A Novel Singular Value Decomposition-Based Denoising Method in 4-Dimensional Computed Tomography of the Brain in Stroke Patients with Statistical Evaluation
Open accessThe DOI matches its Crossref record; 29 of the 32 references it cites were found in a registry or an index; 2 of the 11 sources its claims cite were checked for support.
Abstract
Computed tomography (CT) is a widely used medical imaging modality for diagnosing various diseases. Among CT techniques, 4-dimensional CT perfusion (4D-CTP) of the brain is established in most centers for diagnosing strokes and is considered the gold standard for hyperacute stroke diagnosis. However, because the detrimental effects of high radiation doses from 4D-CTP may cause serious health risks in stroke survivors, our research team aimed to introduce a novel image-processing technique. Our singular value decomposition (SVD)-based image-processing technique can improve image quality, first, by separating several image components using SVD and, second, by reconstructing signal component images to remove noise, thereby improving image quality. For the demonstration in this study, 20 4D-CTP dynamic images of suspected acute stroke patients were collected. Both the images that were and were not processed via the proposed method were compared. Each acquired image was objectively evaluated using contrast-to-noise and signal-to-noise ratios. The scores of the parameters assessed for the qualitative evaluation of image quality improved to an excellent rating (p < 0.05). Therefore, our SVD-based image-denoising technique improved the diagnostic value of images by improving their quality. The denoising technique and statistical evaluation can be utilized in various clinical applications to provide advanced medical services.
Overview
CoChat Summary
TL;DR
This study applied a singular value decomposition (SVD)-based denoising method to 4D brain CT perfusion images from 20 suspected acute stroke patients, reconstructing images after excluding low singular-value components at thresholds of 5–40% of the second singular value. Denoising significantly improved SNR and CNR, with the 25% threshold suggested as a potential optimum for axial basal ganglia images.
Key findings
- SVD denoising significantly increased SNR and CNR at the basal ganglia level; the largest SNR improvement was a 16% increase at the 40% threshold (t = 2.908, p < 0.005), and the largest CNR improvement was a 43% increase at the 40% threshold (t = 6.327, p < 0.001).
- The 25% threshold had the best statistical significance by paired t-test (t = 3.764, p < 0.001), with a 14% SNR increase and a 37% CNR increase (t = 6.080, p < 0.001); the authors considered 25% potentially optimal for balancing noise reduction and diagnostic information.
- The full-rank original image contained 27 singular value components; at the 5% threshold, the 1st–20th singular values were included and the 21st–27th were separated, while at the 40% threshold only the 1st and 2nd singular values were included after separating 25 signal components.
- The abstract reports qualitative image-quality scores improved to an excellent rating (p < 0.05), though the Results section presents quantitative SNR/CNR analyses.
Methods
- Retrospective, IRB-approved study of 20 suspected acute stroke patients (excluded if already diagnosed with ischemic or hemorrhagic stroke on CT) at Kosin University Medical Center, with data acquired from 1 January 2016 to 1 December 2016.
- 4D CTP was performed on a 128-section dual-source CT scanner at 80 kVp and 200 mAs with 0.28 s rotation time, 32 × 1.2 mm collimation, 1.25 s temporal interval, 84 mm coverage, and reconstruction at 5 mm slice thickness with 3 mm interval; contrast injection used 40 cc non-ionic contrast at 6 mL/s followed by 60 cc saline.
- SVD decomposed the reshaped 2D spatio-temporal data matrix; thresholds were set at 5–40% of the second singular value (λ2) in 5% increments, singular values above threshold were retained and those below were set to zero, and denoised datasets were reconstructed. SNR and CNR at the basal ganglia were calculated, and paired t-tests compared original versus each denoised image using SPSS 25.0.
Limitations
- The paper states that the SVD-based technique was not applied to perfusion maps, the sample size was small, and only quantitative image quality was estimated, so diagnostic accuracy still needs assessment.
