Guidelines for wrist-worn consumer wearable assessment of heart rate in biobehavioral research
Open accessAbstract
Researchers have increasingly begun to use consumer wearables or wrist-worn smartwatches and fitness monitors for measurement of cardiovascular psychophysiological processes related to mental and physical health outcomes. These devices have strong appeal because they allow for continuous, scalable, unobtrusive, and ecologically valid data collection of cardiac activity in “big data” studies. However, replicability and reproducibility may be hampered moving forward due to the lack of standardization of data collection and processing procedures, and inconsistent reporting of technological factors (e.g., device type, firmware versions, and sampling rate), biobehavioral variables (e.g., body mass index, wrist dominance and circumference), and participant demographic characteristics, such as skin tone, that may influence heart rate measurement. These limitations introduce unnecessary noise into measurement, which can cloud interpretation and generalizability of findings. This paper provides a brief overview of research using commercial wearable devices to measure heart rate, reviews literature on device accuracy, and outlines the challenges that non-standardized reporting pose for the field. We also discuss study design, technological, biobehavioral, and demographic factors that can impact the accuracy of the passive sensing of heart rate measurements, and provide guidelines and corresponding checklist handouts for future study data collection and design, data cleaning and processing, analysis, and reporting that may help ameliorate some of these barriers and inconsistencies in the literature.
AI snapshot
Written from the paper’s full text.
TL;DR
This paper reviews the use of wrist-worn consumer wearables for heart rate measurement in biobehavioral research and argues that lack of standardized data collection, processing, and reporting threatens reproducibility and generalizability. It provides two reporting checklists covering study design, technological factors, participant characteristics, data cleaning, analysis, and transparency.
Key findings
- Wearables using photoplethysmography (PPG) can provide acceptable heart rate accuracy for research, especially with aggregated data and outlier removal, but not for medical settings; devices slightly underestimate absolute HR versus ECG by −0.9 to −7.2 bpm, with Apple Watch at −1.3 bpm and Fitbit at −9.3 bpm.
- Error ranges were 1.20–6.70% for Apple Watch models and 2.38–21.36% for Fitbit models; any single real-time HR observation can deviate substantially from ECG, even when summary statistics are accurate.
- Between 2008 and 2018, there was a 580% increase in articles containing “wearable” and “heart rate.”
- Accuracy is influenced by motion artifacts and activity, with absolute error during activity reported as 30% higher than at rest; darker skin tone and tattoos can reduce green-light PPG accuracy, and larger wrist circumference or higher BMI are associated with reduced accuracy in some but not all studies.
Methods
- Narrative review and guideline paper; no primary data collection, systematic review, or meta-analysis is described.
- Synthesizes literature on commercial wrist-worn wearable HR accuracy and factors affecting measurement, focusing mainly on Apple Watch and Fitbit while noting Garmin, Samsung, and Microsoft.
- Proposes two checklists (Tables 2 and 3) for descriptive reporting and potential covariates, and recommends open science practices such as preregistration, open code, and open data.
Limitations
- The paper states that the field lacks metadata standards and consistent reporting of device type, firmware, sampling rate, biobehavioral variables, and demographic characteristics, which introduces noise and limits reproducibility and generalizability.
- It states that accuracy evidence is mixed for activity intensity, skin tone, BMI, and wrist circumference, and that proprietary algorithms and lack of raw data prevent standardized processing.
- Evident from the text: this is a narrative review without a systematic search or meta-analysis, and the proposed checklists are not empirically validated in the paper.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 99 cited sources checked · 0 supported · 0 flagged
- Interpretationreasoning
These limitations introduce unnecessary noise into measurement, which can cloud interpretation and generalizability of findings.
“These limitations introduce unnecessary noise into measurement, which can cloud interpretation and generalizability of findings.”
Quote in the paperNot a substantial claim - Interpretationcitation
Cardiac activity may be a biological factor linking mood and anxiety disorders with cardiovascular and metabolic risk and mortality.
“Cardiac activity therefore may be a biological factor that serves as a mechanism linking mood and anxiety disorders with cardiovascular and metabolic risk and mortality14,15.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [14] Kiecolt-Glaser, J. K. & Wilson, S. J. Psychiatric disorders, morbidity, and mortality: tracing mechanistic pathways to accelerated aging. Psychosom. Med. 78, 772–775 (2016). source
Not checked in the free run.
- cites [15] Nemeroff, C. B. & Goldschmidt-Clermont, P. J. Heartache and heartbreak–the link between depression and cardiovascular disease. Nat. Rev. Cardiol. 9, 526–539 (2012). source
Not checked in the free run.
- Premisecitation
The law of initial values holds that an initial value has a large impact on the strength and direction of a response, such that a higher pre-stimulus level leads to a smaller post-stimulus response.
“The law of initial values gives further insight into this issue by highlighting that an initial value has a large impact on the strength and direction of a response, such that a higher pre-stimulus level will lead to a smaller post-stimulus response or, in other words, a smaller response when there is a greater initial value19–21.”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [19] Geenen, R. & Vijver, F. J. R. A simple test of the law of initial values. Psychophysiology 30, 525–530 (1993). source
Not checked in the free run.
- cites [20] Pattyn, N., Migeotte, P.-F., Neyt, X., den Nest, A. & van, Cluydts, R. Comparing real-life and laboratory-induced stress reactivity on cardio-respiratory parameters: differentiation of a tonic and a phasic component. Physiol. Behav. 101, 218–223 (2010). source
Not checked in the free run.
- cites [21] Wilder, J. Basimetric approach (law of initial value) to biological rhythms. Ann. N. Y. Acad. Sci. 98, 1211–1220 (1962). source
Not checked in the free run.
- Interpretationreasoning
Photoplethysmography devices allow for in vivo examinations of the roles that resting heart rate and heart rate dynamics have in medical and psychiatric disorders in real-world settings.
“Overall, PPG devices allow for in vivo examinations of the roles that resting HR and HR dynamics (e.g., reactivity and recovery) have in medical and psychiatric disorders in real-world settings.”
Quote in the paperSubstantial claim - Premisecitation
Current research into heart rate accuracy of wrist-worn wearable devices compares wearable photoplethysmography measurement with a reference method such as electrocardiogram to determine whether measurements between devices are outside clinically important limits of agreement.
“Current research into the HR accuracy of wrist-worn wearable devices attempt to compare the measurement between wearable PPG and a reference method, such as an ECG, in order to determine whether or not measurements between the devices are outside of clinically important limits of agreement, so that devices can supplement, eventually replace, or be used interchangeably37,38.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [37] Bland, J. M. & Altman, D. G. Measuring agreement in method comparison studies. Stat. Meth. Med. Res 8, 135–160 (1999). source
Not checked in the free run.
- cites [38] Nelson, B. W. & Allen, N. B. Accuracy of consumer wearable heart rate measurement during an ecologically valid 24-hour period: intraindividual validation study. JMIR mHealth uHealth 7, e10828 (2019). source
Not checked in the free run.
- Premisecitation
Other studies have used suboptimal reference methods, such as chest straps or pulse oximeters, which themselves have a degree of error varying between 0 and 40 bpm when compared to electrocardiograms.
“other studies have used suboptimal reference methods, such as the chest straps45–47 or pulse oximeters48, which themselves have a degree of error (varying between 0 and 40 bpm when compared to ECGs44).”
Quote in the paperSubstantial claim0 of 5 sources checked- cites [44] Wang, R. et al. Accuracy of wrist-worn heart rate monitors. JAMA Cardiol. 2, 104–106 (2017). source
Not checked in the free run.
- cites [45] Dooley, E. E., Golaszewski, N. M. & Bartholomew, J. B. Estimating accuracy at exercise intensities: a comparative study of self-monitoring heart rate and physical activity wearable devices. JMIR mHealth uHealth 5, e34 (2017). source
Not checked in the free run.
- cites [46] Gillinov, S. et al. Variable accuracy of wearable heart rate monitors during aerobic exercise. Med. Sci. Sports Exerc. 49, 1697–1703 (2017). source
Not checked in the free run.
- cites [47] Stahl, S. E., An, H.-S., Dinkel, D. M., Noble, J. M. & Lee, J.-M. How accurate are the wrist-based heart rate monitors during walking and running activities? Are they accurate enough? BMJ Open Sport Exerc. Med. 2, e000106 (2016). source
Not checked in the free run.
- cites [48] El-Amrawy, F. & Nounou, M. I. Are currently available wearable devices for activity tracking and heart rate monitoring accurate, precise, and medically beneficial? Healthc. Inform. Res. 21, 315–320 (2015). source
Not checked in the free run.
- Interpretationreasoning
Researchers can utilize wearable heart rate to accurately assess heart rate, especially resting heart rate, when using aggregated data across activities and removing outliers.
“These preliminary findings suggest that researchers can utilize wearable HR to accurately assess HR, especially resting HR, when using aggregated data across activities and removing outliers.”
Quote in the paperSubstantial claim - Interpretationreasoning
Suboptimal referencing methods can introduce unnecessary measurement error and potentially undermine findings in wrist-worn wearable heart rate accuracy research.
“This has the potential to introduce a degree of additional error into wrist-worn wearable HR accuracy research as suboptimal referencing methods can introduce unnecessary measurement error and potentially undermine findings.”
Quote in the paperSubstantial claim - Premisecitation
On average, wearables slightly and negligibly underestimate absolute heart rate (−0.9 to −7.2 bpm) compared with gold-standard electrocardiogram.
“Current research verifying wearables to the gold-standard ECG, which uses electrodes to directly measure cardiac muscular contractions from electrical activity of the heart, have shown that on average wearables slightly and negligibly underestimate (−0.9 to −7.2 bpm) absolute HR40,41,44,52”
Quote in the paperSubstantial claim0 of 4 sources checked- cites [40] de Zambotti, M. et al. Measures of sleep and cardiac functioning during sleep using a multi-sensory commercially-available wristband in adolescents. Physiol. Behav. 158, 143–149 (2016). source
Not checked in the free run.
- cites [41] Kroll, R. R., Boyd, J. G. & Maslove, D. M. Accuracy of a wrist-worn wearable device for monitoring heart rates in hospital inpatients: a prospective observational study. J. Med. Internet Res. 18, e253 (2016). source
Not checked in the free run.
- cites [44] Wang, R. et al. Accuracy of wrist-worn heart rate monitors. JAMA Cardiol. 2, 104–106 (2017). source
Not checked in the free run.
- cites [52] Benedetto, S. et al. Assessment of the fitbit charge 2 for monitoring heart rate. PLoS ONE. https://doi.org/10.1371/journal.pone.0192691 (2018). source
Not checked in the free run.
- Premisecitation
The Apple Watch has slightly greater accuracy than Fitbit devices (Apple Watch: −1.3 bpm, Fitbit: −9.3 bpm).
“with the Apple Watch having slightly greater accuracy than Fitbit devices (Apple Watch: −1.3 bpm, Fitbit: −9.3 bpm42–44)”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [42] Shcherbina, A. et al. Accuracy in wrist-worn, sensor-based measurements of heart rate and energy expenditure in a diverse cohort. J. Personalized Med. 7, 3 (2017). source
Not checked in the free run.
- cites [43] Wallen, M. P., Gomersall, S. R., Keating, S. E., Wisløff, U., & Coombes, J. S. Accuracy of heart rate watches: implications for weight management. PLoS ONE 11, https://doi.org/10.1371/journal.pone.0154420 (2016). source
Not checked in the free run.
- cites [44] Wang, R. et al. Accuracy of wrist-worn heart rate monitors. JAMA Cardiol. 2, 104–106 (2017). source
Not checked in the free run.
- Premisecitation
In that study, any individual heart rate observation could deviate from electrocardiogram by significantly large margins that would likely be problematic in medical settings.
“any individual HR observation could deviate from ECG by significantly large margins that would be likely be problematic in medical settings38.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [38] Nelson, B. W. & Allen, N. B. Accuracy of consumer wearable heart rate measurement during an ecologically valid 24-hour period: intraindividual validation study. JMIR mHealth uHealth 7, e10828 (2019). source
Not checked in the free run.
- Premisecitation
Apple Watch models have lower overall error (1.20 to 6.70%) than Fitbit models (2.38 to 21.36%).
“as well as lower overall error with the Apple Watch models ranging from 1.20 to 6.70% error and Fitbit models ranging from 2.38 to 21.36% error39,42,45”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [39] Boudreaux, B. D. et al. Validity of wearable activity monitors during cycling and resistance exercise. Med. Sci. Sports Exerc. 50, 624–633 (2018). source
Not checked in the free run.
- cites [42] Shcherbina, A. et al. Accuracy in wrist-worn, sensor-based measurements of heart rate and energy expenditure in a diverse cohort. J. Personalized Med. 7, 3 (2017). source
Not checked in the free run.
- cites [45] Dooley, E. E., Golaszewski, N. M. & Bartholomew, J. B. Estimating accuracy at exercise intensities: a comparative study of self-monitoring heart rate and physical activity wearable devices. JMIR mHealth uHealth 5, e34 (2017). source
Not checked in the free run.
- Premisecitation
Greater wearable movement is associated with decreased photoplethysmography accuracy.
“which may be due to the fact that greater wearable movement is associated with decreased PPG accuracy as discussed below62.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [62] Menghini, L. et al. Stressing the accuracy: wrist-worn wearable sensor validation over different conditions. Psychophysiology. 56, https://doi.org/10.1111/psyp.13441 (2019). source
Not checked in the free run.
- Interpretationcitation
mHealth research will be particularly susceptible to research practices that degrade quality, consistency, generalizability, and therefore replicability and reproducibility of findings, as data sensors allow vast collection of big data samples vulnerable to publication bias, P-hacking, HARKing, and low power.
“Lastly, from a study design perspective, mHealth research will be particularly susceptible to a number of research practices, which have been shown to degrade the quality, consistency, generalizability, and therefore replicability and reproducibility of findings31 as data sensors in this research field allow for the vast collection of big data samples, which will be vulnerable to some of the “four”
Quote in the paperSubstantial claim0 of 4 sources checked- cites [31] Stupple, A., Singerman, D., & Celi, L. A. The reproducibility crisis in the age of digital medicine. npjDigital Med. 2, https://doi.org/10.1038/s41746-019-0079-z (2019). source
Not checked in the free run.
