Accuracy of a Wrist-Worn Wearable Device for Monitoring Heart Rates in Hospital Inpatients: A Prospective Observational Study

Open access
Kroll, Ryan R, Boyd, J Gordon, Maslove, David M·2016·Journal of Medical Internet Research

Abstract

BACKGROUND: As the sensing capabilities of wearable devices improve, there is increasing interest in their application in medical settings. Capabilities such as heart rate monitoring may be useful in hospitalized patients as a means of enhancing routine monitoring or as part of an early warning system to detect clinical deterioration. OBJECTIVE: To evaluate the accuracy of heart rate monitoring by a personal fitness tracker (PFT) among hospital inpatients. METHODS: We conducted a prospective observational study of 50 stable patients in the intensive care unit who each completed 24 hours of heart rate monitoring using a wrist-worn PFT. Accuracy of heart rate recordings was compared with gold standard measurements derived from continuous electrocardiographic (cECG) monitoring. The accuracy of heart rates measured by pulse oximetry (Spo2.R) was also measured as a positive control. RESULTS: On a per-patient basis, PFT-derived heart rate values were slightly lower than those derived from cECG monitoring (average bias of -1.14 beats per minute [bpm], with limits of agreement of 24 bpm). By comparison, Spo2.R recordings produced more accurate values (average bias of +0.15 bpm, limits of agreement of 13 bpm, P<.001 as compared with PFT). Personal fitness tracker device performance was significantly better in patients in sinus rhythm than in those who were not (average bias -0.99 bpm vs -5.02 bpm, P=.02). CONCLUSIONS: Personal fitness tracker-derived heart rates were slightly lower than those derived from cECG monitoring in real-world testing and not as accurate as Spo2.R-derived heart rates. Performance was worse among patients who were not in sinus rhythm. Further clinical evaluation is indicated to see if PFTs can augment early warning systems in hospitals. TRIAL REGISTRATION: ClinicalTrials.gov NCT02527408; https://clinicaltrials.gov/ct2/show/NCT02527408 (Archived by WebCite at http://www.webcitation.org/6kOFez3on).

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.2196/jmir.6025

Cites

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Some studies have used the gold-standard reference method of electrocardiogram, which directly measures heart rate via electrical impulses from the heart and is used for clinical accuracy.

    “Although some studies have attempted to address wearable HR accuracy using the gold-standard reference method of ECG, which directly measures HR via electrical impulses from the heart and is used for clinical accuracy38–44”
  • On average, wearables slightly and negligibly underestimate absolute heart rate (−0.9 to −7.2 bpm) compared with gold-standard electrocardiogram.

    “Current research verifying wearables to the gold-standard ECG, which uses electrodes to directly measure cardiac muscular contractions from electrical activity of the heart, have shown that on average wearables slightly and negligibly underestimate (−0.9 to −7.2 bpm) absolute HR40,41,44,52”

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

The citation lists haven’t been fetched yet

References 29 · Cited by 123

Authors
Kroll, R. R., Boyd, J. G., & Maslove, D. M.
Affiliations
Queen's University (CA)
Corresponding author
Maslove, David M
Venue
Journal of Medical Internet Research
Published
2016-09-20
Volume
18
Issue
9
Pages
e253
ISSN
1438-8871 (ISSN-L)
OpenAlex
W2479978653
Cited by
123
References
29
Citations by year
2026: 7 · 2025: 8 · 2024: 8 · 2023: 10 · 2022: 19 · 2021: 8 · 2020: 16 · 2019: 16 · 2018: 19 · 2017: 12
FWCI
7.55
Type
journal article
Language
en
Open access
Licence
cc-by
Locations
  • Journal of Medical Internet Research (publishedVersion, cc-by)
  • Europe PMC (PubMed Central) (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
  • Health Sciences › Medicine › Surgery › Healthcare Technology and Patient Monitoring
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
Keywords
wearable devices, heart rate monitoring, wrist-worn sensors, accuracy, hospitalized patients, intensive care unit, electrocardiogram, pulse oximetry, sinus rhythm, early warning systems, clinical deterioration
MeSH terms
Biometry, Female, Heart Rate, Humans, Inpatients, Male, Middle Aged, Monitoring, Physiologic, Prospective Studies, Wrist, Cohort Studies

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review