- The paper states that the threshold value was selected on an experiential basis, with no golden rule because signal and noise boundaries remain vague, and that reducing noise is a trade-off for acquiring diagnostic information.
- Evident from the text: the study used single-center retrospective data, and despite the abstract mentioning qualitative evaluation, the Results focus on quantitative SNR/CNR without a reported reader-performance or diagnostic-accuracy assessment.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures





Verification
Claims the paper makes
2 Valid · 0 Not valid · 6 Not checked · 34 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“The Gaussian noise that is generated by scanning the brain at low radiation doses interferes with the accurate interpretation of these images and with the identification of brain-blood vessels as well as of cerebral parenchyma [14].”
§ A Novel Singular Value Decomposition-Based Denoising Method in 4-Dimensional Computed Tomography of the Brain in Stroke Patients with Statistical Evaluation > 1. Introduction
What the sources say
- Partly supportedcites [14] Lee W.-J. Ahn B.-S. Park Y.-S. Radiation Dose and Image Quality of Low-dose Protocol in Chest CT: Comparison of Standard-dose Protocol J. Radiat. Prot. Res. 2012 37 84 89 10.14407/jrp.2012.37.2.084 sourceExperimental — accuracy not yet measured
Our copy supports only the generic half of the claim: the Abstract (p. 6) and Section 3.1 / Table 1 (p. 2) report that the low-dose protocol (120 kVp, 30 mAs, 2-mm) gave significantly more noise and significantly lower SNR and CNR than the standard-dose protocol, with the conclusion (p. 6) that LDP exposure settings need resetting with image quality in mind. But that finding is confined to chest CT — non-contrast mediastinal images at the level of carina bifurcation in 61 subjects, measured at ascending aorta and infraspinatus muscle (Section 2.1, p. 2), plus water/acrylic phantoms (Section 3.3, p. 4). No passage in our copy concerns the brain, cerebral parenchyma or brain blood vessels, none describes the noise as Gaussian (noise is merely defined as the standard deviation of CT number), and none states that noise interferes with accurate image interpretation. The brain-specific and interpretation-impairing parts of the claim are therefore not supported by this source. (LLM, conf=0.68)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Based on this study, statistical significance was demonstrated in enhancing image quality using 4D CTP datasets.”
§ A Novel Singular Value Decomposition-Based Denoising Method in 4-Dimensional Computed Tomography of the Brain in Stroke Patients with Statistical Evaluation > 4. Discussion
boostedstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on 2 assumptions not shown here.
A result the paper states — it cites no source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Regarding image a9, the 1st and 2nd singular values are predominantly depicted as providing diagnostic information.”
§ A Novel Singular Value Decomposition-Based Denoising Method in 4-Dimensional Computed Tomography of the Brain in Stroke Patients with Statistical Evaluation > 4. Discussion
plainstates an associationSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on 2 assumptions not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“As shown in Figure 3a,b, the removal of larger amounts of Gaussian noise on an image is observed when the image is reconstructed with a smaller number of SCs (i.e., a larger number of low singular values were excluded).”
§ A Novel Singular Value Decomposition-Based Denoising Method in 4-Dimensional Computed Tomography of the Brain in Stroke Patients with Statistical Evaluation > 3. Results > 3.1. SVD-Based Noise Removal
plainstates a causeSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on 2 assumptions not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The scores of the parameters assessed for the qualitative evaluation of image quality improved to an excellent rating (p < 0.05).”
§ A Novel Singular Value Decomposition-Based Denoising Method in 4-Dimensional Computed Tomography of the Brain in Stroke Patients with Statistical Evaluation > Abstract
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on 2 assumptions not shown here.
21 more claims
Create a free account to see all 42 claims and to verify the 9 references still open.
- Probably not a substantial claim. 11
- It may state two separate claims as one. 7
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated major: it would touch a result — a value a result rests on, reported and recomputed p-values on opposite sides of the significance level, two passages that conflict with no sign which is right, or a cited work retracted before this paper was published.
