Analytical performance specifications in clinical chemistry: the holy grail?
Open accessAbstract
The conference in Milan in 2014 resulted in a consensus statement on quality specifications. An EFLM Task and Finish Group was started to deal with issues that remained unresolved and needed further development including total analytical error methods. Most importantly were questions concerning measurement uncertainty (MU) and flaws in the model for calculating the permissible (or allowable) total error (TE). This is related to performance specifications based on biological variation, the second model of the Milan consensus. Performance specifications are closely linked to the work of Westgard and the concept of TE, that depends on the combined effect of the random and systematic errors of the method which is compared to a permissible TE that is in most cases based on biological variation. Where the first models for performance specifications did not include bias, later models did. The concept of bias is however complicated, both in the estimation of the value, and in the integration of bias in a mathematical model. In many fields outside clinical chemistry, the error concept has been abandoned in favour of the MU model. Bias is commonly excluded from this model, where only the uncertainty of bias is taken into account. There is no general consensus on these issues, and challenges remain to integrate the different concepts, both for the definition of performance, of performance specifications as for quality control procedures.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Using biological variation data, APS can be defined as minimum, desirable, and optimum for both bias and imprecision.
“Using biological variation data, the APS can be defined, as minimum, desirable, and optimum for both bias and imprecision97,98 (Table 4 and Supplementary Table S1).”
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Robustness in the Strategy of Scientific Model Building
Box, G. E. P. (1979) · Elsevier eBooks · doi:10.1016/b978-0-12-438150-6.50018-2 · Cited by 1,698 · cited 1×
“It has been stated “all models are wrong, but some models are useful” (31).”
BackgroundSandberg, S., Fraser, C. G., Horvath, A. R., Jansen, R., Jones, G., Oosterhuis, W., Petersen, P. H., Schimmel, H., Sikaris, K., & Panteghini, M. (2015) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2015-0067 · Cited by 563
Self-citationProposals for Setting Generally Applicable Quality Goals Solely Based on Biology
Fraser, C. G., Petersen, P. H., Libeer, J.-C., & Ricós, C. (1997) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1177/000456329703400103 · Cited by 368 · cited 1×
“Fraser (18) adapted this expression, because some measurands are subject to tight homeostatic control (e.”
BackgroundStability study of 81 analytes in human whole blood, in serum and in plasma.
Oddoze, C., Lombard, E., & Portugal, H. (2012) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2012.01.012 · Cited by 302
Criteria for judging precision and accuracy in method development and evaluation
Westgard, J. O., Carey, R. N., & Wold, S. (1974) · Clinical Chemistry · doi:10.1093/clinchem/20.7.825 · Cited by 236 · cited 1×
“In 1974 Westgard, Carey, and Wold (5) introduced the concept of Total Analytical Error to provide a quantitative measure for the acceptability of analytical performance.”
BackgroundCeriotti, F., Fernandez-Calle, P., Klee, G. G., Nordin, G., Sandberg, S., Streichert, T., Vives-Corrons, J.-L., & Panteghini, M. (2017) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2016-0091 · Cited by 181 · cited 3×
“The second principle is based on models that use (components of) biological variation and can be applied for measurands that are in steady state or can be “transformed” to a steady state situation in biological fluids (4).”
BackgroundMethodInfluentialGowans, E. M. S., Petersen, P. H., Blaabjerg, O., & Hørder, M. (1988) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.3109/00365518809088757 · Cited by 162 · cited 8×
“In the same year Gowans et al. published a model based on reference intervals, that formed the basis for many other studies (14).”
BackgroundMethodInfluentialDesirable standards for laboratory tests if they are to fulfill medical needs
Fraser, C. G., & Petersen, P. H. (1993) · Clinical Chemistry · doi:10.1093/clinchem/39.7.1447 · Cited by 109 · cited 1×
“Bias and imprecision specifications are combined to set quality limits based on biological variation: (15,16): pTAE = 0.”
BackgroundThe use of error and uncertainty methods in the medical laboratory
Oosterhuis, W. P., Bayat, H., Armbruster, D., Coşkun, A., Freeman, K. P., Kallner, A., Koch, D. D., MacKenzie, F., Migliarino, G. A., Orth, M., et al. (2017) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2017-0341 · Cited by 100 · cited 2×
“5CVI relates to performance specifications that—according to the TE model—will lead to a sigma metric below 3 that cannot be maintained by internal quality assurance (27).”
