CoChat

Transcutaneous glucose measurement using near-infrared spectroscopy during hypoglycemia.

Gabriely, I, Wozniak, R, Mevorach, M, Kaplan, J, Aharon, Y, Shamoon, H · 1999 · Diabetes Care
Cited by 91

The DOI matches its Crossref record.

Abstract

OBJECTIVE: To analyze a transcutaneous near-infrared spectroscopy system as a technique for in vivo noninvasive blood glucose monitoring during euglycemia and hypoglycemia. RESEARCH DESIGN AND METHODS: Ten nondiabetic subjects and two patients with type 1 diabetes were examined in a total of 27 studies. In each study, the subject's plasma glucose was lowered to a hypoglycemia level (approximately 55 mg/dl) followed by recovery to a glycemic level of approximately 115 mg/dl using an intravenous infusion of insulin and 20% dextrose. Plasma glucose levels were determined at 5-min intervals by standard glucose oxidase method and simultaneously by a near-infrared spectroscopic system. The plasma glucose measured by the standard method was used to create a calibration model that could predict glucose levels from the near-infrared spectral data. The two data sets were correlated during the decline and recovery in plasma glucose, within 10 mg/dl plasma glucose ranges, and were examined using the Clarke Error Grid Analysis. RESULTS: Two sets of 1,704 plasma glucose determinations were examined. The near-infrared predictions during the fall and recovery in plasma glucose were highly correlated (r = 0.96 and 0.95, respectively). When analyzed during 10 mg/dl plasma glucose segments, the mean absolute difference between the near-infrared spectroscopy method and the chemometric reference ranged from 3.3 to 4.4 mg/dl in the nondiabetic subjects and from 2.6 to 3.8 mg/dl in the patients with type 1 diabetes. Using the Error Grid Analysis, 97.7% of all the near-infrared predictions were assigned to the A-zone. CONCLUSIONS: Our findings suggest that the near-infrared spectroscopy method can accurately predict plasma glucose levels during euglycemia and hypoglycemia in humans.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.2337/diacare.22.12.2026

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 91

Metadata

Authors
Gabriely, I., Wozniak, R., Mevorach, M., Kaplan, J., Aharon, Y., & Shamoon, H.
Venue
Diabetes Care
Published
1999-12-01
Volume
22
Issue
12
Pages
2026-2032
ISSN
  • 0149-5992 (ISSN-L)
  • 1935-5548
OpenAlex
W2139939060
Cited by
91
References
0
FWCI
4.25
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Transcutaneous glucose measurement using near-infrared spectroscopy during hypoglycemia. https://app.cochat.ai/papers/10.2337/diacare.22.12.2026
The paper
Gabriely, I., Wozniak, R., Mevorach, M., Kaplan, J., Aharon, Y., & Shamoon, H. (1999). Transcutaneous glucose measurement using near-infrared spectroscopy during hypoglycemia. Diabetes Care. https://doi.org/10.2337/diacare.22.12.2026

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.