- cites [54] Munafò, M. R. et al. A manifesto for reproducible science. Nat. Hum. Behav. 1, https://doi.org/10.1038/s41562-016-0021 (2017). source
Not checked in the free run.
- cites [63] Franco, A., Malhotra, N. & Simonovits, G. Publication bias in the social sciences: unlocking the file drawer. Science 345, 1502–1505 (2014). source
Not checked in the free run.
- cites [64] Button, K. S. et al. Confidence and precision increase with high statistical power. Nat. Rev. Neurosci. 14, 585–585 (2013). source
Not checked in the free run.
- Premisecitation
Simulation studies with intensive longitudinal data suggest that multiple imputation and full-information maximum likelihood reliably produce better parameter estimates than listwise deletion.
“More specifically, simulation studies with intensive longitudinal data suggest that multiple imputation and full-information maximum likelihood reliably produce better parameter estimates than listwise deletion59.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [59] Ji, L., Chow, S.-M., Schermerhorn, A. C., Jacobson, N. C. & Cummings, E. M. Handling missing data in the modeling of intensive longitudinal. Data. Struct. Equ. Model. Multidiscip. J. 25, 715–736 (2018). source
Not checked in the free run.
- Premisecitation
There are a number of study design features related to the wrist that have been shown to have effects on heart rate accuracy.
“There are a number of study design features related to the wrist that have been shown to have effects on HR accuracy.”
Quote in the paperSubstantial claim - Premisecitation
Other data cleaning strategies, such as spectral frequency filters, depend upon the type of sensor being utilized in the wearable sensor.
“Other strategies (e.g. spectral frequency filters), will depend upon the type of the sensor being utilized in the wearable sensor57.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [57] Yang, Z., Zhou, Q., Lei, L., Zheng, K., & Xiang, W. An IoT-cloud based wearable ECG monitoring system for smart healthcare. J. Med. Syst. 40, https://doi.org/10.1007/s10916-016-0644-9 (2016). source
Not checked in the free run.
- Premisecitation
Wrist placement, tightness of device, dominant vs non-dominant hand use, and degree of wrist movement have consistently been shown to influence the accuracy of heart rate measurement.
“In addition to wrist circumference, there are a number of other factors including wrist placement, tightness of device, dominant vs non-dominant hand use, and degree of wrist movement that have consistently been shown to influence the accuracy of HR measurement39”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [39] Boudreaux, B. D. et al. Validity of wearable activity monitors during cycling and resistance exercise. Med. Sci. Sports Exerc. 50, 624–633 (2018). source
Not checked in the free run.
- Interpretationreasoning
These wrist factors are likely some of the most important factors influencing heart rate accuracy.
“These are likely some of the most important factors influencing HR accuracy.”
Quote in the paperSubstantial claim - Premisecitation
Dynamic models of heart rate captured in daily life can be effectively estimated even with very large proportions of missing data (e.g. >70% missingness for each subject) using full-information maximum likelihood estimation.
“Moreover, dynamic models of HR captured in daily life can be effectively estimated even with very large proportions of missing data (e.g. >70% missingness for each subject) using full-information maximum likelihood estimation60.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [60] Jacobson, N. C., Chow, S.-M. & Newman, M. G. The differential time-varying effect model (DTVEM): a tool for diagnosing and modeling time lags in intensive longitudinal data. Behav. Res. Methods 51, 295–315 (2019). source
Not checked in the free run.
- Premisecitation
Research has found higher device error in males compared to females.
“In addition, research has not always reported participant biological sex, but research has found higher device error in males compared to females42, indicating that participant biological sex should be collected and possibly controlled.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [42] Shcherbina, A. et al. Accuracy in wrist-worn, sensor-based measurements of heart rate and energy expenditure in a diverse cohort. J. Personalized Med. 7, 3 (2017). source
Not checked in the free run.
- Premisecitation
Motion artifacts and level of physical activity are two related factors identified as influencing wearable heart rate accuracy.
“two related factors identified as influencing wearable HR accuracy are motion artifacts23 and level of physical activity46.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [23] Mahloko, L. & Adebesin, F. in Responsible Design, Implementation and Use of Information and Communication Technology. Lecture Notes in Computer Science (eds Hattingh, M., et al.) Vol. 12067, 96–107 (Springer International Publishing, Cham, 2020). source
Not checked in the free run.
- cites [46] Gillinov, S. et al. Variable accuracy of wearable heart rate monitors during aerobic exercise. Med. Sci. Sports Exerc. 49, 1697–1703 (2017). source
Not checked in the free run.
- Premisecitation
At rest, wearables can perform similarly to an ECG.
“research has shown that at rest, wearables can perform similarly to an ECG”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [50] Bent, B., Goldstein, B. A., Kibbe, W. A., & Dunn, J. P. Investigating sources of inaccuracy in wearable optical heart rate sensors. npj Digital Med. 3, https://doi.org/10.1038/s41746-020-0226-6 (2020). source
Not checked in the free run.
- Interpretationnone given
Individual differences and variability in biobehavioral and demographic factors have the potential to influence the collection of heart rate during research studies.
“Individual differences and variability in both biobehavioral and demographic factors have the potential to influence the collection of HR during research studies and should therefore be strongly considered in descriptive reporting as well as statistical modeling in studies using HR measurement collected from wearable devices.”
Quote in the paperSubstantial claim - Interpretationreasoning
The current literature indicates that consumer wearable devices can provide acceptable accuracy for heart rate measurement in research settings, especially when scalability and ecological validity are critical considerations that may outweigh the need for gold-standard assessment.
“The current literature indicates that these devices can provide acceptable accuracy for the measurement of HR for research settings, especially when scalability and ecological validity of measurement settings are critical considerations that may outweigh the need to offer gold-standard assessment.”
Quote in the paperSubstantial claim - Interpretationreasoning
Not controlling for firmware version may lead to poorer reproducibility and replicability, as two studies investigating the same device with different firmware versions might come to different conclusions even if all other variables are held constant.
“Not controlling for firmware version may lead to poorer reproducibility and replicability as two studies investigating the same device with two or more different firmware versions might actually come to different conclusions even if all other variables are held constant.”
Quote in the paperSubstantial claim- Rests on an assumption: Wearable algorithms may be altered with firmware updates.
- Rests on an assumption: Most studies fail to report firmware information.
- Premisecitation
Wearable manufacturers use proprietary algorithms to translate PPG signals to heart rate measurements, which may influence accuracy.
“Wearable manufacturers use proprietary algorithms to translate PPG signals to HR measurements, which may influence accuracy23.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [23] Mahloko, L. & Adebesin, F. in Responsible Design, Implementation and Use of Information and Communication Technology. Lecture Notes in Computer Science (eds Hattingh, M., et al.) Vol. 12067, 96–107 (Springer International Publishing, Cham, 2020). source
Not checked in the free run.
- Interpretationnone given
Consumer wearable devices that collect heart rate likely are not yet advanced enough to replace medical-grade ECG anytime soon.
“although consumer wearable devices that collect HR likely are not yet advanced enough to be replacing medical-grade ECG anytime soon”
Quote in the paperNot a substantial claim - Premisecitation
A third study found that measurement during walking, running, and cycling was more accurate for some devices than during sitting.
“a third found that measurement during walking, running, and cycling was more accurate for some devices than during sitting42”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [42] Shcherbina, A. et al. Accuracy in wrist-worn, sensor-based measurements of heart rate and energy expenditure in a diverse cohort. J. Personalized Med. 7, 3 (2017). source
Not checked in the free run.
- Interpretationreasoning
Wearables that use green PPG or a combination of green and red PPG are likely to perform better during movement and physical activity.
“indicating that wearables that use green PPG or a combination of green and red PPG are likely to perform better during movement and physical activity23.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [23] Mahloko, L. & Adebesin, F. in Responsible Design, Implementation and Use of Information and Communication Technology. Lecture Notes in Computer Science (eds Hattingh, M., et al.) Vol. 12067, 96–107 (Springer International Publishing, Cham, 2020). source
Not checked in the free run.
- Premisecitation
Wearable devices are more accurate during rest and low intensity exercise compared to higher intensity exercise, when they can begin to deviate more from ECG recordings.
“a substantial amount of research showing that wearable devices are more accurate during rest and low intensity exercise when compared to higher intensity exercise, when wearable devices can begin to deviate more from ECG recordings38,39,44,46,70–73”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [38] Nelson, B. W. & Allen, N. B. Accuracy of consumer wearable heart rate measurement during an ecologically valid 24-hour period: intraindividual validation study. JMIR mHealth uHealth 7, e10828 (2019). source
Not checked in the free run.
- cites [39] Boudreaux, B. D. et al. Validity of wearable activity monitors during cycling and resistance exercise. Med. Sci. Sports Exerc. 50, 624–633 (2018). source
Not checked in the free run.
- cites [44] Wang, R. et al. Accuracy of wrist-worn heart rate monitors. JAMA Cardiol. 2, 104–106 (2017). source
Not checked in the free run.
- Premisecitation
Research has described a cross-over effect when repetitive motion and activity causes underlying algorithms to mistake the movement cadence for cardiac activity.
“research has described the “cross-over effect” when repetitive motion and activity causes underlying algorithms to mistake the movement cadence for cardiac activity50,75.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [50] Bent, B., Goldstein, B. A., Kibbe, W. A., & Dunn, J. P. Investigating sources of inaccuracy in wearable optical heart rate sensors. npj Digital Med. 3, https://doi.org/10.1038/s41746-020-0226-6 (2020). source
Not checked in the free run.
- cites [75] Valencell, T. Optical heart rate monitoring: what you need to know. Valencell. https://valencell.com/blog/2015/10/optical-heart-rate-monitoring-what-you-need-to-know/. Accessed 13 April, 2020 (2015).
Not checked in the free run.
- Interpretationreasoning
Various technological factors have the potential to influence the collection of heart rate during research studies.
“various technological factors have the potential to influence the collection of HR during research studies and should therefore be strongly considered in descriptive as well as statistical modeling when using wearable devices.”
Quote in the paperSubstantial claim- Rests on an assumption: Various wearable devices and versions on the market may use different hardware, which may introduce differences in heart rate recordings.
- Classified as probably not a substantial claim. 33
- May combine two claims. 38
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Findingnone givenBetween 2008 and 2018 there was a 580% increase in total articles published that contained "wearable" and "heart rate".
“Between 2008 and 2018 there was a 580% increase in total articles published that contained “wearable” and “heart rate”, indicating that this is a burgeoning area of research that may benefit from standardization of procedures to increase generalizability and clinical utility of research findings.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based).
Premisenone givenResearchers have increasingly used consumer wearables or wrist-worn smartwatches and fitness monitors for measurement of cardiovascular psychophysiological processes related to mental and physical health outcomes.
“Researchers have increasingly begun to use consumer wearables or wrist-worn smartwatches and fitness monitors for measurement of cardiovascular psychophysiological processes related to mental and physical health outcomes.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based).
Premisenone givenConsumer wearables allow continuous, scalable, unobtrusive, and ecologically valid data collection of cardiac activity in big data studies.
“These devices have strong appeal because they allow for continuous, scalable, unobtrusive, and ecologically valid data collection of cardiac activity in “big data” studies.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationHeart rate is predominantly influenced by coordination of the sympathetic and parasympathetic branches of the autonomic nervous system, which can be modified by biopsychosocial factors including physical and relational stress, diet, physical fitness, medication use, and substance use.
“HR is predominantly influenced by the coordination of the sympathetic and parasympathetic branches of the autonomic nervous system, which can be modified by biopsychosocial factors including physical and relational stress, diet, physical fitness, medication use, and substance use1; thus, offering a direct and quantifiable connection to the stress diathesis model of health.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [1] Cacioppo J. T., Tassinary L. G., & Berntson G. G. (eds) Handbook of Psychophysiology 4th edn. (Cambridge University Press, 2017). source
Not checked in the free run.
May combine two claims.
PremisecitationCollection of heart rate data in laboratory or medical environments lacks temporal resolution within and across days as heart rate is usually collected during a short recording within a single day, which has precluded research from comprehensively investigating the transition from acute to chronic stress.
“Second, collection of HR data in these environments lacks temporal resolution within and across days as HR is usually collected during a short recording within a single day, which has precluded research from comprehensively investigating the transition from acute to chronic stress22.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [22] Rohleder, N. Stress and inflammation—the need to address the gap in the transition between acute and chronic stress effects. Psychoneuroendocrinology https://doi.org/10.1016/j.psyneuen.2019.02.021 (2019). source
Not checked in the free run.
Probably not a substantial claim: may not pass S4 (based).
Premisenone givenHeart rate research has been constrained by inability to collect temporally detailed and longitudinal heart rate measurements both within and across time.
“(2) inability to collect temporally detailed and longitudinal HR measurements both within and across time (e.g., hours, days, weeks, months)”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationResearchers are increasingly using consumer wearables or wrist-worn smartwatches and fitness monitors for digital measurement of cardiovascular health and psychophysiological stress because these devices allow continuous, scalable, unobtrusive big data collection of cardiac activity in real-world conditions with large samples.
“Researchers are increasingly using consumer wearables or wrist-worn smartwatches and fitness monitors for the digital measurement of cardiovascular health and psychophysiological stress as these devices allow for continuous, scalable, and unobtrusive, “big data” collection of overall cardiac activity in real-world conditions with large samples25–30.”
Quote in the paperSubstantial claim0 of 6 sources checked- cites [25] Bradshaw, B. et al. Influenza surveillance using wearable mobile health devices. In International Society for Disease Surveillance Conference 2019 (2019).
Not checked in the free run.
- cites [26] Natarajan, A., Pantelopoulos, A., Emir-Farinas, H. & Natarajan, P. Heart rate variability with photoplethysmography in 8 million individuals: results and scaling relations with age, gender, and time of day. Preprint at https://www.biorxiv.org/content/10.1101/772285v1 (2019). source
Not checked in the free run.
