Evaluation of Hematocrit Interference With MyStar Extra and Seven Competitive Devices

Open access
Demircik, Filiz, Ramljak, Sanja, Hermanns, Iris, Pfützner, Anke, Pfützner, Andreas · 2015 · Journal of Diabetes Science and Technology

The DOI matches its Crossref record; 23 of the 23 references it cites were found in a registry or an index.

Abstract

Background: In previous studies, meters employing dynamic electrochemistry (DE), have been shown to correct for hematocrit (HCT) interference. This laboratory investigation assessed the HCT stability of MyStar Extra (Sanofi) in comparison to 7 competitive devices (Accu-Chek Aviva Nano & Accu-Chek Performa, Roche Diagnostics; Contour XT and Contour Link, Bayer; FreeStyle Freedom Lite, Abbott; MyLife Pura, Ypsomed; OneTouch Verio Pro, LifeScan). Method: Venous heparinized blood was freshly drawn, immediately aliquoted, and manipulated to contain 3 different blood glucose concentrations (50-80 mg/dL, 150-180 mg/dL, and 350-400 mg/dL) and 5 different HCT levels (20-25%, 30-35%, 40-45%, 50-55%, and 60-65%). After careful oxygenation to normal blood oxygen pressure, each of the 15 different samples was measured 8 times with 2 devices and 2 strip lots of each meter (32 measurements/meter/sample). YSI Stat 2300 served as laboratory reference method. Next to determination of the mean absolute relative deviation (MARD), stability to HCT influence was assumed, when less than 10% difference occurred between the highest and lowest mean glucose deviations in relation to HCT over all tested glucose ranges (HIF: hematocrit interference factor). Results: Four of the devices showed stable performance: Contour XT (MARD: 1.3%/HIF: 6.1%), MyStar Extra (4.7%/7.1%), OneTouch Verio Pro (4.5%/7.3%), and Contour Link (6.3%/9.3%). The 4 other meters were influenced by HCT (Accu-Chek Performa: 4.7%/20.9%, Accu-Chek Aviva Nano: 4.5%/22.4%, FreeStyle Freedom Lite: 4.8%/24.5%; MyLife Pura: 6.4%/28.7%). Conclusions: In this study, all meters showed a good accuracy, but only 50% of them, including MyStar Extra, were shown to reliably correct for potential hematocrit influence on the meter results.

Overview

Integrity signals

Registry

Crossref · 10.1177/1932296814565790

Cites

23 of 23 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Verification

Claims the paper makes

We couldn’t retrieve this paper’s full text, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Citations

Cites

Show all 23 references

Research works printed but not linked: 3

  1. Takubo T, 2000, Rinsho Byori, 48, 207
  2. Pfützner A, 2009, Diabetes Stoffw Herz, 18, 387
  3. Iyengar S, 2007, Diabetes Stoffw Herz, 16, 15

Cited by

See all 22

Metadata

Authors
Demircik, F., Ramljak, S., Hermanns, I., Pfützner, A., & Pfützner, A.
Affiliations
  • Pfützner Science & Health Institute, Mainz, Germany
  • Sciema—Science & Marketing UG, Mainz, Germany
  • University Medical Center of the Johannes Gutenberg University Mainz (DE)
Corresponding author
Pfützner, Andreas
Venue
Journal of Diabetes Science and Technology
Published
2014-12-30
Volume
9
Issue
2
Pages
262-267
ISSN
1932-2968 (ISSN-L)
OpenAlex
W2130554818
Cited by
22
References
23
Citations by year
2026: 1 · 2025: 2 · 2024: 2 · 2023: 2 · 2022: 1 · 2021: 1 · 2020: 2 · 2019: 1 · 2018: 4 · 2017: 2 · 2016: 2 · 2015: 2
FWCI
1.33
Type
journal article
Language
en
Open access
Licence
journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
Locations
  • Journal of Diabetes Science and Technology (publishedVersion)
  • Europe PMC (PubMed Central) (publishedVersion)
  • PubMed Central (publishedVersion)
  • PubMed (publishedVersion)
Topics
  • Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
  • Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
Keywords
hematocrit interference, continuous glucose monitoring, blood glucose meters, meter accuracy, HbA1c
MeSH terms
Blood Glucose, Electrochemistry, Hematocrit, Humans, Blood Glucose Self-Monitoring

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Evaluation of Hematocrit Interference With MyStar Extra and Seven Competitive Devices. https://app.cochat.ai/papers/10.1177/1932296814565790
The paper
Demircik, F., Ramljak, S., Hermanns, I., Pfützner, A., & Pfützner, A. (2015). Evaluation of Hematocrit Interference With MyStar Extra and Seven Competitive Devices. Journal of Diabetes Science and Technology. https://doi.org/10.1177/1932296814565790

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.