In the paper
§ A Novel Singular Value Decomposition-Based Denoising Method in 4-Dimensional Computed Tomography of the Brain in Stroke Patients with Statistical Evaluation > Abstract
“The scores of the parameters assessed for the qualitative evaluation of image quality improved to an excellent rating (p < 0.05).”
§ A Novel Singular Value Decomposition-Based Denoising Method in 4-Dimensional Computed Tomography of the Brain in Stroke Patients with Statistical Evaluation > 4. Discussion
“Finally, we only estimated quantitative image quality; therefore, the diagnostic accuracy needs to be assessed in comparison to that of non-SVD images.”
Citations
Cites
Tissue Plasminogen Activator for Acute Ischemic Stroke
(1995) · New England Journal of Medicine · doi:10.1056/nejm199512143332401 · Cited by 11,654
Feigin, V. L., Lawes, C. M., Bennett, D. A., & Anderson, C. S. (2003) · The Lancet Neurology · doi:10.1016/s1474-4422(03)00266-7 · Cited by 1,932
OFDM channel estimation by singular value decomposition
Edfors, O., Sandell, M., van de Beek, J.-J., Wilson, S. K., & Borjesson, P. O. (1998) · IEEE Transactions on Communications · doi:10.1109/26.701321 · Cited by 1,542
Strategies for Reducing Radiation Dose in CT
McCollough, C. H., Primak, A. N., Braun, N., Kofler, J., Yu, L., & Christner, J. (2009) · Radiologic Clinics of North America · doi:10.1016/j.rcl.2008.10.006 · Cited by 842 · cited 1×
“In 2009, the Food and Drug Administration (FDA) reported that patients who underwent a CT scan of the brain exhibited band-like hair loss and, therefore, issued a warning regarding the excessive radiation dose that patients are subjected to [12].”
BackgroundStroke Epidemiology and Risk Factor Management
Guzik, A., & Bushnell, C. (2017) · Continuum · doi:10.1212/con.0000000000000416 · Cited by 634
Stroke Epidemiology: Advancing Our Understanding of Disease Mechanism and Therapy
Ovbiagele, B., & Nguyen-Huynh, M. N. (2011) · Neurotherapeutics · doi:10.1007/s13311-011-0053-1 · Cited by 432
Denoising by Singular Value Decomposition and Its Application to Electronic Nose Data Processing
Jha, S. K., & Yadava, R. D. S. (2010) · IEEE Sensors Journal · doi:10.1109/jsen.2010.2049351 · Cited by 137 · cited 1×
“The term ‘rank’ refers to the number of singular values added to reconstruct a single image [28,29].”
BackgroundKorn, A., Fenchel, M. C., Bender, B., et al. (2011) · American Journal of Neuroradiology · doi:10.3174/ajnr.a2749 · Cited by 131 · cited 10×
“A MATLAB software application was used to calculate the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) [27].”
MethodInfluentialComparison of CT and MR imaging in ischemic stroke
Vymazal, J., Rulseh, A. M., Keller, J., & Janoušková, L. (2012) · Insights into Imaging · doi:10.1007/s13244-012-0185-9 · Cited by 116 · cited 1×
“Clinically, although MRI is the front-runner among imaging modalities for the detection of affected sites in stroke patients involving no risk of them being exposed to radiation [10], the long examination time that may be required limits its practical clinical use [11].”
BackgroundBrain CT perfusion improves intracranial vessel occlusion detection on CT angiography.
Becks, M., Manniesing, R., Vister, J., Pegge, S., Steens, S., van Dijk, E. V., Prokop, M., & Meijer, F. (2019) · Journal of neuroradiology. Journal de neuroradiologie · doi:10.1016/j.neurad.2018.03.003 · Cited by 89
Research works printed but not linked: 1
- Shlens J. A tutorial on principal component analysis derivation, discussion and singular value decomposition Mar 2003 25 1 16
Cited by
CT image denoising methods for image quality improvement and radiation dose reduction
Sadia, R. T., Chen, J., & Zhang, J. (2024) · Journal of Applied Clinical Medical Physics · doi:10.1002/acm2.14270 · Cited by 91
“106,139,156,168, Dataset (public or institutional) that is used for training/validation,independent dataset indicates those different from public and institutional (independent test).”
