Glucose Biosensors: An Overview of Use in Clinical Practice

Open access
Yoo, Eun-Hyung, Lee, Soo-Youn·2010·Sensors

Abstract

Blood glucose monitoring has been established as a valuable tool in the management of diabetes. Since maintaining normal blood glucose levels is recommended, a series of suitable glucose biosensors have been developed. During the last 50 years, glucose biosensor technology including point-of-care devices, continuous glucose monitoring systems and noninvasive glucose monitoring systems has been significantly improved. However, there continues to be several challenges related to the achievement of accurate and reliable glucose monitoring. Further technical improvements in glucose biosensors, standardization of the analytical goals for their performance, and continuously assessing and training lay users are required. This article reviews the brief history, basic principles, analytical performance, and the present status of glucose biosensors in the clinical practice.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.3390/s100504558

Cites

7 of 7 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • The ADA in 1994 recommended a threshold <10% maximum permissible bias from reference methods for glucose concentrations between 30 and 396 mg/dL, reduced to <5% in 1996.

    “In 1994, the ADA made the first recommendations for the analytical performance of glucose biosensors available, suggesting a threshold <10 % of maximum permissible bias from reference methods for glucose concentrations between 30 and 396 mg/dL. This analytical target was further reduced to <5% in 1996 52.”

    From Glucose biosensors in clinical practice: principles, limits and perspectives of currently used devices (2022) · cites this paper as [52]

    The Analytical Performance Validation section reports the ADA’s 1994 <10% maximum allowable bias recommendation and its reduction to <5% in 1996. It gives the glucose range as 1.6–22.2 mmol/L, approximately 29–400 mg/dL, rather than the claimed 30–396 mg/dL. (LLM, conf=0.84)

    Experimental — accuracy not yet measured
  • Many variables, including hematocrit, hypoxemia, hypotension, temperature, altitude, and humidity may influence the reliability of glucose measurements.

    “Many variables, including hematocrit, hypoxemia, hypotension, temperature, altitude, and humidity may influence the reliability of glucose measurements 52,100.”

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Research works printed but not linked: 102

