Evaluation of Hematocrit Interference With MyStar Extra and Seven Competitive Devices
Open accessThe DOI matches its Crossref record; 23 of the 23 references it cites were found in a registry or an index.
Abstract
Background: In previous studies, meters employing dynamic electrochemistry (DE), have been shown to correct for hematocrit (HCT) interference. This laboratory investigation assessed the HCT stability of MyStar Extra (Sanofi) in comparison to 7 competitive devices (Accu-Chek Aviva Nano & Accu-Chek Performa, Roche Diagnostics; Contour XT and Contour Link, Bayer; FreeStyle Freedom Lite, Abbott; MyLife Pura, Ypsomed; OneTouch Verio Pro, LifeScan). Method: Venous heparinized blood was freshly drawn, immediately aliquoted, and manipulated to contain 3 different blood glucose concentrations (50-80 mg/dL, 150-180 mg/dL, and 350-400 mg/dL) and 5 different HCT levels (20-25%, 30-35%, 40-45%, 50-55%, and 60-65%). After careful oxygenation to normal blood oxygen pressure, each of the 15 different samples was measured 8 times with 2 devices and 2 strip lots of each meter (32 measurements/meter/sample). YSI Stat 2300 served as laboratory reference method. Next to determination of the mean absolute relative deviation (MARD), stability to HCT influence was assumed, when less than 10% difference occurred between the highest and lowest mean glucose deviations in relation to HCT over all tested glucose ranges (HIF: hematocrit interference factor). Results: Four of the devices showed stable performance: Contour XT (MARD: 1.3%/HIF: 6.1%), MyStar Extra (4.7%/7.1%), OneTouch Verio Pro (4.5%/7.3%), and Contour Link (6.3%/9.3%). The 4 other meters were influenced by HCT (Accu-Chek Performa: 4.7%/20.9%, Accu-Chek Aviva Nano: 4.5%/22.4%, FreeStyle Freedom Lite: 4.8%/24.5%; MyLife Pura: 6.4%/28.7%). Conclusions: In this study, all meters showed a good accuracy, but only 50% of them, including MyStar Extra, were shown to reliably correct for potential hematocrit influence on the meter results.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Hematocrit affects the glucose results of some meters, with falsely high glucose results at low hematocrits and falsely low results at high hematocrits (143,144).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [144]
Citations
Cites
American Diabetes Association: clinical practice recommendations 1996.
(1996) · PubMed · Cited by 430
Tang, Z., Lee, J. H., Louie, R. F., & Kost, G. J. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-1135-eodhlo · Cited by 293
Consensus Statement on Self-Monitor ing of Blood Glucose
(1987) · Diabetes Care · doi:10.2337/diacare.10.1.95 · Cited by 248
Louie, R. F., Tang, Z., Sutton, D. V., Lee, J. H., & Kost, G. J. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-0257-pocgt · Cited by 180
S. Karon, B., Griesmann, L., Scott, R., Bryant, S. C., Dubois, J. A., Shirey, T. L., Presti, S., & Santrach, P. J. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2007.0257 · Cited by 132
Thirup, P. (2003) · Sports Medicine · doi:10.2165/00007256-200333030-00005 · Cited by 111
Volkova, N. V., & Arab, L. (2005) · American Journal of Kidney Diseases · doi:10.1053/j.ajkd.2005.09.007 · Cited by 105
Plasma glucose measurement with the Yellow Springs Glucose 2300 STAT and the Olympus AU640
Twomey, P. J. (2004) · Journal of Clinical Pathology · doi:10.1136/jcp.2003.013417 · Cited by 74
Hematocrit Interference of Blood Glucose Meters for Patient Self-Measurement
Ramljak, S., Lock, J. P., Schipper, C., Musholt, P. B., Forst, T., Lyon, M., & Pfützner, A. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700123 · Cited by 72
Self-citationMacdougal, I. (1998) · Nephrology Dialysis Transplantation · doi:10.1093/ndt/13.12.3030 · Cited by 50
Research works printed but not linked: 3
- Takubo T, 2000, Rinsho Byori, 48, 207
- Pfützner A, 2009, Diabetes Stoffw Herz, 18, 387
- Iyengar S, 2007, Diabetes Stoffw Herz, 16, 15
Cited by
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Diabetes Care · doi:10.2337/dci23-0036 · Cited by 351
Sacks, D. B., Arnold, M. A., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Clinical Chemistry · doi:10.1093/clinchem/hvad080 · Cited by 203
Comparative accuracy of 17 point-of-care glucose meters
Ekhlaspour, L., Mondesir, D., Lautsch, N., Balliro, C., Hillard, M., Magyar, K., Radocchia, L. G., Esmaeili, A., Sinha, M., & Russell, S. J. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816672237 · Cited by 127
Gervasi, M., Sisti, D., Amatori, S., Zeppa, S. D., Annibalini, G., Piccoli, G., Vallorani, L., Benelli, P., Rocchi, M., Barbieri, E., et al. (2020) · Journal of the International Society of Sports Nutrition · doi:10.1186/s12970-020-0337-0 · Cited by 47
Histogram analysis for smartphone-based rapid hematocrit determination
Jalal, U. M., Kim, S. C., & Shim, J. S. (2017) · Biomedical Optics Express · doi:10.1364/boe.8.003317 · Cited by 36
Metadata
- Authors
- Demircik, F., Ramljak, S., Hermanns, I., Pfützner, A., & Pfützner, A.
- Affiliations
- Pfützner Science & Health Institute, Mainz, Germany
- Sciema—Science & Marketing UG, Mainz, Germany
- University Medical Center of the Johannes Gutenberg University Mainz (DE)
- Corresponding author
- Pfützner, Andreas
- ORCID
- Hermanns, Iris
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2014-12-30
- Volume
- 9
- Issue
- 2
- Pages
- 262-267
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC4604595
- OpenAlex
- W2130554818
- Semantic Scholar
- f87a6af5c979d950bf047a61ef35fe97b4170d41
- Cited by
- 22
- References
- 23
- Citations by year
- 2026: 1 · 2025: 2 · 2024: 2 · 2023: 2 · 2022: 1 · 2021: 1 · 2020: 2 · 2019: 1 · 2018: 4 · 2017: 2 · 2016: 2 · 2015: 2
- FWCI
- 1.33
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Keywords
- hematocrit interference, continuous glucose monitoring, blood glucose meters, meter accuracy, HbA1c
- MeSH terms
- Blood Glucose, Electrochemistry, Hematocrit, Humans, Blood Glucose Self-Monitoring
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Evaluation of Hematocrit Interference With MyStar Extra and Seven Competitive Devices. https://app.cochat.ai/papers/10.1177/1932296814565790
- The paper
- Demircik, F., Ramljak, S., Hermanns, I., Pfützner, A., & Pfützner, A. (2015). Evaluation of Hematocrit Interference With MyStar Extra and Seven Competitive Devices. Journal of Diabetes Science and Technology. https://doi.org/10.1177/1932296814565790
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.