Detection of Malignant Hepatic Tumors with Ferumoxides-Enhanced MRI: Comparison of Five Gradient-Recalled Echo Sequences with Different TEs
Open accessAbstract
OBJECTIVE: The purpose of our study was to compare the detectability of malignant hepatic tumors on ferumoxides-enhanced MRI using five gradient-recalled echo sequences at different TEs. MATERIALS AND METHODS: Ferumoxides-enhanced MRIs obtained in 31 patients with 50 malignant hepatic tumors (33 hepatocellular carcinomas, 17 metastases) were reviewed retrospectively by three independent offsite radiologists. T1-weighted gradient-recalled echo images with TEs of 1.4 and 4.2 msec; T2*-weighted gradient-recalled echo images with TEs of 6, 8, and 10 msec; and T2-weighted fast spin-echo images of livers were randomly reviewed on a segment-by-segment basis. Observer performance was tested using the McNemar test and receiver operating characteristic analysis for the clustered data. Lesion-to-liver contrast-to-noise ratio was also assessed. RESULTS: Mean lesion-to-liver contrast-to-noise ratios were negative and lower with gradient-recalled echo at 1.4 msec than with the other sequences. Sensitivity was higher (p < 0.05) with gradient-recalled echo at 6, 8, and 10 msec and fast spin-echo sequences (75-83%) than with gradient-recalled echo sequences at 1.4 and 4.2 msec (46-48%), and was higher (p < 0.05) with gradient-recalled echo sequence at 8 msec (83%) than with gradient-recalled echo at 6 msec and fast spin-echo sequences (75-78%). Specificity was comparably high with all sequences (95-98%). The area under the receiver operating characteristic curve (A(z)) was greater (p < 0.05) with gradient-recalled echo at 6, 8, and 10 msec and fast spin-echo sequences (A(z) = 0.91-0.93) than with gradient-recalled echo sequences at 1.4 and 4.2 msec (A(z) = 0.82-0.85). CONCLUSION: In the detection of malignant hepatic tumors, gradient-recalled echo sequences at 8 msec showed the highest sensitivity and had an A(z) value and lesion-to-liver contrast-to-noise ratio comparable with values from gradient-recalled echo sequences at 6 and 10 msec and fast spin-echo sequences.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Gadolinium-based contrast agents show positive contrast enhancement on T1-weighted images, which is clinically generally preferred over negative T2 contrast enhancement mediated by SPION.
“Gadolinium-based contrast agents show positive contrast enhancement on T1-weighted images, which is clinically generally preferred over negative T2 contrast enhancement mediated for instance by SPION [150].”
From Iron oxide nanoparticles: Diagnostic, therapeutic and theranostic applications (2019) · cites this paper as [150]
Cites
The Design and Analysis of Clinical Experiments
Fleiss, J. L. (1999) · Wiley eBooks · doi:10.1002/9781118032923 · Cited by 4,819 · see citing works
Statistical Methods for Rates and Proportions, 2nd ed
GROVE, W. A. (1981) · American Journal of Psychiatry · doi:10.1176/ajp.138.12.1644-a · Cited by 3,756 · see citing works
The measurement of interrater agreement
Fleiss, J. (1981) · Medical Entomology and Zoology · Cited by 686 · see citing works
Senéterre, E., Taourel, P., Bouvier, Y., Pradel, J., BEERS, B. E. V., Daurès, J., Pringot, J., Mathieu, D. G., & Bruel, J. M. (1996) · Radiology · doi:10.1148/radiology.200.3.8756932 · Cited by 310 · see citing works
Ros, P. R., Freeny, P. C., Harms, S. E., Seltzer, S. E., Davis, P. L., Chan, T., Stillman, A. E., Muroff, L. R., Runge, V. M., & Nissenbaum, M. A. (1995) · Radiology · doi:10.1148/radiology.196.2.7617864 · Cited by 277 · see citing works
Hagspiel, K. D., Neidl, K., Eichenberger, A. C., Weder, W., & Marinček, B. (1995) · Radiology · doi:10.1148/radiology.196.2.7617863 · Cited by 250 · see citing works
Reimer, P., Jähnke, N., Fiebich, M., Schima, W., Deckers, F., Marx, C., Holzknecht, N., & Saini, S. (2000) · Radiology · doi:10.1148/radiology.217.1.r00oc31152 · Cited by 188 · see citing works
On the comparison of correlated proportions for clustered data
Obuchowski, N. A. (1998) · Statistics in Medicine · doi:10.1002/(sici)1097-0258(19980715)17:13 · Cited by 167 · see citing works
Tang, Y., Yamashita, Y., Arakawa, A., et al. (1999) · American Journal of Roentgenology · doi:10.2214/ajr.172.6.10350287 · Cited by 151 · see citing works
Liver metastases: safety and efficacy of detection with superparamagnetic iron oxide in MR imaging.
