Impact of Air Exposure Time on the Water Contact Angles of Daily Disposable Silicone Hydrogels
Abstract
The wettability of silicone hydrogel (SiHy) contact lens (CLs) is crucial for the pre-lens tear film stability throughout the day. Therefore, sessile drop and captive bubble setups were used to study the advancing and receding water contact angles (CA) of four SiHy materials: narafilcon A (TE), senofilcon A (AOD), stenfilcon A (MD), and delefilcon A (DT). TE and AOD have 48% and 38% water content, respectively, and no surface coating. MD (54% water) implements “smart chemistry” with just 4.4% bulk silicone content, while DT has >80% water at its surface. These SiHy were subjected to continuous blink-like air exposure (10 s)/rehydration (1s) cycles for 0, 1, 2, 3, 4, 6, 8, 10, 12, 14, and 16 h. The advancing CA, which measures the rehydration propensity of the CL surface, proved to be the most sensitive parameter to discriminate between the samples. The order of performance for the entire time scale was DT > MD >> AOD ≥ TE. The extended desiccation/rehydration cycling increased the differences between the CA of DT and MD compared to AOD and TE. This suggests that the low Si surface content and the high surface hydration are major determinants of SiHy wettability.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Hydrogels with a CA below 90° are considered hydrophilic [23].”
From Honey‐Based Sodium Alginate–Polyvinyl Alcohol Hydrogel Containing Green‐Synthesized Chitosan–Zinc Oxide Nanoparticles for Wound Healing (2025) · cites this paper as [23]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
No references could be resolved yet.
Research works printed but not linked: 43
- Guillon, 2013, Are silicone hydrogel contact lenses more comfortable than hydrogel contact lenses?, Eye Contact Lens, 39, 86, 10.1097/ICL.0b013e31827cb99f · doi:10.1097/icl.0b013e31827cb99f
- Menzies, 2010, The impact of contact angle on the biocompatibility of biomaterials, Optom. Vis. Sci., 87, 387, 10.1097/OPX.0b013e3181da863e · DOI · Paper page
- Craig, 2013, The TFOS International Workshop on Contact Lens Discomfort: Report of the contact lens interactions with the tear film subcommittee, Investig. Ophthalmol. Vis. Sci., 54, TFOS123-56
- Tighe, 2013, A decade of silicone hydrogel development: Surface properties, mechanical properties, and ocular compatibility, Eye Contact Lens, 39, 4, 10.1097/ICL.0b013e318275452b · DOI
- Sankaridurg, 2013, The future of silicone hydrogels, Eye Contact Lens, 39, 125, 10.1097/ICL.0b013e31827d1297 · doi:10.1097/icl.0b013e31827d1297
- Keir, 2013, Wettability and silicone hydrogel lenses: A review, Eye Contact Lens, 39, 100, 10.1097/ICL.0b013e31827d546e · doi:10.1097/icl.0b013e31827d546e
- Lira, 2017, Effect of Lens Care Systems on Silicone Hydrogel Contact Lens Hydrophobicity, Eye Contact Lens, 43, 89, 10.1097/ICL.0000000000000247 · doi:10.1097/icl.0000000000000247
- Fagehi, 2017, Care solution effects on contact lens in vivo wettability, Clin. Exp. Optom., 100, 623, 10.1111/cxo.12518 · doi:10.1111/cxo.12518
- 2014, Comparison of tear film surface quality measured in vivo on water gradient silicone hydrogel and hydrogel contact lenses, Eye Contact Lens, 40, 23, 10.1097/ICL.0000000000000008 · doi:10.1097/icl.0000000000000008
- Mochizuki, 2009, Turnover rate of tear-film lipid layer determined by fluorophotometry, Br. J. Ophthalmol., 93, 1535, 10.1136/bjo.2008.156828 · doi:10.1136/bjo.2008.156828