- cites [27] Perez, M. V. et al. Large-scale assessment of a smartwatch to identify atrial fibrillation. N. Engl. J. Med. 381, 1909–1917 (2019). source
Not checked in the free run.
- cites [28] Sim, I. Mobile devices and health. N. Engl. J. Med. 381, 956–968 (2019). source
Not checked in the free run.
- cites [29] Tison, G. H. et al. Passive detection of atrial fibrillation using a commercially available smartwatch. JAMA Cardiol. 3, 409 (2018). source
Not checked in the free run.
- cites [30] Tison, G. H. et al. Cardiovascular Risk Stratification Using Off-the-Shelf Wearables and a Multi-Task Deep Learning Algorithm. In:; 2018. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Premisenone givenSome recent wrist-worn wearable devices, such as the Apple Watch Series 4 and 5, have allowed for collection of heart rate via electrical impulses on par with a single-lead electrode.
“The latter of these two is the predominant method for data collected with wrist-worn wearable devices, although some more recent wrist-worn wearable devices, such as the Apple Watch Series 4 and 5, have allowed for the collection of HR via electrical impulses on par with a single-lead electrode.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationreasoningResearch has historically been limited to collecting data from medical patients and research participants in novel settings that might not generalize to everyday lives, creating potential for inaccurate baselines and inaccurate measurement of heart rate dynamics.
“In other words, research has been historically limited to collecting data from medical patients and research participants in novel settings that might not generalize to their everyday lives, which creates the potential for inaccurate baselines and therefore inaccurate measurement of HR dynamics (i.e., reactivity/recovery).”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S4 (based).
Premisenone givenHeart rate research has been constrained by prohibitive financial cost of devices for everyday use.
“(3) prohibitive financial cost of devices for everyday use.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationPhotoplethysmography can be used to index heart rate variability, blood pressure, oxygen saturation, and cardiac output, but as of yet has not been widely rolled out in consumer-use devices.
“Furthermore, PPG can be used to index heart rate variability (HRV), blood pressure, oxygen saturation, and cardiac output24, but as of yet have not be widely rolled out in consumer-use devices and therefore data collection other than HR will not be discussed further.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [24] Elgendi, M. On the analysis of fingertip photoplethysmogram signals. Curr. Cardiol. Rev. 8, 14–25 (2012). source
Not checked in the free run.
Probably not a substantial claim: may not pass S4 (based). May combine two claims.
Premisenone givenUntil recently, quantification and measurement of heart rate has been relegated to medical and research settings, largely precluding observation of heart rate during daily life and limiting generalizability of findings to real-world living conditions.
“Until recently, the quantification and measurement of HR has been relegated to medical and research settings, which has largely precluded the observation of HR during patient and participant daily life, thereby limiting the generalizability of findings to real-world living conditions.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationRed LEDs are more commonly used in medical settings, are not absorbed as well by the skin allowing light to transmit deeper and detect multiple biological measures, but have a lower signal-to-noise ratio and are more susceptible to motion artifacts.
“In contrast, red LEDs, which are more commonly used in medical settings, are not absorbed as well by the skin allowing the light to transmit deeper and allowing for detection of multiple biological measures, but it has a lower signal-to-noise ratio and is more susceptible to motion artifacts23.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [23] Mahloko, L. & Adebesin, F. in Responsible Design, Implementation and Use of Information and Communication Technology. Lecture Notes in Computer Science (eds Hattingh, M., et al.) Vol. 12067, 96–107 (Springer International Publishing, Cham, 2020). source
Not checked in the free run.
May combine two claims.
Premisenone givenGreen LED light is a shorter wavelength and has good signal-to-noise ratio and resistance to motion artifacts, but is limited by the amount of light that can pass through tissue, especially darker skin tones.
“For example, green LED light is a shorter wave-length and has been shown to be convenient, because it has good signal-to-noise ratio and is resistant to motion artifacts, but is limited by the amount of light that can pass through tissue, especially darker skin tones.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationreasoningThis novel electrocardiogram technology only allows for discrete, rather than continuous, collection of data, which precludes temporally detailed longitudinal ECG collection at this time.
“Therefore, this novel electrocardiogram (ECG) technology only allows for the discrete, rather than continuous, collection of data, which precludes the ability for temporally detailed longitudinal ECG collection at this time.”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationreasoningHeart rate may be a transdiagnostic biomarker with clinical utility for psychology, psychiatry, and medicine to improve nosology and increase identification of those at risk for onset of psychiatric and physical health disorders.
“In other words, HR may be a transdiagnostic biomarker with clinical utility for psychology, psychiatry, and medicine to improve nosology and increase identification of those at risk for the onset of psychiatric and physical health disorders.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S4 (based).
Premisenone givenHeart rate research has been constrained by lack of ecological validity.
“The foundation of HR research has, until recently, been largely constrained due to (1) lack of ecological validity (i.e., the degree to which a laboratory or medical environment represents actual real-world conditions)”
Quote in the paperNot a substantial claimMay combine two claims.
PremisereasoningWrist-worn wearables provide an opportunity to examine heart rate in real-world environments, across longer periods of time, and with high resolution at low cost, by using photoplethysmography.
“Currently, wrist-worn wearables provide an opportunity to examine HR in real-world environments, across longer periods of time, and with high resolution at a low cost, by using an optical sensor (photoplethysmography; PPG), that allows these devices to collect volumetric changes in blood profusion (i.e., pulse rate) that can serve as a surrogate for HR”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
InterpretationcitationThe reproducibility and replicability crisis is important for digital health and mHealth research, including wrist-worn heart rate accuracy and studies utilizing wearable heart rate data, because this new field allows for large data samples with various sensors that increase the ease for p-hacking.
“Recent information on the reproducibility and replicability crisis has swept the sciences and as such these issues are very important for digital health and mHealth research, which includes wrist-worn HR accuracy and any studies that utilize wearable HR data, as this new field will allow for large data samples31 with various sensors that will increase the ease for p-hacking.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [31] Stupple, A., Singerman, D., & Celi, L. A. The reproducibility crisis in the age of digital medicine. npjDigital Med. 2, https://doi.org/10.1038/s41746-019-0079-z (2019). source
Not checked in the free run.
Classified as probably not a substantial claim.
InterpretationcitationQuestions remain as to the accuracy and validity of data collected from wearables.
“yet questions remain as to the accuracy and validity of data collected from wearables35,36.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [35] Evenson, K. R., Goto, M. M., & Furberg, R. D. Systematic review of the validity and reliability of consumer-wearable activity trackers. Int. J. Behav. Nutr. Phys. Activity. 12, https://doi.org/10.1186/s12966-015-0314-1 (2015). source
Not checked in the free run.
- cites [36] Peake, J. M., Kerr, G., & Sullivan, J. P. A critical review of consumer wearables, mobile applications, and equipment for providing biofeedback, monitoring stress, and sleep in physically active populations. Front. Physiol. 9, https://doi.org/10.3389/fphys.2018.00743 (2018). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationSome studies have used the gold-standard reference method of electrocardiogram, which directly measures heart rate via electrical impulses from the heart and is used for clinical accuracy.
“Although some studies have attempted to address wearable HR accuracy using the gold-standard reference method of ECG, which directly measures HR via electrical impulses from the heart and is used for clinical accuracy38–44”
Quote in the paperNot a substantial claim0 of 7 sources checked- cites [38] Nelson, B. W. & Allen, N. B. Accuracy of consumer wearable heart rate measurement during an ecologically valid 24-hour period: intraindividual validation study. JMIR mHealth uHealth 7, e10828 (2019). source
Not checked in the free run.
- cites [39] Boudreaux, B. D. et al. Validity of wearable activity monitors during cycling and resistance exercise. Med. Sci. Sports Exerc. 50, 624–633 (2018). source
Not checked in the free run.
- cites [40] de Zambotti, M. et al. Measures of sleep and cardiac functioning during sleep using a multi-sensory commercially-available wristband in adolescents. Physiol. Behav. 158, 143–149 (2016). source
Not checked in the free run.
- cites [41] Kroll, R. R., Boyd, J. G. & Maslove, D. M. Accuracy of a wrist-worn wearable device for monitoring heart rates in hospital inpatients: a prospective observational study. J. Med. Internet Res. 18, e253 (2016). source
Not checked in the free run.
- cites [42] Shcherbina, A. et al. Accuracy in wrist-worn, sensor-based measurements of heart rate and energy expenditure in a diverse cohort. J. Personalized Med. 7, 3 (2017). source
Not checked in the free run.
- cites [43] Wallen, M. P., Gomersall, S. R., Keating, S. E., Wisløff, U., & Coombes, J. S. Accuracy of heart rate watches: implications for weight management. PLoS ONE 11, https://doi.org/10.1371/journal.pone.0154420 (2016). source
Not checked in the free run.
- cites [44] Wang, R. et al. Accuracy of wrist-worn heart rate monitors. JAMA Cardiol. 2, 104–106 (2017). source
Not checked in the free run.
May combine two claims.
InterpretationcitationWhile overall summary statistics after outlier detection and data cleaning may be very accurate for research purposes, any single observation in real-time may have a large degree of error, which could be significant for moment-to-moment observations in medical settings.
“This finding indicated that while overall, summary statistics after outlier detection and data cleaning may be very accurate for research purposes, any single observation in real-time may have a large degree of error, which could be significant for moment-to-moment observations in medical settings.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationSome wearable devices, such as the Apple Watch 4 and 5, have gained FDA 510(k) class II clearance for the ECG feature and ability to detect arrhythmias.
“Furthermore, some wearable devices, such as the Apple Watch 4 and 5, have gained FDA 510(k) class II clearance for the ECG feature and ability to detect arrhythmias49”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [49] FDA. Electrocardiographsoftware for Over-the-Counter Use. https://www.accessdata.fda.gov/cdrh_docs/pdf18/DEN180044.pdf (FDA, 2018).
Not checked in the free run.
May combine two claims.
PremisecitationConsumer wearables such as the Apple Watch and Fitbit are more accurate than research-grade wearables, which are much more expensive yet provide the benefit of raw data.
“Furthermore, recent research has indicated that consumer wearables, such as the Apple Watch and Fitbit are more accurate than research-grade wearables, which are much more expensive yet provide the benefit of raw data50.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [50] Bent, B., Goldstein, B. A., Kibbe, W. A., & Dunn, J. P. Investigating sources of inaccuracy in wearable optical heart rate sensors. npj Digital Med. 3, https://doi.org/10.1038/s41746-020-0226-6 (2020). source
Not checked in the free run.
May combine two claims.
PremisecitationIn a study comparing ambulatory electrocardiogram, Apple Watch 3, and Fitbit Charge 2 within individuals across 24 h, devices slightly underestimated heart rate when values were aggregated within and across different daily activities, indicating high heart rate accuracy.
“For example, researchers compared the accuracy of ambulatory ECG, Apple Watch 3, and Fitbit Charge 2 data within an individual across 24 h and found that while devices slightly underestimated HR when values were aggregated within and across different daily activities, which indicated high HR accuracy”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [38] Nelson, B. W. & Allen, N. B. Accuracy of consumer wearable heart rate measurement during an ecologically valid 24-hour period: intraindividual validation study. JMIR mHealth uHealth 7, e10828 (2019). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationMany researchers have begun to push for increased focus on reproducible and open practices in science.
“Therefore, many researchers have begun to push for increased focus on reproducible and open practices in science31,53–55.”
Quote in the paperNot a substantial claim0 of 4 sources checked- cites [31] Stupple, A., Singerman, D., & Celi, L. A. The reproducibility crisis in the age of digital medicine. npjDigital Med. 2, https://doi.org/10.1038/s41746-019-0079-z (2019). source
Not checked in the free run.
- cites [53] Bavel, J. J. V., Mende-siedlecki, P., Brady, W. J. & Reinero, D. A. Contextual sensitivity in scientific reproducibility. Proc. Natl Acad. Sci. USA 113, 6454–6459 (2016). source
Not checked in the free run.
- cites [54] Munafò, M. R. et al. A manifesto for reproducible science. Nat. Hum. Behav. 1, https://doi.org/10.1038/s41562-016-0021 (2017). source
Not checked in the free run.
- cites [55] Tackett, J. L., Brandes, C. M., King, K. M. & Markon, K. E. Psychology’s replication crisis and clinical psychological science. Annu. Rev. Clin. Psychol. 15, 579–604 (2019). source
Not checked in the free run.
Probably not a substantial claim: may not pass S1 (asserted). May combine two claims.
InterpretationcitationWearables detect heart rate with acceptable accuracy in both laboratory and real-world settings for research, but importantly not medical settings.
“These findings have led some researchers to conclude that wearables detect HR with acceptable accuracy in both laboratory and real-world settings for research, but importantly not medical settings38,42 as more research needs to be done.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [38] Nelson, B. W. & Allen, N. B. Accuracy of consumer wearable heart rate measurement during an ecologically valid 24-hour period: intraindividual validation study. JMIR mHealth uHealth 7, e10828 (2019). source
Not checked in the free run.
- cites [42] Shcherbina, A. et al. Accuracy in wrist-worn, sensor-based measurements of heart rate and energy expenditure in a diverse cohort. J. Personalized Med. 7, 3 (2017). source
Not checked in the free run.
May combine two claims.
PremisecitationApple Watch has lower mean difference standard deviation and higher agreement with electrocardiogram than Fitbit devices.
“lower mean difference standard deviation43, and higher agreement with ECG than Fitbit devices39,44.”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [39] Boudreaux, B. D. et al. Validity of wearable activity monitors during cycling and resistance exercise. Med. Sci. Sports Exerc. 50, 624–633 (2018). source
Not checked in the free run.
- cites [43] Wallen, M. P., Gomersall, S. R., Keating, S. E., Wisløff, U., & Coombes, J. S. Accuracy of heart rate watches: implications for weight management. PLoS ONE 11, https://doi.org/10.1371/journal.pone.0154420 (2016). source
Not checked in the free run.