  1. [1]Cowie C.C. Rust K.F. Byrd-Holt D.D. Gregg E.W. Ford E.S. Geiss L.S. Bainbridge K.E. Fradkin J.E. Prevalence of diabetes and high risk for diabetes using hemoglobin A1c criteria in the U.S. population in 1988–2006 Diabetes Care 2010 33 562 568 20067953 PMC2827508
  2. [2]Narayan K.M. Boyle J.P. Geiss L.S. Saaddine J.B. Thompson T.J. Impact of recent increase in incidence on future diabetes burden: U.S., 2005–2050 Diabetes Care 2006 29 2114 2116 16936162
  3. [3]Wild S. Roglic G. Green A. Sicree R. King H. Global prevalence of diabetes: estimates for the year 2000 and projections for 2030 Diabetes Care 2004 27 1047 1053 15111519
  4. [4]Shaw J.E. Sicree R.A. Zimmet P.Z. Global estimates of the prevalence of diabetes for 2010 and 2030 Diabetes Res. Clin. Pract 2010 87 4 14 19896746
  5. [5]American Diabetes Association Diagnosis and classification of diabetes mellitus Diabetes Care 2010 33 S62 69 20042775 PMC2797383
  6. [7]Murata G.H. Shah J.H. Hoffman R.M. Wendel C.S. Adam K.D. Solvas P.A. Bokhari S.U. Duckworth W.C. Intensified blood glucose monitoring improves glycemic control in stable, insulin-treated veterans with type 2 diabetes: the Diabetes Outcomes in Veterans Study (DOVES) Diabetes Care 2003 26 1759 1763 12766106
  7. [8]Skeie S. Kristensen G.B. Carlsen S. Sandberg S. Self-Monitoring of Blood Glucose in Type 1 Diabetes Patients with Insufficient Metabolic Control: Focused Self-Monitoring of Blood Glucose Intervention Can Lower Glycated Hemoglobin A1C J. Diabetes Sci. Technol 2009 3 83 88 20046652 PMC2769849
  8. [9]Tunis S.L. Minshall M.E. Self-monitoring of blood glucose (SMBG) for type 2 diabetes patients treated with oral anti-diabetes drugs and with a recent history of monitoring: cost-effectiveness in the US Curr. Med. Res. Opin 2010 26 151 162 19919376
  9. [10]Boutati E.I. Raptis S.A. Self-monitoring of blood glucose as part of the integral care of type 2 diabetes Diabetes Care 2009 32 Suppl. 2 S205 210 19875553 PMC2811483
  10. [11]Jovanovic L.G. Using meal-based self-monitoring of blood glucose as a tool to improve outcomes in pregnancy complicated by diabetes Endocr. Pract 2008 14 239 247 18308666
  11. [12]O’Kane M.J. Pickup J. Self-monitoring of blood glucose in diabetes: is it worth it? Ann. Clin. Biochem 2009 46 273 282 19454538
  12. [13]The Diabetes Control and Complications Trial Research Group The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus N. Engl. J. Med 1993 329 977 986 8366922
  13. [14]Holman R.R. Paul S.K. Bethel M.A. Matthews D.R. Neil H.A. 10-year follow-up of intensive glucose control in type 2 diabetes N. Engl. J. Med 2008 359 1577 1589 18784090
  14. [16]American Diabetes Association Standards of medical care in diabetes--2010 Diabetes Care 2010 33 S11 61 20042772 PMC2797382
  15. [17]Newman J.D. Turner A.P. Home blood glucose biosensors: a commercial perspective Biosens. Bioelectron 2005 20 2435 2453 15854818
  16. [18]Turner A.P. Biosensors--sense and sensitivity Science 2000 290 1315 1317 11185408
  17. [19]Clark L.C. Jr. Lyons C. Electrode systems for continuous monitoring in cardiovascular surgery Ann. N. Y. Acad. Sci 1962 102 29 45 14021529
  18. [20]Updike S.J. Hicks G.P. The enzyme electrode Nature 1967 214 986 988 6055414
  19. [21]Hiratsuka A. Fujisawa K. Muguruma H. Amperometric biosensor based on glucose dehydrogenase and plasma-polymerized thin films Anal. Sci 2008 24 483 486 18403839
  20. [22]Chambers J.P. Arulanandam B.P. Matta L.L. Weis A. Valdes J.J. Biosensor recognition elements Curr. Issues Mol. Biol 2008 10 1 12 18525101
  21. [23]Iqbal S.S. Mayo M.W. Bruno J.G. Bronk B.V. Batt C.A. Chambers J.P. A review of molecular recognition technologies for detection of biological threat agents Biosens. Bioelectron 2000 15 549 578 11213217
  22. [24]Newman J.D. Turner A.P. Biosensors: Principles and practice Portland Press London, UK 1992 27 147 159 1425600
  23. [25]Habermuller K. Mosbach M. Schuhmann W. Electron-transfer mechanisms in amperometric biosensors Fresenius J. Anal. Chem 2000 366 560 568 11225768
  24. [26]Pearson J.E. Gill A. Vadgama P. Analytical aspects of biosensors Ann. Clin. Biochem 2000 37 Pt 2 119 145 10735356
  25. [27]Thevenot D.R. Toth K. Durst R.A. Wilson G.S. Electrochemical biosensors: recommended definitions and classification Biosens. Bioelectron 2001 16 121 131 11261847
  26. [28]Turner A.P. Chen B. Piletsky S.A. In vitro diagnostics in diabetes: meeting the challenge Clin. Chem 1999 45 1596 1601 10471674
  27. [29]Heller A. Amperometric biosensors Curr. Opin. Biotechnol 1996 7 50 54 8742376