Bellin, M., Zaïm, S., Auberton, E., Sarfati, G., Duron, J. J., Khayat, D., & Grellet, J. (1994) · Radiology · doi:10.1148/radiology.193.3.7972804 · Cited by 145 · see citing works
Larson, R. E., Semelka, R. C., Bagley, A. S., et al. (1994) · Radiology · doi:10.1148/radiology.192.2.8029403 · Cited by 144 · see citing works
Peterson, M. S., Baron, R. L., & Murakami, T. (1996) · Radiology · doi:10.1148/radiology.201.2.8888220 · Cited by 140 · see citing works
Self-citationYamamoto, H., Yamashita, Y., Yoshimatsu, S., et al. (1995) · Radiology · doi:10.1148/radiology.195.1.7892448 · Cited by 118 · see citing works
Hepatic lesions: detection with ferumoxide-enhanced T1-weighted MR imaging.
Oudkerk, M., van den Heuvel, A. G., Wielopolski, P. A., et al. (1997) · Radiology · doi:10.1148/radiology.203.2.9114103 · Cited by 114 · see citing works
Yamashita, Y., Hatanaka, Y., Yamamoto, H., et al. (1994) · Radiology · doi:10.1148/radiology.193.1.8090922 · Cited by 103 · see citing works
Winter, T. C., Freeny, P. C., Nghiem, H. V., Mack, L. A., Patten, R. M., Thomas, C., & Elliott, S. (1993) · American Journal of Roentgenology · doi:10.2214/ajr.161.6.8249724 · Cited by 81 · see citing works
Fretz, C., Elizondo, G., Weissleder, R., Hahn, P. F., Stark, D. D., & Ferrucci, J. T. (1989) · Radiology · doi:10.1148/radiology.172.2.2748820 · Cited by 78 · see citing works
Grangier, C., Tourniaire, J., Mentha, G., Schiau, R., Howart, N., Chachuat, A., Grossholz, M., & Terrier, F. (1994) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-199411000-00008 · Cited by 75 · see citing works
Reimer, P., Müller, M., Marx, C., et al. (1998) · Radiology · doi:10.1148/radiology.209.3.9844683 · Cited by 68 · see citing works
Ward, J., Chen, F., Guthrie, J. A., Wilson, D. J., Lodge, J. P. A., Wyatt, J. I., & Robinson, P. J. (2000) · Radiology · doi:10.1148/radiology.214.1.r00ja21159 · Cited by 68 · see citing works
Detection of Malignant Hepatic Tumors
Matsuo, M., Kanematsu, M., Itoh, K., Ito, K., Maetani, Y. S., Kondo, H., Kako, N., Matsunaga, N., Hoshi, H., & Shiraishi, J. (2001) · American Journal of Roentgenology · doi:10.2214/ajr.177.3.1770637 · Cited by 66 · see citing works
Self-citationMergo, P. J., Helmberger, T. K., Nicolas, A. I., et al. (1996) · American Journal of Roentgenology · doi:10.2214/ajr.166.2.8553952 · Cited by 57 · see citing works
Schwartz, L. H., Seltzer, S. E., Tempany, C. M., et al. (1995) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.1880050516 · Cited by 44 · see citing works
Superparamagnetic iron oxide: Detection of focal liver lesions at high‐field‐strength MR imaging
Duda, S. H., Laniado, M., Kopp, A. F., et al. (1994) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.1880040314 · Cited by 40 · see citing works
BEERS, B. E. V., Lacrosse, M., Jamart, J., et al. (1997) · American Journal of Roentgenology · doi:10.2214/ajr.168.3.9057521 · Cited by 39 · see citing works