- cites [44] Wang, R. et al. Accuracy of wrist-worn heart rate monitors. JAMA Cardiol. 2, 104–106 (2017). source
Not checked in the free run.
May combine two claims.
PremisereasoningHardware differences between devices determine heart rate measurement second to second, with large implications for overall interpretability of the data, such that sampling rate introduces researcher degrees of freedom for calculating bpm.
“Hardware differences between devices determine HR measurement second to second, which has large implications on overall interpretability of the data, such that sampling rate can introduces researcher degrees of freedom for calculating bpm.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S2 (checkable).
InterpretationreasoningFor long-term field studies, information about participant adherence and wear-time should be considered and reported since these issues can affect data quality.
“For long-term field studies, information about participant adherence and wear-time should be considered and reported since these issues can affect data quality.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationOutliers in linear models may strongly impact linear estimators, and some studies may therefore adopt a modeling strategy wherein values are winsorized.
“For example, outliers in linear models may strongly impact linear estimators, and some studies may therefore adopt a modeling strategy wherein values are winsorized (e.g., extreme values can be truncated at the 2.5th and 97.5th quantiles56).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [56] Hastings, C. Jr., Mosteller, F., Tukey, J. W. & Winsor, C. P. Low moments for small samples: a comparative study of order statistics. Ann. Math. Stat. 18, 413–426 (1947). source
Not checked in the free run.
Probably not a substantial claim: may not pass S4 (based). May combine two claims.
Premisenone givenFitbit logs heart rate data at 1 s intervals during exercise tracking and 5 s intervals at other times, whereas the Apple Watch samples heart rate data every 10 min, except during workouts, which are continuous.
“For example, Fitbit logs HR data at 1 s intervals during exercise tracking and 5 s intervals at other times, whereas the Apple Watch samples HR data every 10 min, except during workouts, which are continuous.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable).
InterpretationreasoningData cleaning methods should be customized to the device being used and the models being adopted, rather than adopting a universal pre-processing pipeline, due to device-specific differences and study research questions.
“There is a strong need to customize the methods of data cleaning to the device being used and the models being adopted, rather than adopting a universal pre-processing pipeline due to device-specific differences and based on the study research questions.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable), S4 (based). May combine two claims.
Interpretationnone givenThere are many researcher degrees of freedom in decisions related to study protocol design and data collection, cleaning, and processing that are important to note as standardizing this process may help ameliorate some barriers and inconsistencies in the literature.
“There are many researcher degrees of freedom when it comes to decisions related to study protocol design and data collection, cleaning, and processing that are important to note as standardizing this process may help ameliorate some barriers and inconsistencies in the literature.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationListwise deletion is known to bias estimation and adversely impact the standard error of point estimates across modeling strategies.
“The practice of completely removing persons from studies due to proportions of missing data (i.e. listwise deletion) is known to bias estimation and adversely impact the standard error of point estimates across modeling strategies58.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [58] Graham, J. W. Missing data analysis: making it work in the real world. Annu. Rev. Psychol. 60, 549–576 (2009). source
Not checked in the free run.
Classified as probably not a substantial claim.
InterpretationcitationIf a study is interested in heart rate arrhythmias using wearable assessments, extreme values may be of vital interest and should not be removed or altered.
“However, if a study were interested in HR arrythmias using wearable assessments (such as atrial fibrillation), then these extreme values may be of vital interest and should not be removed or altered27.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [27] Perez, M. V. et al. Large-scale assessment of a smartwatch to identify atrial fibrillation. N. Engl. J. Med. 381, 1909–1917 (2019). source
Not checked in the free run.
Probably not a substantial claim: may not pass S4 (based).
Interpretationnone givenMethods researchers select to clean and process raw data prior to analyses can affect results and conclusions.
“In addition, the methods researchers select to clean and process raw data prior to analyses can also affect results and conclusions and should be reported in detail.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationLarger wrist circumference has been shown to be associated with reduced heart rate accuracy, although these outcomes have not been found in all studies.
“Larger wrist circumference has been shown to be associated with reduced HR accuracy42, although these outcomes have not been found in all studies44,46.”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [42] Shcherbina, A. et al. Accuracy in wrist-worn, sensor-based measurements of heart rate and energy expenditure in a diverse cohort. J. Personalized Med. 7, 3 (2017). source
Not checked in the free run.
- cites [44] Wang, R. et al. Accuracy of wrist-worn heart rate monitors. JAMA Cardiol. 2, 104–106 (2017). source
Not checked in the free run.
- cites [46] Gillinov, S. et al. Variable accuracy of wearable heart rate monitors during aerobic exercise. Med. Sci. Sports Exerc. 49, 1697–1703 (2017). source
Not checked in the free run.
Classified as probably not a substantial claim. May combine two claims.
InterpretationreasoningSome devices report missing heart rate data with zero values; these values should be removed and filled in with missing data rather than winsorized, which would otherwise substantially bias parameter estimates unless the degree of missing heart rate points was trivial.
“Moreover, for some, but not all devices will report missing HR data with zero values and these values should be removed and filled in with missing data rather than winsorized (which would otherwise substantially bias parameter estimates unless the degree of missing HR points was trivial).”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.
Premisenone givenOne issue reported by multiple users on the online Fitbit Community Forum is occasional logging of heart rate values (often higher and more variable than typical, e.g., 100-150 bpm) when the devices are not being worn, suggesting that supplemental non-wear-time diaries may be required for some devices or protocols.
“One issue reported by multiple users on the online Fitbit Community Forum is occasional logging of HR values (often higher and more variable than typical, e.g., 100–150 bpm) when the devices are not being worn, suggesting that supplemental non-wear-time diaries may be required for some devices or protocols.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisereasoningSome devices, such as the Apple Watch, can take multiple heart rate measurements within a minute, allowing researchers to use various calculations such as the mean or median within this minute to calculate bpm, which may alter wearable data values.
“Relatedly, some devices, such as the Apple Watch can take multiple HR measurements within a minute allowing researchers to use various calculations, such as the mean or median within this minute to calculate bpm, which may alter wearable data values.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationAbsolute error during activity is 30% higher than at rest.
“some research showing that absolute error during activity is 30% higher than rest50”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [50] Bent, B., Goldstein, B. A., Kibbe, W. A., & Dunn, J. P. Investigating sources of inaccuracy in wearable optical heart rate sensors. npj Digital Med. 3, https://doi.org/10.1038/s41746-020-0226-6 (2020). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.
InterpretationreasoningLow power likely will not be an issue for mHealth research as large sample sizes will be relatively easy to collect even in samples with low base rates in the population.
“the last of which likely will not be an issue for mHealth research as large samples sizes will be relatively easy to collect even in samples with low base rates in the population.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationreasoningFailure to report or control for participant race and ethnicity in wrist-worn wearable heart rate studies undermines the ability to examine potentially important group differences in health outcomes and decreases the ability of results to generalize to large segments of the population.
“Currently, the vast majority of wrist-worn wearable studies incorporating HR measurements do not report or control for participant race and ethnicity, which undermines the ability to look at potentially important group differences in health outcomes and therefore decreases the ability of results to generalize to large segments of the population.”
Quote in the paperSubstantial claim- Rests on an assumption: The vast majority of wrist-worn wearable studies incorporating heart rate measurements do not report or control for participant race and ethnicity.
May combine two claims.
PremisecitationDarker skin tone (and tattoos) has been found to be associated with reduced wrist-worn heart rate accuracy as they absorb more green light, yet this is not found in all studies.
“Evidence is starting to accumulate on the importance of skin tone when it comes to accuracy and generalizability of wrist-worn wearable studies with darker skin tone (and tattoos) being found to be associated with reduced wrist-worn HR accuracy as they absorb more green light67, yet this is not found in all studies43,50.”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [67] Weiler, D. T., Villajuan, S. O., Edkins, L., Cleary, S. & Saleem, J. J. Wearable heart rate monitor technology accuracy in research: a comparative study between PPG and ECG technology. Proc. Hum. Factors Ergonomics Soc. Annu. Meet. 61, 1292–1296 (2017). source
Not checked in the free run.
- cites [43] Wallen, M. P., Gomersall, S. R., Keating, S. E., Wisløff, U., & Coombes, J. S. Accuracy of heart rate watches: implications for weight management. PLoS ONE 11, https://doi.org/10.1371/journal.pone.0154420 (2016). source
Not checked in the free run.
- cites [50] Bent, B., Goldstein, B. A., Kibbe, W. A., & Dunn, J. P. Investigating sources of inaccuracy in wearable optical heart rate sensors. npj Digital Med. 3, https://doi.org/10.1038/s41746-020-0226-6 (2020). source
Not checked in the free run.
May combine two claims.
PremisecitationWearables that use red light will likely be more effective on darker skin tones because red light is not absorbed by melanin, which may allow for more accurate heart rate measurements.
“wearables that use red light will likely be more effective on darker skin tones as it is not absorbed by melanin, which may allow for more accurate HR measurements23.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [23] Mahloko, L. & Adebesin, F. in Responsible Design, Implementation and Use of Information and Communication Technology. Lecture Notes in Computer Science (eds Hattingh, M., et al.) Vol. 12067, 96–107 (Springer International Publishing, Cham, 2020). source
Not checked in the free run.
May combine two claims.
PremisecitationHigher body mass index, which likely covaries with larger wrist circumference, has been shown to be associated with reduced heart rate accuracy, although these outcomes have not been found in all studies.
“Higher body mass index, which likely covaries with larger wrist circumference as described above, has been shown to be associated with reduced HR accuracy42,62, although again these outcomes have not been found in all studies44,46.”
Quote in the paperNot a substantial claim0 of 4 sources checked- cites [42] Shcherbina, A. et al. Accuracy in wrist-worn, sensor-based measurements of heart rate and energy expenditure in a diverse cohort. J. Personalized Med. 7, 3 (2017). source
Not checked in the free run.
- cites [62] Menghini, L. et al. Stressing the accuracy: wrist-worn wearable sensor validation over different conditions. Psychophysiology. 56, https://doi.org/10.1111/psyp.13441 (2019). source
Not checked in the free run.
- cites [44] Wang, R. et al. Accuracy of wrist-worn heart rate monitors. JAMA Cardiol. 2, 104–106 (2017). source
Not checked in the free run.
- cites [46] Gillinov, S. et al. Variable accuracy of wearable heart rate monitors during aerobic exercise. Med. Sci. Sports Exerc. 49, 1697–1703 (2017). source
Not checked in the free run.
May combine two claims.
PremisecitationHair follicle density and sweat can influence device measurement.
“some research indicates that hair follicle density and sweat can influence device measurement69.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [69] Cosoli, G., Spinsante, S. & Scalise, L. Wrist-worn and chest-strap wearable devices: systematic review on accuracy and metrological characteristics. Measurement 159, 107789 (2020). source
Not checked in the free run.
May combine two claims.
PremisecitationHistorically, science and medicine have paid less attention to recruiting minority groups, which has resulted in blunted health gains in these populations.
“The importance of this point cannot be overstated as historically, science and medicine have paid less attention to recruiting minority groups68, which has resulted in blunted health gains in these populations”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [68] Hall, G. C. N., Yip, T. & Zárate, M. A. On becoming multicultural in a monocultural research world: a conceptual approach to studying ethnocultural diversity. Am. Psychol. 71, 40–51 (2016). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
InterpretationreasoningWearables that use red light will likely be more effective on darker skin tones as it is not absorbed by melanin.
“Furthermore, wearables that use red light will likely be more effective on darker skin tones as it is not absorbed by melanin, which may allow for more”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationreasoningConsumer wearable devices can be used to supplement medical-grade devices and continuously collect heart rate data on research subjects when risk for acute cardiac events is not of immediate concern.
“these devices can be used to supplement medical-grade devices and continuously collect HR data on research subjects when risk for acute cardiac events is not of immediate concern.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationA second study found that accuracy of heart rate measurement was highest during running.
“a second study found that the accuracy of HR measurement was highest during running51”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [51] Xie, J. et al. Evaluating the validity of current mainstream wearable devices in fitness tracking under various physical activities: comparative study. JMIR Mhealth Uhealth 6, e94 (2018). source
Not checked in the free run.
Probably not a substantial claim: may not pass S2 (checkable).
InterpretationreasoningThe recent introduction of consumer wearables has the potential to improve the study of cardiovascular health in medicine and psychophysiological stress in psychiatry and psychology.
“This has the potential to improve the study of cardiovascular health in medicine and psychophysiological stress in psychiatry and psychology.”
Quote in the paperNot a substantial claim- Rests on an assumption: Consumer wearables allow continuous, scalable, unobtrusive, big data collection of cardiac activity in real-world conditions with large samples.
Probably not a substantial claim: may not pass S2 (checkable).
InterpretationreasoningUsing consumer wearables to supplement medical-grade devices has the potential to advance research into cardiovascular health and psychophysiological stress in order to better understand how heart rate influences overall physical and psychological health.
“This has the potential to advance research into cardiovascular health and psychophysiological stress in order to better understand how HR influences overall physical and psychological health.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationPPG green light is less susceptible to motion artifacts, whereas PPG red light is more susceptible to motion artifacts.
“one strength of PPG green light is that it is less susceptible to motion artifacts, whereas PPG red light is more susceptible to motion artifacts”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [23] Mahloko, L. & Adebesin, F. in Responsible Design, Implementation and Use of Information and Communication Technology. Lecture Notes in Computer Science (eds Hattingh, M., et al.) Vol. 12067, 96–107 (Springer International Publishing, Cham, 2020). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationAcross four consumer wearables and two research-grade devices, all were reasonably accurate during rest or during prolonged elevated heart rate, and differences in accuracy tended to arise during changes in activity.
“Other research found that across four consumer wearables and two research-grade devices all were “reasonably accurate” during rest or during prolonged HR that was elevated and that differences in accuracy tended to arise during changes in activity50.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [50] Bent, B., Goldstein, B. A., Kibbe, W. A., & Dunn, J. P. Investigating sources of inaccuracy in wearable optical heart rate sensors. npj Digital Med. 3, https://doi.org/10.1038/s41746-020-0226-6 (2020). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationSome researchers have worked with industry partners to derive measurements of heart rate variability.