  28. [30]Price C.P. Point-of-care testing in diabetes mellitus Clin. Chem. Lab. Med 2003 41 1213 1219 14598871
  29. [31]D’Costa E.J. Higgins I.J. Turner A.P. Quinoprotein glucose dehydrogenase and its application in an amperometric glucose sensor Biosensors 1986 2 71 87 3454651
  30. [32]Slein M.W. D-glucose: Determination with hexokinase and glucose-6-phosphate dehydrogenase Academic Press New York, NY, USA 1963 117
  31. [34]Bankar S.B. Bule M.V. Singhal R.S. Ananthanarayan L. Glucose oxidase--an overview Biotechnol. Adv 2009 27 489 501 19374943
  32. [35]Weibel M.K. Bright H.J. The glucose oxidase mechanism. Interpretation of the pH dependence J. Biol. Chem 1971 246 2734 2744 4324339
  33. [36]Guilbault G.G. Lubrano G.J. An enzyme electrode for the amperometric determination of glucose Anal. Chim. Acta 1973 64 439 455 4701057
  34. [37]Jin W. Wollenberger U. Scheller F.W. PQQ as redox shuttle for quinoprotein glucose dehydrogenase Biol. Chem 1998 379 1207 1211 9792456
  35. [38]Zayats M. Katz E. Baron R. Willner I. Reconstitution of apo-glucose dehydrogenase on pyrroloquinoline quinone-functionalized au nanoparticles yields an electrically contacted biocatalyst J. Am. Chem. Soc 2005 127 12400 12406 16131222
  36. [39]Raitman O.A. Patolsky F. Katz E. Willner I. Electrical contacting of glucose dehydrogenase by the reconstitution of a pyrroloquinoline quinone-functionalized polyaniline film associated with an Au-electrode: an in situ electrochemical SPR study Chem. Commun. (Camb.) 2002 1936 1937 12271682
  37. [40]Bartlett P.N. Whitaker R.G. Strategies for the development of amperometric enzyme electrodes Biosensors 1987 3 359 379 3333622
  38. [41]Bartlett P.N. Simon E. Toh C.S. Modified electrodes for NADH oxidation and dehydrogenase-based biosensors Bioelectrochemistry 2002 56 117 122 12009456
  39. [42]Gorton L. Dominguez E. Electrocatalytic oxidation of NAD(P) H at mediator-modified electrodes J. Biotechnol 2002 82 371 392 11996217
  40. [43]Drury M.I. Timoney F.J. Delaney P. DEXTROSTIX--A RAPID METHOD OF ESTIMATING BLOOD GLUCOSE LEVELS J. Ir. Med. Assoc 1965 56 52 53 14269585
  41. [44]Jensen M.S. Clinical tests with the dextrostix. A new method for rapid blood sugar determination Ugeskr. Laeger 1965 127 709 712 5869863
  42. [45]Korp W. A NEW RAPID BLOOD-SUGAR DETERMINATION AT BEDSIDE (DEXTROSTIX) Wien. Med. Wochenschr 1965 115 435 437 14304577
  43. [46]Schmidt V. Blood sugar determination using the dextrostix Ugesk. Laeger 1965 127 706 709 5869862
  44. [47]Updike S.J. Hicks G.P. Reagentless substrate analysis with immobilized enzymes Science 1967 158 270 272 6053885
  45. [48]Wang J. Electrochemical glucose biosensors Chem. Rev 2008 108 814 825 18154363
  46. [50]Cass A.E. Davis G. Francis G.D. Hill H.A. Aston W.J. Higgins I.J. Plotkin E.V. Scott L.D. Turner A.P. Ferrocene-mediated enzyme electrode for amperometric determination of glucose Anal. Chem 1984 56 667 671 6721151
  47. [51]Frew J.E. Hill H.A. Electrochemical biosensors Anal. Chem 1987 59 933A 944A 3631522
  48. [52]Shichiri M. Kawamori R. Yamasaki Y. Hakui N. Abe H. Wearable artificial endocrine pancrease with needle-type glucose sensor Lancet 1982 2 1129 1131 6128452
  49. [53]Chaubey A. Malhotra B.D. Mediated biosensors Biosens. Bioelectron 2002 17 441 456 11959464
  50. [54]Frew J.E. Hill H.A. Electron-transfer biosensors Philos. Trans. R. Soc. Lond. B. Biol. Sci 1987 316 95 106 2889234
  51. [55]Ballarin B. Cassani M.C. Mazzoni R. Scavetta E. Tonelli D. Enzyme electrodes based on sono-gel containing ferrocenyl compounds Biosens. Bioelectron 2007 22 1317 1322 16846733
  52. [56]Di Gleria K. Hill H.A. McNeil C.J. Green M.J. Homogeneous ferrocene-mediated amperometric immunoassay Anal. Chem 1986 58 1203 1205 3717577
  53. [57]Williams D.L. Doig A.R. Jr. Korosi A. Electrochemical-enzymatic analysis of blood glucose and lactate Anal. Chem 1970 42 118 121 5409504
  54. [58]Hu J. The evolution of commercialized glucose sensors in China Biosens. Bioelectron 2009 24 1083 1089 18929476
  55. [59]Hilditch P.I. Green M.J. Disposable electrochemical biosensors Analyst 1991 116 1217 1220 1816740
  56. [60]Matthews D.R. Holman R.R. Bown E. Steemson J. Watson A. Hughes S. Scott D. Pen-sized digital 30-second blood glucose meter Lancet 1987 1 778 779 2882186
  57. [61]Murray R.W. Ewing A.G. Durst R.A. Chemically modified electrodes. Molecular design for electroanalysis Anal. Chem 1987 59 379A 390A 3565770