BackgroundSelf-citationGross overestimation of total allowable error based on biological variation.
Oosterhuis, W. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2011.165308 · Cited by 69 · cited 2×
“Both maxima of permissible bias and imprecision are added to obtain pTAE, a pragmatic solution first proposed for the use in proficiency testing (3).”
BackgroundSelf-citationQuality goals in external quality assessment are best based on biology.
Fraser, C. G., & Petersen, P. H. (1993) · Scandinavian journal of clinical and laboratory investigation. Supplementum · doi:10.1080/00365519309085446 · Cited by 60 · cited 1×
“Bias and imprecision specifications are combined to set quality limits based on biological variation: (15,16): pTAE = 0.”
BackgroundPermissible limits for uncertainty of measurement in laboratory medicine
Haeckel, R., Wosniok, W., Gurr, E., et al. (2015) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2014-0874 · Cited by 55 · cited 2×
“As an alternative the actual reference interval limits can be used as starting point in the model (20).”
BackgroundMethodAnalytical Goals in Clinical Chemistry: Their Relationship to Medical Care
(1979) · American Journal of Clinical Pathology · doi:10.1093/ajcp/71.6.624 · Cited by 29
The International Bureau of Weights and Measures, 1875-1975
Page, C., & Vigoureux, P. (2017) · doi:10.6028/nbs.sp.420 · Cited by 27 · cited 1×
“Uncertainty methods originated in physical measurements and chemistry (6,7).”
Graphical interpretation of confidence curves in rankit plots
Petersen, P. H., Blaabjerg, O., Andersen, M., et al. (2004) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2004.122 · Cited by 27 · cited 1×
“It should be noted that in most distributions CVG and CVI are log-Gaussian, as are most reference ranges (21).”
BackgroundProposal for the modification of the conventional model for establishing performance specifications
Oosterhuis, W. P., & Sandberg, S. (2015) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2014-1146 · Cited by 26 · cited 3×
“In this model the analytical variation is included as it was when the reference ranges were determined or confirmed (19).”
BackgroundMethodSelf-citationHow to define a significant deviation from the expected internal quality control result
Ceriotti, F., Brugnoni, D., & Mattioli, S. (2015) · Clinical Chemistry and Laboratory Medicine · doi:10.1515/cclm-2014-1149 · Cited by 22 · cited 1×
“It might, however, seem a paradox that when the results of quality assurance measurements are within the limits, in retrospect these results will be well within the pTAE limits with a wide margin (28).”
BackgroundÅsberg, A., Odsæter, I. H., Carlsen, S., et al. (2015) · Clinical Chemistry and Laboratory Medicine · doi:10.1515/cclm-2014-1125 · Cited by 18
Macdonald, R. (2006) · Journal of Laboratory Medicine · doi:10.1515/jlm.2006.018 · Cited by 9
Supplements to a recent proposal for permissible uncertainty of measurements in laboratory medicine
Haeckel, R., Wosniok, W., Gurr, E., et al. (2016) · doi:10.1515/labmed-2015-0112 · Cited by 9 · cited 1×
“Although the contribution of the analytical variation to the total variation will in many cases be small (29), it is a simplification to assume that reference ranges are only determined by biological variation as has been done in many models.”
BackgroundWilliams, A. (2001) · Springer eBooks · doi:10.1007/3-540-27093-0_45 · Cited by 7 · cited 1×
“Laboratory medicine is still struggling to adapt to this long tradition established by physical metrology laboratories (8).”
BackgroundWhat can we learn from traceability in physical measurements
Williams, A. (2000) · Springer eBooks · doi:10.1007/3-540-27093-0_10 · Cited by 5
Flagged first, then by citations
Cited by
Sorted by citations
Van Uytfanghe, K., Ehrenkranz, J., Halsall, D., Hoff, K., Loh, T. P., Spencer, C. A., Köhrle, J., & ATA Thyroid Function Tests Writing Group (2023) · Thyroid® · doi:10.1089/thy.2023.0169 · Cited by 100
“Using biological variation data, the APS can be defined, as minimum, desirable, and optimum for both bias and imprecision 97,98 (Table 4 and Supplementary Table S1).”