“although some researchers have worked with industry partners to derive measurements of HRV26.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [26] Natarajan, A., Pantelopoulos, A., Emir-Farinas, H. & Natarajan, P. Heart rate variability with photoplethysmography in 8 million individuals: results and scaling relations with age, gender, and time of day. Preprint at https://www.biorxiv.org/content/10.1101/772285v1 (2019). source
Not checked in the free run.
May combine two claims.
InterpretationcitationActivity intensity may be less important to device accuracy than the degree of erratic wrist movements and corresponding position of the wearable device during an activity, which cause motion artifacts and may or may not co-occur with more vigorous physical activity.
“Therefore, it is possible that activity intensity may be less important to device accuracy than the degree of erratic wrist movements and corresponding position of the wearable device during an activity74, which cause motion artifacts and may or may not co-occur with”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [74] Parak, J. & Korhonen, I. Evaluation of wearable consumer heart rate monitors based on photopletysmography. Conf. Proc. IEEE Eng. Med. Biol. Soc. 3670-3673 2014, 3670–3673 (2014).
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationSome studies found that accuracy of heart rate measurement was comparable across resting baseline or vigorous activity.
“some studies finding that accuracy of HR measurement was comparable across resting baseline or vigorous activity45”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [45] Dooley, E. E., Golaszewski, N. M. & Bartholomew, J. B. Estimating accuracy at exercise intensities: a comparative study of self-monitoring heart rate and physical activity wearable devices. JMIR mHealth uHealth 5, e34 (2017). source
Not checked in the free run.
Probably not a substantial claim: may not pass S2 (checkable). May combine two claims.
InterpretationreasoningUtilizing these guidelines and referencing them in methods sections has the potential to decrease non-standard reporting, increase reproducibility and replicability, and lead to greater generalizability of results, which may increase clinical utility.
“Utilizing these guidelines and referencing them in methods sections has the potential to decrease non-standard reporting, increase reproducibility and replicability, while also leading to greater generalizability of results, which may increase the clinical utility of findings in this field.”
Quote in the paperNot a substantial claim
The paper relies on this without support.
There is currently a lack of transparency from manufacturers on underlying algorithms and raw data outputs are often not provided.
“Unfortunately, there is currently a lack of transparency from manufacturers on underlying algorithms and raw data outputs are often times not provided, which prevents researchers from dealing with this issue and creating a standardized protocol for transforming raw data to HR.”
The paper relies on this without support.
The vast majority of wrist-worn wearable studies incorporating heart rate measurements do not report or control for participant race and ethnicity.
“Currently, the vast majority of wrist-worn wearable studies incorporating HR measurements do not report or control for participant race and ethnicity, which undermines the ability to look at potentially important group differences in health outcomes and therefore decreases the ability of results to generalize to large segments of the population.”
The paper relies on this without support.
Various wearable devices and versions on the market may use different hardware, which may introduce differences in heart rate recordings.
“there are currently various wearable devices and versions on the market that have the potential to use different hardware, which may introduce differences in HR recordings.”
The paper relies on this without support.
Wearable algorithms may be altered with firmware updates.
“These algorithms, while often simplistic, may be altered with firmware updates”
The paper relies on this without support.
Most studies fail to report firmware information.
“yet most studies fail to report firmware information.”
The paper relies on this without support.
Consumer wearables allow continuous, scalable, unobtrusive, big data collection of cardiac activity in real-world conditions with large samples.
“The recent introduction of consumer wearables or wrist-worn smartwatches and fitness monitors for the digital measurement of cardiovascular health and psychophysiological stress has allowed for the continuous, scalable, and unobtrusive, “big data” collection of overall cardiac activity in real-world conditions with large samples that allow for the collection of passive, large-scale, and ecolog”
The paper relies on this without support.
PPG technology can be used to index blood pressure, oxygen saturation, and cardiac output.
“It is important to note that PPG technology can be used to index blood pressure, oxygen saturation, and cardiac output, which will likely be slowly rolled out into wearable technology moving forward.”
The paper relies on this without support.
PPG can index heart rate variability, which is derived from beat-to-beat interval calculations in the time-domain.
“In addition, PPG can index HRV, which is derived from beat to beat intervals calculations in the time-domain, such as root mean squared of successive differences or the standard deviation of NN intervals, the latter of which is used in the Apple Watch, but only with sporadic recordings throughout the day.”
The paper relies on this without support.
Many commercial wearables do not provide beat-to-beat interval data to users.
“Currently, many commercial wearables do not provide beat to beat interval data to users”
The paper relies on this without support.
Research is heavily context dependent.
“We do this because research is heavily context dependent.”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Algorithm developmental biases stem from selecting features, adjusting training data composition, and model calibrations.
“The accuracy of heart rate measurements affects people with darker skin tones because of algorithm developmental biases that stem from selecting features and adjusting training data composition and model calibrations [23-26].”
From Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring (2025) · cites this paper as [23]
The Biobehavioral and demographic data section reports that darker skin tone has been associated with reduced wrist-worn heart-rate accuracy, while noting that not all studies find this effect. It attributes the potential difference to absorption of green light; it does not identify feature selection, training-data composition, or model calibration as causes of algorithmic bias. No page number is provided in our copy. (LLM, conf=0.82)
Experimental — accuracy not yet measuredSocioeconomic disparities in access to smartwatches may interact with skin-tone-related bias, raising additional equity concerns.
“However, regional variations may exist, and socioeconomic disparities in access to these devices may further interact with skin-tone-related bias, raising additional equity concerns [17-23].”
From Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring (2025) · cites this paper as [23]
The ‘Biobehavioral and demographic data’ section (page not available) discusses reduced heart-rate accuracy on darker skin tones in some studies and the resulting generalizability concerns. It does not address socioeconomic disparities in smartwatch access or whether those disparities interact with skin-tone-related bias. (LLM, conf=0.86)
Experimental — accuracy not yet measuredMore recent research has highlighted the potential presence of skin pigmentation biases in research-grade and consumer PPG-based wearable devices.
“More recent research has also highlighted the potential presence of these biases in research-grade and consumer PPG-based wearable devices [4,27-29]”
From Impact of skin pigmentation on pulse oximetry blood oxygenation and wearable pulse rate accuracy: systematic review and meta-analysis (2024) · cites this paper as [4]
The Biobehavioral and demographic data section (page unavailable) reports that darker skin tone may reduce the accuracy of wrist-worn heart-rate wearables, though not all studies find this. The paper focuses on consumer PPG wearables; the passages do not establish the claim for research-grade devices. (LLM, conf=0.84)
Experimental — accuracy not yet measuredThere is a positive trend in wearable studies, underlining growing interest in wearables in health research.
“Overall, we have identified a positive trend in wearable studies, underlining the growing interest in wearables in health research, in line with other reviews [3,224-226].”
From The impact of wearable technologies in health research: scoping review (2022) · cites this paper as [226]
The accuracy of PPG is impaired by factors that impede light transmission, such as elevated body mass index, tattoos, and higher hair follicle density.
“While safe and inexpensive, the accuracy of PPG is impaired by factors that impede light transmission, such as elevated body mass index, tattoos, and higher hair follicle density [4].”
From Accuracy of heart rate measurement with wrist-worn wearable devices in various skin tones: a systematic review (2023) · cites this paper as [4]
Cites
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1177/096228029900800204 · Cited by 9,619 · see citing works
Molecular Mechanisms of Depression: Perspectives on New Treatment Strategies
Lang, U. E., & Borgwardt, S. (2013) · Cellular Physiology and Biochemistry · doi:10.1159/000350094 · Cited by 8,501 · see citing works
Missing data analysis: making it work in the real world
Graham, J. W. (2009) · Annual Review of Psychology · doi:10.1146/annurev.psych.58.110405.085530 · Cited by 6,163 · see citing works
A manifesto for reproducible science
Munafò, M. R., Nosek, B. A., Bishop, D. V. M., Button, K. S., Chambers, C. D., Percie du Sert, N., Simonsohn, U., Wagenmakers, E.-J., Ware, J. J., & Ioannidis, J. P. A. (2017) · Nature Human Behaviour · doi:10.1038/s41562-016-0021 · Cited by 3,397 · see citing works
Walker, E. R., McGee, R. E., & Druss, B. G. (2015) · JAMA Psychiatry · doi:10.1001/jamapsychiatry.2014.2502 · Cited by 2,976 · see citing works
Cacioppo, J. T., Tassinary, L. G., Berntson, G. G., Shan, H., Mason, P., Kolb, B., Whishaw, I. Q., Geuter, S., Lindquist, M. A., Wager, T. D., et al. (2016) · Cambridge University Press eBooks · doi:10.1017/9781107415782 · Cited by 2,032 · see citing works
Large-Scale Assessment of a Smartwatch to Identify Atrial Fibrillation
Perez, M. V., Mahaffey, K. W., Hedlin, H., Rumsfeld, J. S., Garcia, A., Ferris, T., Balasubramanian, V., Russo, A. M., Rajmane, A., Cheung, L., et al. (2019) · New England Journal of Medicine · doi:10.1056/nejmoa1901183 · Cited by 1,917 · see citing works
Publication bias in the social sciences: Unlocking the file drawer
Franco, A., Malhotra, N., & Simonovits, G. (2014) · Science · doi:10.1126/science.1255484 · Cited by 1,542 · see citing works
Systematic review of the validity and reliability of consumer-wearable activity trackers
Evenson, K. R., Goto, M. M., & Furberg, R. D. (2015) · International Journal of Behavioral Nutrition and Physical Activity · doi:10.1186/s12966-015-0314-1 · Cited by 1,232 · see citing works
On the Analysis of Fingertip Photoplethysmogram Signals
Elgendi, M. (2012) · Current Cardiology Reviews · doi:10.2174/157340312801215782 · Cited by 1,206 · see citing works
EFFECTS OF BLOOD-PRESSURE MEASUREMENT BY THE DOCTOR ON PATIENT'S BLOOD PRESSURE AND HEART RATE
Mancia, G. (1983) · The Lancet · doi:10.1016/s0140-6736(83)92244-4 · Cited by 746 · see citing works
Investigating sources of inaccuracy in wearable optical heart rate sensors
Bent, B., Goldstein, B. A., Kibbe, W. A., & Dunn, J. P. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0226-6 · Cited by 673 · see citing works
Shcherbina, A., Mattsson, C., Waggott, D., Salisbury, H., Christle, J., Hastie, T., Wheeler, M., & Ashley, E. (2017) · Journal of Personalized Medicine · doi:10.3390/jpm7020003 · Cited by 625 · see citing works
Quality-space theory in olfaction
Young, B. D., Keller, A., & ROSENTHAL, D. M. (2014) · Frontiers in Psychology · doi:10.3389/fpsyg.2014.00001 · Cited by 580 · see citing works
Passive Detection of Atrial Fibrillation Using a Commercially Available Smartwatch
Tison, G. H., Sanchez, J. M., Ballinger, B., Singh, A., Olgin, J. E., Pletcher, M. J., Vittinghoff, E., Lee, E. S., Fan, S. M., Gladstone, R. A., et al. (2018) · JAMA Cardiology · doi:10.1001/jamacardio.2018.0136 · Cited by 566 · see citing works
Peake, J. M., Kerr, G., & Sullivan, J. P. (2018) · Frontiers in Physiology · doi:10.3389/fphys.2018.00743 · Cited by 523 · see citing works
Alvares, G. A., Quintana, D. S., Hickie, I. B., & Guastella, A. J. (2016) · Journal of Psychiatry and Neuroscience · doi:10.1503/jpn.140217 · Cited by 503 · see citing works
