Comparison of the accuracy of the hemocue glucose analyzer with the Yellow Springs Instrument glucose oxidase analyzer, particularly in hypoglycemia
The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
Objective: We aimed to assess the accuracy of the HemoCue Beta-glucose analyzer (HemoCue) and its correlation with the Yellow Springs Instrument (YSI 2300 STAT; YSI) glucose oxidase analyzer, in particular for hypoglycemic values. Design and methods: Samples were taken from 24 volunteers during hyperinsulinemic glucose clamp studies. Glucose concentrations were determined immediately with the HemoCue in whole blood and with the YSI in plasma from the same sample. After correction for the difference between whole blood and plasma, the paired plasma glucose concentrations were analyzed with various statistical methods. Results: A total of 500 paired glucose values were obtained, 209 of which were in the hypoglycemic range. Mean±s.e. values were 4.85±0.004 mmol/l for the HemoCue (range 1.87–16.17) and 4.81±0.004 mmol/l for the YSI (range 1.88–15.00; P = 0.80). In the hypoglycemic region, values were 3.26±0.004 mmol/l for the HemoCue (range 1.87–5.17) and 3.22±0.003 mmol/l for the YSI (range 1.88–4.20; P = 0.59). Regression analyses were HemoCue = 1.019(YSI) −0.0577 mmol/l, with r = 0.9787 for all values; for hypoglycemic values the HemoCue = 1.1169(YSI) −0.3393 mmol/l, with r = 0.8798. Using Altman’s residual plot, the difference was 0.03±0.0009 mmol/l, with 18 (3.6%) paired values outside the 95% limits of agreement (−0.82 to 0.89 mmol/l). In the hypoglycemic range, the difference was 0.04±0.001 mmol/l, with six (2.9%) values outside the 95% limits of agreement (−0.71 to 0.79 mmol/l). In error grid analysis, one value was in zone D (0.2%) and five values (1%) were in zone B; 98.8% were within zone A. Conclusions: Determination of glucose with the HemoCue system had very good correlation with the YSI system in a broad range of glycemia and also for hypoglycemic values. We believe that these methods can be used interchangeably for research and clinical purposes in adults.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Comparing methods of measurement: why plotting difference against standard method is misleading
Bland, J. M., & Altman, D. G. (1995) · The Lancet · doi:10.1016/s0140-6736(95)91748-9 · Cited by 2,343
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435
Validation of home blood glucose meters with respect to clinical and analytical approaches
Brunner, G. A., Ellmerer, M., Sendlhofer, G., Wutte, A., Trajanoski, Z., Schaupp, L., Quehenberger, F., Wach, P., Krejs, G. J., & Pieber, T. R. (1998) · Diabetes Care · doi:10.2337/diacare.21.4.585 · Cited by 119
Sell-Monitoring of Blood Glucose
(1996) · Diabetes Care · doi:10.2337/diacare.19.1.s62 · Cited by 89
Newer portable glucose meters—analytical improvement compared with previous generation devices?
Weitgasser, R., Gappmayer, B., & Pichler, M. (1999) · Clinical Chemistry · doi:10.1093/clinchem/45.10.1821 · Cited by 85
(1994) · doi:10.2337/diacare.17.6.595 · Cited by 32
Cited by
Translating the A1C assay into estimated average glucose values
Nathan, D. M., Kuenen, J., Borg, R., Zheng, H., Schoenfeld, D., Heine, R. J., & for the A1c-Derived Average Glucose (ADAG) Study Group (2008) · Diabetes Care · doi:10.2337/dc08-0545 · Cited by 1,812
“(The Hemocue measurements, recognized as providing values that are equivalent to laboratory measurements, were used primarily to calibrate the CGM [26].”
MethodThe Accuracy of Point-of-Care Glucose Measurements
Rebel, A., Rice, M. A., & Fahy, B. G. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600228 · Cited by 144
Ranawana, V., Monro, J. A., Mishra, S., & Henry, C. J. K. (2010) · Nutrition Research · doi:10.1016/j.nutres.2010.02.004 · Cited by 129
Cichosz, S. L., Frystyk, J., Hejlesen, O. K., Tarnow, L., & Fleischer, J. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814528838 · Cited by 88
“The blood samples were analyzed with a HemoCue Glucose 201+ glucose analyzer (HemoCue, Sweden), which has been shown to have very good correlation with the YSI system.(25) Hypoglycemia was induced by a single subcutaneous bolus of insulin aspart (NovoRapid; NovoNordisk A/S, Bagsværd, Denmark); bolus size was determine…”
MethodThondre, P. S., Shafat, A., & Clegg, M. E. (2013) · British Journal Of Nutrition · doi:10.1017/s0007114513001682 · Cited by 57
Metadata
- Authors
- Stork, A. D. M., Kemperman, H., Erkelens, D. W., & Veneman, T. F.
- Venue
- European Journal of Endocrinology
- Publisher
- Oxford University Press (OUP)
- Published
- 2005-08
- Volume
- 153
- Issue
- 2
- Pages
- 275-281
- ISSN
- 0804-4643 (ISSN-L)
- 1479-683X
- OpenAlex
- W2133004806
- Cited by
- 86
- References
- 7
- FWCI
- 2.01
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.