MethodOttka, C., Vapalahti, K., Puurunen, J., Vahtera, L., & Lohi, H. (2021) · Veterinary Clinical Pathology · doi:10.1111/vcp.12954 · Cited by 22
Outcome-based analytical performance specifications: current status and future challenges
Horvath, A. R., Bell, K., Ceriotti, F., Jones, G. R. D., Loh, T. P., Lord, S. J., & Sandberg, S. (2024) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2024-0125 · Cited by 23
Cho, J., Seo, D. M., & Uh, Y. (2019) · Annals of Laboratory Medicine · doi:10.3343/alm.2020.40.3.201 · Cited by 16
Performance specifications for sodium should not be based on biological variation
Oosterhuis, W. P., Coşkun, A., Sandberg, S., et al. (2023) · Clinica Chimica Acta · doi:10.1016/j.cca.2023.117221 · Cited by 15
Self-citationHollestelle, M. J., Ruinemans‐Koerts, J., Idema, R. N., Meijer, P., & de Maat, M. P. M. (2019) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2018-0934 · Cited by 7
Milinković, N., Jovičić, S. Ž., & Ignjatović, S. (2020) · Critical Reviews in Clinical Laboratory Sciences · doi:10.1080/10408363.2020.1784838 · Cited by 6
“accredited laboratories to evaluate MU [40].”
MethodBiological variation of major gut-derived uremic toxins in the serum of healthy adult cats
Summers, S., Quimby, J. M., Yao, L., Hess, A. M., Broeckling, C. D., & Lappin, M. R. (2021) · Journal of Veterinary Internal Medicine · doi:10.1111/jvim.16043 · Cited by 6
Panteghini, M. (2025) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2025-1014 · Cited by 4
Characteristics of a novel NMR-based metabolomics platform for dogsPreprint
Ottka, C., Vapalahti, K., Puurunen, J., Vahtera, L., & Lohi, H. (2019) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/871285 · Cited by 4
A software tool for applying Bayes' theorem in medical diagnostics
Chatzimichail, T., & Hatjimihail, A. T. (2024) · BMC Medical Informatics and Decision Making · doi:10.1186/s12911-024-02721-x · Cited by 4
“We should, however, keep in mind that "all models will be based on assumptions and can only approach complex reality" [88], as "all models are wrong, but some models are useful" [89].”
Hamid, H., Al-Wafi, M., Ahmida, M., et al. (2025) · Cureus · doi:10.7759/cureus.77236 · Cited by 1
Does Reproducibility Drive Clinical Accuracy?
Emancipator, K. (2021) · American Journal of Clinical Pathology · doi:10.1093/ajcp/aqaa267 · Cited by 1
From error to uncertainty: A search for variabilities in the laboratory reports
Goswami, K., Sahoo, S., & Mukhopadhyay, K. (2023) · Indian Journal of Experimental Biology · doi:10.56042/ijeb.v61i07.2284
Shipkova, M., Wieland, E., & Schütz, E. (2026) · Therapeutic Drug Monitoring · doi:10.1097/ftd.0000000000001490
Elsharkawi, S. M. (2026) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2026.111208
Metadata
- Authors
- Oosterhuis, W. P.
- Affiliations
- Zuyderland Medisch Centrum (NL)
- Corresponding author
- Oosterhuis, Wytze P.
- ORCID
- Oosterhuis, Wytze P.
- Venue
- Journal of Laboratory and Precision Medicine
- Publisher
- AME Publishing Company
- Published
- 2017-09
- Volume
- 2
- Pages
- 78-78
- ISSN
- 2519-9005 (ISSN-L)
- OpenAlex
- W2758988133
- Cited by
- 16
- References
- 22
- Citations by year
- 2026: 2 · 2025: 2 · 2024: 2 · 2023: 3 · 2021: 3 · 2020: 1 · 2019: 3
- FWCI
- 0.81
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by-nc-nd
- Locations
- Journal of Laboratory and Precision Medicine (publishedVersion, cc-by-nc-nd)
- Topics
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Physics and Astronomy › Instrumentation › Scientific Measurement and Uncertainty Evaluation
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Keywords
- analytical performance specifications, total error, measurement uncertainty, biological variation, total allowable error, bias, Westgard rules, quality control, clinical chemistry
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review