Interchangeability of blood gas, electrolyte and metabolite results measured with point-of-care, blood gas and core laboratory analyzers
The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
Background: The random use of point-of-care, blood gas and core laboratory analyzers to measure electrolytes and metabolites increases the variability in test results. This study was designed to determine whether these results performed on whole blood (point-of-care and blood gas) and plasma (core laboratory) platforms are interchangeable without a risk of clinically relevant discrepancies. Methods: The interchangeability of the blood gas analysis, electrolytes, glucose, lactate and hemoglobin results performed with three stat platforms (i-STAT, Radiometer ABL 825, RapidLab 865) and two core laboratory platforms (Roche Modular P800 and Sysmex XE-2100) were evaluated using samples from critically ill patients. Results: For pH, pCO 2, pO 2 and Ca 2+, good correlation (r-values 0.96–1.00) was observed for all comparative analyzers and the biases were within clinically acceptable limits. Potassium, sodium, glucose, lactate and hemoglobin measured with stat analyzers was highly correlated with measurements performed using the laboratory analyzers, r-values 0.89–1.00 and slopes 0.83–1.07. Mean differences with significant bias (p<0.0001) were found for sodium with blood gas analyzers and hemoglobin with i-STAT. Conclusions: The blood gas, K+, Na+, Ca 2+, glucose and lactate results measured with stat and core laboratory analyzers can be used in different clinical settings for critical care management. However, when monitoring small changes in sodium concentrations, the use of single analyzer is encouraged to avoid analytical differences (acceptance limit ±2%) and misinterpretation of results measured with multiple analyzers. The users of the i-STAT at low hemoglobin values overdiagnose anaemia. Thus, prior to transfusion, the use of hemoglobin concentrations measured with laboratory analyzers is preferable.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Cited by
Allardet-Servent, J., Lebsir, M., Dubroca, C., Fabrigoule, M., Jordana, S., Signouret, T., Castanier, M., Thomas, G., Soundaravelou, R., Lepidi, A., et al. (2017) · PLOS ONE · doi:10.1371/journal.pone.0169593 · Cited by 64
Point-of-care testing: where is the evidence? A systematic survey
Pecoraro, V., Germagnoli, L., & Banfi, G. (2014) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2013-0386 · Cited by 56
Point of care blood gases with electrolytes and lactates in adult emergencies
Kapoor, D., Srivastava, M., & Singh, P. (2014) · International Journal of Critical Illness and Injury Science · doi:10.4103/2229-5151.141411 · Cited by 49
Metadata
- Authors
- Leino, A., & Kurvinen, K.
- Venue
- cclm
- Publisher
- Walter de Gruyter GmbH
- Published
- 2011-04-20
- Volume
- 49
- Issue
- 7
- Pages
- 1187-1191
- ISSN
- 1434-6621 (ISSN-L)
- 1437-4331
- OpenAlex
- W2023576715
- Cited by
- 74
- References
- 1
- FWCI
- 3.66
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.