Noninvasive Glucose Monitoring: Comprehensive Clinical Results
Abstract
CONTEXT: Intensive diabetes management using frequent blood glucose measurements to guide therapy has been shown to significantly improve short- and long-term outcomes. Development of a device that makes possible frequent, automatic, painless, and accurate measurements of glucose would facilitate intensive management. OBJECTIVE: To determine the accuracy of the GlucoWatch automatic glucose biographer (Cygnus Inc) compared with that of serial blood glucose measurements. DESIGN: Multicenter comparative study of the GlucoWatch biographer and the HemoCue blood glucose analyzer (Aktiebolaget Leo) performed between August 29 and October 17, 1998. Participants wore up to 2 biographers during the 15-hour study session and performed 2 fingersticks per hour for comparative blood glucose measurements. The biographers were calibrated with a single HemoCue measurement after a 3-hour warm-up period. Diet and insulin were manipulated to produce a broad glycemic range during the study. SETTING: Controlled clinical environment at 2 diabetes centers and 3 contract research organizations in the United States. PARTICIPANTS: A total of 92 subjects (mean [SD] age, 42.1 [15.1] years; 59.8% women) with type 1 or 2 diabetes requiring treatment with insulin. MAIN OUTCOME MEASURES: Mean error, mean absolute error, correlation, slope, and intercept using Deming regression, and clinical significance of differences between biographer readings and blood glucose measurements using the Clarke error grid. RESULTS: Results showed close tracking of blood glucose over a range of 2.2 to 22.2 mmol/L (40-400 mg/dL) for up to 12 hours using a single point calibration. The biographer readings lagged behind serial blood glucose values by a mean of 18 minutes. An analysis of 2167 data pairs shows a linear relationship (r = 0.88; slope = 1.03; intercept = -0.33 mmol/L [-6 mg/dL]) between biographer readings and serial glucose measurements. The mean absolute error between the 2 measurements was 15.6% (mean error [SD], -0.07 [1.82] mmol/L [-1 [33] mg/dL]), and 96.8% of the data fell in the therapeutically relevant regions of the error grid analysis. CONCLUSION: These results demonstrate close agreement between GlucoWatch biographer readings and blood glucose measurements using repeated fingerstick blood samples. The automatic, frequent, and noninvasive measurements obtained with the biographer provides more information about glucose levels than the current standard of care.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“One of the first commercial wearable sensing devices, a wrist-worn glucose monitor, used iontophoresis to extract ISF for analysis77.”
From End-to-end design of wearable sensors (2022) · cites this paper as [77]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667
Turner, R. C., Fox, C., Matthews, McElroy, H. H., Cull, C. A., Holman, R. R., Neil, P., Hadden, D. R., Wright, D., Manley, E., et al. (1998) · The Lancet · Cited by 8,419
Ohkubo, Y., Kishikawa, H., Araki, E., Miyata, T., Isami, S., Motoyoshi, S., Kojima, Y., Furuyoshi, N., & Shichiri, M. (1995) · Diabetes Research and Clinical Practice · doi:10.1016/0168-8227(95)01064-k · Cited by 3,079
DRAIZE, J. H., WOODARD, G., & CALVERY, H. O. (1944) · The Journal of Pharmacology and Experimental Therapeutics · doi:10.1016/s0022-3565(25)08751-8 · Cited by 2,602
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435
Measurement of glucose in diabetic subjects using noninvasive transdermal extraction