  58. [62]Zhang W. Li G. Third-generation biosensors based on the direct electron transfer of proteins Anal. Sci 2004 20 603 609 15116955
  59. [63]Gregg B.A. Heller A. Cross-linked redox gels containing glucose oxidase for amperometric biosensor applications Anal. Chem 1990 62 258 263 2305956
  60. [64]Lin Y. Yantasee W. Wang J. Carbon nanotubes (CNTs) for the development of electrochemical biosensors Front. Biosci 2005 10 492 505 15574386
  61. [65]Riklin A. Katz E. Willner I. Stocker A. Buckmann A.F. Improving enzyme-electrode contacts by redox modification of cofactors Nature 1995 376 672 675 7651516
  62. [66]Khan G.F. Ohwa M. Wernet W. Design of a stable charge transfer complex electrode for a third-generation amperometric glucose sensor Anal. Chem 1996 68 2939 2945 8794929
  63. [67]Palmisano F. Zambonin P.G. Centonze D. Quinto M. A disposable, reagentless, third-generation glucose biosensor based on overoxidized poly(pyrrole)/tetrathiafulvalene-tetracyanoquinodimethane composite Anal. Chem 2002 74 5913 5918 12498183
  64. [68]Azevedo A.M. Martins V.C. Prazeres D.M. Vojinovic V. Cabral J.M. Fonseca L.P. Horseradish peroxidase: a valuable tool in biotechnology Biotechnol. Annu. Rev 2003 9 199 247 14650928
  65. [69]Rubio Retama J. Lopez Cabarcos E. Mecerreyes D. Lopez-Ruiz B. Design of an amperometric biosensor using polypyrrole-microgel composites containing glucose oxidase Biosens. Bioelectron 2004 20 1111 1117 15556356
  66. [70]Vidal J.C. Garcia E. Castillo J.R. Electropolymerization of pyrrole and immobilization of glucose oxidase in a flow system: influence of the operating conditions on analytical performance Biosens. Bioelectron 1998 13 371 382 9642772
  67. [71]Wu J. Qu Y. Mediator-free amperometric determination of glucose based on direct electron transfer between glucose oxidase and an oxidized boron-doped diamond electrode Anal. Bioanal. Chem 2006 385 1330 1335 16773297
  68. [72]Albisser A.M. Leibel B.S. Ewart T.G. Davidovac Z. Botz C.K. Zingg W. Schipper H. Gander R. Clinical control of diabetes by the artificial pancreas Diabetes 1974 23 397 404 4598090
  69. [74]Csoregi E. Schmidtke D.W. Heller A. Design and optimization of a selective subcutaneously implantable glucose electrode based on “wired” glucose oxidase Anal. Chem 1995 67 1240 1244 7733469
  70. [75]Henry C. Getting under the skin: implantable glucose sensors Anal. Chem 1998 70 594A 598A 9737202
  71. [76]Schmidtke D.W. Freeland A.C. Heller A. Bonnecaze R.T. Measurement and modeling of the transient difference between blood and subcutaneous glucose concentrations in the rat after injection of insulin Proc. Natl. Acad. Sci. USA 1998 95 294 299 9419369 PMC18205
  72. [77]Rebrin K. Steil G.M. Can interstitial glucose assessment replace blood glucose measurements? Diabetes Technol. Ther 2000 2 461 472 11467349
  73. [78]Gross T.M. Bode B.W. Einhorn D. Kayne D.M. Reed J.H. White N.H. Mastrototaro J.J. Performance evaluation of the MiniMed continuous glucose monitoring system during patient home use Diabetes Technol. Ther 2000 2 49 56 11467320
  74. [79]Cox M. An overview of continuous glucose monitoring systems J. Pediatr. Health Care 2009 23 344 347 19720271
  75. [80]Hashiguchi Y. Sakakida M. Nishida K. Uemura T. Kajiwara K. Shichiri M. Development of a miniaturized glucose monitoring system by combining a needle-type glucose sensor with microdialysis sampling method. Long-term subcutaneous tissue glucose monitoring in ambulatory diabetic patients Diabetes Care 1994 17 387 396 8062605
  76. [81]Poscia A. Mascini M. Moscone D. Luzzana M. Caramenti G. Cremonesi P. Valgimigli F. Bongiovanni C. Varalli M. A microdialysis technique for continuous subcutaneous glucose monitoring in diabetic patients (part 1) Biosens. Bioelectron 2003 18 891 898 12713912
  77. [82]Wentholt I.M. Vollebregt M.A. Hart A.A. Hoekstra J.B. DeVries J.H. Comparison of a needle-type and a microdialysis continuous glucose monitor in type 1 diabetic patients Diabetes Care 2005 28 2871 2876 16306547
  78. [83]Nielsen J.K. Freckmann G. Kapitza C. Ocvirk G. Koelker K.H. Kamecke U. Gillen R. Amann-Zalan I. Jendrike N. Christiansen J.S. Koschinsky T. Heinemann L. Glucose monitoring by microdialysis: performance in a multicentre study Diabet. Med 2009 26 714 721 19573121
  79. [85]Bode B. Beck R.W. Xing D. Gilliam L. Hirsch I. Kollman C. Laffel L. Ruedy K.J. Tamborlane W.V. Weinzimer S. Wolpert H. Sustained benefit of continuous glucose monitoring on A1C, glucose profiles, and hypoglycemia in adults with type 1 diabetes Diabetes Care 2009 32 2047 2049 19675193 PMC2768224