- Georgiev, 2017, Structure-function relationship of tear film lipid layer: A contemporary perspective, Exp. Eye Res., 163, 17, 10.1016/j.exer.2017.03.013 · doi:10.1016/j.exer.2017.03.013
- Read, 2011, Dynamic contact angle analysis of silicone hydrogel contact lenses, J. Biomater. Appl., 26, 85, 10.1177/0885328210363505 · doi:10.1177/0885328210363505
- Mann, 2013, Contact lens interactions with the tear film, Exp. Eye Res., 117, 88, 10.1016/j.exer.2013.07.013 · DOI
- Wolffsohn, 2018, Effect of contact lens surface properties on comfort, tear stability and ocular physiology, Contact Lens Anterior Eye, 41, 117, 10.1016/j.clae.2017.09.009 · doi:10.1016/j.clae.2017.09.009
- Lorentz, 2012, The impact of intermittent air exposure on lipid deposition, Optom. Vis. Sci., 89, 1574, 10.1097/OPX.0b013e31826c6508 · doi:10.1097/opx.0b013e31826c6508
- Campbell, 2013, Applicability of contact angle techniques used in the analysis of contact lenses, part 1: Comparative methodologies, Eye Contact Lens, 39, 254, 10.1097/ICL.0b013e31828ca174 · DOI
- Musgrave, C.S.A., and Fang, F. (2019). Contact Lens Materials: A Materials Science Perspective. Materials, 12. · DOI · Paper page
- Ratner, B.D. (2013). Correlation, Materials Properties, Statistics and Biomaterials Science. Biomaterials Science, Academic Press. [3rd ed.]. · DOI
- Cohen, 2012, Multiple rank test for pairwise comparisons, Contemporary Developments in Bayesian Analysis and Statistical Decision Theory: A Festschrift for William, E. Strawderman, Volume 8, 57
- Lee, 2016, Alternatives to P value: Confidence interval and effect size, Korean J. Anesthesiol., 69, 555, 10.4097/kjae.2016.69.6.555 · DOI · Paper page
- Svitova, 2011, Wettability conundrum: Discrepancies of soft contact lens performance in vitro and in vivo, Eur. Phys. J. Spec. Top., 197, 295, 10.1140/epjst/e2011-01471-6 · doi:10.1140/epjst/e2011-01471-6
- Morgan, 2007, In vitro water wettability of silicone hydrogel contact lenses determined using the sessile drop and captive bubble techniques, J. Biomed. Mater. Res. A, 83, 496
- 1985, Wetting: Statics and dynamics, RevModPhys, 57, 827
- Lin, 2010, Contact lenses wettability in vitro: Effect of surface-active ingredients, Optom. Vis. Sci., 87, 440, 10.1097/OPX.0b013e3181dc9a1a · DOI
- Bracco, G., and Holst, B. (2013). Contact Angle and Wetting Properties. Surface Science Techniques, Springer. · DOI · Paper page
- Curtin, 2013, Characterization of contact lenses through oxygen permeability, equilibrium water content, and silicone content, Investig. Ophthalmol. Vis. Sci., 54, 490
- Rex, J., Knowles, T., Zhao, X., Lemp, J., Maissa, C., and Perry, S.S. (2018). Elemental Composition at Silicone Hydrogel Contact Lens Surfaces. Eye Contact Lens. · doi:10.1016/j.clae.2018.04.097
- Tonge, 2001, The ex vivo wettability of soft contact lenses, Curr. Eye Res., 23, 51, 10.1076/ceyr.23.1.51.5418 · DOI
- Paradiso, 2016, Controlled Release of Antibiotics From Vitamin E-Loaded Silicone-Hydrogel Contact Lenses, J. Pharm. Sci., 105, 1164, 10.1016/S0022-3549(15)00193-8 · DOI
- Johnson, 2016, Inspired by the science of tears, Optician, 251, 32
- Diec, 2012, The clinical performance of lenses disposed of daily can vary considerably, Eye Contact Lens, 38, 313, 10.1097/ICL.0b013e31826586c3 · DOI
