Comparison of the Polar S810i monitor and the ECG for the analysis of heart rate variability in the time and frequency domains
Open accessThe DOI matches its Crossref record; 15 of the 15 references it cites were found in a registry or an index.
Abstract
The aim of the present study was to compare heart rate variability (HRV) at rest and during exercise using a temporal series obtained with the Polar S810i monitor and a signal from a LYNX(R) signal conditioner (BIO EMG 1000 model) with a channel configured for the acquisition of ECG signals. Fifteen healthy subjects aged 20.9 +/- 1.4 years were analyzed. The subjects remained at rest for 20 min and performed exercise for another 20 min with the workload selected to achieve 60% of submaximal heart rate. RR series were obtained for each individual with a Polar S810i instrument and with an ECG analyzed with a biological signal conditioner. The HRV indices (rMSSD, pNN50, LFnu, HFnu, and LF/HF) were calculated after signal processing and analysis. The unpaired Student t-test and intraclass correlation coefficient were used for data analysis. No statistically significant differences were observed when comparing the values analyzed by means of the two devices for HRV at rest and during exercise. The intraclass correlation coefficient demonstrated satisfactory correlation between the values obtained by the devices at rest (pNN50 = 0.994; rMSSD = 0.995; LFnu = 0.978; HFnu = 0.978; LF/HF = 0.982) and during exercise (pNN50 = 0.869; rMSSD = 0.929; LFnu = 0.973; HFnu = 0.973; LF/HF = 0.942). The calculation of HRV values by means of temporal series obtained from the Polar S810i instrument appears to be as reliable as those obtained by processing the ECG signal captured with a signal conditioner.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Heart Rate Variability: Standards of Measurement, Physiological Interpretation, and Clinical Use
Electrophysiology, T. F. O. T. E. S. O. C. T. N. A. (1996) · Circulation · doi:10.1161/01.cir.93.5.1043 · Cited by 17,159
Heart rate variability: Standards of measurement, physiological interpretation, and clinical use
Malik, M., Bigger, J. T., Camm, A. J., Kleiger, R. E., Malliani, A., Moss, A. J., & Schwartz, P. J. (1996) · European Heart Journal · doi:10.1093/oxfordjournals.eurheartj.a014868 · Cited by 15,289
Heart Rate Monitoring: Applications and Limitations
Achten, J., & Jeukendrup, A. E. (2003) · Sports Medicine · doi:10.2165/00007256-200333070-00004 · Cited by 1,166
Heart Rate Variability in Athletes
Aubert, A. E., Seps, B., & Beckers, F. (2003) · Sports Medicine · doi:10.2165/00007256-200333120-00003 · Cited by 911
Validity of the Polar S810 Heart Rate Monitor to Measure R-R Intervals at Rest
GAMELIN, F. X., BERTHOIN, S., & BOSQUET, L. (2006) · Medicine & Science in Sports & Exercise · doi:10.1249/01.mss.0000218135.79476.9c · Cited by 509
Rezk, C. C., Marrache, R. C. B., Tinucci, T., Mion, D., & Forjaz, C. L. M. (2006) · European Journal of Applied Physiology · doi:10.1007/s00421-006-0257-y · Cited by 349
Stauss, H. M. (2003) · American Journal of Physiology-Regulatory, Integrative and Comparative Physiology · doi:10.1152/ajpregu.00452.2003 · Cited by 318
Kingsley, M., Lewis, M. J., & Marson, R. E. (2005) · International Journal of Sports Medicine · doi:10.1055/s-2004-817878 · Cited by 230
Agreement of two different methods for measurement of heart rate variability
Radespiel‐Tröger, M., Rauh, R. D., Mahlke, C., Gottschalk, T., & M�ck-Weymann, M. (2003) · Clinical Autonomic Research · doi:10.1007/s10286-003-0085-7 · Cited by 163
Validity of the Polar S810 to Measure R-R Intervals in Children
Gamelin, F.-X., Baquet, G., Berthoin, S., & Bosquet, L. (2008) · International Journal of Sports Medicine · doi:10.1055/s-2007-964995 · Cited by 128
Cited by
Heart rate variability: are you using it properly? Standardisation checklist of procedures
Catai, A. M., Pastre, C. M., Godoy, M. F. D., Silva, E. D., Takahashi, A. C. D. M., & Vanderlei, L. C. M. (2020) · Brazilian Journal of Physical Therapy · doi:10.1016/j.bjpt.2019.02.006 · Cited by 423
“Heart rate monitors, which are practical and low-cost devices compared to other devices, are used to collect beat-to-beat heart rate and, consequently, analyse HRV indexes.(39) However, heart rate monitors have limited correlation to 24-h holter-ECG.”
