 40
- References
- 2
- FWCI
- 1.27
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.