- Lira, 2008, Comparative study of silicone-hydrogel contact lenses surfaces before and after wear using atomic force microscopy, J. Biomed. Mater. Res. B Appl. Biomater., 85, 361, 10.1002/jbm.b.30954 · doi:10.1002/jbm.b.30954
- Santos, 2008, Bacterial adhesion to worn silicone hydrogel contact lenses, Optom. Vis. Sci., 85, 520, 10.1097/OPX.0b013e31817c92f3 · doi:10.1097/opx.0b013e31817c92f3
- Lorentz, 2007, The impact of lipid on contact angle wettability, Optom. Vis. Sci., 84, 946, 10.1097/OPX.0b013e318157a6c1 · DOI
- Cheng, 2004, Wettability of silicone-hydrogel contact lenses in the presence of tear-film components, Curr. Eye Res., 28, 93, 10.1076/ceyr.28.2.93.26231 · DOI
- Muntz, 2015, Tear exchange and contact lenses: A review, J. Optom., 8, 2, 10.1016/j.optom.2014.12.001 · DOI
- Downie, 2017, Tear film evaluation and management in soft contact lens wear: A systematic approach, Clin. Exp. Optom., 100, 438, 10.1111/cxo.12597 · DOI
- Chauhan, 2002, Settling and deformation of a thin elastic shell on a thin fluid layer lying on a solid surface, J. Colloid Interface Sci., 245, 187, 10.1006/jcis.2001.8008 · doi:10.1006/jcis.2001.8008
- Nichols, 2003, The effect of eye closure on the post-lens tear film thickness during silicone hydrogel contact lens wear, Cornea, 22, 539, 10.1097/00003226-200308000-00010 · doi:10.1097/00003226-200308000-00010
- Chen, 2010, Ultrahigh-resolution measurement by optical coherence tomography of dynamic tear film changes on contact lenses, Investig. Ophthalmol. Vis. Sci., 51, 1988, 10.1167/iovs.09-4389 · doi:10.1167/iovs.09-4389
- Stalder, 2006, A Snakebased Approach to Accurate Determination of Both Contact Points and Contact Angles, Colloids Surf. A, 286, 92, 10.1016/j.colsurfa.2006.03.008 · doi:10.1016/j.colsurfa.2006.03.008
- Yoshioka, 2002, KyPlot—A User-oriented Tool for Statistical Data Analysis and Visualization, Comput. Stat., 17, 425, 10.1007/s001800200117 · DOI
- Hammer, 2001, PAST: Paleontological Statistics Software Package for Education and Data Analysis, Palaeontol. Electron., 4, 9
Cited by
Shahabadi, N., Zendehcheshm, S., & Khademi, F. (2025) · Bioinorganic Chemistry and Applications · doi:10.1155/bca/9989560 · Cited by 6
Metadata
- Authors
- Eftimov, P., Yokoi, N., Peev, N., & Georgiev, G. A.
- Affiliations
- Department of Cytology, Histology and Embryology, Faculty of Biology, St. Kliment Ohridski University of Sofia, Sofia 1164, Bulgaria
- Department of Ophthalmology, Kyoto Prefectural University of Medicine, Kyoto 602-8566, Japan
- Department of Optics and Spectroscopy, Faculty of Physics, St. Kliment Ohridski University of Sofia, Sofia 1164, Bulgaria
- Venue
- International Journal of Molecular Sciences
- Publisher
- MDPI AG
- Published
- 2019-03-15
- Volume
- 20
- Issue
- 6
- Pages
- 1313
- ISSN
- 1422-0067
- Cited by
- 29
- References
- 0
- Language
- en
- Funders
- Alcon Japan Ltd. Japan (17)
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Impact of Air Exposure Time on the Water Contact Angles of Daily Disposable Silicone Hydrogels. https://app.cochat.ai/papers/10.3390/ijms20061313
- The paper
- Eftimov, P., Yokoi, N., Peev, N., & Georgiev, G. A. (2019). Impact of Air Exposure Time on the Water Contact Angles of Daily Disposable Silicone Hydrogels. International Journal of Molecular Sciences. https://doi.org/10.3390/ijms20061313
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.