O’Connor, M.-F., Bower, J. E., Cho, H. J., Creswell, J. D., Dimitrov, S., Hamby, M. E., Hoyt, M. A., Martin, J. L., Robles, T. F., Sloan, E. K., et al. (2009) · Brain, Behavior, and Immunity · doi:10.1016/j.bbi.2009.04.005 · Cited by 500 · see citing works
Sim, I. (2019) · New England Journal of Medicine · doi:10.1056/nejmra1806949 · Cited by 491 · see citing works
An IoT-cloud Based Wearable ECG Monitoring System for Smart Healthcare
Yang, Z., Zhou, Q., Lei, L., Zheng, K., & Xiang, W. (2016) · Journal of Medical Systems · doi:10.1007/s10916-016-0644-9 · Cited by 490 · see citing works
Rohleder, N. (2019) · Psychoneuroendocrinology · doi:10.1016/j.psyneuen.2019.02.021 · Cited by 439 · see citing works
Heartache and heartbreak—the link between depression and cardiovascular disease
Nemeroff, C. B., & Goldschmidt-Clermont, P. J. (2012) · Nature Reviews Cardiology · doi:10.1038/nrcardio.2012.91 · Cited by 424 · see citing works
Quintana, D. S., Alvares, G. A., & Heathers, J. A. J. (2016) · Translational Psychiatry · doi:10.1038/tp.2016.73 · Cited by 420 · see citing works
Contextual sensitivity in scientific reproducibility
Van Bavel, J. J., Mende-Siedlecki, P., Brady, W. J., & Reinero, D. A. (2016) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1521897113 · Cited by 398 · see citing works
Accuracy of wrist-worn heart rate monitors
Wang, R., Blackburn, G., Desai, M., Phelan, D., Gillinov, L., Houghtaling, P., & Gillinov, M. (2017) · JAMA Cardiology · doi:10.1001/jamacardio.2016.3340 · Cited by 386 · see citing works
Low Moments for Small Samples: A Comparative Study of Order Statistics
Hastings, C., Mosteller, F., Tukey, J. W., & Winsor, C. P. (1947) · The Annals of Mathematical Statistics · doi:10.1214/aoms/1177730388 · Cited by 366 · see citing works
Variable Accuracy of Wearable Heart Rate Monitors during Aerobic Exercise
GILLINOV, S., ETIWY, M., WANG, R., BLACKBURN, G., PHELAN, D., GILLINOV, A. M., HOUGHTALING, P., JAVADIKASGARI, H., & DESAI, M. Y. (2017) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000001284 · Cited by 345 · see citing works
Accuracy of Heart Rate Watches: Implications for Weight Management
Wallen, M. P., Gomersall, S. R., Keating, S. E., Wisløff, U., & Coombes, J. S. (2016) · PLOS ONE · doi:10.1371/journal.pone.0154420 · Cited by 336 · see citing works
Zhang, D., Shen, X., & Qi, X. (2015) · Canadian Medical Association Journal · doi:10.1503/cmaj.150535 · Cited by 329 · see citing works
Xie, J., Wen, D., Liang, L., Jia, Y., Gao, L., & Lei, J. (2018) · JMIR mHealth and uHealth · doi:10.2196/mhealth.9754 · Cited by 310 · see citing works
Nelson, B. W., & Allen, N. B. (2019) · JMIR mHealth and uHealth · doi:10.2196/10828 · Cited by 310 · see citing works
Self-citationRadin, J. M., Wineinger, N. E., Topol, E. J., & Steinhubl, S. R. (2020) · The Lancet Digital Health · doi:10.1016/s2589-7500(19)30222-5 · Cited by 300 · see citing works
El-Amrawy, F., & Nounou, M. I. (2015) · Healthcare Informatics Research · doi:10.4258/hir.2015.21.4.315 · Cited by 287 · see citing works
Stahl, S. E., An, H.-S., Dinkel, D. M., Noble, J. M., & Lee, J.-M. (2016) · BMJ Open Sport & Exercise Medicine · doi:10.1136/bmjsem-2015-000106 · Cited by 276 · see citing works
Basimetric approach (law of initial value) to biological rhythms
Wilder, J. (1962) · Annals of the New York Academy of Sciences · doi:10.1111/j.1749-6632.1962.tb30629.x · Cited by 269 · see citing works
Dooley, E. E., Golaszewski, N. M., & Bartholomew, J. B. (2017) · JMIR mHealth and uHealth · doi:10.2196/mhealth.7043 · Cited by 242 · see citing works
Qiu, S., Cai, X., Sun, Z., Li, L., Zügel, M., Steinacker, J. M., & Schümann, U. (2017) · Journal of the American Heart Association · doi:10.1161/jaha.117.005505 · Cited by 233 · see citing works
Heart rate variability with photoplethysmography in 8 million individuals: a cross-sectional study
Natarajan, A., Pantelopoulos, A., Emir-Farinas, H., & Natarajan, P. (2020) · The Lancet Digital Health · doi:10.1016/s2589-7500(20)30246-6 · Cited by 229 · see citing works
de Zambotti, M., Baker, F. C., Willoughby, A. R., Godino, J. G., Wing, D., Patrick, K., & Colrain, I. M. (2016) · Physiology & Behavior · doi:10.1016/j.physbeh.2016.03.006 · Cited by 222 · see citing works
Real-world heart rate norms in the Health eHeart study
Avram, R., Tison, G. H., Aschbacher, K., Kuhar, P., Vittinghoff, E., Butzner, M., Runge, R., Wu, N., Pletcher, M. J., Marcus, G. M., et al. (2019) · npj Digital Medicine · doi:10.1038/s41746-019-0134-9 · Cited by 217 · see citing works
Validity of Wearable Activity Monitors during Cycling and Resistance Exercise
BOUDREAUX, B. D., HEBERT, E. P., HOLLANDER, D. B., WILLIAMS, B. M., CORMIER, C. L., NAQUIN, M. R., GILLAN, W. W., GUSEW, E. E., & KRAEMER, R. R. (2018) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000001471 · Cited by 205 · see citing works
Psychology's Replication Crisis and Clinical Psychological Science
Tackett, J. L., Brandes, C. M., King, K. M., & Markon, K. E. (2019) · Annual Review of Clinical Psychology · doi:10.1146/annurev-clinpsy-050718-095710 · Cited by 196 · see citing works
Kemp, A. H., Brunoni, A. R., Santos, I. S., Nunes, M. A., Dantas, E. M., Carvalho de Figueiredo, R., Pereira, A. C., Ribeiro, A. L. P., Mill, J. G., Andreão, R. V., et al. (2014) · American Journal of Psychiatry · doi:10.1176/appi.ajp.2014.13121605 · Cited by 189 · see citing works
Wigton, R., Radua, J., Allen, P., Averbeck, B., Meyer-Lindenberg, A., McGuire, P., Shergill, S. S., & Fusar-Poli, P. (2015) · Journal of Psychiatry and Neuroscience · doi:10.1503/jpn.130289 · Cited by 179 · see citing works
Hall, G. C. N., Yip, T., & Zárate, M. A. (2016) · American Psychologist · doi:10.1037/a0039734 · Cited by 173 · see citing works
Kandola, A., Ashdown-Franks, G., Stubbs, B., Osborn, D. P. J., & Hayes, J. F. (2019) · Journal of Affective Disorders · doi:10.1016/j.jad.2019.07.088 · Cited by 172 · see citing works
Stressing the accuracy: Wrist‐worn wearable sensor validation over different conditions
Menghini, L., Gianfranchi, E., Cellini, N., Patron, E., Tagliabue, M., & Sarlo, M. (2019) · Psychophysiology · doi:10.1111/psyp.13441 · Cited by 157 · see citing works
Validation of Biofeedback Wearables for Photoplethysmographic Heart Rate Tracking.
(2016) · PubMed Central · Cited by 150 · see citing works
Assessment of the Fitbit Charge 2 for monitoring heart rate
Benedetto, S., Caldato, C., Bazzan, E., Greenwood, D. C., Pensabene, V., & Actis, P. (2018) · PLOS ONE · doi:10.1371/journal.pone.0192691 · Cited by 149 · see citing works
Evaluation of wearable consumer heart rate monitors based on photopletysmography
Parak, J., & Korhonen, I. (2014) · 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/embc.2014.6944419 · Cited by 148 · see citing works
Comparative evaluation of heart rate-based monitors: Apple Watch vs Fitbit Charge HR
Bai, Y., Hibbing, P. R., Mantis, K., & Welk, G. J. (2017) · Journal of Sports Sciences · doi:10.1080/02640414.2017.1412235 · Cited by 147 · see citing works
Validation of photoplethysmography as a method to detect heart rate during rest and exercise
Spierer, D. K., Rosen, Z., Litman, L. L., & Fujii, K. (2015) · Journal of Medical Engineering & Technology · doi:10.3109/03091902.2015.1047536 · Cited by 139 · see citing works
The reproducibility crisis in the age of digital medicine
Stupple, A., Singerman, D., & Celi, L. A. (2019) · npj Digital Medicine · doi:10.1038/s41746-019-0079-z · Cited by 137 · see citing works
Kroll, R. R., Boyd, J. G., & Maslove, D. M. (2016) · Journal of Medical Internet Research · doi:10.2196/jmir.6025 · Cited by 123 · see citing works
Cosoli, G., Spinsante, S., & Scalise, L. (2020) · Measurement · doi:10.1016/j.measurement.2020.107789 · Cited by 122 · see citing works
Low, C. A., Salomon, K., & Matthews, K. A. (2009) · Psychosomatic Medicine · doi:10.1097/psy.0b013e3181ba18ed · Cited by 121 · see citing works
Self-citationConfidence and precision increase with high statistical power
Button, K. S., Ioannidis, J. P. A., Mokrysz, C., Nosek, B. A., Flint, J., Robinson, E. S. J., & Munafò, M. R. (2013) · Nature Reviews Neuroscience · doi:10.1038/nrn3475-c4 · Cited by 121 · see citing works
Kemp, A. H., Quintana, D. S., Quinn, C. R., Hopkinson, P., & Harris, A. W. F. (2014) · Frontiers in Psychology · doi:10.3389/fpsyg.2014.01387 · Cited by 94 · see citing works
Gorny, A. W., Liew, S. J., Tan, C. S., & Müller-Riemenschneider, F. (2017) · JMIR mHealth and uHealth · doi:10.2196/mhealth.8233 · Cited by 94 · see citing works
The impact of posttraumatic stress disorder on blood pressure and heart rate in a veteran population
Paulus, E. J., Argo, T. R., & Egge, J. A. (2013) · Journal of Traumatic Stress · doi:10.1002/jts.21785 · Cited by 91 · see citing works
Weiler, D. T., Villajuan, S. O., Edkins, L., Cleary, S., & Saleem, J. J. (2017) · Proceedings of the Human Factors and Ergonomics Society Annual Meeting · doi:10.1177/1541931213601804 · Cited by 87 · see citing works
Latvala, A., Kuja-Halkola, R., Rück, C., D’Onofrio, B. M., Jernberg, T., Almqvist, C., Mataix-Cols, D., Larsson, H., & Lichtenstein, P. (2016) · JAMA Psychiatry · doi:10.1001/jamapsychiatry.2016.2717 · Cited by 82 · see citing works
The Accuracy of Heart Rate Monitoring by Some Wrist-Worn Activity Trackers
Cadmus‐Bertram, L., Gangnon, R. E., Wirkus, E. J., Thraen-Borowski, K. M., & Gorzelitz, J. (2017) · Annals of Internal Medicine · doi:10.7326/l16-0353 · Cited by 80 · see citing works
Khan, H., Kunutsor, S. K., Kalogeropoulos, A. P., Georgiopoulou, V. V., Newman, A. B., Harris, T. B., Bibbins-Domingo, K. B., Kauhanen, J., Gheorghiade, M., Fonarow, G. C., et al. (2015) · Journal of the American Heart Association · doi:10.1161/jaha.114.001364 · Cited by 74 · see citing works
Handling Missing Data in the Modeling of Intensive Longitudinal Data
Ji, L., Chow, S.-M., Schermerhorn, A. C., Jacobson, N. C., & Cummings, E. M. (2018) · Structural Equation Modeling: A Multidisciplinary Journal · doi:10.1080/10705511.2017.1417046 · Cited by 74 · see citing works
Self-citationJacobson, N. C., Chow, S.-M., & Newman, M. G. (2019) · Behavior Research Methods · doi:10.3758/s13428-018-1101-0 · Cited by 63 · see citing works
Self-citationExtending the Passive-Sensing Toolbox: Using Smart-Home Technology in Psychological Science
Nelson, B. W., & Allen, N. B. (2018) · Perspectives on Psychological Science · doi:10.1177/1745691618776008 · Cited by 60 · see citing works
Self-citationA simple test of the Law of Initial Values
GEENEN, R., & VAN DE VIJVER, F. J. R. (1993) · Psychophysiology · doi:10.1111/j.1469-8986.1993.tb02076.x · Cited by 52 · see citing works
Responsible Design, Implementation and Use of Information and Communication Technology
Hattingh, M., Matthee, M. C., Smuts, H., Pappas, I., Dwivedi, Y. K., & Mäntymäki, M. (2020) · Lecture notes in computer science · doi:10.1007/978-3-030-44999-5 · Cited by 51 · see citing works
Altered autonomic arousal in psychosis: An analysis of vulnerability and specificity
Clamor, A., Hartmann, M. M., Köther, U., Otte, C., Moritz, S., & Lincoln, T. M. (2014) · Schizophrenia Research · doi:10.1016/j.schres.2014.02.006 · Cited by 42 · see citing works
Hunkin, H., King, D. L., & Zajac, I. T. (2019) · Clinical Psychology Science and Practice · doi:10.1111/cpsp.12290 · Cited by 37 · see citing works
Pattyn, N., Migeotte, P.-F., Neyt, X., den Nest, A. V., & Cluydts, R. (2010) · Physiology & Behavior · doi:10.1016/j.physbeh.2010.04.037 · Cited by 32 · see citing works
(2020) · Nature Human Behaviour · doi:10.1038/s41562-020-0818-9 · Cited by 29 · see citing works
Mahloko, L., & Adebesin, F. (2020) · Lecture Notes in Computer Science · doi:10.1007/978-3-030-45002-1_9 · Cited by 23 · see citing works
Tison, G. H., Singh, A. C., Ohashi, D. A., Hsieh, J. T., Ballinger, B. M., Olgin, J. E., Marcus, G. M., & Pletcher, M. J. (2017) · Circulation · doi:10.1161/circ.136.suppl_1.21042 · Cited by 19 · see citing works
Influenza Surveillance Using Wearable Mobile Health Devices
Konty, K., Bradshaw, B., Ramirez, E. R., Lee, W.-N., Signorini, A., & Foschini, L. (2019) · Online Journal of Public Health Informatics · doi:10.5210/ojphi.v11i1.9758 · Cited by 16 · see citing works
Psychiatric disorders, morbidity, and mortality: tracing mechanistic pathways to accelerated aging
Kiecolt-Glaser, J. K., & Wilson, S. J. (2016) · Psychosomatic Medicine · doi:10.1097/psy.0000000000000373 · Cited by 14 · see citing works
Clinic and ambulatory heart rate in sustained and white-coat hypertension
Pierdomenico, S. D., Bucci, A., Lapenna, D., Cuccurullo, F., & Mezzetti, A. (2001) · Blood Pressure Monitoring · doi:10.1097/00126097-200110000-00003 · Cited by 9 · see citing works
Natarajan, A., Pantelopoulos, A., Emir-Farinas, H., & Natarajan, P. (2019) · doi:10.1101/772285 · Cited by 4 · see citing works
Flagged first, then by citations
Research works printed but not linked: 1
- Latvala, A. et al. Association of resting heart rate and blood pressure in late adolescence with subsequent mental disorders. JAMA Psychiatry 41, 89–104 (2016).
