Continuous Subcutaneous Glucose Monitoring in Diabetic Patients
Abstract
OBJECTIVE—To evaluate the accuracy of a new subcutaneous glucose sensor (Glucoday; A. Menarini Diagnostics) compared with venous blood glucose measurement in type 1 and type 2 diabetic patients. RESEARCH DESIGN—A multicenter study was performed in 70 diabetic patients. A microdialysis fiber was inserted subcutaneously into the periumbelical region and perfused with a buffer solution. Glucose concentrations in the dialysate were then measured every 3 min by the glucose sensor over a 24-h period, during which nine venous blood samples were also collected throughout the day. RESULTS—Both the insertion of the fiber and the wearing of the device were well tolerated by the patients. Subcutaneous glucose levels were well correlated with venous glucose measurements (r = 0.9, P < 0.001) over a wide range (40–400 mg/dl) for up to 24 h, with a single-point calibration. An analysis of 381 data pairs showed a linear relationship between the GlucoDay and serial venous blood glucose levels, and 97% of the data fell in the A and B regions of the error grid analysis. Percentage bias between the GlucoDay and the blood venous levels was −2.0% in the hypoglycemic range (<70 mg/dl), 6.9% in the euglycemic range (70–180 mg/dl), and 11.2% in the hyperglycemic range (>180 mg/dl). CONCLUSIONS—The GlucoDay system demonstrated high reliability and reported values that closely agreed with venous blood glucose measurements. The system was well tolerated and thus constitutes a relatively easy method to monitor glucose excursions in diabetic patients.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Electrochemical Glucose Biosensors
Wang, J. (2008) · Chemical Reviews · doi:10.1021/cr068123a · Cited by 3,498
Electrochemical Glucose Sensors and Their Applications in Diabetes Management
Heller, A., & Feldman, B. (2008) · Chemical Reviews · doi:10.1021/cr068069y · Cited by 1,707
Klonoff, D. C. (2005) · Diabetes Care · doi:10.2337/diacare.28.5.1231 · Cited by 724
Glucose sensors: a review of current and emerging technology
Oliver, N. S., Toumazou, C., Cass, A. E. G., & Johnston, D. G. (2009) · Diabetic Medicine · doi:10.1111/j.1464-5491.2008.02642.x · Cited by 584
Sparacino, G., Zanderigo, F., Corazza, S., Maran, A., Facchinetti, A., & Cobelli, C. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2006.889774 · Cited by 338
Self-citation
Metadata
- Authors
- Maran, A., Crepaldi, C., Tiengo, A., Grassi, G., Vitali, E., Pagano, G., Bistoni, S., Calabrese, G., Santeusanio, F., Leonetti, F., Ribaudo, M., Di Mario, U., Annuzzi, G., Genovese, S., Riccardi, G., Previti, M., Cucinotta, D., Giorgino, F., Bellomo, A., Giorgino, R., Poscia, A., & Varalli, M.
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 2002-02-01
- Volume
- 25
- Issue
- 2
- Pages
- 347-352
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- OpenAlex
- W2093168969
- Cited by
- 198
- References
- 0
- FWCI
- 9.42
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Continuous Subcutaneous Glucose Monitoring in Diabetic Patients. https://app.cochat.ai/papers/10.2337/diacare.25.2.347
- The paper
- Maran, A., Crepaldi, C., Tiengo, A., Grassi, G., Vitali, E., Pagano, G., Bistoni, S., Calabrese, G., Santeusanio, F., Leonetti, F., Ribaudo, M., Di Mario, U., Annuzzi, G., Genovese, S., Riccardi, G., Previti, M., Cucinotta, D., Giorgino, F., Bellomo, A., . . . Varalli, M. (2002). Continuous Subcutaneous Glucose Monitoring in Diabetic Patients. Diabetes Care. https://doi.org/10.2337/diacare.25.2.347
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.