Tamada, J. A., Bohannon, N. J. V., & Potts, R. O. (1995) · Nature Medicine · doi:10.1038/nm1195-1198 · Cited by 182
Self-citationReverse Iontophoresis: Noninvasive Glucose Monitoring in Vivo in Humans
Rao, G. S., Guy, R. H., Glikfeld, P., LaCourse, W. R., Leung, L., Tamada, J. A., Potts, R. O., & Azimi, N. (1995) · Pharmaceutical Research · doi:10.1023/a:1016271301814 · Cited by 138
Self-citationMeyerhoff, C., Bischof, F., Sternberg, F., Zier, H., & Pfeiffer, E. F. (1992) · Diabetologia · doi:10.1007/bf02221686 · Cited by 126
Schmidtke, D. W., Freeland, A. C., Heller, A., & Bonnecaze, R. T. (1998) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.95.1.294 · Cited by 120
Validation of home blood glucose meters with respect to clinical and analytical approaches
Brunner, G. A., Ellmerer, M., Sendlhofer, G., Wutte, A., Trajanoski, Z., Schaupp, L., Quehenberger, F., Wach, P., Krejs, G. J., & Pieber, T. R. (1998) · Diabetes Care · doi:10.2337/diacare.21.4.585 · Cited by 119
Thomé-Duret, V., Reach, G., Gangnerau, M. N., Lemonnier, F., Jc, K., Zhang, Y., Hu, Y., & Wilson, G. S. (1996) · Analytical Chemistry · doi:10.1021/ac960069i · Cited by 118
Pickup, J. C., Shaw, G. W., & Claremont, D. J. (1989) · Diabetologia · doi:10.1007/bf00265097 · Cited by 114
Accuracy of home blood glucose meters during hypoglycemia
Trajanoski, Z., Brunner, G. A., Gfrerer, R., Wach, P., & Pieber, T. R. (1996) · Diabetes Care · doi:10.2337/diacare.19.12.1412 · Cited by 85
Bolinder, J., Toft, E., Ungerstedt, U., & Arner, P. (1997) · Diabetes Care · doi:10.2337/diacare.20.1.64 · Cited by 71
Design and Simulation of a Reverse Iontophoretic Glucose Monitoring Device
Kurnik, R. T., Berner, B., Tamada, J. A., & Potts, R. O. (1998) · Journal of The Electrochemical Society · doi:10.1149/1.1838924 · Cited by 53
Self-citationKurnik, R. T. (1999) · Sensors and Actuators B Chemical · doi:10.1016/s0925-4005(99)00239-7 · Cited by 41
Evaluation of a Second‐generation Electrochemical Blood Glucose Monitoring System
Tieszen, K. L., Burton, S., Dornan, T., Matthews, D. R., & McMurray, J. R. (1995) · Diabetic Medicine · doi:10.1111/j.1464-5491.1995.tb00449.x · Cited by 5
Cited by
Prausnitz, M. R., & Langer, R. (2008) · Nature Biotechnology · doi:10.1038/nbt.1504 · Cited by 3,517
Current status and future potential of transdermal drug delivery
Prausnitz, M. R., Mitragotri, S., & Langer, R. (2004) · Nature Reviews Drug Discovery · doi:10.1038/nrd1304 · Cited by 1,462
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Kirkman, M. S., Lernmark, A., Metzger, B. E., & Nathan, D. M. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2010.161596 · Cited by 1,360
End-to-end design of wearable sensors
Ates, H. C., Nguyen, P. Q., Gonzalez-Macia, L., Morales-Narváez, E., Güder, F., Collins, J. J., & Dincer, C. (2022) · Nature Reviews Materials · doi:10.1038/s41578-022-00460-x · Cited by 1,306
Transdermal drug delivery: overcoming the skin’s barrier function
Naik, A., Kalia, Y. N., & Guy, R. H. (2000) · Pharmaceutical Science & Technology Today · doi:10.1016/s1461-5347(00)00295-9 · Cited by 720
Metadata
- Authors
- Tamada, J. A., Garg, S., Jovanovic, L., Pitzer, K. R., Fermi, S., Potts, R. O., & and the Cygnus Research Team
- Venue
- JAMA
- Publisher
- American Medical Association (AMA)
- Published
- 1999-11-17
- Volume
- 282
- Issue
- 19
- Pages
- 1839
- ISSN
- 0002-9955 (ISSN-L)
- 0098-7484
- OpenAlex
- W2088032108
- Cited by
- 282
- References
- 18
- FWCI
- 11.70
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.