MethodSelf-citationValidity of the Polar V800 heart rate monitor to measure RR intervals at rest
Giles, D., Draper, N., & Neil, W. (2016) · European Journal of Applied Physiology · doi:10.1007/s00421-015-3303-9 · Cited by 373
The role of heart rate variability in sports physiology
DONG, J.-G. (2016) · Experimental and Therapeutic Medicine · doi:10.3892/etm.2016.3104 · Cited by 319
Weippert, M., Kumar, M., Kreuzfeld, S., Arndt, D., Rieger, A., & Stoll, R. (2010) · European Journal of Applied Physiology · doi:10.1007/s00421-010-1415-9 · Cited by 300
Can Wearable Devices Accurately Measure Heart Rate Variability? A Systematic Review
Georgiou, K., Larentzakis, A. V., Khamis, N. N., Alsuhaibani, G. I., Alaska, Y. A., & Giallafos, E. J. (2018) · Folia Medica · doi:10.2478/folmed-2018-0012 · Cited by 244
Dobbs, W. C., Fedewa, M. V., MacDonald, H. V., Holmes, C. J., Cicone, Z. S., Plews, D. J., & Esco, M. R. (2019) · Sports Medicine · doi:10.1007/s40279-019-01061-5 · Cited by 189
Evaluation of wearable consumer heart rate monitors based on photopletysmography
Parak, J., & Korhonen, I. (2014) · 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/embc.2014.6944419 · Cited by 148
de Rezende Barbosa, M. P. D. C., Silva, N. T. D., de Azevedo, F. M., Pastre, C. M., & Vanderlei, L. C. M. (2016) · Clinical Physiology and Functional Imaging · doi:10.1111/cpf.12203 · Cited by 129
“Other studies that evaluated the Polar S810iTM (Polar Electro OY) (Kingsley et al., 2005; Gamelin et al., 2006; Vanderlei et al., 2008; Porto & Junqueira, 2009) to correlate the rates of both domains, with the indexes obtained by ECG, also found high correlation coefficients between the devices.”
MethodResultSelf-citationFlatt, A. A., & Esco, M. R. (2013) · Journal of Human Kinetics · doi:10.2478/hukin-2013-0071 · Cited by 115
Validity and Reliability of the Apple Watch for Measuring Heart Rate During Exercise
Khushhal, A., Nichols, S., Evans, W., Gleadall-Siddall, D., Page, R., O'Doherty, A., Carroll, S., Ingle, L., & Abt, G. (2017) · Sports Medicine International Open · doi:10.1055/s-0043-120195 · Cited by 92
“Surrogate measures including HR monitors that connect wirelessly to an in-situ chest strap have been successfully validated against 12-lead ECG devices for measuring HR and heart rate variability at rest and during exercise [10, 14, 18].”
Ellis, R. J., Zhu, B., Koenig, J., Thayer, J. F., & Wang, Y. (2015) · Physiological Measurement · doi:10.1088/0967-3334/36/9/1827 · Cited by 85
“The combined advantages of easy and robust electrocardiogram (ECG) recording, affordable and valid ambulatory monitoring devices (e.g. Vanderlei et al 2008), and intuitive and comprehensive open-source analysis software (e.g. Tarvainen et al 2009) suggest that this growth will continue well into…”
BackgroundBenjamim, C. J. R., Kliszczewicz, B., Garner, D. M., Cavalcante, T. C. F., da Silva, A. A. M., Santana, M. D. R., & Valenti, V. E. (2020) · Journal of the American College of Nutrition · doi:10.1080/07315724.2019.1705201 · Cited by 39
Suh, S., Ellis, R. J., Sollers, J. J., Thayer, J. F., Yang, H.-C., & Emery, C. F. (2013) · Journal of Psychosomatic Research · doi:10.1016/j.jpsychores.2013.02.007 · Cited by 28
“Two previous validation studies [34,35] have compared the S810i with standard ECG recordings during supine rest, walking, and exercise.”
Caffeine slows heart rate autonomic recovery following strength exercise in healthy subjects
Benjamim, C. J. R., Monteiro, L. R. L., Pontes, Y. M. D. M., Silva, A. A. M. D., Souza, T. K. M. D., Valenti, V. E., Garner, D. M., & Cavalcante, T. C. F. (2021) · Revista Portuguesa de Cardiologia (English Edition) · doi:10.1016/j.repce.2020.07.021 · Cited by 24
“In this study, we investigate the autonomic modulation of heart rhythm via heart rate variability (HRV), which is a simple and non-invasive method to assess neural control of the heart.(12)”
Method
Metadata
- Authors
- Vanderlei, L. C. M., Silva, R. A., Pastre, C. M., Azevedo, F. M., & Godoy, M. F.
- Venue
- Brazilian Journal of Medical and Biological Research
- Publisher
- FapUNIFESP (SciELO)
- Published
- 2008-09-30
- Volume
- 41
- Issue
- 10
- Pages
- 854-859
- ISSN
- 0100-879X (ISSN-L)
- 1414-431X
- OpenAlex
- W2117887112
- Cited by
- 296
- References
- 15
- FWCI
- 8.36
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.