Comparison of three hand dynamometers in relation to the accuracy and precision of the measurements
Open accessThe DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
BACKGROUND: Given the variety of available hand-held dynamometers and their different handle shapes, reliability studies are needed. OBJECTIVES: To compare the accuracy and reliability between three different hand-held dynamometers and analyze the influence of their handles on grip strength. METHODS: The tests were performed with the Jamar® dynamometer, the Takei® dynamometer and the EMG System Manual Transducer with modified handle. Eighteen healthy volunteers aged 20.0±1.3 years without history of musculoskeletal disorders or trauma in the evaluated limbs were included. Data normality was tested using the Shapiro-Wilk test. To verify possible differences between the dynamometers, repeated measures ANOVA was administered, followed by Tukey post-hoc tests. Reliability between measurements was evaluated using intraclass correlation coefficient (ICC) and agreement was tested using Bland and Altman plots. The dynamometers calibration process was evaluated using linear regressions. RESULTS: We observed statistically significant differences on the female group between the Jamar® and the Takei® dynamometers (females p<0.001 and males p=0.022) and the EMG System Manual Transducer (female p<0.001 and males p=0.007). However, the Takei® dymamometer and the EMG System Manual Transducer were similar for both female (p=0.161) and male groups (p=0.850). Although acceptable values of intraclass correlation coefficients between measurements were identified, low agreement between the Jamar® dynamometer and all other instruments was found. CONCLUSIONS: The results demonstrated that there is an influence of the dynamometer's handle shapes on the measurements of grip strength. Furthermore, the results demonstrated the need for previous calibration of this type of instrument.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Statistical methods for assessing agreement between two methods of clinical measurement
Bland, J. M., & Altman, D. G. (2010) · International Journal of Nursing Studies · doi:10.1016/j.ijnurstu.2009.10.001 · Cited by 5,220
Lauretani, F., Russo, C. R., Bandinelli, S., Bartali, B., Cavazzini, C., Di Iorio, A., Corsi, A. M., Rantanen, T., Guralnik, J. M., & Ferrucci, L. (2003) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00246.2003 · Cited by 1,946 · cited 1×
“The grips are identified by anterior (1) and posterior (2) aspects that support the fingers and the thenar and hypothenar eminences.”
BackgroundHand-Grip Dynamometry Predicts Future Outcomes in Aging Adults
Bohannon, R. W. (2008) · Journal of Geriatric Physical Therapy · doi:10.1519/00139143-200831010-00002 · Cited by 948
Cited by
Grip Strength across the Life Course: Normative Data from Twelve British Studies
Dodds, R. M., Syddall, H. E., Cooper, R., Benzeval, M., Deary, I. J., Dennison, E. M., Der, G., Gale, C. R., Inskip, H. M., Jagger, C., et al. (2014) · PLoS ONE · doi:10.1371/journal.pone.0113637 · Cited by 1,221
Rijk, J. M., Roos, P. R., Deckx, L., van den Akker, M., & Buntinx, F. (2016) · Geriatrics & Gerontology International · doi:10.1111/ggi.12508 · Cited by 376
Differences in handgrip strength protocols to identify sarcopenia and frailty - a systematic review
Sousa-Santos, A. R., & Amaral, T. F. (2017) · BMC Geriatrics · doi:10.1186/s12877-017-0625-y · Cited by 273
Handgrip strength: a comparison of values obtained from the NHANES and NIH Toolbox studies
Bohannon, R. W., Wang, Y.-C., Yen, S.-C., & Grogan, K. A. (2019) · American Journal of Occupational Therapy · doi:10.5014/ajot.2019.029538 · Cited by 72
“These include, but are not limited to, differences in the hand dynamometer used (e.g., Takei vs. Jamar; Amaral et al., 2012), dynamometer handle position (Trampisch et al., 2012), upper limb position (e.g., shoulder, elbow; Desrosiers et al., 1995; Oxford, 2000; Su et al., 1994), and posture (e.g.,…”
MethodLópez-Samanes, Á., Gómez Parra, A., Moreno-Pérez, V., & Courel-Ibáñez, J. (2020) · Nutrients · doi:10.3390/nu12010188 · Cited by 39
Ha, Y.-C., Yoo, J.-I., Park, Y.-J., Lee, C. H., & Park, K.-S. (2018) · Journal of Bone Metabolism · doi:10.11005/jbm.2018.25.4.243 · Cited by 28
“Amaral et al.[12] compared the accuracy and precision of measurements of the grip strength of 3 different dynamometers.”
Metadata
- Authors
- Amaral, J. F., Mancini, M., & Novo Júnior, J. M.
- Venue
- Brazilian Journal of Physical Therapy
- Publisher
- FapUNIFESP (SciELO)
- Published
- 2012-06
- Volume
- 16
- Issue
- 3
- Pages
- 216-224
- ISSN
- 1413-3555 (ISSN-L)
- OpenAlex
- W1977209005
- Cited by
- 110
- References
- 4
- FWCI
- 3.91
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.