Kuwatsuru, R., Brasch, R. C., Mühler, A., et al. (1997) · Radiology · doi:10.1148/radiology.202.1.8988202 · Cited by 33 · see citing works
Scharf, J., Hoffmann, V., Lehnert, T., et al. (1998) · Radiology · doi:10.1148/radiology.207.1.9530300 · Cited by 19 · see citing works
Bonk, R. T., Schmiedl, U. P., Yuan, C., Nelson, J. A., BLACK, C. D. V., & Ladd, D. L. (2000) · Journal of Magnetic Resonance Imaging · doi:10.1002/1522-2586(200006)11:6 · Cited by 10 · see citing works
Tanimoto, A., Wakabayashi, G., Shinmoto, H., et al. (2005) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.20263 · Cited by 10 · see citing works
Larkman (2001) · doi:10.1002/1522-2586(200102)13:2
Flagged first, then by citations
Cited by
Sorted by citations
Iron oxide nanoparticles: Diagnostic, therapeutic and theranostic applications
Dadfar, S. M., Roemhild, K., Drude, N. I., von Stillfried, S., Knüchel, R., Kiessling, F., & Lammers, T. (2019) · Advanced Drug Delivery Reviews · doi:10.1016/j.addr.2019.01.005 · Cited by 1,181 · see citing works
- Authors
- Matsuo, M., Kanematsu, M., Itoh, K., Murakami, T., Maetani, Y., Kondo, H., Goshima, S., Kako, N., Hoshi, H., Konishi, J., Moriyama, N., & Nakamura, H.
- Affiliations
- Gifu University Hospital (JP)
- Kyoto University (JP)
- The University of Osaka (JP)
- National Cancer Center Hospital (JP)
- Venue
- American Journal of Roentgenology
- Publisher
- American Roentgen Ray Society
- Published
- 2004-01
- Volume
- 182
- Issue
- 1
- Pages
- 235-242
- ISSN
- 0002-9580 (ISSN-L)
- 0361-803X
- 1546-3141
- OpenAlex
- W2160395888
- Cited by
- 17
- References
- 30
- Citations by year
- 2026: 1 · 2019: 1 · 2017: 1 · 2014: 1
- FWCI
- 1.69
- Type
- journal article
- Language
- en
- Open access
- green
- pure.teikyo.jp
- Licence
- cc-by-nc-nd
- Locations
- American Journal of Roentgenology (publishedVersion)
- PubMed (publishedVersion)
- Institutional Repositories DataBase (IRDB) (submittedVersion, cc-by-nc-nd)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › MRI in cancer diagnosis
- Health Sciences › Medicine › Hepatology › Hepatocellular Carcinoma Treatment and Prognosis
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
- Keywords
- ferumoxides, hepatocellular carcinoma, liver metastasis, gradient echo, echo time, fast spin echo, T2-weighted imaging, T1-weighted imaging, tumor detection sensitivity, MRI sequence comparison, liver MRI, diagnostic accuracy, observer performance
- MeSH terms
- Aged, Aged, 80 and over, Contrast Media, Dextrans, Digestive System Neoplasms, Female, Carcinoma, Hepatocellular, Humans, Iron, Liver Neoplasms, Male, Middle Aged, Oxides, Retrospective Studies, Sensitivity and Specificity, Time Factors, Reproducibility of Results, Echo-Planar Imaging, Ferrosoferric Oxide, Magnetite Nanoparticles
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review