BackgroundA Hybrid Approach for CT Image Noise Reduction Combining Method Noise-CNN and Shearlet Transform
Katta, S., Singh, P., Garg, D., & Diwakar, M. (2024) · Biomedical & Pharmacology Journal · doi:10.13005/bpj/2991 · Cited by 33
Gradient-Based Pulsed Excitation and Relaxation Encoding in Magnetic Particle Imaging
Jia, G., Huang, L., Wang, Z., Liang, X., Zhang, Y., Zhang, Y., Miao, Q., Hu, K., Li, T., Wang, Y., et al. (2022) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2022.3193219 · Cited by 27
“Singular value decomposition (SVD) was applied on the system matrix [29] to determine the solution of Equation 8 for 1D image reconstruction [30].”
MethodNaeem, K., Kim, B. H., Yoon, D., et al. (2021) · Applied Sciences · doi:10.3390/app11041928 · Cited by 14
“It has been effectively demonstrated for data reconstruction [27] and denoising [28] applications of the magnetic resonance imaging (MRI).”
BackgroundSelf-citationArtificial Intelligence Based Medical Imaging: An Overview
Hong, J.-Y., Hyun, P. S., & Jung, Y.-J. (2020) · Bangsaseon gisul gwahak/Journal of radiological science and technology · doi:10.17946/jrst.2020.43.3.195 · Cited by 14
Self-citation
Metadata
- Authors
- Yang, W., Hong, J.-Y., Kim, J.-Y., Paik, S.-H., Lee, S. H., Park, J.-S., Lee, G., Kim, B. M., & Jung, Y.-J.
- Affiliations
- Dong-A University Hospital (KR)
- Dongseo University (KR)
- Inje University Ilsan Paik Hospital (KR)
- Korea University (KR)
- Samsung Medical Center (KR)
- Corresponding author
- Kim, Beop Min
- Jung, Young-Jin
- ORCID
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2020-05-28
- Volume
- 20
- Issue
- 11
- Pages
- 3063
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC7309118
- OpenAlex
- W3031827868
- Cited by
- 17
- References
- 32
- Citations by year
- 2025: 2 · 2024: 6 · 2023: 3 · 2022: 2 · 2021: 2 · 2020: 2
- FWCI
- 0.87
- Type
- journal article
- Language
- en
- Open access
- gold
- dx.doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- National Research Foundation of Korea (2015M3C7A1029034)
- Korea Health Industry Development Institute (KHIDI)
- Topics
- Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Image and Signal Denoising Methods
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Image Enhancement Techniques
- Keywords
- singular value decomposition, image denoising, brain perfusion imaging, acute stroke, radiation dose reduction, image quality, signal-to-noise ratio, contrast-to-noise ratio
- MeSH terms
- Algorithms, Brain, Humans, Image Processing, Computer-Assisted, Tomography, X-Ray Computed, Stroke, Signal-To-Noise Ratio
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-01, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-01 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (qwen3.8-omni-flash).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A Novel Singular Value Decomposition-Based Denoising Method in 4-Dimensional Computed Tomography of the Brain in Stroke Patients with Statistical Evaluation. https://app.cochat.ai/papers/10.3390/s20113063
- The paper
- Yang, W., Hong, J.-Y., Kim, J.-Y., Paik, S.-H., Lee, S. H., Park, J.-S., Lee, G., Kim, B. M., & Jung, Y.-J. (2020). A Novel Singular Value Decomposition-Based Denoising Method in 4-Dimensional Computed Tomography of the Brain in Stroke Patients with Statistical Evaluation. Sensors. https://doi.org/10.3390/s20113063
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.