TLD-100 post-irradiation fading characteristics according to IEC 62387:2020 standard

Open access
Stanković Petrović, Jelena, Krajinović, Marko, Kržanović, Nikola, Živanović, Miloš, Kojić, Andrea, Božović, Predrag·2021·Book of Abstracts

Abstract

The results of the post-irradiation fading of whole body dosemeters based on two TLD-100TM (Thermo Scientific™ Harshaw™, USA) detectors are presented. The dosemeters are regularly used by accredited individual monitoring service (IMS) at Vinca Institute of Nuclear Science (VINS), Belgrade, Serbia. The testing of post-irradiation fading was carried out according to International Electrotechnical Commission (IEC) 62387:2020 standard. The irradiations are done in S-Cs137 field at secondary standard dosimetry laboratory (SSDL) at VINS, according to International Organization for Standardization document ISO 4037-3:2019. The chosen reference personal dose equivalent value was 3 mSv. The research was carried out in the period from October 2020 to February 2021, and it lasted 128 days. The frequency of irradiations was approximately 7 days, while 3 periods between irradiations were longer than 10 days, due to COVID-19 pandemic. The irradiations were performed at different dates; thus, all dose readouts were done on the same day to prevent influence of reader’s instabilities. The irradiated dosemeters were stored at the same room where the average temperature was nearly 20°C. There were 14 groups with 6 dosemeters for irradiation and 2 dosemeters for natural background radiation level correction. The reader was Harshaw TLD™ Model 6600 Plus Automated Reader (Thermo Fisher Scientific, USA). The whole glow curve was used for dose estimation (all of 200 channels). The time as influence quantity was considered to be of type F, thus the range of relative response was limiting factor in the analysis. The results showed that the maximum measurement time tmax is 72 days for deep dose (Hp(10)) detector, and 85 days for shallow dose (Hp(0.07)) detector. Thus, the standard’s requirement is satisfied, as it is required minimum of 30 days. The relative response range for all of the 14 groups was from 0.82 to 1.14 and from 0.83 to 1.17, for deep and shallow dose, respectively. The research has limitation as the irradiations were organized aligned to COVID-19 working schedule. One of the consequences of this timetable is lack of 7 days’ time point, thus the values from 16 days’ time point was used as referential.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.21175/rad.abstr.book.2021.31.12

Cites

41 of 41 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

1 retracted

Show all 41 references

Metadata

Authors
Stanković Petrović, J., Krajinović, M., Kržanović, N., Živanović, M., Kojić, A., & Božović, P.
Venue
Book of Abstracts
Publisher
RAD Centre
Published
2021
OpenAlex
W4213346811
Cited by
273
References
41
Type
conference abstract
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: TLD-100 post-irradiation fading characteristics according to IEC 62387:2020 standard. https://app.cochat.ai/papers/10.21175/rad.abstr.book.2021.31.12
The paper
Stanković Petrović, J., Krajinović, M., Kržanović, N., Živanović, M., Kojić, A., & Božović, P. (2021). TLD-100 post-irradiation fading characteristics according to IEC 62387:2020 standard. Book of Abstracts. https://doi.org/10.21175/rad.abstr.book.2021.31.12

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.