Abstract

OBJECTIVES. From January 1, 2005, through December 31, 2005, the Food and Drug Administration received 5 adolescent death reports associated with the use of insulin pumps, raising concerns about use of this device in this age group. To understand better the types of infusion pump–related problems in adolescents, we performed a comprehensive evaluation of insulin and patient-controlled analgesic pump–related adverse events reported for adolescents that were received by the Food and Drug Administration from 1996 to 2005. METHODS. A search for medical device adverse event reports from January 1, 1996 through December 31, 2005, involving insulin pumps or patient-controlled analgesic pumps used by patients who were aged 12 to 21 years was conducted in the Food and Drug Administration's Manufacturer and User Facility Device Experience Database. Reports were reviewed for demographic characteristics, type of adverse event, and patient morbidity, and potential contributory factors were classified from narratives in the reports. RESULTS. A total of 1674 reports were identified: 1594 for insulin pumps and 53 for patient-controlled analgesic pumps. In reports of insulin pump events, there were 13 reported deaths, 2 reports that indicated possible suicide attempts, and several additional reports indicating severe hypoglycemic or hyperglycemic events that seemed to be device-related. A total of 102 (6.4%) insulin-pump reports highlighted factors that may have contributed to the adverse event, including problems associated with compliance, education, sports-related activities, and dropping or damaging the pump. Eighty-two percent of cases involving the insulin pump resulted in hospitalization. Half of the reports involving patient-controlled analgesic pumps indicated that the patient received an excess of medication; tampering and noncompliance were evident in some cases. CONCLUSIONS. Adolescents are a special population who deserve careful consideration of risk and benefit for use of device technology. Studies need to further identify safety problems in this age group.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1542/peds.2007-1707

Cites

3 of 3 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Cited by

See all 53

Metadata

Authors
Cope, J. U., Morrison, A. E., & Samuels-Reid, J.
Venue
Pediatrics
Published
2008-05-01
Volume
121
Issue
5
Pages
e1133-e1138
ISSN
  • 0031-4005 (ISSN-L)
  • 1098-4275
OpenAlex
W2028545243
Cited by
53
References
3
FWCI
4.79
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.