  80. [87]Montagnana M. Lippi G. Guidi G.C. Continuous glucose monitoring and type 1 diabetes N. Engl. J. Med 2009 360 190 author reply 191–192. 19129536
  81. [88]Klonoff D.C. Noninvasive blood glucose monitoring Diabetes Care 1997 20 433 437 9051401
  82. [89]Oliver N.S. Toumazou C. Cass A.E. Johnston D.G. Glucose sensors: a review of current and emerging technology Diabet. Med 2009 26 197 210 19317813
  83. [90]Rabinovitch B. March W.F. Adams R.L. Noninvasive glucose monitoring of the aqueous humor of the eye: Part I. Measurement of very small optical rotations Diabetes Care 1982 5 254 258 7172992
  84. [91]Goetz M.J. Jr. Cote G.L. Erckens R. March W. Motamedi M. Application of a multivariate technique to Raman spectra for quantification of body chemicals IEEE Trans. Biomed. Eng 1995 42 728 731 7622156
  85. [92]Gabriely I. Wozniak R. Mevorach M. Kaplan J. Aharon Y. Shamoon H. Transcutaneous glucose measurement using near-infrared spectroscopy during hypoglycemia Diabetes Care 1999 22 2026 2032 10587837
  86. [93]MacKenzie H.A. Ashton H.S. Spiers S. Shen Y. Freeborn S.S. Hannigan J. Lindberg J. Rae P. Advances in photoacoustic noninvasive glucose testing Clin. Chem 1999 45 1587 1595 10471673
  87. [94]Larin K.V. Eledrisi M.S. Motamedi M. Esenaliev R.O. Noninvasive blood glucose monitoring with optical coherence tomography: a pilot study in human subjects Diabetes Care 2002 25 2263 2267 12453971
  88. [95]D’Orazio P. Burnett R.W. Fogh-Andersen N. Jacobs E. Kuwa K. Kulpmann W.R. Larsson L. Lewenstam A. Maas A.H. Mager G. Naskalski J.W. Okorodudu A.O. Approved IFCC recommendation on reporting results for blood glucose: International Federation of Clinical Chemistry and Laboratory Medicine Scientific Division, Working Group on Selective Electrodes and Point-of-Care Testing (IFCC-SD-WG-SEPOCT) Clin. Chem. Lab. Med 2006 44 1486 1490 17163827
  89. [96]Montagnana M. Caputo M. Giavarina D. Lippi G. Overview on self-monitoring of blood glucose Clin. Chim. Acta 2009 402 7 13 19167374
  90. [97]Goldstein D.E. Little R.R. Lorenz R.A. Malone J.I. Nathan D. Peterson C.M. Sacks D.B. Tests of glycemia in diabetes Diabetes Care 2004 27 1761 1773 15220264
  91. [98]American Diabetes Association American Diabetes Association: clinical practice recommendations 1996 Diabetes Care 1996 19 S1 118 8689936
  92. [102]Heinemann L. Measurement quality of blood glucose meters: is there a need for an institution with an unbiased view? J. Diabetes Sci. Technol 2007 1 178 180 19888404 PMC2771470
  93. [103]Barreau P.B. Buttery J.E. Effect of hematocrit concentration on blood glucose value determined on Glucometer II Diabetes Care 1988 11 116 118 3383731
  94. [104]Rao L.V. Jakubiak F. Sidwell J.S. Winkelman J.W. Snyder M.L. Accuracy evaluation of a new glucometer with automated hematocrit measurement and correction Clin. Chim. Acta 2005 356 178 183 15936315
  95. [108]Schleis T.G. Interference of maltose, icodextrin, galactose, or xylose with some blood glucose monitoring systems Pharmacotherapy 2007 27 1313 1321 17723085
  96. [109]Flore K.M. Delanghe J.R. Analytical interferences in point-of-care testing glucometers by icodextrin and its metabolites: an overview Perit. Dial. Int 2009 29 377 383 19602602
  97. [110]Janssen W. Harff G. Caers M. Schellekens A. Positive interference of icodextrin metabolites in some enzymatic glucose methods Clin. Chem 1998 44 2379 2380 9799776
  98. [111]Oyibo S.O. Pritchard G.M. McLay L. James E. Laing I. Gokal R. Boulton A.J. Blood glucose overestimation in diabetic patients on continuous ambulatory peritoneal dialysis for end-stage renal disease Diabet. Med 2002 19 693 696 12147153
  99. [112]Tang Z. Du X. Louie R.F. Kost G.J. Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyzer Am. J. Clin. Pathol 2000 113 75 86 10631860
  100. [113]Cartier L.J. Leclerc P. Pouliot M. Nadeau L. Turcotte G. Fruteau-de-Laclos B. Toxic levels of acetaminophen produce a major positive interference on Glucometer Elite and Accu-chek Advantage glucose meters Clin. Chem 1998 44 893 894 9554511
  101. [114]Solnica B. Naskalski J.W. Sieradzki J. Analytical performance of glucometers used for routine glucose self-monitoring of diabetic patients Clin. Chim. Acta 2003 331 29 35 12691861
  102. [115]Vesper H.W. Myers G.L. Approaches for improving glucose monitor measurements for self-monitoring of blood glucose: from measurement harmonization to external quality assessment programs J. Diabetes Sci. Technol 2007 1 153 157 19888399 PMC2771459