Cited by
Sorted by citations
The impact of wearable technologies in health research: scoping review
Huhn, S., Axt, M., Gunga, H.-C., Maggioni, M. A., Munga, S., Obor, D., Sié, A., Boudo, V., Bunker, A., Sauerborn, R., et al. (2022) · JMIR mHealth and uHealth · doi:10.2196/34384 · Cited by 379 · see citing works
Machine Learning for Healthcare Wearable Devices: The Big Picture
Sabry, F., Eltaras, T., Labda, W., Alzoubi, K., & Malluhi, Q. (2022) · Journal of Healthcare Engineering · doi:10.1155/2022/4653923 · Cited by 321 · see citing works
Self-Powered Sensing in Wearable Electronics─A Paradigm Shift Technology
Tang, W., Sun, Q., & Wang, Z. L. (2023) · Chemical Reviews · doi:10.1021/acs.chemrev.3c00305 · Cited by 318 · see citing works
Quigley, K. S., Gianaros, P. J., Norman, G. J., Jennings, J. R., Berntson, G. G., & de Geus, E. J. C. (2024) · Psychophysiology · doi:10.1111/psyp.14604 · Cited by 253 · see citing works
Teixeira, E., Fonseca, H., Diniz‐Sousa, F., Veras, L., Boppre, G., de Oliveira, J. J. F., Pinto, D., Alves, A. J., Barbosa, A., Mendes, R., et al. (2021) · Geriatrics · doi:10.3390/geriatrics6020038 · Cited by 172 · see citing works
A Review of Wearable Multi-Wavelength Photoplethysmography
Ray, D., Collins, T., Woolley, S., & Ponnapalli, P. (2023) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2021.3121476 · Cited by 155 · see citing works
Mansi, S. A., Barone, G. B., Forzano, C., Pigliautile, I., Ferrara, M., Pisello, A. L., & Arnesano, M. (2021) · Measurement · doi:10.1016/j.measurement.2021.109872 · Cited by 146 · see citing works
AI-Driven Wearable Bioelectronics in Digital Healthcare
Huang, G., Chen, X., & Liao, C. (2025) · Biosensors · doi:10.3390/bios15070410 · Cited by 136 · see citing works
Scardulla, F., Cosoli, G., Spinsante, S., Poli, A., Iadarola, G., Pernice, R., Busacca, A., Pasta, S., Scalise, L., & D’Acquisto, L. (2023) · Measurement · doi:10.1016/j.measurement.2023.113150 · Cited by 120 · see citing works
Mühlen, J. M., Stang, J. S., Skovgaard, E. L., Júdice, P. B., Molina‐García, P., Johnston, W., Sardinha, L. B., Ortega, F. B., Caulfield, B., Bloch, W., et al. (2021) · British Journal of Sports Medicine · doi:10.1136/bjsports-2020-103148 · Cited by 118 · see citing works
Chee, N. I. Y. N., Ghorbani, S., Golkashani, H. A., Leong, R. L. F., Ong, J. L., & Chee, M. W. L. (2021) · Nature and Science of Sleep · doi:10.2147/nss.s286070 · Cited by 108 · see citing works
Consumer Wearable Health and Fitness Technology in Cardiovascular Medicine
Petek, B. J., Al‐Alusi, M. A., Moulson, N., Grant, A. J., Besson, C., Guseh, J. S., Wasfy, M. M., Grémeaux, V., Churchill, T. W., & Baggish, A. L. (2023) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2023.04.054 · Cited by 106 · see citing works
Roos, L. G., & Slavich, G. M. (2023) · Brain Behavior and Immunity · doi:10.1016/j.bbi.2023.08.008 · Cited by 100 · see citing works
Cosoli, G., Spinsante, S., Scardulla, F., D’Acquisto, L., & Scalise, L. (2021) · Measurement · doi:10.1016/j.measurement.2021.109243 · Cited by 98 · see citing works
Magnon, V., Vallet, G. T., Benson, A. C., Mermillod, M., Chaussé, P., Lacroix, A., Bouillon‐Minois, J., & Dutheil, F. (2022) · Cortex · doi:10.1016/j.cortex.2022.07.008 · Cited by 97 · see citing works
The Future of Stress Management: Integration of Smartwatches and HRV Technology
Jerath, R., Syam, M., & Ahmed, S. (2023) · Sensors · doi:10.3390/s23177314 · Cited by 96 · see citing works
Koerber, D., Khan, S., Shamsheri, T., Kirubarajan, A., & Mehta, S. (2023) · Journal of Racial and Ethnic Health Disparities · doi:10.1007/s40615-022-01446-9 · Cited by 91 · see citing works
Use of technology and social media in dementia care: Current and future directions
Shu, S., & Woo, B. K. P. (2021) · World Journal of Psychiatry · doi:10.5498/wjp.v11.i4.109 · Cited by 88 · see citing works
Scardulla, F., D’Acquisto, L., Colombarini, R., Hu, S., Pasta, S., & Bellavia, D. (2020) · Sensors · doi:10.3390/s20185052 · Cited by 72 · see citing works
Ueafuea, K., Boonnag, C., Sudhawiyangkul, T., Leelaarporn, P., Gulistan, A., Chen, W., Mukhopadhyay, S. C., Wilaiprasitporn, T., & Piyayotai, S. (2020) · IEEE Sensors Journal · doi:10.1109/jsen.2020.3046259 · Cited by 72 · see citing works
FLIRT: A feature generation toolkit for wearable data
Föll, S., Maritsch, M., Spinola, F., Mishra, V., Barata, F., Kowatsch, T., Fleisch, E., & Wortmann, F. (2021) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2021.106461 · Cited by 72 · see citing works
Consumer escapism: Scale development, validation, and physiological associations
Orazi, D. C., Mah, K. Y., Derksen, T., & Murray, K. B. (2023) · Journal of Business Research · doi:10.1016/j.jbusres.2023.113805 · Cited by 65 · see citing works
Nissen, M., Slim, S., Jaeger, K. M., Flaucher, M., Huebner, H., Danzberger, N., Fasching, P. A., Beckmann, M. W., Gradl, S., & Eskofier, B. M. (2022) · JMIR Formative Research · doi:10.2196/33635 · Cited by 62 · see citing works
Physiological markers of traffic-related stress during active travel
Bigazzi, A. Y., Ausri, F., Peddie, L., Fitch, D. T., & Puterman, E. M. (2021) · Transportation Research Part F Traffic Psychology and Behaviour · doi:10.1016/j.trf.2021.12.003 · Cited by 53 · see citing works
Wearable biosensors for cardiovascular monitoring leveraging nanomaterials
Chen, X., Manshaii, F., Tioran, K., Wang, S., Zhou, Y., Zhao, J., Yang, M., Yin, X., Liu, S., & Wang, K. (2024) · Advanced Composites and Hybrid Materials · doi:10.1007/s42114-024-00906-6 · Cited by 50 · see citing works
Birenboim, A., Helbich, M., & Kwan, M. (2021) · Computers Environment and Urban Systems · doi:10.1016/j.compenvurbsys.2021.101650 · Cited by 49 · see citing works
Promises and Perils of Consumer Mobile Technologies in Cardiovascular Care
Varma, N., Han, J. K., PASSMAN, R. S., Rosman, L., Ghanbari, H., Noseworthy, P. A., Silva, J. N. A., Deshmukh, A. J., Sanders, P., Hindricks, G., et al. (2024) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2023.11.024 · Cited by 48 · see citing works
Gonzalez, D. A. Z., Richards, D., & BILGIN, A. A. B. (2020) · International Journal of Human-Computer Studies · doi:10.1016/j.ijhcs.2020.102579 · Cited by 47 · see citing works
Amourizi, F., Dashtian, K., Zare‐Dorabei, R., & Hajati, S. (2024) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2024.136731 · Cited by 47 · see citing works
A Review of Wearable Optical Fiber Sensors for Rehabilitation Monitoring
Li, X., Li, Y., Wei, H., Wang, C., & Liu, B. (2024) · Sensors · doi:10.3390/s24113602 · Cited by 45 · see citing works
Research on digital technology use in cardiology: bibliometric analysis
Yeung, A. W. K., Kulnik, S. T., Parvanov, E. D., Fassl, A., Eibensteiner, F., Völkl-Kernstock, S., Kletecka-Pulker, M., Crutzen, R., Gutenberg, J., Höppchen, I., et al. (2022) · Journal of Medical Internet Research · doi:10.2196/36086 · Cited by 44 · see citing works
Demographic Imbalances Resulting From the Bring-Your-Own-Device Study Design
Cho, P. J., Yi, J., Ho, E., Shandhi, M. M. H., Dinh, Y., Patil, A., Martin, L., Singh, G., Bent, B., Ginsburg, G. S., et al. (2021) · JMIR mhealth and uhealth · doi:10.2196/29510 · Cited by 44 · see citing works
Nelson, B. W., Sheeber, L., Pfeifer, J. H., & Allen, N. B. (2020) · Journal of Child Psychology and Psychiatry · doi:10.1111/jcpp.13264 · Cited by 43 · see citing works
Self-citationUse of Consumer Wearables in Health Research: Issues and Considerations
Dobson, R., Stowell, M., Warren, J., Tane, T., Ni, L., Gu, Y., McCool, J. P., & Whittaker, R. (2023) · Journal of Medical Internet Research · doi:10.2196/52444 · Cited by 42 · see citing works
Detection and Monitoring of Viral Infections via Wearable Devices and Biometric Data
Goergen, C. J., Tweardy, M. J., Steinhubl, S. R., Wegerich, S., Singh, K., Mieloszyk, R. J., & Dunn, J. (2021) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-103020-040136 · Cited by 41 · see citing works
Are Activity Wrist-Worn Devices Accurate for Determining Heart Rate during Intense Exercise?
Martín-Escudero, P., Cabanas, A. M., Dotor-Castilla, M. L., Galindo-Canales, M., Miguel‐Tobal, F., Fernández, C., Fuentes, M. E., & Giannetti, R. (2023) · Bioengineering · doi:10.3390/bioengineering10020254 · Cited by 40 · see citing works
Huhn, S., Matzke, I., Koch, M., Gunga, H., Maggioni, M. A., Sié, A., Boudo, V., Ouédraogo, W. A., Compaoré, G., Bunker, A., et al. (2022) · Frontiers in Public Health · doi:10.3389/fpubh.2022.972177 · Cited by 39 · see citing works
Jia, S., Gao, H., Xue, Z., & Meng, X. (2022) · Biosensors · doi:10.3390/bios12111057 · Cited by 39 · see citing works
Hehlmann, M. I., Schwartz, B., Lutz, T., Penedo, J. M. G., Rubel, J. A., & Lutz, W. (2021) · Frontiers in Psychiatry · doi:10.3389/fpsyt.2021.613085 · Cited by 38 · see citing works
Wearable and telemedicine innovations for Olympic events and elite sport
Muñiz-Pardos, B., Angeloudis, K., Guppy, F. M., Keramitsoglou, I., Sutehall, S., Bosch, A. N., Tanisawa, K., Hosokawa, Y., Ash, G. I., Schobersberger, W., et al. (2021) · The Journal of Sports Medicine and Physical Fitness · doi:10.23736/s0022-4707.21.12752-5 · Cited by 38 · see citing works
Heart rate variability in marketing research: A systematic review and methodological perspectives
Kakaria, S., Alcañiz, J. E. B., Catrambone, V., & Valenza, G. (2022) · Psychology and Marketing · doi:10.1002/mar.21734 · Cited by 37 · see citing works
Singh, S., Bennett, M. R., Chen, C., Shin, S., Ghanbari, H., & Nelson, B. W. (2024) · Journal of Medical Internet Research · doi:10.2196/62769 · Cited by 36 · see citing works
“This has enabled consumers to continuously and unobtrusively track health status, generating information that is becoming commonly used in health care settings and as a source of research end points [3,4].”
InfluentialSelf-citationDepression Recognition Using Daily Wearable-Derived Physiological Data
Shui, X., Xu, H. J., Tan, S., & Zhang, D. (2025) · Sensors · doi:10.3390/s25020567 · Cited by 35 · see citing works
Shui, X., Chen, Y., Hu, X., Wang, F., & Zhang, D. (2023) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2023.3253820 · Cited by 34 · see citing works
Helmer, P., Hottenrott, S., Rodemers, P., Leppich, R., Helwich, M., Pryss, R. C., Kranke, P., Meybohm, P., Winkler, B. E., & Sammeth, M. (2022) · Journal of Medical Internet Research · doi:10.2196/42359 · Cited by 33 · see citing works
Vanhatalo, S., Liedes, H., & Pennanen, K. (2022) · Foods · doi:10.3390/foods11070964 · Cited by 33 · see citing works
Areia, C. M., Santos, M., Vollam, S., Pimentel, M. A. F., Young, L., Román, C., Ede, J., Piper, P., King, E., Gustafson, O. D., et al. (2021) · Journal of Medical Internet Research · doi:10.2196/27547 · Cited by 32 · see citing works
Artificial Intelligence in Endurance Sports: Metabolic, Recovery, and Nutritional Perspectives
Grivas, G. V., & Safari, K. (2025) · Nutrients · doi:10.3390/nu17203209 · Cited by 30 · see citing works
Zheng, X. H., Liu, Z., Liu, J., Hu, C., Du, Y., Li, J., Pan, Z., & Ding, K. (2025) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.4c22895 · Cited by 29 · see citing works
Budig, M., Keiner, M., Stoohs, R. A., Hoffmeister, M., & Höltke, V. (2021) · Sensors · doi:10.3390/s22010180 · Cited by 29 · see citing works
Heart Rate Monitoring Using PPG With Smartphone Camera
Ayesha, A. H., Qiao, D., & Zulkernine, F. H. (2021) · IEEE International Conference on Bioinformatics and Biomedicine (BIBM) · doi:10.1109/bibm52615.2021.9669735 · Cited by 27 · see citing works
Myths and mythconceptions: what does it mean to be a programming language, anyhow?
