An observational study to assess if automated diabetic retinopathy image assessment software can replace one or more steps of manual imaging grading and to determine their cost-effectiveness

Open access
Tufail, Adnan, Kapetanakis, Venediktos V, Salas-Vega, Sebastian, Egan, Catherine, Rudisill, Caroline, Owen, Christopher G, Lee, Aaron, Louw, Vern, Anderson, John, Liew, Gerald, Bolter, Louis, Bailey, Clare, Sadda, SriniVas, Taylor, Paul, Rudnicka, Alicja R · 2016 · Health Technology Assessment
Cited by 133

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Abstract

Background: Diabetic retinopathy screening in England involves labour-intensive manual grading of retinal images. Automated retinal image analysis systems (ARIASs) may offer an alternative to manual grading. Objectives: To determine the screening performance and cost-effectiveness of ARIASs to replace level 1 human graders or pre-screen with ARIASs in the NHS diabetic eye screening programme (DESP). To examine technical issues associated with implementation. Design: Observational retrospective measurement comparison study with a real-time evaluation of technical issues and a decision-analytic model to evaluate cost-effectiveness. Setting: A NHS DESP. Participants: Consecutive diabetic patients who attended a routine annual NHS DESP visit. Interventions: Retinal images were manually graded and processed by three ARIASs: iGradingM (version 1.1; originally Medalytix Group Ltd, Manchester, UK, but purchased by Digital Healthcare, Cambridge, UK, at the initiation of the study, purchased in turn by EMIS Health, Leeds, UK, after conclusion of the study), Retmarker (version 0.8.2, Retmarker Ltd, Coimbra, Portugal) and EyeArt (Eyenuk Inc., Woodland Hills, CA, USA). The final manual grade was used as the reference standard. Arbitration on a subset of discrepancies between manual grading and the use of an ARIAS by a reading centre masked to all grading was used to create a reference standard manual grade modified by arbitration. Main outcome measures: Screening performance (sensitivity, specificity, false-positive rate and likelihood ratios) and diagnostic accuracy [95% confidence intervals (CIs)] of ARIASs. A secondary analysis explored the influence of camera type and patients’ ethnicity, age and sex on screening performance. Economic analysis estimated the cost per appropriate screening outcome identified. Results: A total of 20,258 patients with 102,856 images were entered into the study. The sensitivity point estimates of the ARIASs were as follows: EyeArt 94.7% (95% CI 94.2% to 95.2%) for any retinopathy, 93.8% (95% CI 92.9% to 94.6%) for referable retinopathy and 99.6% (95% CI 97.0% to 99.9%) for proliferative retinopathy; and Retmarker 73.0% (95% CI 72.0% to 74.0%) for any retinopathy, 85.0% (95% CI 83.6% to 86.2%) for referable retinopathy and 97.9% (95% CI 94.9 to 99.1%) for proliferative retinopathy. iGradingM classified all images as either ‘disease’ or ‘ungradable’, limiting further iGradingM analysis. The sensitivity and false-positive rates for EyeArt were not affected by ethnicity, sex or camera type but sensitivity declined marginally with increasing patient age. The screening performance of Retmarker appeared to vary with patient’s age, ethnicity and camera type. Both EyeArt and Retmarker were cost saving relative to manual grading either as a replacement for level 1 human grading or used prior to level 1 human grading, although the latter was less cost-effective. A threshold analysis testing the highest ARIAS cost per patient before which ARIASs became more expensive per appropriate outcome than human grading, when used to replace level 1 grader, was Retmarker £3.82 and EyeArt £2.71 per patient. Limitations: The non-randomised study design limited the health economic analysis but the same retinal images were processed by all ARIASs in this measurement comparison study. Conclusions: Retmarker and EyeArt achieved acceptable sensitivity for referable retinopathy and false-positive rates (compared with human graders as reference standard) and appear to be cost-effective alternatives to a purely manual grading approach. Future work is required to develop technical specifications to optimise deployment and address potential governance issues. Funding: The National Institute for Health Research (NIHR) Health Technology Assessment programme, a Fight for Sight Grant (Hirsch grant award) and the Department of Health’s NIHR Biomedical Research Centre for Ophthalmology at Moorfields Eye Hospital and the University College London Institute of Ophthalmology.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.3310/hta20920

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 2 references

Cited by

See all 133

Metadata

Authors
Tufail, A., Kapetanakis, V. V., Salas-Vega, S., Egan, C., Rudisill, C., Owen, C. G., Lee, A., Louw, V., Anderson, J., Liew, G., Bolter, L., Bailey, C., Sadda, S., Taylor, P., & Rudnicka, A. R.
Venue
Health Technology Assessment
Published
2016-12
Volume
20
Issue
92
Pages
1-72
ISSN
  • 1366-5278 (ISSN-L)
  • 2046-4924
OpenAlex
W2564554913
Cited by
133
References
2
FWCI
5.52
Type
journal article
Language
en
Open access
diamond

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: An observational study to assess if automated diabetic retinopathy image assessment software can replace one or more steps of manual imaging grading and to determine their cost-effectiveness. https://app.cochat.ai/papers/10.3310/hta20920
The paper
Tufail, A., Kapetanakis, V. V., Salas-Vega, S., Egan, C., Rudisill, C., Owen, C. G., Lee, A., Louw, V., Anderson, J., Liew, G., Bolter, L., Bailey, C., Sadda, S., Taylor, P., & Rudnicka, A. R. (2016). An observational study to assess if automated diabetic retinopathy image assessment software can replace one or more steps of manual imaging grading and to determine their cost-effectiveness. Health Technology Assessment. https://doi.org/10.3310/hta20920

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.