Cited by

Sorted by citations

Authors
Yoo, E.-H., & Lee, S.-Y.
Affiliations
  • Konyang University (KR)
  • Samsung Medical Center (KR)
Corresponding author
Lee, Soo-Youn
Venue
Sensors
Publisher
MDPI AG
Published
2010-05-04
Volume
10
Issue
5
Pages
4558-4576
ISSN
1424-8220 (ISSN-L)
OpenAlex
W2018584517
Cited by
1,137
References
7
Citations by year
2026: 50 · 2025: 84 · 2024: 96 · 2023: 102 · 2022: 126 · 2021: 106 · 2020: 86 · 2019: 88 · 2018: 83 · 2017: 54 · 2016: 68 · 2015: 44 · 2014: 41 · 2013: 47 · 2012: 31
FWCI
4.92
Type
journal article
Language
en
Open access
Locations
  • Sensors (publishedVersion, cc-by)
  • Europe PMC (PubMed Central) (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
Topics
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
  • Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrochemical sensors and biosensors
  • Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
Keywords
glucose biosensor, blood glucose monitoring, diabetes management, point-of-care testing, continuous glucose monitoring, non-invasive glucose monitoring, analytical performance, clinical practice
MeSH terms
Chemistry, Clinical, Glucose, Humans, Blood Glucose Self-Monitoring, Biosensing Techniques, Point-of-Care Systems, History, 20th Century

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review