Shaw, M. (2020) · Proceedings of the ACM on Programming Languages · doi:10.1145/3480947 · Cited by 27 · see citing works
Hurley, O. A. (2021) · Frontiers in Psychology · doi:10.3389/fpsyg.2021.621283 · Cited by 27 · see citing works
The relationship between wearable-derived sleep features and relapse in Major Depressive Disorder
Matcham, F., Carr, E., Meyer, N., White, K. M., Oetzmann, C., Leightley, D., Lamers, F., Siddi, S., Cummins, N., Annas, P., et al. (2024) · Journal of Affective Disorders · doi:10.1016/j.jad.2024.07.136 · Cited by 25 · see citing works
Embedding Temporal Convolutional Networks for Energy-efficient PPG-based Heart Rate Monitoring
Burrello, A., Pagliari, D. J., Rapa, P. M., Semilia, M., Risso, M., Polonelli, T., Poncino, M., Luca Benini, C., & Benatti, S. (2022) · ACM Transactions on Computing for Healthcare · doi:10.1145/3487910 · Cited by 25 · see citing works
Tabassum, S. M., Ruesch, A., Acharya, D., Yang, J., Relander, F. A. J., Scammon, B., Wolf, M. S., Rakkar, J., Clark, R. S. B., McDowell, M. M., et al. (2022) · Journal of neurosurgery · doi:10.3171/2022.9.jns221203 · Cited by 25 · see citing works
Using Smartwatches in Stress Management, Mental Health, and Well-Being: A Systematic Review
Kapogianni, N.-A., Sideraki, A., & Anagnostopoulos, C. (2025) · Algorithms · doi:10.3390/a18070419 · Cited by 23 · see citing works
Meaningful engagement: A crossfunctional framework for digital therapeutics
Gabe, Flannery, J., Vierra, E., Koepsell, X., Berglund, E., Miller, I., & Lake, J. I. (2022) · Frontiers in Digital Health · doi:10.3389/fdgth.2022.890081 · Cited by 23 · see citing works
Alfonso, C., Garcia-Gonzalez, M. A., Parrado, E., Gil-Rojas, J., Ramos-Castro, J. J., & Capdevila, L. (2022) · Scientific Reports · doi:10.1038/s41598-022-18356-9 · Cited by 23 · see citing works
Sinichi, M., Gevonden, M. J., & Krabbendam, L. (2025) · Psychophysiology · doi:10.1111/psyp.70004 · Cited by 22 · see citing works
Harvie, H. M. K., Jain, B., Nelson, B. W., Knight, E. L., Roos, L. E., & Giuliano, R. J. (2021) · Stress · doi:10.1080/10253890.2021.1995714 · Cited by 22 · see citing works
Self-citationHRS White Paper on Clinical Utilization of Digital Health Technology
Wan, E. Y., Ghanbari, H., AKOUM, N. W., Attia, Z. I., Asirvatham, S. J., Chung, E. H., Dagher, L., Al‐Khatib, S. M., Mendenhall, G. S., McManus, D. D., et al. (2021) · Cardiovascular Digital Health Journal · doi:10.1016/j.cvdhj.2021.07.001 · Cited by 21 · see citing works
Hunter, A., Leckie, T., Coe, O., Hardy, B., Fitzpatrick, D., Gonçalves, A., Standing, M.-K., Koulouglioti, C., Richardson, A., & Hodgson, L. E. (2022) · JMIR Rehabilitation and Assistive Technologies · doi:10.2196/25494 · Cited by 21 · see citing works
Bio-inspired ultra-thin microfluidics for soft sweat-activated batteries and skin electronics
Wu, M., Shi, R., Zhou, J., Wong, T. H., Yao, K., Li, J., Huang, X., Li, D., Gao, Y., Liu, Y., et al. (2022) · Journal of Materials Chemistry A · doi:10.1039/d2ta01154a · Cited by 21 · see citing works
Ali, M. A., Abdelsallam, A., Rasslan, A., & Rabee, A. (2024) · International Journal of Physical Education Fitness and Sports · doi:10.54392/ijpefs2421 · Cited by 20 · see citing works
Çetin, A. E., Öztürk, Y., Hanosh, O., & Ansari, R. (2021) · Digital Signal Processing · doi:10.1016/j.dsp.2021.103247 · Cited by 20 · see citing works
Lambe, L. J., & Craig, W. (2022) · International Journal of Bullying Prevention · doi:10.1007/s42380-022-00128-8 · Cited by 20 · see citing works
Designing daily-life research combining experience sampling method with parallel data
Velozo, J. D. C., Habets, J., George, S. V., Niemeijer, K., Minaeva, O., Hagemann, N., Herff, C., Kuppens, P., Rintala, A., Vaessen, T., et al. (2022) · Psychological Medicine · doi:10.1017/s0033291722002367 · Cited by 20 · see citing works
Budig, M., Stoohs, R. A., & Keiner, M. (2022) · Sensors · doi:10.3390/s22239540 · Cited by 20 · see citing works
Heterogeneity in Extended Reality Influences Procedural Knowledge Gain and Operation Training
Yong, J., Wei, J., Wang, Y., Dang, J., Lei, X., & Lu, W. (2023) · IEEE Transactions on Learning Technologies · doi:10.1109/tlt.2023.3286612 · Cited by 20 · see citing works
Bryan, A. C., Heinz, M. V., Salzhauer, A. J., Price, G. D., Tlachac, M., & Jacobson, N. C. (2024) · Biomedical Materials & Devices · doi:10.1007/s44174-023-00150-4 · Cited by 20 · see citing works
Self-citationDamme, K. S. F., Vargas, T., Walther, S., Shankman, S. A., & Mittal, V. A. (2024) · Translational Psychiatry · doi:10.1038/s41398-024-02794-2 · Cited by 19 · see citing works
Deriving Accurate Nocturnal Heart Rate, rMSSD and Frequency HRV from the Oura Ring
Liang, T., Yılmaz, G., & Soon, C. S. (2024) · Sensors · doi:10.3390/s24237475 · Cited by 19 · see citing works
Nelson, B. W., Flannery, J., Flournoy, J. C., Duell, N., Prinstein, M. J., & Telzer, E. H. (2021) · Journal of Child Psychology and Psychiatry · doi:10.1111/jcpp.13471 · Cited by 19 · see citing works
Self-citationPatient Engagement in a Multimodal Digital Phenotyping Study of Opioid Use Disorder
Campbell, C. I., Chen, C.-H., Adams, S. R., Asyyed, A. H., Athale, N., Does, M. B., Hassanpour, S., Hichborn, E. G., Jackson-Morris, M., Jacobson, N. C., et al. (2023) · Journal of Medical Internet Research · doi:10.2196/45556 · Cited by 19 · see citing works
Self-citationConsumer Wearable Sleep Trackers
Chiang, A. A.-P., & Khosla, S. (2023) · Sleep Medicine Clinics · doi:10.1016/j.jsmc.2023.05.005 · Cited by 19 · see citing works
A feedforward neural network for drone accident prediction from physiological signals
Sakib, M. N., Chaspari, T., & Behzadan, A. H. (2021) · Smart and Sustainable Built Environment · doi:10.1108/sasbe-12-2020-0181 · Cited by 18 · see citing works
Accuracy of wrist-worn wearable devices for determining exercise intensity
Ho, W.-T., Yang, Y.-J., & Li, T.-C. (2022) · Digital Health · doi:10.1177/20552076221124393 · Cited by 18 · see citing works
Nature-Based Feasibility Intervention to Influence Mitigation Strategies for Perceived Stress
Vermeesch, A. L., Coro, A., Mattes, K., Ostendorff, D., Olson, E. T., & Garrigues, L. (2022) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph191912277 · Cited by 18 · see citing works
Insights into vascular physiology from sleep photoplethysmography
Yılmaz, G., Ong, J. L., Ling, L. H., & Chee, M. W. L. (2023) · SLEEP · doi:10.1093/sleep/zsad172 · Cited by 18 · see citing works
A Reconfigurable Nonvolatile Memory Architecture for Prolonged Wearable Health Monitoring Devices
Gargari, M. A., Eslami, N., & Moaiyeri, M. H. (2024) · IEEE Transactions on Consumer Electronics · doi:10.1109/tce.2024.3399223 · Cited by 18 · see citing works
Tanade, C., Khan, N. S., Rakestraw, E., Ladd, W., Draeger, E. W., & Randles, A. P. (2024) · npj Digital Medicine · doi:10.1038/s41746-024-01216-3 · Cited by 18 · see citing works
Lu, J. K., Sijm, M., Janssens, G. E., Goh, J., & Maier, A. B. (2023) · GeroScience · doi:10.1007/s11357-023-00815-4 · Cited by 17 · see citing works
Malarvizhi, C. A., Manzoor, S. R., & Haque, R. (2024) · International Journal of Service Science Management Engineering and Technology · doi:10.4018/ijssmet.349911 · Cited by 17 · see citing works
Ibrahim, N. S., Rampal, S., Lee, W. L., Pek, E. W., & Suhaimi, A. (2023) · Cardiovascular Engineering and Technology · doi:10.1007/s13239-023-00693-z · Cited by 16 · see citing works
Spatial proximity as a behavioral marker of relationship dynamics in older adult couples
Ogolsky, B. G., Mejía, S. T., Chronopoulou, A., Dobson, K., Maniotes, C. R., Rice, T. M., Hu, Y., Theisen, J. C., & Leite, C. C. M. (2021) · Journal of Social and Personal Relationships · doi:10.1177/02654075211050073 · Cited by 16 · see citing works
Wiltshire, C., Wanna, C., Stenson, A. F., Minton, S. T., Reda, M., Davie, W., Hinrichs, R. C., Winters, S., France, J. M., & Jovanović, T. (2022) · Behaviour Research and Therapy · doi:10.1016/j.brat.2022.104036 · Cited by 16 · see citing works
Paroxysmal Atrial Fibrillation in Horses: Pathophysiology, Diagnostics and Clinical Aspects
Kjeldsen, S. T., Nissen, S. D., Buhl, R., & Hopster-Iversen, C. C. S. (2022) · Animals · doi:10.3390/ani12060698 · Cited by 15 · see citing works
A template and tutorial for preregistering studies using passive smartphone measures
Langener, A. M., Siepe, B. S., Elsherif, M. M., Niemeijer, K., Andresen, P. K., Akre, S., Bringmann, L. F., Cohen, Z. D., Choukas, N. R., Drexl, K., et al. (2024) · Behavior Research Methods · doi:10.3758/s13428-024-02474-5 · Cited by 15 · see citing works
Self-citationNeudorfer, M., Kumar, D., Smeddinck, J. D., Kulnik, S. T., Niebauer, J., Treff, G., & Sareban, M. (2024) · Scandinavian Journal of Medicine and Science in Sports · doi:10.1111/sms.14756 · Cited by 15 · see citing works
Garcia, L., Ferguson, S. E., Facio, L., Schary, D. P., & Guenther, C. H. (2023) · Mental Health & Prevention · doi:10.1016/j.mhp.2023.200280 · Cited by 14 · see citing works
Dropping the E: The potential for integrating e-mental health in psychotherapy
Daele, T. V., Best, P., Bernaerts, S., Assche, E. V., & Witte, N. A. J. D. (2021) · Current Opinion in Psychology · doi:10.1016/j.copsyc.2021.02.007 · Cited by 14 · see citing works
Nelson, B. W., Harvie, H. M. K., Jain, B., Knight, E. L., Roos, L. E., & Giuliano, R. J. (2023) · Psychosomatic Medicine · doi:10.1097/psy.0000000000001178 · Cited by 13 · see citing works
Self-citationXing, Y., Zhong, S., Aronson, S. L., Rausa, F. M., Webster, D. E., Crouthamel, M., & Wang, L. (2024) · Digital Biomarkers · doi:10.1159/000536499 · Cited by 13 · see citing works
Peaks and troughs: Are heart rate cues more attractive to tourists?
Lv, X., Feng, P., Chen, Q., Huang, X., & Fu, X. (2024) · Tourism Management · doi:10.1016/j.tourman.2024.105098 · Cited by 13 · see citing works
Implicit and explicit measures of food emotions
de Wijk, R. A., & Noldus, L. P. J. J. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-821124-3.00006-5 · Cited by 13 · see citing works
Nulty, A. K., Chen, E. S., & Thompson, A. L. (2022) · Digital Health · doi:10.1177/20552076221084461 · Cited by 13 · see citing works
Moon, E., Yang, M. L., Seon, Q., & Mantere, O. (2021) · Current Psychiatry Reports · doi:10.1007/s11920-021-01291-3 · Cited by 13 · see citing works
Potter, L. N., Yap, J. R. T., Dempsey, W., Wetter, D. W., & Nahum-Shani, I. B. (2023) · Prevention Science · doi:10.1007/s11121-023-01495-4 · Cited by 12 · see citing works
Wearable Devices and Explainable Unsupervised Learning for COVID-19 Detection and Monitoring
Hasasneh, A., Hijazi, H., Talib, M. W. A., Afadar, Y., Nassif, A. B., & Nasir, Q. (2023) · Diagnostics · doi:10.3390/diagnostics13193071 · Cited by 12 · see citing works
- Authors
- Nelson, B. W., Low, C. A., Jacobson, N., Areán, P., Torous, J., & Allen, N. B.
- Affiliations
- University of Oregon (US)
- University of Pittsburgh (US)
- Dartmouth College (US)
- University of Washington (US)
- Beth Israel Deaconess Medical Center (US)
- Corresponding author
- Nelson, Benjamin W.
- Venue
- npj Digital Medicine
- Published
- 2020-06-26
- Volume
- 3
- Issue
- 1
- Pages
- 90-90
- ISSN
- 2398-6352 (ISSN-L)
- PMC
- PMC7320189
- OpenAlex
- W3037389028
- Semantic Scholar
- 5f08e91b734b990fbd995109628c651537b5b8c5
- Cited by
- 241
- References
- 78
- Citations by year
- 2026: 22 · 2025: 30 · 2024: 54 · 2023: 46 · 2022: 39 · 2021: 42 · 2020: 7
- FWCI
- 2.80
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Locations
- npj Digital Medicine (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Minerva Access (University of Melbourne) (submittedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- National Cancer Institute (K07CA204380)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Health Sciences › Medicine › Physiology › Physical Activity and Health
- Keywords
- heart rate measurement, consumer wearable devices, smartwatches, wrist-worn sensors, measurement accuracy, device standardization, reporting guidelines, skin tone, sampling frequency, reproducibility
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review